Steris Anesthesia / Respiratory Rack Operating instructions

Type
Operating instructions
i
Introduction Operator Manual 920515-083
This manual contains important information on the proper use and
routine maintenance of the AN Vision SC Manifold Rack. All
personnel involved in the use and routine maintenance of this
equipment must carefully review and comply with the Safety
Precautions and instructions contained in this manual. These
instructions are important to protect the health and safety of
personnel using this AN Vision SC Manifold Rack to process items in
the Reliance®/Hamo® Vision® Single-Chamber Washer/Disinfector or
the Vision® Single-Chamber Washer/Disinfector, and should be
retained in a conveniently accessible area for quick reference.
Advisory IMPORTANT: A listing of the Safety Precautions to be observed
when using and servicing this equipment can be found in SECTION 1
of this manual. Do not use the equipment until you have become
familiar with this information.
STERIS does not intend, recommend or represent in any way that
this AN Vision SC Manifold Rack be used in the Reliance/Hamo
Vision Single-Chamber Washer/Disinfector or the Vision Single-
Chamber Washer/Disinfector for the terminal disinfection or
sterilization of any regulated medical device. The AN Vision SC
Manifold Rack is intended only to perform an initial step in the
processing of soiled, reusable medical devices in the washer/
disinfector. If medical devices will be contacting blood or
compromised tissues, such devices must be terminally processed
in accordance with Good Hospital Practices before each use in
human patients.
STERIS does not recommend processing flexible endoscopes, single-
use instruments, rigid telescopes (optical lens), fiber-optic cables,
multi-part instruments that cannot be disassembled, heat-sensitive
devices and coagulating electrodes using the AN Vision SC Manifold
Rack in the washer/disinfector. Processing these types of instruments
may result in instrument or equipment damage as well as improperly
cleaned items. All items processed using the AN Vision SC Manifold
Rack in the washer/disinfector must withstand temperatures of up to
240°F (116°C).
This AN Vision SC Manifold Rack is specifically designed to only hold
items as outlined in this manual. If there is any doubt about a specific
material or product, contact the instrument manufacturer for
recommended washing techniques.
Any alteration of this equipment not authorized or performed by
STERIS will void the warranty. Alteration of this equipment which
could adversely affect its operation and efficacy may violate national,
state and local laws or regulations.
©2022, STERIS Corporation. All rights reserved. Printed in Canada.
A WORD FROM STERIS
ii
920515-083 Operator Manual Introduction
Serious incidents that have occurred in relation to this medical
device should be reported to the manufacturer and competent
authority in the country where the incident occurred.
Indications For Use The AN Vision SC Manifold Rack is designed to be used in the
Reliance/Hamo Vision Single-Chamber Washer/Disinfector for the
efficient cleaning, rinsing, drying and intermediate level disinfection
of different types of anesthesia tubing, bags, masks and respiratory
goods.
IMPORTANT: The AN Vision SC Manifold Rack is designed to be
used only with the Anesthesia/Respiratory Goods cycle.
Service Information A thorough preventive maintenance program is essential to help
ensure safe and proper washer/disinfector and rack operation. This
manual contains a routine maintenance procedure which should be
followed for satisfactory equipment performance.
Customers are encouraged to contact STERIS concerning extended
service maintenance agreements to give your washer/disinfector
and rack planned maintenance, which will help ensure equipment
performance according to factory specifications. STERIS maintains a
global staff of well-equipped, qualified service technicians to provide
this service, as well as professional repair services. Please contact
STERIS for details.
Technical
Specifications
The following specifications apply to the AN Vision SC Manifold
Rack:
Overall Size: 25.5 x 25.5 x 22" (648 x 648 x 559 mm)
Weight: 28.5 lb (13 kg)
Material: Stainless steel, Plastic, Silicone, Polypropylene
The AN Vision SC Manifold Rack includes (refer to Figure 2-1):
Large hollow jets (x 12);
Medium hollow jets (x 12);
Small hollow jets (x 12);
Hooks (x 8);
Spring stoppers (large hollow jets) (x 12);
Helicline for anesthesia tubing (x 1);
Adapters for anesthesia tubing (secured in bottom of rack) (x 8);
Additional adapters for anesthesia tubing (x 8).
iii
Introduction Operator Manual 920515-083
Certification The Reliance/Hamo Vision Single-Chamber Washer/Disinfector,
compatible with the AN Vision SC Manifold Rack, meets the
applicable requirements of certain standards. Refer to washer/
disinfector Operator Manual (P920508-556) for the current list of
standards.
The Vision Single-Chamber Washer/Disinfector, compatible with the
AN Vision SC Manifold Rack, meets the applicable requirements of
certain standards. Refer to washer/disinfector Operator Manual
(P920518-029) for the current list of standards.
iv
920515-083 Operator Manual Introduction
Manufactured by:
STERIS Canada ULC
490 boulevard Armand-Paris
Québe c ( Q u ébec)
Canada G1C 8A3
The base language of this
document is ENGLISH. Any
translation must be made from
the base language document.
European Community Representative:
STERIS Ireland Limited
IDA Business and Technology Park
Tullamore
County Offaly
R35 X865
Ireland
Sales and Service -- Americas, Oceania and Asia:
STERIS Corporation
5960 Heisley Road
Mentor, Ohio 44060
440-354-2600 • 800-548-4873
Sales and Service -- Europe, Middle-East and Africa:
STERIS
116 Avenue de Magudas, 33185
Le Haillan
FRANCE
Support Service, by email:
Support_EMEA_IPT_SE@steris.com
All other requests, by phone: +33 5 56 93 94 94
Website:
www.steris.com
v
Table of Contents Operator Manual 920515-083
A WORD FROM STERIS............................................................................................i
Advisory............................................................................................................................................................ i
Indications For Use...........................................................................................................................................ii
Service Information...........................................................................................................................................ii
Technical Specifications...................................................................................................................................ii
Certification......................................................................................................................................................iii
1 SAFETY PRECAUTIONS ......................................................................................1-1
2 LOADING INSTRUCTIONS ...................................................................................2-1
2.1 AN Vision SC Manifold Rack – Components ......................................................................................2-1
2.2 Preparing Instruments ........................................................................................................................ 2-2
2.3 Loading Rack ..................................................................................................................................... 2-2
2.3.1 General ........................................................................................................................................ 2-2
2.3.2 DIN Tray....................................................................................................................................... 2-3
2.3.3 Helicline, Hollow Jets and Hooks ................................................................................................ 2-3
3 OPERATING INSTRUCTIONS ..............................................................................3-1
4 UNLOADING INSTRUCTIONS ..............................................................................4-1
4.1 Unloading AN Vision SC Manifold Rack............................................................................................. 4-1
4.2 Visual Inspection ................................................................................................................................ 4-2
5 ROUTINE MAINTENANCE AND REPLACEMENT PARTS .....................................5-1
5.1 Routine Maintenance.......................................................................................................................... 5-1
5.2 Replacement Parts ............................................................................................................................. 5-2
TABLE OF CONTENTS
Section
Number Description Page
vi
920515-083 Operator Manual List of Illustrations
Section 2: Loading Instructions
2-1 AN Vision SC Manifold Rack – Components ......................................................................................2-1
2-2 Example of Anesthesia Items and Respiratory Goods....................................................................... 2-2
2-3 Example of DIN Tray Loaded in Rack ................................................................................................ 2-3
2-4 Example of Anesthesia Items and/or Respiratory Goods on Large Hollow Jets ................................ 2-3
2-5 Example of Anesthesia Items and/or Respiratory Goods on Medium Hollow Jets ............................ 2-4
2-6 Example of Anesthesia Items and/or Respiratory Goods on Small Hollow Jets ................................ 2-4
2-7 Example of Anesthesia Items and/or Respiratory Goods on Hooks .................................................. 2-5
2-8 Adapters in Bottom of Rack................................................................................................................ 2-5
2-9 Example of Anesthesia Tubing With Adapters on Helicline ............................................................... 2-5
2-10 Example of Loaded AN Vision SC Manifold Rack ............................................................................ 2-6
LIST OF TABLES
Table
Number Description Page
5 ROUTINE MAINTENANCE AND REPLACEMENT PARTS .....................................5-1
5.1 Replacement Parts ................................................................................................................................ 5-2
LIST OF ILLUSTRATIONS
Figure
Number Description Page
1-1
Safety Precautions Operator Manual 920515-083
1
The following Safety Precautions must be observed when operating or servicing the AN Vision SC Manifold Rack.
WARNING indicates the potential for personal injury and CAUTION indicates the potential for damage to
equipment. For emphasis, certain Safety Precautions are repeated throughout the manual. It is important to
review ALL Safety Precautions before operating or servicing the glassware washer.
Strictly following these Safety Precautions enhances the Customer’s ability to safely and effectively utilize the unit
and helps avoid improper maintenance methods which may damage the unit or render it unsafe. It is important to
understand that these Safety Precautions are not exhaustive; Customers are encouraged to develop their own
safety policies and procedures to enhance and complement these Safety Precautions.
WARNING – PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD:
WARNING – BURN HAZARD:
CAUTION – POSSIBLE EQUIPMENT DAMAGE:
Always load or unload AN Vision SC Manifold Rack on appropriate loading cart or surface.
Regularly scheduled preventive maintenance is required for safe and reliable operation of this
equipment. Contact STERIS to schedule preventive maintenance.
Repairs and adjustments to this equipment must be made only by STERIS or STERIS-trained service
personnel. Repairs and adjustments performed by unqualified personnel or installation of unauthorized
parts could cause personal injury, result in improper equipment performance, void the warranty or result
in costly damage. Contact STERIS regarding service options.
AN Vision SC Manifold Rack and items are extremely hot when cycle is completed. Allow rack and
items to cool to room temperature and always wear appropriate Personal Protective Equipment (PPE)
before reaching into chamber.
When cycle is completed, wait until water flow stops before opening door and unloading washer/
disinfector. Hot water or steam may be sprayed through door opening if door is opened too soon. Wear
appropriate Personal Protective Equipment (PPE) whenever reaching into chamber.
To avoid equipment damage and maximize processing efficiency, always load soiled items as instructed
in this AN Vision SC Manifold Rack Operator Manual.
To prevent voiding the warranty or damaging the equipment, use only authorized STERIS replacement
parts.
Use nonabrasive cleaners when cleaning AN Vision SC Manifold Rack. Follow directions on
containers and rub in a back-and-forth motion (in same direction as surface grain). Abrasive cleaners
will damage stainless steel.
SAFETY PRECAUTIONS
1-2
920515-083 Operator Manual Safety Precautions
Healthcare
Capital Equipment
2
2-1
Loading Instructions Operator Manual 920515-083
2.1 AN Vision SC
Manifold Rack –
Components
See Figure 2-1.
Before using the AN Vision SC manifold rack, it is important to
become familiar with the location of the rack components.
IMPORTANT: A listing of the Safety Precautions to be observed when using and servicing this equipment
can be found in SECTION 1 of this manual. Do not use the equipment until you have become familiar with
this information.
Figure 2-1. AN Vision SC Manifold Rack – Components
Large
Hollow Jets
(x 12)
Small
Hollow Jets
(x 12)
Medium
Hollow Jets
(x 12)
Adapters
(x 8)
Spring
Stoppers
(x 12)
Helicline
Hooks
(x 8)
Additional
Adapters
(x 8)
LOADING INSTRUCTIONS
2-2
920515-083 Operator Manual Loading Instructions
2.2 Preparing
Instruments
See Figure 2-2.
Disassemble all anesthesia tubing, bags, masks and respiratory
goods prior to loading them in the AN Vision SC Manifold Rack.
2.3 Loading Rack
2.3.1 General Anesthesia items and respiratory goods are loaded on the AN Vision
SC Manifold Rack components.
A DIN tray can also be used to load smaller anesthesia items and
respiratory goods.
Figure 2-2. Example of Anesthesia Items and
Respiratory Goods
WARNING – PERSONAL INJURY
AND/OR EQUIPMENT DAMAGE
HAZARD: Always load or
unload AN Vision SC Manifold
Rack on appropriate loading
cart or surface.
CAUTION – POSSIBLE
EQUIPMENT DAMAGE: To
avoid equipment damage and
maximize processing efficiency,
always load soiled items as
instructed in this AN Vision SC
Manifold Rack Operator Manual.
2-3
Loading Instructions Operator Manual 920515-083
2.3.2 DIN Tray See Figure 2-3.
To load the AN Vision SC Manifold Rack using a DIN tray, proceed
as follows:
1. Pull out DIN tray from rack.
2. Place small anesthesia items and respiratory goods in DIN tray.
IMPORTANT: Do not place any tubing in DIN Tray.
3. Place DIN tray back in position in rack.
2.3.3 Helicline, Hollow Jets
and Hooks
See Figure 2-1 and Figures 2-4 to 2-10.
To load tubing as well as anesthesia items and respiratory goods
using the helicline, the hollow jets and the hooks, proceed as follows:
1. Place desired items on large hollow jets (see Figure 2-4).
2. Spring stoppers can be used to adjust height of items to be
processed (see Figure 2-1).
NOTE: Ensure the hollow jets are not shorter than the processed
items to provide maximum coverage during the Anesthesia/
Respiratory Goods cycle.
Figure 2-3. Example of DIN Tray Loaded in Rack
Small Tube
Connected
to Hollow
Jet
Figure 2-4. Example of Anesthesia Items and/
or Respiratory Goods on Large Hollow Jets
2-4
920515-083 Operator Manual Loading Instructions
NOTE: Some items are equipped with small tubes. These small
tubes can be connected to hollow jets to ensure complete
coverage of tubes during the cycle. Ensure other extremity of
each tube is free for evacuation (see Figure 2-4).
3. Place desired items on medium hollow jets (see Figure 2-5).
NOTE: Ensure the hollow jets are not shorter than the processed
items to provide maximum coverage during the Anesthesia/
Respiratory Goods cycle.
4. Place desired items on small hollow jets (see Figure 2-6).
NOTE: Ensure the hollow jets are not shorter than the processed
items to provide maximum coverage during the Anesthesia/
Respiratory Goods cycle.
Figure 2-5. Example of Anesthesia Items and/
or Respiratory Goods on Medium Hollow Jets
Figure 2-6. Example of Anesthesia Items and/
or Respiratory Goods on Small Hollow Jets
2-5
Loading Instructions Operator Manual 920515-083
5. Place desired items on hooks (see Figure 2-7).
6. To place desired anesthesia tubing on helicline, proceed as
follows:
a. To obtain maximum coverage during cycle, connect one
extremity of anesthesia tubing on adapter located in bottom
of rack (see Figure 2-8).
b. Place tubing on helicline.
c. To obtain maximum coverage during cycle, insert one of
additional adapters in other extremity of anesthesia tubing
(see Figures 2-1 and 2-9).
Figure 2-7. Example of Anesthesia Items
and/or Respiratory Goods on Hooks
Figure 2-8. Adapters in Bottom of Rack
Figure 2-9. Example of Anesthesia Tubing
with Adapters on Helicline
2-6
920515-083 Operator Manual Loading Instructions
7. AN Vision SC manifold rack is loaded and anesthesia items and
respiratory goods are now ready to be processed (see
Figure 2-10).
Refer to Reliance®/Hamo® Vision® Single-Chamber Washer/
Disinfector Operator Manual (P920508-556) or to the Vision®
Single-Chamber Washer/Disinfector Operator Manual (P920518-
029) for information about processing Anesthesia/Respiratory
Goods cycle.
3
3-1
Operating Instructions Operator Manual 920515-083
OPERATING INSTRUCTIONS
For information about the
Anesthesia/Respiratory
Goods Cycle, refer to the
Reliance®/Hamo® Vision®
Single-Chamber Washer/
Disinfector Operator Manual
(P920508-556)
or to the
Vision® Single-Chamber
Washer/Disinfector
Operator Manual
(P920518-029)
3-2
920515-083 Operator Manual Operating Instructions
Healthcare
Capital Equipment
4
4-1
Unloading Instructions Operator Manual 920515-083
4.1 Unloading AN
Vision SC Manifold
Rack
IMPORTANT: When Anesthesia/Respiratory Goods cycle is
completed, allow load to cool before removing AN Vision SC Manifold
Rack from washer/disinfector chamber.
To unload AN Vision SC Manifold Rack, proceed as follows:
1. Remove AN Vision SC Manifold Rack from washer/disinfector
chamber and place rack on loading cart.
2. If desired, step back to pull loading cart away from washer/
disinfector.
3. Items need to be inspected before removing items from AN
Vision SC Manifold Rack (see SECTION 4.2, VISUAL INSPECTION,
for more details).
IMPORTANT: STERIS recommends unloading AN Vision SC
Manifold Rack on an appropriate surface or work table.
4. When inspection is completed and all anesthesia items and
respiratory goods are properly cleaned, carefully empty rack.
5. Slide out DIN tray from bottom of rack.
6. Carefully empty tray.
IMPORTANT: A listing of the Safety Precautions to be observed when using and servicing this equipment
can be found in SECTION 1 of this manual. Do not use the equipment until you have become familiar with
this information.
WARNING – PERSONAL INJURY
AND/OR EQUIPMENT DAMAGE
HAZARD: Always load or
unload AN Vision SC Manifold
Rack on appropriate loading
cart or surface.
WARNING – BURN HAZARD:
AN Vision SC Manifold Rack
and items are extremely hot
when cycle is completed.
Allow rack and items to cool
to room temperature and
always wear appropriate
Personal Protective
Equipment (PPE) before
reaching into chamber.
When cycle is completed,
wait until water flow stops
before opening door and
unloading washer/
disinfector. Hot water or
steam may be sprayed
through door opening if
door is opened too soon.
Wear appropriate Personal
Protective Equipment (PPE)
whenever reaching into
chamber.
UNLOADING INSTRUCTIONS
4-2
920515-083 Operator Manual Unloading Instructions
4.2 Visual Inspection IMPORTANT: When Anesthesia/Respiratory Goods cycle is
completed, allow load to cool before removing AN Vision SC manifold
rack from washer/disinfector chamber.
IMPORTANT: Good Hospital Practices dictate all instruments be
inspected for visible debris after processing in washer. Any
instrument with visible debris must be reprocessed until clean and
free of visible debris prior to terminal processing. Failure to reprocess
until all visible debris have been removed may impede terminal
processing.
1. Remove AN Vision SC manifold rack from washer/disinfector
chamber and place rack on loading cart.
2. Once items are cool, check each item to detect residues or
debris.
3. Any item that has not been properly cleaned must be
reprocessed in washer/disinfector until clean.
4. See SECTION 4.1, UNLOADING AN VISION SC MANIFOLD RACK, to
properly unload AN Vision SC manifold rack.
IMPORTANT: STERIS recommends unloading AN Vision SC
manifold rack on an appropriate surface or work table.
WARNING – PERSONAL INJURY
AND/OR EQUIPMENT DAMAGE
HAZARD: Always load or
unload AN Vision SC Manifold
Rack on appropriate loading
cart or surface.
WARNING – BURN HAZARD:
AN Vision SC Manifold Rack
and items are extremely hot
when cycle is completed.
Allow rack and items to cool
to room temperature and
always wear appropriate
Personal Protective
Equipment (PPE) before
reaching into chamber.
When cycle is completed,
wait until water flow stops
before opening door and
unloading washer/
disinfector. Hot water or
steam may be sprayed
through door opening if
door is opened too soon.
Wear appropriate Personal
Protective Equipment (PPE)
whenever reaching into
chamber.
5
5-1
Routine Maintenance and Replacement Parts Operator Manual 920515-083
5.1 Routine
Maintenance
NOTE: AN Vision SC manifold rack must be inspected regularly to
ensure no sediment has accumulated in the rack components.
1. Inspect for damage all connections, caps, adapters and hollow
jets.
2. If necessary, clean any clogged rack component using a fine
wire (approximately wire gauge of a paper clip).
3. Inspect caps. Replace if torn, discolored or damaged.
IMPORTANT: A listing of the Safety Precautions to be observed when using and servicing this equipment
can be found in SECTION 1 of this manual. Do not use the equipment until you have become familiar with
this information.
WARNING – PERSONAL INJURY
AND/OR EQUIPMENT DAMAGE
HAZARD:
Regularly scheduled preventive
maintenance is required for
safe and reliable operation of
this equipment. Contact
STERIS to schedule preventive
maintenance.
Repairs and adjustments to
this equipment must be made
only by STERIS or STERIS-
trained service personnel.
Repairs and adjustments
performed by unqualified
personnel or installation of
unauthorized parts could
cause personal injury, result
in improper equipment
performance, void the
warranty or result in costly
damage. Contact STERIS
regarding service options.
CAUTION – POSSIBLE
EQUIPMENT DAMAGE: Use
nonabrasive cleaners when
cleaning AN Vision SC
Manifold Rack. Follow
directions on containers and
rub in a back-and forth motion
(in same direction as surface
grain). Abrasive cleaners will
damage stainless steel.
ROUTINE MAINTENANCE AND
REPLACEMENT PARTS
5-2
920515-083 Operator Manual Routine Maintenance and Replacement Parts
5.2 Replacement Parts To order replacement parts, proceed as follows:
1. Include description and part number as listed in Table 5-1 on
your order.
2. Specify model of your rack on your order (AN Vision SC manifold
rack).
3. Send your order directly to STERIS.
NOTE: Use only STERIS authorized parts on this equipment.
Use of unauthorized parts will void the warranty.
CAUTION – POSSIBLE EQUIPMENT
DAMAGE: To prevent voiding
the warranty or damaging the
equipment, use only authorized
STERIS replacement parts.
Table 5-1. Replacement Parts
Description Part Number
HOLLOW JET, M6 x 3 x 50 mm Long 117076-991
HOLLOW JET, M6 x 4 x 102 mm Long 117076-989
HOLLOW JET, M6 x 4 x 162 mm Long 117076-986
HOLLOW JET, M8, For Balloon 117076-985
HOLLOW JET, M8, For Bellow 117076-982
SUPPORT, Star, Polypropylene 117077-647
SUPPORT, Glassware, Polypropylene 117077-646
PLUG, Respiratory Therapy Tube 117076-994
ADAPTOR, Respiratory Therapy Tube 117904-441
PROTECTOR, M-2/M-8, Glassware 117045-265
STOPPER, AN Vision SC manifold rack 117049-215
STOPPER, Spring, Spindle, 5/16 117910-406
DIN TRAY, AN Vision SC Manifold Rack 117077-693
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20

Steris Anesthesia / Respiratory Rack Operating instructions

Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI