Steris Harmonyair Equipment Boom Operating instructions

Type
Operating instructions
10038023
Revision G
HarmonyAIR®
Equipment Boom
Operator Manual
2
10038023_G HarmonyAIR®Equipment Boom Operator Manual
TABLE OF CONTENTS
Section Number Description Page
1 A Word from STERIS...................................................................................................................................... 4
1.1 Introduction - A word from STERIS Corporation...................................................................... 4
1.2 Indications for Use...................................................................................................................... 4
1.3 Service Informaton ..................................................................................................................... 4
1.4 Advisory....................................................................................................................................... 4
1.5 Addresses.................................................................................................................................... 5
2 Safety Precautions ........................................................................................................................................ 6
3 Symbols ........................................................................................................................................................ 11
4 Technical Specifications.............................................................................................................................. 13
4.1 Application Specific................................................................................................................... 13
4.2 HarmonyAir Equipment Boom................................................................................................. 15
4.3 HarmonyAIR Equipment Boom with Supply Column/Supply Head ...................................... 20
4.4 Essential Performance ............................................................................................................. 20
4.5 Environmental Conditions........................................................................................................ 20
4.6 Electrical Requirements ........................................................................................................... 21
4.7 Accessories ............................................................................................................................... 21
5 Component Identification ........................................................................................................................... 22
5.1 Component Identification (Single Fixed Arm System)............................................................ 22
5.2 Component Identification (Dual Fixed Arm System) .............................................................. 23
5.3 Component Identification (Spring Arm System) ..................................................................... 24
5.4 Component Identification (Motorized Arm System)............................................................... 25
5.5 Component Identification (Central Tandem Fixed Arm System) ........................................... 26
5.6 Component Identification (Central Tandem Motorized Arm System)................................... 27
5.7 Component Identification (Central Tandem Spring Arm System) ......................................... 28
6 Operating Instructions ................................................................................................................................ 29
6.1 Introduction-Operating Instructions ....................................................................................... 29
6.2 Supply Column Load Capacity.................................................................................................. 29
6.3 Small Shelf Load Capacity ........................................................................................................ 29
6.4 Large Shelf Load Capacity ........................................................................................................ 29
6.5 Preparing Unit for Initial Use (Column) for Fixed Arm System.............................................. 30
6.6 Preparing Unit for Initial Use (Column) for Motorized Arm System...................................... 32
6.7 Preparing Unit for Initial Use (Shelf) for Fixed Arm System .................................................. 34
6.8 Preparing Unit for Initial Use (Shelf) for Motorized Arm System.......................................... 36
6.9 Supply Head Load Capacity...................................................................................................... 37
6.10 Preparing Unit for Initial Use (Supply Head for Fixed Arm System)...................................... 38
6.11 Preparing Unit for Initial Use (Supply Head for Spring Arm/Fixed Arm System) ................. 40
6.12 Preparing Unit for Initial Use (Supply Head for Motorized Arm System) ............................. 42
6.13 Operating the Unit .................................................................................................................... 44
6.14 Gas Supply Connectors............................................................................................................. 46
6.15 Hose Assembly Operating Instructions................................................................................... 48
6.16 Means of Isolation - Disconnecting Power and Gas Connections ........................................ 49
6.17 Data Plate Label........................................................................................................................ 50
6.18 Label Location on Supply Column............................................................................................ 51
6.19 Label Location on Supply Head................................................................................................ 52
3
Operator Manual HarmonyAIR®Equipment Boom 10038023_G
7 Cleaning the Equipment.............................................................................................................................. 53
7.1 Disinfecting/Cleaning ............................................................................................................... 53
7.2 Disinfect Surfaces..................................................................................................................... 54
8 Troubleshooting........................................................................................................................................... 55
8.1 Troubleshooting........................................................................................................................ 55
8.2 Insufficient Gas Supply or Audible Hissing Sound.................................................................. 56
8.3 Arm Only Moves with Difficulty or Drifts................................................................................ 56
8.4 LED Brake Indicators do not Light ........................................................................................... 56
8.5 Brakes do not Release or are Continually Disengaged .......................................................... 56
9 Routine Maintenance .................................................................................................................................. 57
9.1 Preventive Maintenance .......................................................................................................... 57
9.2 Stop Adjustment Column and Bearing(s)............................................................................... 58
9.3 Testing Requirements (Canada Only)...................................................................................... 58
9.4 Replacement Parts ................................................................................................................... 58
10 Waste Management Guidelines................................................................................................................. 59
11 Appendix EMC Compliance Technical Data............................................................................................... 60
1 — A Word from STERIS
4
10038023_G HarmonyAIR®Equipment Boom Operator Manual
A Word from STERIS
1.1 Introduction - A word from STERIS Corporation
This manual contains important information on proper use and maintenance of the HarmonyAIR Equipment Boom. All
personnel involved in the use and maintenance of this equipment must carefully review and comply with the warnings,
cautions, and instructions contained in this manual. These instructions are important to protect the health and safety of
personnel operating a HarmonyAIR Equipment Boom and should be retained in a conveniently accessible area for quick
reference.
STERIS is able to provide Customer training. Contact STERIS Customer Service for more information.
Complete instructions for uncrating and connecting utilities, as well as equipment drawings, have been furnished. If
missing, contact STERIS for replacement copies, giving the serial number and model numbers of the unit.
STERIS carries a complete line of accessories for use with this equipment management system. A STERIS
representative will gladly review these accessories with you. Any ancillary equipment (i.e., monitors, video recorders,
etc.) that is to be used with the HarmonyAIR Equipment Boom must comply with all applicable medical equipment
standards.
1.2 Indications for Use
For use in hospital facilities where there is a need for ready access to medical gases, electrical power and audio/visual
data services. HarmonyAIR Equipment Boom organize surgical equipment such as endoscopy equipment, laparoscopic
equipment, cautery devices, insufflators, etc. HarmonyAIR Equipment Boom units help hospitals to improve efficiency
and safety, and to maximize total room workflow by ergonomically centralizing all surgical support equipment:
As individual unit or in combination with the ceiling supply units of the HarmonyAIR Equipment Management
System.
For communicating and exchanging data, e.g., telephone, nurse's call button or computer networking.
For the ergonomic positioning of medical equipment during use.
For supplying medical equipment with electrical power and medical gases.
To provide additional instrument trays, shelves and supports.
The intended user for this equipment includes OR, ER/Trauma Room, Ambulatory Surgery Center and ICU staff.
NOTE:
There are no known contradictions to using this unit.
1.3 Service Informaton
You are encouraged to contact STERIS concerning our comprehensive annual maintenance agreements. Under the
terms of these agreements, preventive maintenance, adjustments and replacement of worn parts are done on a
scheduled basis to help ensure equipment performance at peak capability and to help avoid untimely or costly
interruptions. STERIS maintains a global staff of well-equipped, factory-trained technicians to provide this service, as
well as repair services. Please contact STERIS for details.
1.4 Advisory
A listing of the safety precautions to be observed when operating and servicing this equipment can be found inSection
2, of this manual. Do not operate or service the equipment until you have become familiar with this information.
Any alteration of this equipment not authorized or performed by STERIS that could affect its operation will void the
warranty and could violate national, state and local regulations.
1 — A Word from STERIS
5
Operator Manual HarmonyAIR®Equipment Boom 10038023_G
1.5 Addresses
STERIS Corporation
2720 Gunter Park East
Montgomery, AL 36109 USA
Class 1 Equipment
Ordinary Equipment (enclosed equipment without
protection against ingress of water)
Equipment not suitable for use in the presence of a
flammable anesthetic mixture with air or with oxygen or
nitrous oxide.
Suitable for continuous operation.
Canada Only:
Nonflammable medical gas pipeline systems shall
comply with CSA Z7396-1.
Electrical installations in healthcare facilities shall comply
with CSA C22.1 and CAN/CSA-Z32.
Anesthetic gas scavenging systems shall comply with
CSA Z7396.2.
STERIS Corporation, Montgomery, Alabama is an ISO 13485 certified facility.
The base language of this document is ENGLISH. Any translations must be made from the base language
document.
2 — Safety Precautions
6
10038023_G HarmonyAIR®Equipment Boom Operator Manual
Safety Precautions
The following Safety Precautions must be observed when operating or servicing this HarmonyAIR Equipment Boom.
WARNING indicates the potential for personal injury and CAUTION indicates the potential for damage to equipment. For
emphasis, certain Safety Precautions are repeated throughout the manual. It is important to review all Safety
Precautions before operating or servicing the unit.
Strictly following these Safety Precautions enhances the Customer's ability to use the unit safely and effectively and
helps avoid improper maintenance methods which may damage the unit or render it unsafe. It is important to
understand that these Safety Precautions are not exhaustive; Customers are encouraged to develop their own safety
policies and procedures to enhance and complement these Safety Precautions.
WARNING
PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD
Repairs and adjustments to the unit should be made only by STERIS or STERIS-trained,
fully qualified personnel. Work by inexperienced or unqualified persons or the
installation of unauthorized parts could cause personal injury, invalidate the warranty or
result in costly damage.
When installing clinical instrumentation, securely attach instrumentation to platform
and support shelves/accessories. Instrumentation could fall off platform and shelves
during unit positioning if not fastened securely.
Avoid patient injury and anesthesia head damage. Do not place objects on top of
anesthesia head.
Use of this equipment adjacent to other equipment should be avoided due to risk of
improper operation of the powered device. Observe equipment used adjacent to this
equipment for proper operation prior to use.
This equipment is intended for use by healthcare professionals only. This equipment
may cause radio interference or may disrupt the operation of nearby equipment. It may
be necessary to take mitigation measures, such as re-orienting the equipment or
shielding the location.
2 — Safety Precautions
7
Operator Manual HarmonyAIR®Equipment Boom 10038023_G
WARNING
PERSONAL INJURY HAZARD
Before using on a patient, check for proper equipment operation and acceptable gas
flows.
Do not position equipment boom over patient.
Nitrogen regulator is intended as a means of supplying power to pneumatic surgical
tools only. Nitrogen regulator is not to be used for medical purposes (e.g., anesthesia).
Connecting electrical equipment to the multiple socket outlet (MSO) effectively leads to
creating a medical electrical (ME) system, and the result can be a reduced level of
safety.
The compressed air supply to the ceiling supply unit must be turned off for the repair
procedure. Inform the user in advance before turning off the compressed air supply.
The incoming ac power supply must be interrupted temporarily during repair. To avoid
endangering the patients, including those on life support mechanisms, it is critical that
all users connected to the same mains power supply be informed so that appropriate
precautions can be taken.
Install shelf according to the instructions provided in this manual. Improperly installed
shelf can cause a pinching hazard.
WARNING
FIRE HAZARD
Avoid any risk of fire. Do not use flammable cleaning agents when cleaning or
disinfecting the HarmonyAIR Equipment Management System.
Avoid dust/lint buildup that could start a fire in the system. Contact STERIS or STERIS-
trained technician to remove dust/lint as necessary.
2 — Safety Precautions
8
10038023_G HarmonyAIR®Equipment Boom Operator Manual
WARNING
EXPLOSION HAZARD
The HarmonyAIR Equipment Boom must not be installed where flammable gases or
anesthesia will be used. This equipment is not designed for use in a hazardous location
as defined in National Fire Protection Association (NFPA) Flammable Anesthetics Code.
WARNING
ELECTRIC SHOCK AND BURN HAZARD
To avoid the risk of electric shock, this equipment must only be connected to a supply
mains protective earth.
Disconnect all utilities to the HarmonyAIR Equipment Boom before servicing. Do not
service the equipment management system unless all utilites have been properly locked
out. Always follow lockout-tagout and electrical safety-related work practice standards.
CAUTION
POSSIBLE EQUIPMENT DAMAGE
Any use of the apparatus requires full understanding and strict observation of these
instructions. The apparatus is only to be used for purposes specified here.
The apparatus must be inspected and serviced at six month intervals by trained service
personnel to maintain basic safety, essential performance and EMC at acceptable levels.
Refer to maintenance manual (P764335-769). Maintain complete service records.
Repair and general overhaul of the apparatus may only be carried out by STERIS or
STERIS-trained service personnel. STERIS recommends that a service contract be
obtained with STERIS and that all repairs also be carried out by STERIS. Only authentic
STERIS spare parts may be used for maintenance. Observe Maintenance Intervals
section.
No modification allowed to this equipment other than as performed or approved by
STERIS.
2 — Safety Precautions
9
Operator Manual HarmonyAIR®Equipment Boom 10038023_G
Use of any disinfectant solution OTHER than those listed here may cause discoloration
or deformation on system components:
Coverage® Plus Germicidal Surface Wipes Disinfecting/Deodorizing/Cleaning Wipes
Quaternary Ammonium Compound (Quats) with Ethanol solvent
Quaternary Ammonium Compound (Quats) with Isopropyl Alcohol (IPA) solvent
Quaternary Ammonium Compound (Quats) with IPA + 2-Butoxyethanol solvent
Quaternary Ammonium Compound (Quats) + Biguanide
H2O2 (Hydrogen Peroxide)
Neutral Cleaners
Hypochlorite (3 to 6% solution diluted 1:10 in water)
The use of H2O2 + PAA (Hydrogen Peroxide + Peracetic Acid) is strongly discouraged for
use on all STERIS products.
Always follow manufacturer instructions for concentrations and use of cleaning
products.
Do not use STERIS Stainless Steel Cleaner to clean the surface of the label found on the
nitrogen and/or instrument air regulator panel assemblies. Using this cleaner will result
in the label lifting at the edges.
Only personnel authorized to operate the equipment and properly trained in the use of
the hose assembly should use this product.
Keep the hose assembly in a clean area when not being used.
Verify all connections are tightened and free of leaks prior to use.
Use only an oxygen-safe leak detector when testing for leaks.
DO NOT gas sterilize with ethylene oxide (EO).
DO NOT clean with pungent hydrocarbons.
Connection to the gas source must be done with fittings appropriately indexed to the
latch assembly on the supply column/head.
2 — Safety Precautions
10
10038023_G HarmonyAIR®Equipment Boom Operator Manual
Cables or accessories other than those supplied by STERIS may affect EMC
performance.
Portable RF communications equipment, including peripherals such as antenna cables
and external antennas, should be used no closer than 12" (305 mm) to any part of the
equipment boom, including cables specified by STERIS. Degradation of equipment
performance could result.
For standard EMS units the total weight capacity on supply column (consisting of
shelves, equipment, and accessories) not to exceed 450 lb. (204 kg.). DO NOT
OVERLOAD.
For Central Tandem EMS units, total weight capacity on supply column (consisting of
shelves, equipment, and accessories) not to exceed 285 lb. (129 kg.). DO NOT
OVERLOAD.
Small shelf weighs 5.7 lb. (2.6 kg.). Maximum load capacity of small shelf is 44 lb. (20
kg.). DO NOT OVERLOAD.
Large shelf weighs 18 lb. (8.2 kg.). Maximum load capacity of large shelf is 110 lb. (50
kg.). DO NOT OVERLOAD.
Total weight capacity of supply head (consisting of accessories mounted to the
distance tube above the supply head) not to exceed 70 lb. (32 kg). DO NOT
OVERLOAD.
For equipment booms with electromechanical brakes, avoid brake pad and internal
bearing damage:
Do not articulate the suspension arm(s) without first depressing the brake release
button(s).
After depressing brake release button(s), the suspension arms, supply column, or
supply head can be repositioned in the room.
Bring the system to a complete stop before releasing the brake button(s).
NOTE:
The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11
class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment might
not offer adequate protection to radio-frequency communication services. The user might need to take mitigation
measures, such as relocating or re-orienting the equipment.
3 — Symbols
11
Operator Manual HarmonyAIR®Equipment Boom 10038023_G
Symbols
The following symbols may appear on the HarmonyAIR Equipment Boom:
Table 3-1. Definition of Symbols
Symbol Definition
Protective Earth (Ground)
Consult manual for further instructions
General Warning
Consult instructions before use
Serial Number of Unit
Equipment or Reorder Number
Symbol indicating the legal manufacturer name and address
Symbol indicating the date of manufacture (yyyy/mm)
Symbol indicating This Way Up
3 — Symbols
12
10038023_G HarmonyAIR®Equipment Boom Operator Manual
Table 3-1 Definition of Symbols (Continued)
Symbol Definition
Symbol indicating high voltage.
Maximum Load Hazard (maximum load limit given on label)
V~ Voltage Rating of Unit, Alternating Current
AAmperage Rating of Unit
Hz Frequency Rating of Unit
Table 3-2. Warning Labels
Warning Label - Disconnect Supply Before Servicing
Warning Label - Brake Control System Power Switch
4 — Technical Specifications
13
Operator Manual HarmonyAIR®Equipment Boom 10038023_G
Technical Specifications
4.1 Application Specific
The HarmonyAIR®Equipment Boom’s minimal footprint enables it to be easily placed at various locations in OR or ICU
rooms without monopolizing valuable space. The HarmonyAIR®Equipment Boom supplies gases, electrical power, and
acts as a shelving unit for various types of medical equipment. Surgical staff in operating rooms, trauma centers, critical
care rooms, clinics, and outpatient surgery centers will be the primary users.
Intended Medical Indication:
The HarmonyAIR Equipment Boom can be used in a variety of applications, including diagnostic procedures performed
in outpatient surgery centers, operating rooms, intensive care units, and major acute care suites.
Intended Patient Population:
There are no restrictions on patient populations.
Intended part of the body or type of tissue applied to or interacted with:
This device will not directly interact with the patient but will interact with the caregiver. The hands, arms, and torso
may be used to interact with the device.
Intended user profiles:
Surgical staff members are the intended operators of this device. The Operator Manual is written at a level consistent
with this intended user.
Intended conditions of use:
This equipment is designed to be used in operating rooms, trauma rooms, exam rooms, and critical care areas of
hospitals, clinics, physician offices and outpatient surgery centers that meets the environment requirements of the
operator manual. This equipment is suitable for continuous operation.
Operating principle:
Electromechanical brake operation for lateral movement:
Column/head lateral movement control is accomplished releasing either of two electromechanical brakes.
The brakes are controlled by two pushbuttons on the column handle, accessory shelf handle, or on the supply
head handle. The green button is for the upper arm brake, and the blue button for the lower arm brake.
Motorized Arm operation for vertical movement:
The motorized arm allows the head/column to be adjusted up or down by activating a motor. The motor is
activated to move the head up by pressing and holding the up arrow button on the handle. Conversely, the
motor is activated to move the head down by pressing and holding the down arrow button on the handle. In
either case, releasing the button will stop movement of the head.
The available range of travel can be adjusted by repositioning two limit switched installed along the drive
screw. The adjustment can be accessed under the covers at the second bearing.
Gas Lines:
Each gas hose line has a special marked hose and fitting for each gas outlet based on the DISS system.
The gas outlets will not allow crossing of lines.
Each outlet is marked with type of gas.
Nitrogen and instrument air have pressure regulation and a pressure meter for control of these gases.
The pressure meter is ranged so that the readings are in the middle third of the meter range. The accuracy is
±2%.
4 — Technical Specifications
14
10038023_G HarmonyAIR®Equipment Boom Operator Manual
Electrical Power:
Medical equipment power is divided for multiple outlets on different electrical panel circuit breakers (customer
supplied).
Each circuit number is identified on the main label and the rating for that breaker.
A special circuit is provided for Laser equipment when so configured — normally 230 VAC and is restricted to
one receptacle.
Ambient Light:
Supply Heads that have brakes will also be equipped with a light on the bottom surface of the unit for user
convenience. The light can be set to off, low, medium, or high illumination.
The light is controlled by two pushbuttons on the head handles. The + turns the light on or increases its
illumination, while the – button decreases illumination or turns the light off.
4 — Technical Specifications
15
Operator Manual HarmonyAIR®Equipment Boom 10038023_G
4.2 HarmonyAir Equipment Boom
Figure 4-1. HarmonyAIR Equipment Boom
4 — Technical Specifications
16
10038023_G HarmonyAIR®Equipment Boom Operator Manual
Figure 4-2. HarmonyAIR Equipment Boom
4 — Technical Specifications
17
Operator Manual HarmonyAIR®Equipment Boom 10038023_G
Figure 4-3. HarmonyAIR Equipment Boom
4 — Technical Specifications
18
10038023_G HarmonyAIR®Equipment Boom Operator Manual
Figure 4-4. HarmonyAIR Equipment Boom
4 — Technical Specifications
19
Operator Manual HarmonyAIR®Equipment Boom 10038023_G
Figure 4-5. HarmonyAIR Equipment Boom
4 — Technical Specifications
20
10038023_G HarmonyAIR®Equipment Boom Operator Manual
4.3 HarmonyAIR Equipment Boom with Supply Column/Supply Head
HarmonyAIR Equipment Boom Single Arm or Double Arm
with Supply Column/Supply Head
Ceiling mounting:
Suspended ceiling depth up to 47" (1200 mm) mounting on concrete ceiling with heavy-duty anchor
Maximum moment on structural ceiling 3024 ft-lb (4100 N•m) for single arm HarmonyAIR Equipment Boom
6417 ft-lb (8700 N•m) for double arm HarmonyAIR Equipment Boom
Rotation range of joints: Gradually adjustable of up to max. 330°
4.4 Essential Performance
A. Uninterrupted delivery of power to receptacles as supplied to the device input with less than 3% voltage loss within
the system.
B. Uninterrupted deliver of gas flow to the gas outlets with gas flow rate at least 5 cfm when gas line is pressurized at
160 psi.
NOTE: For extreme cases of EM disturbance, the user may see increased noise on the ac receptacles.
4.5 Environmental Conditions
Transport and Storage
NOTE:
The following storage conditions apply for up to 15 weeks.
Temperature: 13 to 158°F (11 to +70°C)
Relative Humidity: 10 to 75%
Barometric Pressure: 0.5 to 1.05 atm STP atmospheric pressure (500 to 1060 hPa)
Store only in closed or covered spaces. After 15 weeks, the values for the Operational Location Requirements
(below) apply.
Operational Location Requirements
The equipment is not suitable for operation in areas where there is an explosion hazard.
The HarmonyAIR Equipment Boom is not suitable for use in the presence of a flammable anesthetic mixture with
air or with oxygen or nitrous oxide.
The ambient temperature during operation must be kept between 20 and 25°C (68 and 77°F)
Relative humidity should stay within a range of 30 to 75%.
Atmospheric pressure should stay within a range of 0.7 to 1.05 atm STP atmospheric pressure (700 to 1060 hPa).
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65

Steris Harmonyair Equipment Boom Operating instructions

Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI