Quick Setup Guide
InPro 8000 Series
1 Introduction
ThankyouforbuyingthisInPro
®
8000Seriesturbiditysen-
sorfromMETTLERTOLEDO.InPro8000Seriesareoptical
sensors which measure the concentration of suspended
particlesinaliquidmediumonthebasisoftheprincipleof
backscatteredlight.ThesensorisoperatedwiththeM800
transmitterfromMETTLERTOLEDO.
2 Safety instructions
h
Notice:Before every start-up, the sensor must be
checkedfor:
– Damagetotheconnections,fastenings,etc.
– Leakage
– Perfectfunctioning
– Authorizationforuseinconjunctionwithotherplant
equipmentandresources.
h
Notice: The manufacturer/supplier accepts no re-
sponsibilityforanydamagecausedbyunauthorized
attachments orfor the incorporationof spareparts
whicharenotofMETTLERTOLEDOprovenance.The
riskisborneentirelybytheoperator.
a
Caution:Beforethesensorisputintooperation,the
operatormusthavealreadyclariedthatuseofthe
sensorinconjunctionwiththeotherassociatedequip-
mentandresourcesisfullyauthorized.
a
Caution:Adefectivesensormayneitherbeinstalled
norputintooperation.Faultycontainmentorinstalla-
tionoutofconformancewithregulationsandinstruc-
tionscanleadtotheescapeofmediumortopressure
surges (explosion),potentially harmfulboth to per-
sonsandtotheenvironment.
a
Caution:BeforeinstallationinEx-zones,pleaseread
andobservetheguidelines(frominstructionmanuals)
andadheretosiderequirements.
a
Caution:Theberopticcableshaveacoreofhard
cladsilica(HCS)bers.Whenlayingberopticca-
bles, the bend radiusmustalways exceed at least
10cm(4").Fiberopticcablesmayneverbeshortened
independently.FittingoftheSMAplugrequiresspecial
tooling.Thetransparentplasticcaponthesensoris
toberemovedbeforeuse.Anyformationofdeposits
ofsolidparticlesonthesurfaceistobeavoided.
a
Warning:Beforedismantlingthesensororcommenc-
inganymaintenanceworkonit,ensurethattheequip-
mentinwhichthesensorisinstalledisinasafecon-
dition(depressurize,noexplosionrisk,empty,rinse,
vent,etc.).Themetalbodyoftheturbiditysensorsas
wellasthesafetyweld-insocketsMETTLERTOLEDO
housingInFit
®
type76*-***andInTrac
®
7**-***are,
ifnecessary,tobeincludedintotherecurringpressure
testoftheunit.
3 Product description
InPro8100/8200seriesareATEXcertiedtobeinstalledin
Ex-classiedarea.Pleaseobservetheprecautionsandspe-
cialconditionsforsafeusebeforeinstallingthesensorinan
Ex-classiedarea.