Mettler Toledo InTap Portable oDO Analyzer User manual

Type
User manual

This manual is also suitable for

InTap Portable oDO Analyzer
30 461 735 A
cs Návodkpoužití 3
da Brugervejledning 17
de Benutzerhandbuch 31
en User Manual 45
es Manualdeusuario 59
 Käyttöohje 73
fr Guidedel‘utilisateur 87
hu Felhasználóiútmutató 101
it Manualeperl‘utente 115
ja ユーザマニ 129
ko 사용자 매뉴얼 143
nl Handleiding 157
pl Podręcznikużytkownika 171
pt Manualdousuário 185
ru Руководство пользователя 199
sv Användarmanual 213
th คู่มือผู้ใช ้ 227
zh 简明用户手册 245
User Manual
InTap Portable oDO Analyzer
2 InTap Portable oDO Analyzer • 30 461 735 A
©Mettler-ToledoGmbH06/2018
Subjecttotechnicalchanges.PrintedinSwitzerland.30461735A
10 InTap Portable oDO Analyzer • 30 461 735 A
5 Struktura menu
cs
24 InTap Portable oDO Analyzer • 30 461 735 A
5 Menustruktur
da
38 InTap Portable oDO Analyzer • 30 461 735 A
5 Menüstruktur
de
45InTap Portable oDO Analyzer • 30 461 735 A
User Manual
InTap Portable oDO Analyzer
Content
1 Safety and Intended Use 46
2 TechnicalDataandFurtherInformation 47
3 DesignandConnections 48
4 DisplayandOperatingElements 50
5 Menu Structure 52
6 Commissioning 53
7 Calibration 54
8 MaintenanceandRepair 56
9 Disposal 57
10 Certicatesandapprovals 58
en
InProisaregisteredtrademarkoftheMETTLERTOLEDOGroup
inSwitzerlandandafurthertwelvecountries.
ISMisaregisteredtrademarkoftheMETTLERTOLEDOGroup
inSwitzerland,USA,theEuropeanUnionandafurtherve
countries.
InTap,iSenseandOptoCaparetrademarksoftheMETTLER
TOLEDOGroup.
46 InTap Portable oDO Analyzer • 30 461 735 A
1 Safety and Intended Use
TheInTap™portableoDOanalyzermustonlybeconnected,commissioned,andmaintained
byqualifiedspecialistsinfullcompliancewiththeinstructionsinthisUserManual,theappli-
cablenormsandlegalregulations.
ThespecialistmusthavereadandunderstoodthisUserManualandmustfollowtheinstruc-
tionsitcontains.
TheReferenceManualprovidesfurtherinformationonthedeviceandcanbedownloadedvia
Internet“www.mt.com/InTap“.
TheInTapportableoDOanalyzershouldbeoperatedonlybypersonnelfamiliarwiththedevice
andwhoarequalifiedforsuchwork.
Intended Use
The InTap portable oDO analyzer is a portable measurement device for dissolved oxygen
measurementandforcalibrationanddataloggingofopticaldissolvedoxygenISM
®
sensors
(InPro
®
6970i)installedintheprocess.TheInTapandthesensorscommunicateviaBlue-
tooth.Forthis,theseparateBluetoothinterfaceT100(ordernumber30432819)hastobe
mountedontoeachsensor.
TheInTapisdesignedforuseinthebeverageindustries,innon-hazardousareas.TheInTap
isintendedtobeusedindoor.
ThefollowingInTapvariantsareavailable:
InTapportableoDOanalyzer(metricconnectors):ordernumber30425550
InTapportableoDOanalyzer(imperialinch/USconnectors):ordernumber30457912
en
47InTap Portable oDO Analyzer • 30 461 735 A
2 Technical Data and Further Information
Themostimportanttechnicaldatasuchassupplyvoltagearegivenonthenameplate.The
nameplateisplacedonthebottomofthedevice.Payparticularattentiontothetechnicaldata
onthenameplate.
Measurementparameter OpticalDissolvedOxygenSaturation
Measuringrange 0to2000ppb
Accuracy
1)
(sensor) ≤±(1%+2ppb)
Responsetimet
98
at+25°C(+77°F)
(airtoN
2
)
<20s
Powersupply
1)
External:12VDC,2.5A
Internal:Li-ionbattery45.4Wh
Batterylifetime Up to 24 hours
DatastorageUSB2.0 3GB
Operatingtemperature –5to+45°C(+23to+113°F)withmedia
Upto+80°C(+176°F)forcleaningwith2%NaOH
Withpoweradapteronlyupto+40°C(+104°F)
Storagetemperature –20to50°C(–4to+122°F)
Operatingpressure 0to6bar(0to87psi)
Designpressure 10bar(145psi)
Protectionrating IP 67
Hoseconnection Variantwithmetricconnectors:Swagelock6mm
Variantwithimperialinch/USconnectors:1/4"NPT
Weight 3.5kg(7.7lbs)
Dimensions(height3width3depth)
280mm3170mm3150mm(11.0"36.7"35.9")
Altitude Max.2000m
Relativehumidity 0to95%non-condensing
1) Measurementspecificationbasedonsensorspecification
en
48 InTap Portable oDO Analyzer • 30 461 735 A
3 Design and Connections
1
4
5
6
10
2
7 89
3
1 Monochrome4"touchdisplay
2 On/Offbutton
3 Housingwithmeasurementcell,O
2
InTapsensor,valve,battery,HWboard
4 Connectionforhose“Mediumin”
5 Connectionforhose“Mediumout”
6 Rotaryknobforadjustingmediumflowrate
7 USBstickfordatalogging
8 USBtypeBconnectionforsoftwareupdates
9 USBtypeAconnection
10 ConnectionforAC/DCadapterforloadingbattery
en
49InTap Portable oDO Analyzer • 30 461 735 A
Scope of delivery
Powersupply30W:AC/DCadapterwithACadaptersetEU,US,AU,UK
USBstick,plugged-intheUSBconnection
Supply voltage (Power)
Youhavethefollowingpossibilitiestopowerthedevice:
eitherviatheinternalbattery
orviathesuppliedAC/DCadapterconnectedto100to240VAC/50to60Hz.
Hoses “Medium In” and “Medium Out”
Youcanconnectthefollowinghosesto“MediumIn”and“MediumOut”:
InTap portable oDO analyzer Hoses
Description Order number Description Order number
Withmetricconnectors 30425550 6mm 30422575
Withimperialinch/
US connectors
30457912 1/4"NPT 30432818
Adjusting medium flow rate
Theflowrateisadjustedwiththerotaryknob.Theflowratedependsonthemediumpressure
Theflowrateincreaseswithincreasingmediumpressure.
1. Turnrotaryknobclockwisetotheendstop.
2. Turnrotaryknobanti-clockwisetoadjusttheflowrate.Therotaryknobhasaclick-stop
positionevery15°.E.g.foraflowrateofapproximately200ml/minat2baryouhave
toturntherotaryknobtothe30°position.Thatmeansyouhear2clicks.
Forthediagram“Adjustingtheflowrate”refertotheReferenceManual.
en
50 InTap Portable oDO Analyzer • 30 461 735 A
4 Display and Operating Elements
\ CAL
Calibrate InTap Sensor
Calibrate In-Line Sensor
InTap
200
A
ppbO2
25.0
°C
DLI
ACT
0 d
0.0 d
\ CAL \ Calibrate InTap Sensor
Chan
Unit
Methods
Options
InTap
% air
2-Point
Options
Verify Cal
InTap
200
A
ppbO2
25.0
°C
DLI
ACT
0 d
0.0 d
1 2
34
6 7
55
1 Homescreen
2 Measurementscreen
3 Menuscreen,heree.g.“Calibrationmenu”
4 Parametrizationscreen,heree.g.“CalibrationInTap”
5 NavigationtoHomescreen
6 Navigationtonext-highermenulevel,hereMeasurementscreen
7 Navigationtonext-highermenulevel,hereCalibrationmenuscreen
en
51InTap Portable oDO Analyzer • 30 461 735 A
Display or operating
element
Description Display or operating
element
Description
USBstickisconnected. EnterTrenddisplayscreen.
Displayofthecurrentbattery
status
EnterISMmenu.
TheInTapispoweredviathe
AC/DCadapter.
c
EnterCalibrationmenu.
Failure:Nobatteryorbattery
isempty.
C
EnterConfigurationmenu.
Failure:Batteryislow
(<5%).TheInTapisnot
poweredviatheAC/DC
adapter.
EnterMeasuringPoint(MP)
menu.
Failure:USBstickisnot
connectedordamaged.
H
ReturntoMenuscreen.
Failure:Leakage,high
temperature,batteryfaulty
etc.RefertotheReference
Manual.
Enternext-lowermenulevel.
EnterMessagesscreen.
p
Returntonext-highermenu
level.
F
Entermenuscreen.
< >
Changebetweenpages
withinonemenulevel.
s
Enterstartscreen.
en
52 InTap Portable oDO Analyzer • 30 461 735 A
5 Menu Structure
en
53InTap Portable oDO Analyzer • 30 461 735 A
6 Commissioning
TheInTapissuppliedreadyforuse.Afterswitchingon,youcanimmediatelyusetheInTapas
ameasuringdeviceorasameasuringdevicewithdataloggingfunction.
Atthefactorythedataloggingissettoautomaticmode.Usetheparameter“LoggingMode”to
changetomanualmode(CAL\CONFIG\MeasurementMode).
Notethatwithfactorysettings,thedisplayisswitchedoffaftersomeminutes,butthemeasure-
mentwillcontinue.Usethemenu“DisplaySetup”tochangethesettings(CAL\CONFIG\Dis-
playSetup).
ForcommunicationwithanIn-Linesensor(processcalibration)aseparateBluetoothinterface
T100(ordernumber30432819)hastobemountedontoeachsensorandthesensorhas
tobepairedwiththeInTap.
Pairing InTap with In-Line Sensors
Prerequisite
TheIn-LinesensoriswithintheBluetoothrangeoftheInTap.
MP – Select
New MP
New BT device recognized 00
MP_01
MP_02
MP_03
Edit
Edit
Edit
MP_01 is selected.
1
2
3
4
1 A new Bluetooth device is within the Bluetooth
rangeandwasrecognized.Thesensorneedstobe
parametrized.
2 TheIn-Linesensor“MP_01”isselected.The
sensoriswithintheBluetoothrangeandalready
parameterized.
3 TheIn-Linesensor“MP_02”isnotselected.The
sensoriswithintheBluetoothrangeandalready
parameterized.
4 TheIn-Linesensor“MP_03”isnotselected.The
sensorisalreadyparameterizedbutnotwithinthe
Bluetoothrange.
en
54 InTap Portable oDO Analyzer • 30 461 735 A
1. Enter“MeasuringPoint(MP)”menu.SeeChapter4onPage50.
Thescreen“MPSelect”withtheitem“NewBTdevicerecognized00”isshown.
2. Tap“+”fortheitem“NewBTdevicerecognized00”.
3. Configurethenewmeasuringpoint(MP).
4. Tap“OK”
Thescreen“MPSelect”isshown.
5. Activatetheradiobutton.
6. Returntothenext-highermenulevel.
7. Answerthequestion“Savechanges?”with“Yes”.
TheMeasurementScreenfortheselectedmeasuringpointisshown.Aftersomeseconds
themeasurementresultsareshown.Tap“Save”tosavetheresultsontheUSBstick.Tap
“Repeat”togetnewvaluesfromthesensor.Tap“Cancel”tocanceltheprocedure.
7 Calibration
Youcanalso calibratethedismounted InTapsensororthe dismountedIn-LineSensorvia
iSense™software.SeeOperatingInstructionsoftheiSensesoftware.
Assoonasthecalibrationisinprogressnoothercalibrationcanbestarted.
Calibration Menu c
Aftereverysuccessfulcalibrationdifferentoptionsareavailable.If”Adjust”or”Calibrate”is
chosen,themessage”Calibrationsavedsuccessfully!”isdisplayed.Press”Done”.
Option ISM sensors
Adjust Calibrationvaluesarestoredinthesensorandusedforthemeasurement.
Additionally,thecalibrationvaluesarestoredinthecalibrationhistory.
Calibrate Calibrationvaluesarestoredinthecalibrationhistoryfordocumentation,butnot
beusedforthemeasurement.Thecalibrationvaluesfromthelastvalid
adjustmentarefurtherusedforthemeasurement.
Cancel Calibrationvaluesarediscarded.
en
55InTap Portable oDO Analyzer • 30 461 735 A
Calibrating an In-Line Sensor
Prerequisite
TheIn-LinesensorispairedwiththeInTapandiswithintheBluetoothrangeoftheInTap.
1. Enter“CalibrateIn-LineSensor”menu.Menupath:CAL>CalibrateIn-LineSensor
Thescreen“MPSelect”isshown.
2. SelecttheMP(In-Linesensor)youwanttocalibrate.Thereforeactivatethecorrespond
-
ingradiobutton.
3. Returntothenext-highermenulevel.
Thescreen“CalibrateIn-LineSensor”isshown.
4. Settheoptions“CalPressure”,“Rel.Humidity”and“Stability”.
5. Tap“Cal”.
6. Confirmthemessage“Thereisa“autologging”procedurerunning”with“Yes”.
7. Answerthequestion“FirstcalibrationafterOptoCap™change”.
ThecurrentvaluesfortheInTapsensorandtheIn-Linesensorareshown.
8. Tap“Next”.
ThevaluesforP100andP0areshown.
9. Tap“Adjust”or“Calibrate”.
10.Tap“Done”tofinishcalibrationoftheIn-Linesensor.
Calibrating the InTap Sensor
h
Note:Useasairgasdry,cleanandoil-freecompressedair.UseaszerogasN
2
or CO
2
witha
purityofatleast99.9%.Forbestresultsuseazerogaswithapurityof99.995%.
1. Connectthecompressedairtotheconnection“MediumIn”.
2. PurgetheInTapfor5minuteswithcompressedairtoremovetheexcesshumidityoutof
themeasurementcell.Ensureasteadymediumflowwithoutpressurebuild-upinthe
measurementcell.Forthis,performtwoturnswiththerotaryknob.SeeChapter3on
Page48.
3. Turnoffcompressedair.
4. Enter“CalibrateInTapSensor”menu.Menupath:CAL>CalibrateInTapSensor
5. Settheunits.
en
56 InTap Portable oDO Analyzer • 30 461 735 A
6. Settheoptions“CalPressure”,“Rel.Humidity”and“Stability”.
7. Tap“Cal”.
8. Confirmthemessage“Thereisa“autologging”procedurerunning”with“Yes”.
9. Answerthequestion“FirstcalibrationafterOptoCapchange”correctly.
Themessage“ConnectthegastoInTapinlet.”isshown.
10.Turnonthecompressedair.
11.Ensureasteadymediumflowwithoutpressurebuild-upinthemeasurementcell.
12.Tap“Next”.
Thecurrentvalueisshown.
Ifthecriteriaaremet,themessage“Pleasechangegas.”isshown.
13.Connectthesecondreferencegas(zerogas)totheconnection“MediumIn”.
14.Tap“Next”.
Thecurrentvalueisshown.
Ifthestabilitycriteriaaremet,thevaluesforP100andP0areshown.
15.Tap“Adjust”or“Calibrate”.
16.Tap“Done”tofinishcalibrationoftheInTapsensor.
8 Maintenance and Repair
Cleanthesurfaceswithasoftdampclothanddrythesurfaceswithaclothcarefully.
Replacing the OptoCap of the InTap Sensor (Sensors Sensing Element)
Typically,theOptoCaphastobereplacedafter6to12months.Thisintervaldependsonthe
aggressiveness(CIP/SIP)oftheapplicationconditions.
Fordismantlingthesensorrefertothe“ExchangingtheInTapSensor“section.Forreplacingthe
OptoCaprefertotheUserManual“InPro6000OpticalO
2
Sensors“.
Exchanging the InTap Sensor
1. Ifnecessary,exchangethesensorasshowninthefollowingfigure.
2. Tightencoverscrewswithamaximumtorqueof5Nm.
3. Calibratesensor.SeeChapter7onPage54.
en
57InTap Portable oDO Analyzer • 30 461 735 A
Replacing the Battery
ThebatterymayonlybereplacedbyMETTLERTOLEDO.Looseningthescrewsoftheinner
coverwillleadtotheexpirationofanywarrantyclaims.
9 Disposal
Environmental Protection
Wasteelectricalproductsshouldnotbedisposedofwithhouseholdwaste.Pleaserecycle
wherefacilitiesexist.CheckwithyourLocalAuthorityorretailerforrecyclingadvice.
en
58 InTap Portable oDO Analyzer • 30 461 735 A
10 Certicates and approvals
Europeandirectives TheEUDeclarationofconformityispartofthedelivery.
Safety standards CAN/CSA–C22.2No.61010-1-12
ANSI/ULStd.No.61010-1(3rdEdition)
Overvoltage Category II
FCC(USA) CFR47FCCPart15
ANSIC63.4-2014
IC(Canada) ICES-003,Issue6
ANSIC63.4-2014
en
66 InTap Portable oDO Analyzer • 30 461 735 A
5 Estructura de menús
es
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227
  • Page 228 228
  • Page 229 229
  • Page 230 230
  • Page 231 231
  • Page 232 232
  • Page 233 233
  • Page 234 234
  • Page 235 235
  • Page 236 236
  • Page 237 237
  • Page 238 238
  • Page 239 239
  • Page 240 240
  • Page 241 241
  • Page 242 242
  • Page 243 243
  • Page 244 244
  • Page 245 245
  • Page 246 246
  • Page 247 247
  • Page 248 248
  • Page 249 249
  • Page 250 250
  • Page 251 251
  • Page 252 252
  • Page 253 253
  • Page 254 254
  • Page 255 255
  • Page 256 256
  • Page 257 257
  • Page 258 258
  • Page 259 259
  • Page 260 260

Mettler Toledo InTap Portable oDO Analyzer User manual

Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI