Roche COBAS TaqMan User manual

Type
User manual
COBAS
®
TaqMan
®
Analyzer
Instrument Manual
Version 3.0
For use with the AMPLILINK Software, Version 3.4.2
COBAS® TaqMan® Analyzer
Roche Diagnostics
2 Instrument Manual · Version 3.0
Publication information
Edition notice
This publication is intended for operators of the COBAS
®
TaqMan
®
Analyzer.
Some of the tests identified for use with the COBAS
®
TaqMan
®
Analyzer may not be
approved for use in all countries. Contact a Roche representative for tests that are
commercially available in your country.
Every effort has been made to ensure that all the information contained in this
publication is correct at the time of publishing. However, Roche Molecular Systems,
Inc. may need to update the publication information as output of product
surveillance activities, leading to a new version of this publication.
Where to find information The Instrument Manual contains all information about the product, including the
following:
o Routine operation
o Maintenance
o Safety
o Troubleshooting information
o A software reference
o Configuration information
o Background information
CAUTION
General attention
To avoid serious or fatal injury, ensure that you are familiar with the system and safety
information before you use the system.
r Pay particular attention to all safety precautions.
r Always follow the instructions in this publication.
r Do not use the instrument in a way that is not described in this publication.
r Store all publications in a safe and easily accessible place.
Training Do not carry out operation tasks or maintenance actions unless you have received
training from Roche Diagnostics. Leave tasks that are not described in the user
documentation to trained Roche Service representatives.
Publication
version
Software version Revision date Change description
1.0 3.2 March 2007
1.1 3.2 and 3.3 September 2008 Update to Roche Corporate Identity and AMPLILINK Software
Version 3.3 Series
1.2 3.2 and 3.3 November 2012 Address change
Update laser information and warnings
Cleaning procedures updated
General changes and adaptations
1.3 3.2 and 3.3 June 2013 Hardware error messages updated
2.0 3.3 and 3.4 June 2015 Update to AMPLILINK Software Version 3.4 Series
3.0 3.3 and 3.4 January 2020
u What is new in Instrument Manual version 3.0 (p. 11)
Table 1 Revision history
Roche Diagnostics
Instrument Manual · Version 3.0 3
COBAS® TaqMan® Analyzer
Images The screenshots and hardware images in this publication have been added exclusively
for illustration purposes. Configurable and variable data in screenshots, such as tests,
results, or path names visible therein must not be used for laboratory purposes.
Warranty Any customer modification to the system renders the warranty or service agreement
null and void.
For conditions of warranty, contact your local sales representative or refer to your
warranty contract partner.
Always leave software updates to a Roche Service representative or perform such
updates with their assistance.
Copyright © 2007-2020, F. Hoffmann La-Roche Ltd. All rights reserved.
License information
COBAS
®
TaqMan
®
Analyzer software is protected by contract law, copyright law, and
international treaties. COBAS
®
TaqMan
®
Analyzer contains a user license between F.
Hoffmann-La Roche Ltd. and a license holder, and only authorized users may access
the software and use it. Unauthorized use and distribution may result in civil and
criminal penalties.
Open-source and commercial
software
COBAS
®
TaqMan
®
Analyzer may include components or modules of commercial or
open-source software. For further information on the intellectual property and other
warnings, as well as licenses pertaining to the software programs included in
COBAS
®
TaqMan
®
Analyzer, refer to the electronic distribution included with this
product.
This open-source and commercial software and COBAS
®
TaqMan
®
Analyzer as a
whole can constitute a device regulated in accordance with applicable law. For more
detailed information, refer to the corresponding user documentation and labeling.
Note that the respective authorization is no longer valid according to the
corresponding legislation should any unauthorized changes be made to
COBAS
®
TaqMan
®
Analyzer.
Trademarks The following trademarks are acknowledged:
AMPERASE, AMPLILINK, AMPLIPREP, COBAS, and TAQMAN are trademarks of
Roche.
All other trademarks are the property of their respective owners.
Feedback Every effort has been made to ensure that this publication fulfills the intended use. All
feedback on any aspect of this publication is welcome and is considered during
updates. Contact your Roche representative, should you have any such feedback.
Approvals
The COBAS
®
TaqMan
®
Analyzer meets the requirements laid down in:
Directive 98/79/EC of the European Parliament and of the Council of 27 October
1998 on in vitro diagnostic medical devices.
Directive 2011/65/EU of the European Parliament and of the Council of 8 June 2011
on the restriction of the use of certain hazardous substances in electrical and
electronic equipment.
Compliance with the applicable directive(s) is provided by means of the Declaration
of Conformity.
The following marks demonstrate compliance:
COBAS® TaqMan® Analyzer
Roche Diagnostics
4 Instrument Manual · Version 3.0
Lead This product includes an internal constituent containing a Substance of Very High
Concern (SVHC), Lead (CAS 7439-92-1), in a concentration above 0.1% weight by
weight, as identified under REACH and added to the Candidate List.
There is no direct exposure to the substance when the instrument is operated
according to the instructions for use.
Contact addresses
For in vitro diagnostic use.
Complies with the provisions of the applicable EU directives.
Issued by Underwriters Laboratories, Inc. (UL) for Canada and
the US.
CUS
®
Roche Molecular Systems, Inc.
1080 US Highway 202 South
Branchburg, NJ 08876
USA
Made in Switzerland
Roche Diagnostics GmbH
Sandhofer Strasse 116
68305 Mannheim
Germany
Roche Diagnostics
Instrument Manual · Version 3.0 5
COBAS® TaqMan® Analyzer
Table of contents
Publication information 2
Contact addresses 4
Table of contents 5
Intended use 7
Symbols 7
What is new in Instrument Manual version 3.0 8
What is new in Instrument Manual version 2.0 8
System description
1 General safety information
Safety classifications 13
Safety precautions 14
Safety summary 16
Safety labels on the system 20
Disposal of the analyzer 23
2 Overview
Housing 27
Technical specifications 28
Installation 30
3 Instrument
Methodology 37
Covers and panels 39
LEDs 43
Work area 44
Hardware overview 45
Operation
4 Operation
Overview 55
Safety information 56
Workf low 57
Setup and operation 58
Maintenance
5 Service
Overview 65
Safety information 66
General cleaning 67
Daily maintenance 69
Periodic maintenance 70
Replacing the fan filters 76
Replacing the lamp 78
Troubleshooting
6 Troubleshooting
Overview 85
Error messages 87
Glossary
7 Glossary
Index
Index 95
Revisions
Roche Diagnostics
6 Instrument Manual · Version 3.0
COBAS® TaqMan® Analyzer
Roche Diagnostics
Instrument Manual · Version 3.0 7
COBAS® TaqMan® Analyzer
Intended use
The COBAS
®
TaqMan
®
Analyzer is an automated amplification and detection system
for nucleic acids using 5' nuclease technology.
The docking station connects the COBAS
®
AmpliPrep Instrument and the
COBAS
®
TaqMan
®
Analyzer to fully automate sample preparation, amplification, and
detection.
The COBAS
®
TaqMan
®
Analyzer and docking station are intended to be used by
laboratory professionals trained in laboratory techniques and by instruction on the
use of the system.
This manual is to be used in conjunction with the Instrument Manual for the
COBAS
®
AmpliPrep Instrument and the Application Manual for the AMPLILINK
Software Version 3.4.2.
NOTICE
Notice
r This manual is only to be used for COBAS
®
TaqMan
®
Analyzer with serial number
450001.
r This manual is only to be used in combination with AMPLILINK Software Version 3.4.2.
r This manual is not to be used for applications performed using versions of the
AMPLILINK Software earlier than 3.4.2 Series.
Symbols
Symbols used in the publication
GTIN icon used on the instrument
Symbol Explanation
p Start of procedure
s End of procedure
o List item
u Cross-reference
q Tip
Electrical and electronic equipment marked with this symbol are
covered by the European directive WEEE.
The symbol denotes that the equipment must not be disposed of in
the municipal waste system.
Table 2 Symbols used in the publication
Symbol Explanation
Global Trade Item Number
Table 3 GTIN icon used on the instrument
COBAS® TaqMan® Analyzer
Roche Diagnostics
8 Instrument Manual · Version 3.0
What is new in Instrument Manual version 3.0
Main cover When opening the main cover, the analyzer is automatically powered off. After
closing the main cover, the analyzer is powered on again and initializes.
All instructional information in this publication has been adapted accordingly.
Access protection Access protection has been improved by introducing several safety switches, an
interlock functionality on the power supply, and access protection flaps at the
entrance of the sample rack loading area.
Lead A REACH statement about lead was added.
u Lead (p. 4)
Shar ps, rough edges, and/or
moving parts
A warning statement about sharps, rough edges, and/or moving parts was added.
u Sharps, rough edges, and/or moving parts (p. 16)
Laser class 1 The laser transmitter of the barcode scanner on the rack platform is now classified as
IEC 60825-1 class 1 laser product and is less harmful. Consequently, all warning
statements about the loss of sight due to staring into the laser beam have been
removed.
COBAS
®
Taq Man
®
48 Analyzer
All information related to the COBAS
®
TaqMan
®
48 Analyzer has been removed from
this publication. The AMPLILINK Software Version 3.4.2 is not compatible with the
COBAS
®
TaqMan
®
48 Analyzer.
Compatibility o This publication is only to be used for COBAS
®
TaqMan
®
Analyzer with serial
number 450001.
o This publication is only to be used in combination with AMPLILINK Software
Version 3.4.2.
What is new in Instrument Manual version 2.0
Changes in the fro nt mat ter and
preface
o Front matter and preface have been updated according to the new Roche
Diagnostics standards.
o Company name adapted in copyright information.
o GTIN (Global Trade Item Number) symbol added to the list of symbols.
Error messages Hardware error messages updated.
u For details, see H ardware error messages (p. 87)
General changes in this doc ument o AMPLILINK Software Version updated: 3.2 Series deleted and 3.4 Series added.
o Term “data station” replaced with “control unit.
o Any reference to the COBAS
®
AMPLICOR
®
removed.
o Information about the Windows operating system adapted.
System description
1 General safety information .............................................................................................................. 11
2 Overview .......................................................................................................................................... 25
3 Instrument........................................................................................................................................ 35
Roche Diagnostics
Instrument Manual · Version 3.0 11
COBAS® TaqMan® Analyzer 1 General safety information
Table of co nte nts
General safety information 1
In this chapter, you will find information on the safe operation of the
COBAS
®
TaqMan
®
Analyzer.
In this chapter
Chapter
1
Safety classifications.............................................................................................................13
Safety precautions.................................................................................................................14
Operator qualification...................................................................................................14
Safe and proper use of the analyzer .............................................................................14
Miscellaneous safety precautions.................................................................................15
Safety summary ....................................................................................................................16
Warning messages..........................................................................................................16
Sharps, rough edges, and/or moving parts...........................................................16
Electrical safety.........................................................................................................16
Optical safety ............................................................................................................16
Biohazardous materials ...........................................................................................17
Waste..........................................................................................................................17
Explosion and fire risk ............................................................................................17
Caution messages...........................................................................................................18
Mechanical safety.....................................................................................................18
Insoluble contaminants in samples........................................................................18
Notices .............................................................................................................................19
Moving parts.............................................................................................................19
Circuit breakers and fuses.......................................................................................19
Spillage.......................................................................................................................19
Safety labels on the system ..................................................................................................20
Safety information for laser transmitter......................................................................22
Disposal of the analyzer.......................................................................................................23
Roche Diagnostics
12 Instrument Manual · Version 3.0
1 General safety information COBAS® TaqMan® Analyzer
Table of co nte nts
Roche Diagnostics
Instrument Manual · Version 3.0 13
COBAS® TaqMan® Analyzer 1 General safety information
Safety classifications
Safety classifications
This section explains how precautionary information is presented in this manual.
The safety precautions and important user notes are classified according to the
ANSI Z535.6 Standard. Familiarize yourself with the following icons and their
meanings:
Generic safety precaution
r The safety alert symbol by itself without a signal word is used to promote awareness to
hazards which are generic or to direct the reader to safety information provided
elsewhere in the document.
These symbols and signal words are used for specific hazards:
WARNING
WARNING
r Indicates a hazardous situation which, if not avoided, could result in death or serious
injury.
CAUTION
CAUTION
r Indicates a hazardous situation which, if not avoided, could result in minor or moderate
injury.
NOTICE
NOTICE
r Indicates a hazardous situation which, if not avoided, could result in damage to
equipment.
u For more information about the product safety labels, see Safety labels on the system
(p. 20)
Important information which is not safety relevant is indicated by the following
symbol:
q
Tip
Indicates additional information on correct use of the analyzer or useful tips.
Roche Diagnostics
14 Instrument Manual · Version 3.0
1 General safety information COBAS® TaqMan® Analyzer
Safety precautions
Safety precautions
Generic safety precaution
r Particular attention must be paid to the following safety precautions. If these safety
precautions are ignored, the operator may suffer serious or fatal injury. Each
precaution is important.
Operator qualification
Operators are required to have a sound knowledge of relevant guidelines and
standards as well as the information and procedures contained in the Instrument and
Application Manuals.
o Do not carry out operation and maintenance unless you have been trained.
o Carefully follow the procedures specified in the Instrument and Application
Manuals for the operation and maintenance of the analyzer.
o Leave maintenance, installation, or service that is not described in the Instrument
and Application Manuals to trained Roche Service personnel.
o Follow standard laboratory practices, especially when working with biohazardous
material.
Safe and proper use of the analyzer
Personal protective equipment o Be sure to wear appropriate protective equipment, including, but not limited to,
safety glasses with side shields, lab coat, and approved disposable gloves.
o Wear a face shield if there is a chance of splash or splatter.
Accuracy/precision of measured
results
An incorrect measuring result may lead to an error in diagnosis, therefore posing
danger to the patient.
o For proper use of the analyzer, measure control samples and monitor the analyzer
during operation.
o Do not use reagents and consumables that have exceeded their expiration date,
otherwise inaccurate data may be obtained.
o For diagnostic purposes always assess the results in conjunction with the patients
medical history, clinical examination, and other findings.
o Each laboratory must verify that analyzer and reagent performance meets the
published specifications.
Installation o Installation must be performed by trained Roche Service personnel only.
o Leave installation that is not described in the Instrument and Application
Manuals to trained Roche Service personnel. Follow the described installation
instructions carefully.
Correct use
Use the COBAS
®
TaqMan
®
Analyzer only for measuring liquid samples with the
provided reagents.
Roche Diagnostics
Instrument Manual · Version 3.0 15
COBAS® TaqMan® Analyzer 1 General safety information
Safety precautions
Operating conditions o Operation outside of the specified ranges may lead to incorrect results or
malfunction of the analyzer (see Technical specifications (p. 28)).
o Use the analyzer indoors only and avoid heat and humidity (see Tec hnic al
specifications (p. 28)).
o Make sure that the analyzer’s ventilation openings remain unobstructed at all
times.
o Perform maintenance according to the specified intervals to maintain the
operating conditions of the analyzer.
o Keep the Instrument Manual in a safe place to ensure that it is not damaged and
remains available for use. This manual must be easily accessible at all times.
Approved part s Use of non-approved parts or devices may result in malfunction of the analyzer and
may render the warranty null and void. Only use parts and devices approved by
Roche Diagnostics.
Miscellaneous safety precautions
Power interruption A power failure or momentary drop in voltage may damage the analyzer or lead to
data loss. Perform regularly backups of measurement results. Operation with an
uninterruptible power supply (UPS) is recommended.
Electromagnetic fields Do not use this device in close proximity to sources of strong electromagnetic
radiation (e.g. unshielded intentional RF sources), as these can interfere with the
proper operation.
Relocation and transportation Do not attempt to relocate or transport the analyzer or docking station. Leave
relocation and transportation to personnel trained or authorized by Roche.
u For additional details, see Disposal of the analyzer (p. 23)
Roche Diagnostics
16 Instrument Manual · Version 3.0
1 General safety information COBAS® TaqMan® Analyzer
Safety summary
Safety summary
This safety summary contains the most important general warning and caution
messages. Additionally, you will find specific safety information at the beginning of
Chapter 4 Operation (p. 53) and Chapter 5 Service (p. 63).
Warning messages
WARNING
List of warning messages
r Before operating the analyzer, read the warning messages contained in this summary
carefully. Failure to observe them may result in death or serious injury.
Sharps, rough edges, and/or moving parts
WARNING
Personal injury and infection due to sharps, rough edges, and/or moving parts
Good Laboratory Practice can reduce the risk of injury. Be aware of your laboratory
environment, well-prepared, and follow the instructions for use. Some areas of the
instrument may have sharps, rough edges, and/or moving parts.
r Wear personal protective equipment to minimize the risk of injury from bodily contact
with such parts, especially in less accessible areas, or while cleaning the instrument.
r Your personal protective equipment should be appropriate to the degree and type of
potential hazard, e.g. suitable lab gloves, eye protection, lab coat, and footwear.
Electrical safety
WARNING
Electrical shock by electronic equipment
r Do not attempt to work in any electronic compartment.
r Do not remove any cover of the analyzer other than those specified in this Instrument
Manual.
r Do not open the side panels while the analyzer is connected to the power supply.
r Do not touch any parts of the analyzer other than those specified. Especially do not
touch any power supply parts.
r Never remove the middle grounding prong from the power cable or defeat its purpose
by using an ungrounded adapter.
r Installation, service, and repair must only be performed by personnel authorized and
qualified by Roche.
r Observe the system safety labels as illustrated in section Safety labels on the system
(p. 20).
Optical safety
WARNING
Damage to the eyes due to staring into light of the fluorometer lamp
r Looking into the light of the fluorometer lamp may severely damage your eyesight.
r Always wear dark safety glasses when looking at the light of the fluorometer lamp.
Roche Diagnostics
Instrument Manual · Version 3.0 17
COBAS® TaqMan® Analyzer 1 General safety information
Safety summary
Biohazardous materials
WARNING
Infection by samples and associated materials
Samples analyzed with this analyzer are treated to inactivate potentially biohazardous
material. However, as with all human-derived specimens, universal precautions should be
taken when handling and processing samples.
Contact with samples containing material of human origin may result in infection. All
materials and mechanical components associated with samples of human origin are
potentially biohazardous.
r Follow standard laboratory practices, especially when working with biohazardous
material.
r Keep the main cover and the load panel closed and in place while the analyzer is
operating.
r Be sure to wear appropriate protective equipment, including, but not limited to, safety
glasses with side shields, lab coat, and approved disposable gloves.
r Wear a face shield if there is a chance of splash or splatter.
r If any biohazardous material is spilled, wipe it up immediately and apply disinfectant.
See General cleaning (p. 67) for additional cleaning recommendation.
r If sample or waste solution comes into contact with your skin, immediately apply
disinfectant, and then wash it off with soap and water. Consult a physician.
Waste
WARNING
Infection by biohazardous waste
Contact with waste may result in infection. All materials and mechanical components
associated with the waste systems are potentially biohazardous.
r Be sure to wear protective equipment. Take extra care when working with protective
gloves; these can easily be pierced or cut which can lead to infection.
r If any biohazardous material is spilled, wipe it up immediately and apply disinfectant.
r If waste solution comes into contact wit your skin, wash it off immediately with water
and apply a disinfectant. Consult a physician.
r Observe the system safety labels as illustrated in section Safety labels on the system
(p. 20).
WARNING
Contamination of the environment by waste solution and solid waste
The waste of the analyzer is potentially biohazardous.
r When disposing of any waste generated by the analyzer, do so according to the
appropriate local regulations.
u For information about disposal of the analyzer, see Dis po sal of the analyzer (p. 23)
Explosion and fire risk
WARNING
Explosion through sparks
Hazard of explosion through sparks.
r Keep all potentially flammable or explosive material (e.g. anesthetic gas) away from the
analyzer.
Roche Diagnostics
18 Instrument Manual · Version 3.0
1 General safety information COBAS® TaqMan® Analyzer
Safety summary
WARNING
Fire risk through usage of sprays
Spraying liquid on the power supply parts can cause a short circuit and result in a fire.
r Keep the cover closed while the analyzer is connected to the main power supply, and
do not use sprays in the vicinity of the analyzer.
r During fire-fighting operations disconnect the COBAS
®
TaqMan
®
Analyzer from the
main power supply.
Caution messages
CAUTION
List of caution messages
r Before operating the analyzer, read the caution messages contained in this summary
carefully. Failure to observe them may result in minor or moderate injury.
Mechanical safety
CAUTION
Personal injury due to contact with moving parts
Transfer head moves rapidly during operation. The main cover is locked during operation,
protecting the user from moving parts.
r Keep the main cover and the load panel closed and in place while the analyzer is
operating.
r Keep the docking station front cover closed when operating a docked system.
r Do not touch any parts of the analyzer other than those specified. Keep away from
moving parts during analyzer operation.
r During operation and maintenance of the analyzer, proceed according to the
instructions.
r Observe the system safety labels as illustrated in section Safety labels on the system
(p. 20).
Insoluble contaminants in samples
CAUTION
Incorrect results and interruption of analysis due to samples with insoluble
contaminants
r Insoluble contaminants in samples and bubbles or films inside a sample container may
cause clogging, pipetting volume shortage, or block optical measurements.
r When loading samples on the instrument, make sure they contain no insoluble
contaminants such as fibrin or dust.
Roche Diagnostics
Instrument Manual · Version 3.0 19
COBAS® TaqMan® Analyzer 1 General safety information
Safety summary
Notices
NOTICE
List of notices
r Before operating the analyzer, read the notices contained in this summary carefully.
Failure to observe them may result in damage to equipment.
Moving parts
NOTICE
Damage to the analyzer due to contact with moving parts
Contact with moving parts may damage components.
r Keep all covers closed and in place while the analyzer is operating.
r Do not touch any parts of the analyzer other than those specified. Keep away from
moving parts during analyzer operation.
Circuit breakers and fuses
NOTICE
Damage to the analyzer due to improper use
r Should one of the analyzer circuit breakers or fuses blow, do not attempt to operate the
analyzer before contacting Roche Service.
Spillage
NOTICE
Malfunction due to spilled liquid
Any liquid spilled on the analyzer may result in malfunction or damage of the analyzer.
r Do not place samples, reagents, or any other liquid on the surface of the analyzer.
r If liquid does spill on the analyzer, wipe it up immediately and apply disinfectant. Be
sure to wear protective equipment.
Roche Diagnostics
20 Instrument Manual · Version 3.0
1 General safety information COBAS® TaqMan® Analyzer
Safety labels on the system
Safety labels on the system
Warning labels have been placed on the analyzer to draw your attention to areas of
potential hazard. The labels and their definitions are listed below according to their
location on the analyzer.
The safety labels on the analyzer comply with the following standards: ANSI Z535,
IEC 61010-1, IEC 60417-DB-3M, or ISO 7000.
q
If the labels are damaged, they must be replaced by Roche Service personnel. For
replacement labels contact your local Roche representative.
Safety label on system Location and meaning
Caution
Refer to accompanying documents. The triangular
symbol on the back of the analyzer is a general
reminder that users read the safety information
contained in this manual. Users must be able to
identify specific hazards and take appropriate
action to avoid them.
Refer to the Instrument Manual for instructions
on safe operation.
Laser transmitter warning
Situated behind the load panel to designate the
presence of a laser transmitter.
Do not stare into the laser transmitter.
u For details, see Safety information for laser
transmitter (p. 22)
Laser transmitter warning
Situated on Barcode Scanner to designate the
presence of a laser transmitter.
u For details, see Safety information for laser
transmitter (p. 22)
Warning - Fire Hazard
For continued protection replace only
with same type and rating of fuse
according to the fuse rate label on the
instrument.
Avertissement-Risque d'incendie
Afin d'assurer une protection constante
toujours remettre un fusible du même
calibre.
Electrical warning
Situated on the power supply. Only replace a fuse
with a fuse of the same type and rating.
Refer to the Instrument Manual for instructions
on safe operation.
Table 1-1 Safety labels on the system
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98