Roche cobas s 201 system User manual

Type
User manual

The Roche cobas s 201 system is a fully automated sample-to-result system for molecular diagnostic testing, designed to provide fast and accurate results for a variety of tests, including infectious disease testing, genetic testing, and oncology testing. With its ability to process up to 96 samples in a single run and deliver results in as little as 3 hours, the cobas s 201 system can help laboratories improve efficiency and reduce turnaround time for molecular testing.

The Roche cobas s 201 system is a fully automated sample-to-result system for molecular diagnostic testing, designed to provide fast and accurate results for a variety of tests, including infectious disease testing, genetic testing, and oncology testing. With its ability to process up to 96 samples in a single run and deliver results in as little as 3 hours, the cobas s 201 system can help laboratories improve efficiency and reduce turnaround time for molecular testing.

04/2006, version 1.0 0.1
0
0
cobas s 201 System 0
Operator’s Manual Supplement
for Use of the Docking Station
P/N: 04875982018-01
0.2 04/2006, version 1.0
This page intentionally left blank.
Preface
04/2006, version 1.0 0.3
Table of Contents
Table of Contents .................................................................................0.3
Revision History ...................................................................................0.4
How to Use this Supplement ..............................................................0.6
Conventions Used in This Supplement ............................................0.7
Glossary ...............................................................................................0.10
Operational Precautions and Limitations ......................................0.12
..............................................................................................................0.12
Warnings .............................................................................................0.13
Intended Use .......................................................................................0.13
1. Workflow
Overview ...............................................................................................1.1
Perform Startup Procedures ...............................................................1.2
Load Reagents .......................................................................................1.4
Load Consumables ...............................................................................1.6
Create Orders ........................................................................................1.8
Load SK24 Racks ..................................................................................1.9
Start the Run .......................................................................................1.10
Remove Used Consumables .............................................................1.13
Review Run Messages ........................................................................1.13
Review and Accept Results ...............................................................1.13
2. Docking Station
Overview ...............................................................................................2.1
Installation ............................................................................................2.2
Methodology .........................................................................................2.3
Loading Empty K-carriers ..................................................................2.5
Docking Station Maintenance ............................................................2.6
3. Index
0.4 04/2006, version 1.0
Revision History
The Docking Station is manufactured by:
Manual Revision Revision Date
1.0 April 2006
Manufacturer Roche Instrument Center AG
CH-6343 Rotkreuz
Switzerland
Distributed by
Roche Diagnostics
Indianapolis, IN 46256 USA
(For Technical Assistance call the
Roche Response Center
toll-free 1-800 526 1247)
Roche Diagnostics
H7V 4A2 Laval, Quebec
(For Technical Assistance call:
Pour toute assistance technique,
appeler le: 1-877 273 3433)
Roche Diagnostics (Schweiz) AG
CH-6343 Rotkreuz
Roche Diagnostics
F-38240 Meylan
Roche Diagnostics GmbH
D-68298 Mannheim, Germany
Distributore in Italia:
Roche Diagnostics SpA
Piazza Durante 11
I-20131 Milano
Roche Diagnostics S.L.
E-08006 Barcelona
Distribuidor em Portugal:
Roche Farmacêutica Química, Lda
P-2700 Amadora
l
Declaration of Conformity
The cobas s 201 System Docking Station complies with the requirements
of the Directive 98/79/EC of the European Parliament and of the Council
of 27 October 1998 on In-vitro Diagnostic Medical Devices.
Preface
04/2006, version 1.0 0.5
Questions or comments regarding the contents of this manual can be
directed to the address below or to your Roche representative.
Roche Molecular Systems, Inc.
4300 Hacienda Drive
Pleasanton, California 94588-2722
ROCHE, AMPERASE, AMPLILINK, COBAS, AMPLIPREP and
TAQMAN are trademarks of Roche.
ROCHE RESPONSE CENTER is a servicemark of Roche.
Microlab is a registered trademark of Hamilton Company.
Microsoft, Windows, and Windows XP are either registered trademarks
or trademarks of Microsoft Corporation in the United States and/or other
countries.
AMPLILINK Software:
The Source Code version of the executable code (the "Software") is
available under the terms of the Interbase Public License Version 1.0. The
user may obtain a copy of the license at:
http://interbase.com/IPL.html.
The Software is offered solely under the terms of the Interbase Public
License Version 1.0.
Copyright © 2006, Roche Molecular Systems, Inc. All rights reserved.
0.6 04/2006, version 1.0
How to Use this Supplement
This supplement is to be used in conjunction with the cobas s 201 System
Operator’s Manual.
Before using the Docking Station, it is essential that the Warnings,
Cautions, and Safety Requirements that are contained in that manual and
in this supplement are read and understood by the user.
Refer to the section titled Symbols on the following page for a description
of the warning symbols and caution symbols that appear in this
supplement.
This supplement contains the following sections:
Preface (Preface) summarizes safety precautions necessary when
operating the system.
Section 1 (Workflow) outlines the procedures that are different
during routine operation of the cobas
s 201 with the
Docking Station.
Section 2 (Docking Station) describes the major components of the
Docking Station.
Preface
04/2006, version 1.0 0.7
Conventions Used in This Supplement
Text Conventions To convey information readily and consistently, certain text conventions
are used throughout this supplement. These text conventions are as
follows:
Text convention Used for
Numbered lists Numbered procedures that should be
implemented in order during an operation:
1 Turn power switch on.
2 Click on File.
Bold typeface, lower case Emphasis of any term or component that is
being described:
A profile is a user-defined series of tests to
be performed.
Italics Designating another section of this
Operator’s Manual that should be consulted:
Details concerning order entry are outlined
in the Orders section of this Manual.
0.8 04/2006, version 1.0
Symbols Certain symbols are used throughout this supplement to provide a ready
visual reference. These symbols are as follows:
Symbol Used for
Information note. Designates a note that
provides additional information concerning
the current topic or procedure.
Important note. Designates an important
note that must be reviewed and understood.
Danger. Indicates a direct danger that, if not
avoided, may result in death or serious
injury.
Warning. Indicates a possible hazardous
situation that, if not avoided, may result in
death or serious injury.
Caution. Indicates a possible hazardous
situation which, if not avoided, may result in
personal injury and/or in damage to the
system.
Warning Potential Biohazardous Material.
Indicates a possible hazardous situation due
to biohazardous material that may result in
death or serious injury.
Warning Laser Transmitter. Designates the
presence of a laser transmitter. Do not stare
into the laser transmitter.
Warning Moving Parts. Indicates a possible
hazardous situation. Keep hands away from
moving parts.
Preface
04/2006, version 1.0 0.9
The following symbols appear on the instrument type plate.
Electrical and electronic equipment marked
with this symbol are covered by the
European directive WEEE. This symbol
denotes that the equipment must not be
disposed of in the municipal waste system.
(WEEE) Directive 2002/96/EC of the
European Parliment and the Council of 27
January 2003 on waste electrical and
electronic equipment
Danger Area. The triangular warning
symbol on the back of the analyzer is a
general reminder that users read the safety
information contained in this manual. Users
must be able to identify specific hazards and
take appropriate action to avoid them.
Symbol Used for
Caution: Refer to accompanying documents.
Manufacturer of device.
0.10 04/2006, version 1.0
Glossary
The terminology that is used in this supplement is generally the same as
that used in clinical laboratories. Certain specific terms, however, are used
for description of an operation or a component.
Refer to the cobas s 201 System Operator’s Manual for a summary
of terms that are used with this system and their definitions.
A summary of the terms used in this Supplement and their definitions is
provided below:
K-carrier Reusable metallic holder that accommodates
up to 24 K-tubes that contain PCR-ready
samples for analysis on the COBAS
®
TaqMan
®
Analyzer.
K-tube Kinetic tube. The disposable sample tube that
is used to contain a prepared sample. K-tubes
come prepackaged in a covered K-tube rack.
Multicarrier Holds a K-carrier for transport from the
COBAS
®
AmpliPrep Instrument to the
COBAS TaqMan Analyzer by the Docking
Station and vice versa.
SK24 rack Reusable rack that holds S-tubes or K-tubes
in barcode clips. The SK24 rack loads and
unloads samples throughout the pipetting,
extraction, and amplification/detection
process.
Thermal Cycler Programmable temperature block in the
COBAS TaqMan Analyzer. The thermal
cycler rapidly changes temperature according
to the amplification profile defined in the test
definition files.
Preface
04/2006, version 1.0 0.11
Units
The following measurement abbreviations are used:
°C Degrees Centigrade
°F Degrees Fahrenheit
CAN Canada.
CE Conformité Européenne
cm Centimeters
CSA Canadian Standards Association
EN European Normative
FCC Federal Communications Commission
IEC International Electrotechnical Commission
in Inches
kg Kilogram
lb Pounds (weight)
LED Light Emitting Diode
μL microliters(10
-6
liters)
UL Underwriters Laboratories (USA)
VA Volt-Ampères (power consumption)
VAC Volts (alternating current)
0.12 04/2006, version 1.0
Operational Precautions and Limitations
Safety Requirements
The safety precautions that are necessary when installing, operating
and servicing the Docking Station are summarized in the following
sections of this manual. It is important that you carefully read and
understand the statements regarding safety contained in these sections.
This information should also be made available for new employees and
for future reference.
Disposal Recommendations
All electrical and electronic products should be disposed of separately
from the municipal waste system. Proper disposal of your old appliance
prevents potential negative consequences for the environment and
human health.
Disposal of the Instrument
The instrument must be treated as biologically contaminated-
hazardous waste. Decontamination (i.e., a combination of processes,
including cleaning, disinfection and/or sterilization) is required before
reuse, recycling or disposal.
Dispose the instrument according to local and/or labor regulations. For
more information, contact your local Roche Support personnel.
Disposal of Control Unit Components
Components of your control unit such as the computer, monitor,
keyboard, etc. which are marked with the crossed-out wheeled bin
symbol are covered by the European Directive 2002/96/EC (WEEE).
These items must be disposed of via designated collection facilities
appointed by government or local authorities.
For more information about disposal of your old product, please
contact your city office, waste disposal service or your local Roche
Support personnel.
Constraint
It is left to the responsible laboratory organization to determine
whether control unit components are contaminated or not. If
contaminated, treat in the same way as the instrument.
Preface
04/2006, version 1.0 0.13
Warnings
Intended Use
The Docking Station connects the COBAS AmpliPrep Instrument to the
COBAS TaqMan Analyzer and fully automates sample preparation,
amplification, and detection.
This supplement contains instructions for using the Docking Station with
the cobas s 201 system. This supplement should be used in conjunction
with the cobas s 201 System Operator’s Manual.
WARNING
For service, contact your Roche Service representative, who is trained
and qualified to perform the necessary servicing.
WARNING
Never open the Docking Station while a run is in progress. Failure to
follow this procedure can result in potential injury to the operator.
WARNING
After the Docking Station is installed by a Roche Service representative,
the COBAS AmpliPrep Instrument and the COBAS TaqMan Analyzer
must not be moved. Movement of the docked system can cause the
Docking Station to become non-operational.
Electrical safety
The Docking Station is designed in accordance with safety standard
EN/IEC 61010-1. Grounding of the instrument and those surfaces the
user can come into contact with is provided by a grounded cable in
accordance with Protection Class I (IEC).
Users should not perform any servicing.
0.14 04/2006, version 1.0
04/2006, version 1.0 1.1
Workflow 1
Overview
The modified workflow for processing pooled samples on a COBAS
®
AmpliPrep Instrument that is linked to a COBAS
®
TaqMan
®
Analyzer by
means of the Docking Station is shown in
Figure 1.1.
Workflow procedures that are described in this supplement replace
the Extraction, Amplification, and Detection workflow procedures
described in section 2 (Quick Start Guide: Workflow) of the cobas s
201 System Operator’s Manual when a Docking Station is used.
The COBAS AmpliPrep Instrument and the COBAS TaqMan
Analyzer must be connected to the same Data Station for
AMPLILINK software.
Figure 1.1
Extraction, Amplification, Detection Workflow when Using the Docking Station
1.2 04/2006, version 1.0
Perform Startup Procedures
The Data Station for AMPLILINK software is normally left on,
except for a daily restart. If the Data Station is powered off, refer to
the cobas s 201 System Operator’s Manual for startup procedures.
COBAS AmpliPrep Instrument
When using the Docking Station, an empty K-carrier must be
present in each of the two K-carrier park positions of the COBAS
AmpliPrep Instrument. Refer to
page 2.5 for the procedure to place
K-carriers in these positions.
The COBAS AmpliPrep Instrument is normally left in Standby
mode. If the instrument is powered off, refer to the cobas s 201
System Operator’s Manual for startup procedures.
1 Check the status of the wash reagent reservoir. Replace if necessary.
A reservoir icon on the Systems window System tab is red if the
corresponding reservoir is empty.
2 Empty the waste container if it requires emptying.
The waste icon on the Systems window System tab is red if the
waste level is too high.
3 Perform any maintenance that is listed in the Systems window Service Due
tab.
The background on the Systems icon turns yellow and an oil can
appears on the icon when service is due. Perform the
required service before starting the next run.
The system is automatically primed after each service is performed.
Workflow
04/2006, version 1.0 1.3
COBAS TaqMan Analyzer
When using the Docking Station, four empty K-carriers must be
present in positions 9 to 12 of the Reload Area of the COBAS
TaqMan Analyzer. Refer to
page 2.6 for the procedure to place
K-carriers in these positions.
The COBAS TaqMan Analyzer is normally left in Standby mode. If
the analyzer is powered off, refer to the cobas s 201 System
Operator’s Manual for startup procedures.
1 Place 24 empty and capped K-tubes into a K-carrier, and then place the
K-carrier in park position 2 on the left side of the COBAS TaqMan
Analyzer (
page 2.6).
The analyzer must be powered off to perform this step.
If a batch contains fewer than six K-tubes, the COBAS TaqMan
Analyzer uses these empty K-tubes to fill positions 7, 8, 12, 13, 17,
and 18 in the K-carrier. These positions must be occupied to
balance the K-carrier in the thermal cycler.
2 Perform any maintenance that is listed in the Systems window Service Due
tab before starting the next run.
The background on the Systems icon turns yellow and an oil can
appears on the icon when service is due. Perform the
required service before starting the next run.
1.4 04/2006, version 1.0
Load Reagents
The reagents that are required depend upon the cassette status check and
the daily workload.
Reagents should be removed from the refrigerator prior to use.
Reagents should be loaded onto the COBAS AmpliPrep
Instrument immediately after they are removed from storage to
avoid condensation on the barcode labels. Allow the reagents to
equilibrate on the instrument for at least thirty minutes before the
first sample is to be processed.
Do not pool reagents from different lots or from different bottles of
the same lot.
Do not mix reagents (including cassettes) from different kits.
Do not use more than one lot number of reagents for a processing
run on the COBAS AmpliPrep Instrument. A single lot of reagents
must be used for processing all specimens and controls in a batch.
Do not open COBAS AmpliPrep Instrument cassettes and
exchange, mix, remove, or add bottles.
Do not use a kit after its expiration date.
Do not use a kit on more than one COBAS AmpliPrep Instrument.
Do not freeze reagents.
Never attempt to remove a rack when the LED for that rack is red.
1 Obtain the magnetic glass particle (MGP) cassettes that are required.
Reloading of MGP cassettes is not possible during a run due to the
shaking of the rack. The rack containing the MGP cassette is
unlocked after no more tests remain for that cassette, and a new
MGP cassette can be loaded at that time.
2 Place the MGP cassettes on a single reagent rack.
3 Obtain the other cassettes that are required.
4 Place these cassettes on other (up to four) reagent racks.
Other reagent cassettes that are required to complete the orders in
the worklist can be loaded on up to four additional reagent racks.
These reagent racks are loaded in reagent rack positions B through
E, which are not shaken.
Workflow
04/2006, version 1.0 1.5
More than one cassette may be needed for a particular reagent to
complete a run. When the first cassette is depleted, the system will
automatically switch to the next cassette.
5 Open the load panel.
6 Slide the reagent rack containing the magnetic glass particle cassettes onto
rack position A. Slide the rack up to the first pin.
7 Wait for the reagent rack LED to begin flashing green. Then continue
sliding the rack to the back of the instrument. The LED turns orange.
8 Load the other reagent racks on rack positions B through E. When a
complete kit of reagent cassettes has been loaded, all of the reagent rack
LEDs turn solid green.
The COBAS AmpliPrep Instrument scans the reagent rack barcode label
and the reagent cassette barcode labels as each rack is loaded onto the
instrument.
The AMPLILINK software updates the Systems window Cassettes/Samples
tab with the scanned information.
Reagents are supplied in a kit package that includes eight cassettes.
Reagent cassettes from the same kit package must all be loaded
together. If any cassette is missing the remaining cassettes for that
assay are Blocked.
9 Close the load panel.
WARNING
Transfer Head 1 contains a class 1 laser diode. Do not stare into the
laser transmitter.
Figure 1.2
Rack Platform
Shaker Rack
Reagent Racks
SK24 Racks
(A)
(B-E)
(F-H)
1.6 04/2006, version 1.0
Load Consumables
Determine the load requirements. One SPU, one K-tube, and one K-tip
are needed for each S-tube order.
SPUs
1 Obtain the required number of SPUs (up to three full SPU racks can be
loaded).
2 Remove the SPUs from their packaging. Inspect each SPU for defects.
3 Place the SPUs in the SPU racks.
Always make sure that the tips are properly seated. Press down on
the SPU splash guard (not the sample tips) to properly seat the SPU
in the rack. Make sure the sample tips are even.
4 Open the load panel.
5 Position the end of a SPU rack at the front of one of the SPU load
positions. Then slide the SPU rack towards the rear of the instrument
onto the load positions J, K, or L. When the rack is loaded properly, the
SPU LED is green.
When positioning the SPU rack, be sure that the single wells of the
SPU rack are to the right, that the left side of the SPU rack is under
the lip, and that all SPUs are seated properly.
The COBAS AmpliPrep Instrument counts the SPUs and the
Systems window System tab updates supply status as each SPU rack
is loaded onto the system.
Figure 1.3
Filling a SPU Rack
Double Wells
Single Wells
Sample Tip
Splash Guard
WARNING
Transfer Head 1 contains a class 1 laser diode. Do not stare into the
laser transmitter.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36

Roche cobas s 201 system User manual

Type
User manual

The Roche cobas s 201 system is a fully automated sample-to-result system for molecular diagnostic testing, designed to provide fast and accurate results for a variety of tests, including infectious disease testing, genetic testing, and oncology testing. With its ability to process up to 96 samples in a single run and deliver results in as little as 3 hours, the cobas s 201 system can help laboratories improve efficiency and reduce turnaround time for molecular testing.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI