Midmark PowerVac Installation guide

Type
Installation guide

This manual is also suitable for

Midmark PowerVac G is a powerful dental vacuum system designed to meet the demands of modern dental practices. It features a compact design, quiet operation, and a variety of customizable options to suit your specific needs. With its strong suction and reliable performance, the PowerVac G ensures efficient removal of fluids and debris during dental procedures. It is suitable for use with various dental equipment, including high-volume evacuators, saliva ejectors, and handpiece coolant lines.

Midmark PowerVac G is a powerful dental vacuum system designed to meet the demands of modern dental practices. It features a compact design, quiet operation, and a variety of customizable options to suit your specific needs. With its strong suction and reliable performance, the PowerVac G ensures efficient removal of fluids and debris during dental procedures. It is suitable for use with various dental equipment, including high-volume evacuators, saliva ejectors, and handpiece coolant lines.

1
© 2015 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
004-1130-00 Rev. A (11/24/16)
Applies to Models:
Power Vac “G”
Power Vac (Std)
Classic Series Wet Ring
Power Max
Special Tools:
none
Language of origin: English
Intended Use
To provide general dental vacuum plumbing layout recommendations per industry standards. Installation should be conducted by
qualied dental/plumbing professionals.
Limitations of Recommendations
This Midmark document is designed to provide general vacuum plumbing recommendations commonly used in the dental industry. Midmark does not have
expertise in building construction or detailed information about a specic buildings construction. Therefore, Midmark is unable to provide recommendations for
specic pipe and tting locations.
Vacuum Plumbing Layout Recommendations
Equipment Alert
Vacuum systems must be installed per local plumbing and electrical codes.
midmark.com
2
© 2015 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
004-1130-00 Rev. A (11/24/16)
Power Vac “G” Intake (Suction) Line
Single Models G3, G5, G7 Twin Models G6, G10, G14
Type PVC Sch 40 Pipe Recommended - for Main Trunk and Branch Lines
Line Size Refer to: "Site Requirement Layout", located in this document.
Pump Termination 1 1/2” 2”
Power Vac “std” Intake (Suction) Line
Single Models P3, P5, P7 Twin Models P6, P10, P14
Type PVC Sch 40 Pipe Recommended - for Main Trunk and Branch Lines
Line Size Refer to: "Site Requirement Layout", located in this document.
Pump Termination 1 1/2” 2”
Wet-ring Intake (Suction) Line
Single Models CV3, CV5 Twin Models CV6, CV10
Type PVC Sch 40 Pipe Recommended - for Main Trunk and Branch Lines
Line Size 1“ Hose 1 1/2” hose
Pump Termination 3/4” NPT connection 1 1/4’” NPT connection
Power Max Intake (Suction) Line
Model PM2 Model PM3 Model PM4
Type Type K or L Copper Tubing - *Note: Do Not use compression ttings for vacuum lines.
Line Size 1” 1 1/4” 1 1/4”
Operatory Termination 5/8” OD copper
Vacuum Plumbing Specications
* Device/Conditions User Rating
High-volume Ecavuation (HVE) 1
Saliva Ejector 1/2
Hand Wash Station 1
Nitrous Oxide Scavenger 1/2
Main Vacuum line 50 ft. (15m) 1
Branch Vacuum Line 50ft. (15m) beyond
main line
1
Number of Users
(See Note)
Vacuum line pipe size
(inches)
1 3/4
2 3/4
3 1
4 1
5 1 1/4
6 1 1/4
7 1 1/2
8 1 1/2
9 1 1/2
10 2
11 2
12 2
* NOTE:
The number of devices and/or plumbing conditons
will add to the total number of users.
midmark.com
3
© 2015 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
004-1130-00 Rev. A (12/11/15)
Below oor or at oor level plumbing
General Recommendations
Vacuum Piping to be Type M Copper or Schedule 40 PVC per product specs and local code.
Main trunk lines must have a minimum slope of 1/4 inch per 10 feet towards the pump.
Use “drain,waste and vent”(DWV), or 45 deg. elbows.
Down size the main trunk line after every few junction boxes or branches in order to maintain sufcient suction in operatoires farthest from
the vacuum.
Every 50 feet of main trunk line should be considered a user.
Attempts should be made to limit branch lines, orient main trunk lines below junction boxes or cabinets.
Some applications may require the main trunk line to be split, attempts should be made to equalize ow in both directions.
Always stagger branch lines from main trunk, do NOT use 4-way cross ttings.
Large applications may require multiple vacuums and dividing the operatoires into zones in order to maintain sufcient suction.
Power Vac (“G” / “STD”) specic recommendations
6 feet maximum height from main trunk line to the pump.
When using in an oral surgery environment we recommend using the surgical operatory recovery room packages to collect everything and
prevent build up/clogs in the lines and vacuum tank.
Wet-ring specic recommendations
4 feet maximum height from main trunk line to the pump.
midmark.com
4
© 2015 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
004-1130-00 Rev. A (12/11/15)
AA2252
Equipment Alert
When performing positive pressure leak tests to validate plumbing installation, verify vacuum systems are not connected to office piping.
1 1/2”
Branch Line “Typical”
Equipment
Room
Main
Trunk Line
1 1/2”
1 1/4”
6’
12’
12’
3/4”
Reducer Bushing
3/4” PVC
Floor Level
45° Elbows
Main Line
To Junction Box / Inlet (Branch Plumbing)
To Vacuum Pump Inlet
Line Termination at Vacuum Pump
Centerline
Of Floor
Level Piping
Low Restriction Trap (6” Max.)
Riser Pipe - 24” Max. / 10” Min.
Main Line From Operatories
Below Floor - Sample Plumbing Layout
3/4”
1”
Optional
Clean Out
midmark.com
5
© 2015 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
004-1130-00 Rev. A (12/11/15)
1 1/2”
Branch Line
“Typical”
Equipment
Room
Main
Trunk Line
1 1/4”
Below Floor - Sample Plumbing Layout (Split Trunk Line)
3/4”
1”
2”
2”
2”
1 1/4”
1”
Additional Recommendations
In certain situations it is necessary to split the main
trunk line in different directions. When doing this take
into consideration the potential volume on each line
and size them accordingly. Use 45 deg. fittings flow-
ing back towards the vacuum.
Optional
Clean Out
Optional
Clean Out
midmark.com
6
© 2015 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
004-1130-00 Rev. A (12/11/15)
Below Floor - Large Application, Multiple Pump, Multiple Zones
1 1/2”
All Branch Lines
“Typical” 3/4”
Equipment
Room
1 1/4”
3/4”
Additional Recommendations
Large applications may require multiple vacuums and dividing the operatoires into zones in order to maintain sufcient suction. Attempts should be made
to equalize ow in both directions and size plumbing according to potential volume.
2”
Optional
Clean Out
2”
1 1/2”
1”
3/4”
3/4”
1”
1 1/2”
1 1/4”
1”
3/4”
1 1/4”
1 1/2”
1”
3/4”
1 1/2”
1 1/4”
1”
ZONE 1
ZONE 2
ZONE 3
Main
Trunk Line
Optional
Clean Out
Optional
Clean Out
Optional
Clean Out
3/4”
3/4”
3/4”
Main
Trunk Line
Color Page
midmark.com
7
© 2015 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
004-1130-00 Rev. A (12/11/15)
1 1/2”
Branch Line
“Typical”
Equipment
Room
Main
Trunk Line
At Floor Level - Sample Plumbing Layout (Split Trunk Line)
3/4”
1”
1 1/4”
1”
1”
3/4”
1 1/4”
Additional Recommendations
In certain situations it is necessary to run the main
trunk line at oor level along the wall. When doing this
follow the recommendation for plumbing below oor
level. Avoid running lines through doorways and other
trafc areas.
Optional
Clean Out
Optional
Clean Out
midmark.com
8
© 2015 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
004-1130-00 Rev. A (12/11/15)
Overhead Plumbing
General Recommendations
Vacuum Piping to be Type M Copper or Schedule 40 PVC per product specs and local code.
Main trunk lines must have a minimum slope of 1/4 inch per 10 feet towards the pump.
Use “drain,waste and vent”(DWV), or 45 deg. elbows.
Down size the main trunk line after every few junction boxes or branches in order to maintain sufcient suction.
Every 50 feet of main trunk line should be considered a user.
Attempts should be made to limit branch lines, orient main trunk lines in close proximity to the vertical vacuum lines
Some applications may require the main trunk line to be split, attempts should be make to equalize ow in both directions.
Micro Mini Mufer / Vacuum Relief Valve (VRV) to be Installed at top of vent line in each operatory.
Each Micro Mini Mufer / Vacuum Relief Valve VRV is considered 1/2 of a user.
Maximum height of riser to the main trunk line trap is 10 feet. (A loss of 1” hg per 18” of rise limits height of trunk line.)
It may be necessary to upgrade to the next size vacuum for systems with over head plumbing depending on the ofce size and usage.
All risers need to be plumbed into the top of the main trunk line through a trap.
We do not recommend combining overhead pluming with below oor plumbing due to the different performance characteristics.
When using in a oral surgery environment we recommend using the surgical operatory recovery room packages to collect everything and
prevent build up/clogs in the lines and vacuum tank.
Operation efciencies must be considered I.E.: holding vacuum tips open longer in operatory’s to allow ow up walls to help prevent gurgling
in plumbing
midmark.com
9
© 2015 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
004-1130-00 Rev. A (12/11/15)
Overhead plumbing - Sample Plumbing Layout
3/4” Inverted
P-Trap
Trunk Lines: Follow Sizing
Guide Under Floor Installation
and Add One User
77005027 Micro Mini Mufer / VRV
to be Installed at Top of Vent Line
in Each Operatory to Keep Air Flow
Pushing Liquids Up the Wet Line
After HVE or SE is Closed
Vacuum Piping to be Type M or Schedule 40 PVC Maintain a 1/4”
Slope per 10 feet of Piping (toward pump). Always Stagger Branch
Lines From Main Trunk, Do Not Use 4-way Cross Fittings.
Equipment Alert
Subfloor plumbing is preferred. Operatory
utilization to be modified to allow flow up walls.
Up size pump as per number of operatories. At least 1/2
user per VRV and 8 ft. riser. More than 8 ft. riser requires
more pump capacity.
Scavengers Should
be Installed on the
Vent Line to Prevent
Backow of Fluids
Into Scavenger
All Vertical Wet
lines are 1/2”
Optional
Clean Out
All Vertical Vent Lines are 1/2”
Installed with VRV Above Inverted
P-Trap. Allows Air Into System to
Minimize Gurgling in Vacuum Lines
Use 45
0
Y’s and
elbows only
1/2”
1/2”
1 1/2”
1 1/4”
1 ”
3/4”
VRV is 1/8” MPT a 1/8” FPT tting
will need to be soldered to the top
of the 1/2” vent line.
midmark.com
10
© 2015 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
004-1130-00 Rev. A (12/11/15)
Overhead plumbing - Sample Plumbing Layout (Split Trunk Line)
Equipment Alert
Subfloor plumbing is preferred. Operatory
utilization to be modified to allow flow up walls.
Up size pump as per number of operatories. At least 1/2
user per VRV and 8 ft. riser. More than 8 ft. riser requires
more pump capacity.
1 1/2”
1 ”
1 1/4”
1 ”
3/4”
Scavengers Should
be Installed on the
Vent Line to Prevent
Backow of Fluids
into Scavenger
3/4” Inverted
P-Trap
All Vertical Vent Lines are 1/2”
Installed with VRV Above Inverted
P-Trap. Allows Air Into System to
Minimize Gurgling in Vacuum Lines
All Vertical Wet
lines are 1/2”
Trunk Lines: Follow Sizing
Guide Under Floor Installation
and Add One User
Vacuum Piping to be Type M or Schedule 40 PVC Maintain a 1/4”
Slope per 10 feet of Piping (toward pump). Always Stagger Branch
Lines From Main Trunk, Do Not Use 4-way Cross Fittings.
Use 45
0
Y’s and
elbows only
1/2”
1/2”
1/2”
Optional
Clean Out
77005027 Micro Mini Mufer / VRV
to be Installed at Top of Vent Line
in Each Operatory to Keep Air Flow
Pushing Liquids Up the Wet Line
After HVE or SE is Closed
midmark.com
11
© 2015 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
004-1130-00 Rev. A (12/11/15)
Oral Surgery Plumbing Recommendations
General Recommendations (Power Max)
Vacuum Piping to be Type M Copper or Schedule 40 PVC per product specs and local code.
A Surgical Operatory Recovery Package is required for all Power Max applications.
All liquids and solids should be collected in the Surgical Canister and disposed of per local codes prior to reaching the vacuum lines.
Down size the main trunk line after every few junction boxes or branches in order to maintain sufcient suction.
Every 50 feet of main trunk line should be considered a user.
Attempts should be made to limit branch lines, orient main trunk lines close junction boxes or cabinets.
Some applications may require the main trunk line to be split, attempts should be make to equalize ow in both directions.
There is no loss of suction when overhead pluming is used in conjunction with the Surgical Operatory Recovery Package due to no liquids or
solids moving through the vacuum lines.
Power Vac (“G” / “std”) specic recommendations
When using in a oral surgery environment we highly recommend using the surgical operatory recovery room packages to collect everything
and prevent build up/clogs in the lines and vacuum tank.
When electing not to use the Surgical Operatory Recovery Package and collecting waste in the Power Vac collection canister, thorough line
cleaning should be done after every procedure and the end of the day. The collection tank will also need rinsed periodically. When using the
Power Vac style vacuum in this application the lines will have to be constructed with “drain,waste and vent”(DWV), or 45 deg. elbows (follow
Overhead or Through the Floor specications previously stated in this guide).
midmark.com
12
© 2015 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
004-1130-00 Rev. A (12/11/15)
AA206600
1 ”
1 1/4”
At Each Termination
“Surgical Operatory
Recovery Package “
5/8” Branch
Line “Typical”
Vacuum Piping to be Type M or Schedule 40 PVC Maintain a 1/4”
Slope per 10 feet of Piping (toward pump). Always Stagger Branch
Lines From Main Trunk, Do Not Use 4-way Cross Fittings.
Oral Surgery - Sample Plumbing Layout (Split Trunk Line)
Main
Trunk Line
1 1/4”
Equipment
Room
Equipment Alert
Subfloor plumbing is preferred. Operatory
utilization to be modified to allow flow up walls.
Upsize pump as per number of operatories. At least 1/2 user
per VRV and 8 ft. riser. More than 8 ft. riser requires more
pump capacity.
midmark.com
13
© 2015 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
004-1130-00 Rev. A (12/11/15)
midmark.com
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13

Midmark PowerVac Installation guide

Type
Installation guide
This manual is also suitable for

Midmark PowerVac G is a powerful dental vacuum system designed to meet the demands of modern dental practices. It features a compact design, quiet operation, and a variety of customizable options to suit your specific needs. With its strong suction and reliable performance, the PowerVac G ensures efficient removal of fluids and debris during dental procedures. It is suitable for use with various dental equipment, including high-volume evacuators, saliva ejectors, and handpiece coolant lines.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI