Roche cobas s 201 system User manual

Type
User manual
COBAS
®
TaqMan
®
Analyzer
Instrument Manual
Version 2.0
For use with the cobas s 201 system
Operator’s Manual and the
AMPLILINK Software Version 3.3
Series
COBAS
®
TaqMan
®
Analyzer
Roche Diagnostics
2 Instrument Manual · Version 2.0
Document information
Revision history
Edition notice This publication is for users of the COBAS
®
TaqMan
®
Analyzer.
Some of the tests identified for use with the COBAS
®
TaqMan
®
Analyzer may not be
approved for use in all countries. Contact a Roche representative for tests that are
commercially available in your country.
Every effort has been made to ensure that all the information contained in this
manual is correct at the time of printing. However, Roche Diagnostics reserves the
right to make any changes necessary without notice as part of ongoing product
development.
Any customer modification to the analyzer will render the warranty or service
agreement null and void.
Software updates are done by Roche Service representatives.
Intended use The COBAS
®
TaqMan
®
Analyzer is an automated amplification and detection system
for nucleic acids using 5' nuclease technology.
The docking station connects the COBAS
®
AmpliPrep Instrument and the COBAS
®
TaqMan
®
Analyzer to fully automate sample preparation, amplification, and
detection.
The COBAS
®
TaqMan
®
Analyzer and docking station are intended to be used by
laboratory professionals trained in laboratory techniques and by instruction on the
use of the system.
This manual is to be used in conjunction with the Instrument Manual for the
COBAS
®
AmpliPrep Instrument and the cobas s 201 system Operator’s Manual.
This manual is not to be used for applications performed using versions of the
AMPLILINK software earlier than 3.3 Series.
Copyright © 2007-2014, Roche Diagnostics International Ltd. All rights reserved.
Trademarks ROCHE, AMPLILINK, AMPERASE, COBAS, AMPLIPREP, TAQMAN, and
COBAS S are trademarks of Roche.
Microsoft and Windows are either registered trademarks or trademarks of Microsoft
Corporation in the United States and/or other countries.
Manual Version Software Version Revision date Changes
1.0 3.2 March 2007
1.1 3.3 August 2010 Update to Roche Corporate
Identity and AMPLILINK
software version 3.3
1.2 3.3 February 2013 Address change
Update laser information and
warnings
Cleaning procedures updated
General changes and
adaptations
2.0
3.3 September 2014 Support for PDM SW4.0.
Roche Diagnostics
Instrument Manual · Version 2.0 3
COBAS
®
TaqMan
®
Analyzer
Instrument approvals The COBAS® TaqMan® Analyzer meets the protection requirements laid down in
Directive 98/79/EC of the European Parliament and of the Council of 27 October
1998 on In-vitro Diagnostics Medical Devices. Compliance is provided by means of
the Declaration of Conformity.
Regulatory compliance is demonstrated by the following marks:
Contact addresses
Complies with the IVD directive 98/79/EC.
CUS
®
Issued by Underwriters Laboratories, Inc. (UL) for Canada and the
US.
Roche Molecular Systems, Inc.
1080 US Highway 202 South
Branchburg, NJ 08876
USA
Made in Switzerland
Roche Diagnostics GmbH
Sandhofer Strasse 116
68305 Mannheim
Germany
COBAS
®
TaqMan
®
Analyzer
Roche Diagnostics
4 Instrument Manual · Version 2.0
Roche Diagnostics
Instrument Manual · Version 2.0 5
COBAS
®
TaqMan
®
Analyzer
Table of contents
Document information 2
Contact addresses 3
Table of contents 5
Preface 7
How to use this manual 7
Conventions used in this manual 8
System Description Part A
1 General safety information
Safety classifications A–5
Safety precautions A–6
Safety summary A–8
Safety labels on the system A–13
Disposal of the analyzer A–16
2 Overview
Technical specifications A–19
Installation A–21
3 Instrument
Methodology A–27
Covers and panels A–29
LEDs A–33
Work area A–34
Hardware overview A–35
Operation Part B
4Operation
Overview B–5
Safety information B–5
Workflow B–6
Setup and operation B–7
Maintenance Part C
5Service
Overview C–5
Safety information C–5
General cleaning C–7
Daily maintenance C–9
Periodic maintenance C–10
Troubleshooting Part D
6 Troubleshooting
Overview D–5
Error messages D–7
Glossary and Index Part E
Glossary E–3
Index E–5
Revisions Part F
Roche Diagnostics
6 Instrument Manual · Version 2.0
COBAS
®
TaqMan
®
Analyzer
Roche Diagnostics
Instrument Manual · Version 2.0 7
COBAS
®
TaqMan
®
Analyzer
Preface
The COBAS
®
TaqMan
®
Analyzer is an automated amplification and detection system
for nucleic acids using 5' nuclease technology.
The docking station connects the COBAS
®
AmpliPrep Instrument and the COBAS
®
TaqMan
®
Analyzer to fully automate sample preparation, amplification, and
detection.
The COBAS
®
TaqMan
®
Analyzer and docking station are intended to be used by
laboratory professionals trained in laboratory techniques and by instruction on the
use of the system.
This manual is to be used in conjunction with the Instrument Manual for the
COBAS
®
AmpliPrep Instrument and the cobas s 201 system Operator’s Manual.
How to use this manual
To help you find information quickly, there is a table of contents at the beginning of
the book and each chapter. In addition, a complete index can be found at the end.
The Instrument Manual for the COBAS
®
TaqMan
®
Analyzer contains the following
chapters:
Preface Summarizes the conventions used in this manual.
Safety Summarizes safety precautions that are necessary when installing, operating, and
servicing the analyzer.
Overview Provides an introduction to the COBAS
®
TaqMan
®
Analyzer and docking station. The
analyzer specifications and procedures for installation and setup of the analyzer are
described.
Instrument Describes the COBAS
®
TaqMan
®
Analyzer in detail. The major components of the
analyzer are listed, and the function and operation of each component is described.
Operation Outlines the procedures used during routine operation of the COBAS
®
TaqMan
®
Analyzer.
Service Reviews the steps required for regular service of the COBAS
®
TaqMan
®
Analyzer.
Troubleshooting Describes the procedures for resolving problems during operation.
Glossary and Index Provides reference and look-up material.
NOTICE
o This manual is only to be used in combination with AMPLILINK software version 3.3
Series.
o This manual is not to be used for applications performed using versions of the
AMPLILINK Software earlier than 3.3 Series.
o Keep this Instrument Manual in a safe place to ensure that it is not damaged and
remains available for use.
o This Instrument Manual should be easily accessible at all times.
COBAS
®
TaqMan
®
Analyzer
Roche Diagnostics
8 Instrument Manual · Version 2.0
Conventions used in this manual
This section explains the formatting conventions used in this manual.
Text conventions To convey information readily and consistently, certain text conventions are used
throughout this manual. These text conventions are as follows:
Symbols Visual cues are used to help locate and interpret information in this manual quickly.
The following symbols are used:
Screen shots The screen representations shown in this publication are for illustrative purposes
only. The screens do not necessarily show valid data.
Text convention Used for
Numbered lists
Numbered procedures that should be implemented sequentially
during an operation:
1
Turn power switch on.
2
Click File.
Italics
Designating another section of this manual that should be consulted:
e
For additional details, see To perform daily maintenance on
page C-9
>
Separates command options in a multi-level command:
Select File > Shut Down from the menu bar.
Symbol Used for
a Start of procedure
End of procedure
o List item
e
Cross-reference
Tip
Danger alert symbol
Warning alert symbol
Caution and general safety alert symbol
Electrical and electronic equipment marked with this symbol are
covered by the European directive WEEE.
The symbol denotes that the equipment must not be disposed of in
the municipal waste system.
Global Trade Item Number. Used on the type plate of the instrument.
Roche Diagnostics
Instrument Manual · Version 2.0 9
COBAS
®
TaqMan
®
Analyzer
Units The following measurement abbreviations are used:
cm Centimeters
Hz Hertz
in Inches
kg Kilograms
lb Pounds (weight)
VA Volt-Ampères (power consumption)
VAC Volts (alternating current)
COBAS
®
TaqMan
®
Analyzer
Roche Diagnostics
10 Instrument Manual · Version 2.0
1 General safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
2 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-17
3 Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-25
System Description
A
Roche Diagnostics
Instrument Manual · Version 2.0 A-3
COBAS
®
TaqMan
®
Analyzer 1 General safety information
Table of contents
General safety information
In this chapter, you will find information on the safe operation of the COBAS
®
TaqMan
®
Analyzer.
Safety classifications ........................................................................................................ A-5
Safety precautions ........................................................................................................... A-6
Operator qualification .............................................................................................. A-6
Safe and proper use of the analyzer ........................................................................ A-6
Miscellaneous safety precautions ............................................................................ A-7
Safety summary ............................................................................................................... A-8
Warning messages .................................................................................................... A-8
Electrical safety .................................................................................................... A-8
Optical safety ....................................................................................................... A-8
Biohazardous materials ...................................................................................... A-9
Waste .................................................................................................................... A-9
Explosion and fire risk ..................................................................................... A-10
Caution messages .................................................................................................... A-11
Mechanical safety .............................................................................................. A-11
Insoluble contaminants in samples ................................................................ A-11
Notices ...................................................................................................................... A-12
Moving parts ..................................................................................................... A-12
Circuit breakers and fuses ............................................................................... A-12
Spillage ............................................................................................................... A-12
Safety labels on the system ........................................................................................... A-13
Safety information for laser transmitter .............................................................. A-14
Disposal of the analyzer ............................................................................................... A-16
In this chapter
Chapter
1
Roche Diagnostics
A-4 Instrument Manual · Version 2.0
1 General safety information COBAS
®
TaqMan
®
Analyzer
Table of contents
Roche Diagnostics
Instrument Manual · Version 2.0 A-5
COBAS
®
TaqMan
®
Analyzer 1 General safety information
Safety classifications
Safety classifications
This section explains how precautionary information is presented in this manual.
The safety precautions and important user notes are classified according to the
ANSI Z535.6 Standard. Familiarize yourself with the following icons and their
meanings:
These symbols and signal words are used for specific hazards:
e
For more information about the product safety labels, see Safety labels on the system on
page A-13
Important information which is not safety relevant is indicated by the following
symbol:
The safety alert symbol by itself without a signal word is used to promote awareness to
hazards which are generic or to direct the reader to safety information provided elsewhere
in the document.
DANGER
DANGER
Indicates a hazardous situation which, if not avoided, will result in death or serious injury.
WARNING
WARNING
Indicates a hazardous situation which, if not avoided, could result in death or serious
injury.
CAUTION
CAUTION
Indicates a hazardous situation which, if not avoided, could result in minor or moderate
injury.
NOTICE
NOTICE
Indicates a hazardous situation which, if not avoided, could result in damage to equipment.
Tip
Indicates additional information on correct use of the analyzer or useful tips.
Roche Diagnostics
A-6 Instrument Manual · Version 2.0
1 General safety information COBAS
®
TaqMan
®
Analyzer
Safety precautions
Safety precautions
Operator qualification
Operators are required to have a sound knowledge of relevant guidelines and
standards as well as the information and procedures contained in the Instrument
Manuals.
o Do not carry out operation and maintenance unless you have been trained.
o Carefully follow the procedures specified in the Instrument Manuals for the
operation and maintenance of the analyzer.
o Leave maintenance, installation, or service that is not described in the Instrument
Manuals to trained Roche Service personnel.
o Follow standard laboratory practices, especially when working with biohazardous
material.
Safe and proper use of the analyzer
Personal protective equipment o Be sure to wear appropriate protective equipment, including, but not limited to,
safety glasses with side shields, lab coat, and approved disposable gloves.
o Wear a face shield if there is a chance of splash or splatter.
Accuracy/precision of measured
results
An incorrect measuring result may lead to an error in diagnosis, therefore posing
danger to the patient.
o For proper use of the analyzer, measure control samples and monitor the analyzer
during operation.
o Do not use reagents and consumables that have exceeded their expiration date,
otherwise inaccurate data may be obtained.
o For diagnostic purposes always assess the results in conjunction with the patient’s
medical history, clinical examination, and other findings.
o Each laboratory must verify that analyzer and reagent performance meets the
published specifications.
Installation o Installation must be performed by trained Roche Service personnel only.
o Leave installation that is not described in the Instrument Manuals to trained
Roche Service personnel. Follow the described installation instructions carefully.
Correct use Use the COBAS
®
TaqMan
®
Analyzer only for measuring liquid samples with the
provided reagents.
Particular attention must be paid to the following safety precautions. If these safety
precautions are ignored, the operator may suffer serious or fatal injury. Each precaution is
important.
Roche Diagnostics
Instrument Manual · Version 2.0 A-7
COBAS
®
TaqMan
®
Analyzer 1 General safety information
Safety precautions
Operating conditions o Operation outside of the specified ranges may lead to incorrect results or
malfunction of the analyzer (see Technical specifications on page A-19).
o Use the analyzer indoors only and avoid heat and humidity (see Technical
specifications on page A-19).
o Make sure that the analyzer’s ventilation openings remain unobstructed at all
times.
o Perform maintenance according to the specified intervals to maintain the
operating conditions of the analyzer.
o Keep the Instrument Manual in a safe place to ensure that it is not damaged and
remains available for use. This manual must be easily accessible at all times.
Approved parts Use of non-approved parts or devices may result in malfunction of the analyzer and
may render the warranty null and void. Only use parts and devices approved by
Roche Diagnostics.
Miscellaneous safety precautions
Power interruption A power failure or momentary drop in voltage may damage the analyzer or lead to
data loss. Perform regularly backups of measurement results. Operation with an
uninterruptible power supply (UPS) is recommended.
Electromagnetic fields Do not use this device in close proximity to sources of strong electromagnetic
radiation (e.g. unshielded intentional RF sources), as these can interfere with the
proper operation.
Relocation and transportation Do not attempt to relocate or transport the analyzer or docking station. Leave
relocation and transportation to personnel trained or authorized by Roche.
e
For additional details, see Disposal of the analyzer on page A-16
Roche Diagnostics
A-8 Instrument Manual · Version 2.0
1 General safety information COBAS
®
TaqMan
®
Analyzer
Safety summary
Safety summary
This safety summary contains the most important general warning and caution
messages. Additionally, you will find specific safety information at the beginning of
Chapter Operation and Chapter Service
.
Warning messages
Electrical safety
Optical safety
WARNING
List of warning messages
Before operating the analyzer, read the warning messages contained in this summary
carefully. Failure to observe them may result in death or serious injury.
Electrical shock by electronic equipment
o Do not attempt to work in any electronic compartment.
o Do not remove any cover of the analyzer other than those specified in this Instrument
Manual.
o Do not open the side panels while the analyzer is connected to the power supply.
o Do not touch any parts of the analyzer other than those specified. Especially do not
touch any power supply parts.
o Never remove the middle grounding prong from the power cable or defeat its purpose
by using an ungrounded adapter.
o Installation, service, and repair must only be performed by personnel authorized and
qualified by Roche.
o Observe the system safety labels as illustrated in section Safety labels on the system on
page A-13.
Loss of sight due to staring into laser beam
The on-board barcode reader contains a class 2 laser diode.
Do not stare into the laser transmitter beam as your eyesight may be severely damaged.
e
For position of the laser transmitter, see Safety information for laser transmitter on
page A-14
Damage to the eyes due to staring into light of the fluorometer lamp
Looking into the light of the fluorometer lamp may severely damage your eyesight.
Always wear dark safety glasses when looking at the light of the fluorometer lamp.
Roche Diagnostics
Instrument Manual · Version 2.0 A-9
COBAS
®
TaqMan
®
Analyzer 1 General safety information
Safety summary
Biohazardous materials
Waste
e
For information about disposal of the analyzer, see Disposal of the analyzer on page A-16
Infection by samples and associated materials
Samples analyzed with this analyzer are treated to inactivate potentially biohazardous
material. However, as with all human-derived specimens, universal precautions should be
taken when handling and processing samples.
Contact with samples containing material of human origin may result in infection. All
materials and mechanical components associated with samples of human origin are
potentially biohazardous.
o Follow standard laboratory practices, especially when working with biohazardous
material.
o Keep the main cover and the load panel closed and in place while the analyzer is
operating.
o When working with open main cover while the analyzer is powered on (e.g. for
cleaning or maintenance), always put the analyzer in Maintenance mode or in
Shutdown status first.
o Be sure to wear appropriate protective equipment, including, but not limited to, safety
glasses with side shields, lab coat, and approved disposable gloves.
o Wear a face shield if there is a chance of splash or splatter.
o If any biohazardous material is spilled, wipe it up immediately and apply disinfectant.
See General cleaning on page C-7 for additional cleaning recommendation.
o If sample or waste solution comes into contact with your skin, immediately apply
disinfectant, and then wash it off with soap and water. Consult a physician.
Infection by biohazardous waste
Contact with waste may result in infection. All materials and mechanical components
associated with the waste systems are potentially biohazardous.
o Be sure to wear protective equipment. Take extra care when working with protective
gloves; these can easily be pierced or cut which can lead to infection.
o If any biohazardous material is spilled, wipe it up immediately and apply disinfectant.
o If waste solution comes into contact with your skin, wash it off immediately with water
and apply a disinfectant. Consult a physician.
o Observe the system safety labels as illustrated in section Safety labels on the system on
page A-13.
Contamination of the environment by waste solution and solid waste
The waste of the analyzer is potentially biohazardous.
o When disposing of any waste generated by the analyzer, do so according to the
appropriate local regulations.
Roche Diagnostics
A-10 Instrument Manual · Version 2.0
1 General safety information COBAS
®
TaqMan
®
Analyzer
Safety summary
Explosion and fire risk
Explosion through sparks
Hazard of explosion through sparks.
o Keep all potentially flammable or explosive material (e.g. anesthetic gas) away from the
analyzer.
Fire risk through usage of sprays
Spraying liquid on the power supply parts can cause a short circuit and result in a fire.
o Keep the cover closed while the analyzer is connected to the main power supply, and
do not use sprays in the vicinity of the analyzer.
o During fire-fighting operations disconnect the COBAS
®
TaqMan
®
Analyzer from the
main power supply.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96