AMPLILINK Software
Application Manual
Version 1.0
Software Version 3.4 Series
AMPLILINK Software
Software Version 3.4 Series
Roche Diagnostics
2 Application Manual · Version 1.0
Publication information
Edition notice This publication is intended for operators of the AMPLILINK Software Version 3.4
Series.
Some of the tests identified for use with the AMPLILINK Software Version 3.4 Series,
may not be approved for use in all countries. Contact a Roche representative for tests
that are commercially available in your country.
Every effort has been made to ensure that all the information contained in this
manual is correct at the time of publishing. However, Roche Molecular Systems, Inc.
may need to update the manual information as output of product surveillance
activities, leading to a new version of this manual.
CAUTION
General attention
To avoid incorrect results, ensure that you are familiar with the instructions and safety
information.
r Pay particular attention to all safety notices.
r Always follow the instructions in this publication.
r Do not use the software in a way that is not described in this publication.
r Store all publications in a safe and easily retrievable place.
Training Do not carry out operation tasks or maintenance actions unless you have received
training from Roche Diagnostics. Leave tasks that are not described in the user
documentation to trained Roche Service representatives.
Screenshots The screenshots in this publication have been added exclusively for illustration
purposes. Configurable and variable data, such as tests, results, or path names visible
therein must not be used for laboratory purposes.
Warranty Any customer modification to the system renders the warranty or service agreement
null and void.
For conditions of warranty, contact your local sales representative or refer to your
warranty contract partner.
Always leave software updates to a Roche Service representative or perform such
updates with their assistance.
Copyright © 2001-2015, F. Hoffmann-La Roche Ltd. All rights reserved.
Trademarks The following trademarks are acknowledged:
AMPLILINK, COBAS, COBAS P, AMPLIPREP, and TAQMAN are trademarks of
Roche.
All other trademarks are the property of their respective owners.
Publication
version
Software version Revision date Change description
1.0 3.4 June 2015 This is the first version of the AMPLILINK Software Version 3.4
Series Application Manual. The changes in the Application
Manual reflect the update from Software Version 3.3 Series to
Software Version 3.4 Series.
Table 1 Revision history
Roche Diagnostics
Application Manual · Version 1.0 3
AMPLILINK Software
Software Version 3.4 Series
Open source license notification
and licenses
The AMPLILINK Software might contain components or modules that are open
source or commercial software programs. For copyright, other notices, and licensing
information regarding such software programs included with the AMPLILINK
Software please see the CD/DVD provided with the product or the folder “Licensing
Information” located on the hard disk drive D of the AMPLILINK control unit.
Feedback Every effort has been made to ensure that this publication fulfills the intended use. All
feedback on any aspect of this publication is welcome and is considered during
updates. Contact your Roche representative, should you have any such feedback.
Approvals The AMPLILINK Software meets the requirements of the Directive 98/79/EC.
Contact addresses
Complies with the IVD directive 98/79/EC.
For in vitro diagnostic use.
Roche Molecular Systems, Inc.
1080 US Highway 202 South
Branchburg, NJ 08876
USA
Made in Switzerland
Roche Diagnostics GmbH
Sandhofer Strasse 116
68305 Mannheim
Germany
AMPLILINK Software
Software Version 3.4 Series
Roche Diagnostics
4 Application Manual · Version 1.0
Roche Diagnostics
Application Manual · Version 1.0 5
AMPLILINK Software
Software Version 3.4 Series
Table of contents
Publication information 2
Contact addresses 3
Table of contents 5
Intended use 7
Compatibility 7
How to use this publication 7
Symbols and abbreviations 8
What is new in version 1.0 of the publication
for AMPLILINK Software version 3.4 10
System description
1 General safety information
Safety classifications 17
Safety precautions 18
Safety summary 19
2 Introduction
System overview 25
Client-server architecture 26
Primary tube pipetting 27
Multi-result tests 29
About patient IDs, order numbers, and
sample IDs 30
3 Software
Introduction to the AMPLILINK Software 35
Working with lists 45
Working with the mouse and the keyboard 55
Using Help 57
Operation
4 Workflows
Introduction 69
Safety information 71
cobas p 630 instrument 72
COBAS
®
AmpliPrep Instrument -
COBAS
®
TaqMan
®
Analyzer 80
COBAS
®
AmpliPrep Instrument -
COBAS
®
TaqMan
®
48 Analyzer 90
COBAS
®
AmpliPrep Instrument - Offline
amplification and detection 100
Manual preparation - COBAS
®
TaqMan
®
48
Analyzer 106
Stop an instrument or analyzer during a run 113
5 Operation
Safety information 119
Starting up, shutting down, and restarting 120
System handling 123
Order handling 125
Result handling 153
Report handling 167
Operator administration 168
AMPLILINK Monitor 170
AMPLILINK Archive Viewer 172
Maintenance
6 General maintenance
Managing service actions 183
Troubleshooting
7 Flags
General flag description 191
List of all AMPLILINK Software flags 196
cobas p 630 instrument flags 200
COBAS
®
AmpliPrep Instrument flag remarks 203
COBAS
®
TaqMan
®
Analyzer and
COBAS
®
TaqMan
®
48 Analyzer flags with flag
remark 210
COBAS
®
TaqMan
®
Analyzer and
COBAS
®
TaqMan
®
48 Analyzer flags without
flag remark 224
AMPLILINK Software flag remarks 235
8 General troubleshooting
Troubleshooting 239
List of messages 244
Software description
9 General
Logon to the AMPLILINK Software 263
Menu description 265
AMPLILINK Information (dialog box) 274
Unsupported characters 275
10 Messages
Messages window 279
Tasks in Messages window 285
11 Overview
Overview window 289
Tasks in Overview window 296
Roche Diagnostics
6 Application Manual · Version 1.0
AMPLILINK Software
Software Version 3.4 Series
12 Systems
Systems window 299
COBAS
®
AmpliPrep Instrument 301
COBAS
®
TaqMan
®
Analyzer 310
COBAS
®
TaqMan
®
48 Analyzer 315
Run Log tab 320
Service Due tab 322
Service All tab 324
Service Log tab 326
Counters: Instrument (dialog box) 327
Firmware Information (dialog box) 328
Tasks in Systems window 329
13 Orders
Orders window 335
Order origin 350
Order status 351
Rack order status 352
Sample status 353
Test type 354
14 Results
Results window 357
15 Configuration
Configuration window 373
General Configuration 376
Maintenance Configuration 378
Operator Access Configuration 387
Barcode Configuration 391
Systems configuration 393
Lab Configuration 395
Contacts Configuration 400
LIS Configuration 402
Report Layout Configuration 412
Processing Configuration 420
Test Definitions Configuration 424
Export Definitions 428
Windows configuration 434
Tasks in Configuration window 442
16 Memo Pad
Memo Pad window 455
Glossary
17 Glossary
Index
Index 465
Revisions
Roche Diagnostics
Application Manual · Version 1.0 7
AMPLILINK Software
Software Version 3.4 Series
Intended use
As a component of the COBAS
®
AmpliPrep Instrument/COBAS
®
TaqMan
®
Analyzer
2
nd
generation system, the AMPLILINK Software is used to integrate all instruments
and analyzers offering a single user interface for PCR testing. AMPLILINK Software
supports various instrument configurations, data management, and analysis of test
results from clinical specimens using PCR technology.
Test results can be automatically transferred to an external host.
The AMPLILINK Software is setup as a client/server architecture. This allows the use
of several decentralized AMPLILINK control units each controlling individual
instruments and/or analyzers.
The AMPLILINK Software also handles the test definition file (TDF) for each assay it
supports, captures sample identification, absorbance results, and interpretations and
comments from the data output for the instruments it supports and presents the data
in the user interface.
Compatibility
AMPLILINK Software Software Version 3.4 Series is compatible with the following
control units:
o rp5700
o rp5800
The following control units are not supported:
o VL420
o d530
o Evo 510
o dc7100
o dc7600
How to use this publication
To help you to find information quickly, there is a table of contents at the beginning
of the book and at the beginning of every chapter. In addition, a complete index and a
glossary of terms can be found at the end.
This manual is organized to reflect the major functions supported by the
AMPLILINK Software. It contains the following chapters.
Preface Lists the conventions, symbols, and units used in the manual.
Chapter 1 General safety
information
Provides safety information.
u Chapter General safety information
AMPLILINK Software
Software Version 3.4 Series
Roche Diagnostics
8 Application Manual · Version 1.0
Chapter 2 Introduction Gives an introduction to the AMPLILINK Software and introduces some basic
concepts.
u Chapter Introduction
Chapter 3 Software Describes how to use the AMPLILINK Software.
u Chapter Software
Chapter 4 Workflows Provides instructions for routine operation of the cobas p 630 instrument,
COBAS
®
AmpliPrep Instrument, COBAS
®
TaqMan
®
Analyzer, and
COBAS
®
TaqMan
®
48 Analyzer.
u Chapter Workflows
Chapter 5 Operation Provides detailed instructions on startup, and shutdown, orders handling, results
handling, and using the AMPLILINK Archive Viewer.
u Chapter Operation
Chapter 6 General maintenance Provides instructions on managing service actions.
u Chapter General maintenance
Chapter 7 Flags Explains the flags and gives recommended actions.
u Chapter Flags
Chapter 8 General troubleshooting Describes the procedures for resolving problems during operation.
u Chapter General troubleshooting
Chapters 9-16 Software
description
Describes the AMPLILINK Software in detail.
u Chapter General
Chapter Messages
Chapter Overview
Chapter Systems
Chapter Orders
Chapter Results
Chapter Configuration
Chapter Memo Pad
Symbols and abbreviations
Symbols used in the publication
Symbol Explanation
p Start of procedure
s End of procedure
o List item
u Cross-reference
q Tip
Danger alert symbol
Table 2 Symbols used in the publication
Roche Diagnostics
Application Manual · Version 1.0 9
AMPLILINK Software
Software Version 3.4 Series
Abbreviations The following abbreviations are used:
Warning alert symbol
Caution and general safety alert symbol
Electrical and electronic equipment marked with this symbol are
covered by the European directive WEEE.
The symbol denotes that the equipment must not be disposed of in
the municipal waste system.
Symbol Explanation
Table 2 Symbols used in the publication
Abbreviation Definition
AFI Absolute Fluorescence Increase
BLS Baseline Subtraction Elbow (algorithm)
BOD Begin of Day
cm Centimeters
C/mL
cp/mL
Copies per milliliter
DNA Deoxyribonucleic Acid
DP Detection Position
DSLM Double Sigmoid Levenberg Marquardt (flag group)
EOD End of Day
GZ Grey Zone
Hz Hertz
IC Internal Control
in Inches (2.54 cm)
IU/mL International Units per milliliter
kg Kilogram
lb Pounds (weight)
LAN Local Area Network
LIS Laboratory Information System
mL
milliliters (10-
3
liters)
MGP Magnetic Glass Particle
NEG Negative
OD Optical Density
POS Positive
QS Quantitation Standard
RFI Relative Fluorescence Increase
SPU Sample Processing Unit
TC Thermal Cycler
TDF Test Definition File
Table 3 Abbreviations
AMPLILINK Software
Software Version 3.4 Series
Roche Diagnostics
10 Application Manual · Version 1.0
What is new in version 1.0 of the publication for AMPLILINK
Software version 3.4
Changes in the front matter and
preface
Front matter and preface have been updated according to the new Roche Diagnostics
standards.
General safety information o Section Software and data security was updated to reflect the latest Roche
Diagnostics standards.
u For details, see Software and data security (p. 21)
o Firewall recommendation (FortiGate-40C firewall) added.
u For details, see Software and data security (p. 21)
Control unit o Compatibility list for control units updated.
u For details, see Compatibility (p. 7)
o Term "data station" changed to "control unit".
Data encryption and data
masking
Information about data encryption and data masking added.
u For details, see Data encryption and data masking (p. 26)
Barcode clips Notes about the proper insertion of the barcode clips into the sample rack positions
have been added. A clicking sound indicates that the barcode clip snaps into place.
Startup procedure Further details about the start up procedure have been added.
u For details, see Perform start up procedures (p. 81) and Starting up, shutting down, and
restarting (p. 120)
Patient first and last name It is not possible to enter more than 30 characters for the patient’s last and first name.
Sample ID The length of the sample ID is limited to 20 characters.
Result upload Only reviewed results (results that are either accepted or rejected) can be uploaded to
the LIS. It is not possible to upload results before the Operator has reviewed them.
The Result Upload dialog box was restructured and has new options about partial
result upload for CT, RFI, and AFI values for COBAS
®
TaqMan
®
Analyzer and
COBAS
®
TaqMan
®
48 Analyzer results.
u For details, see Result Upload tab (p. 404)
TNAI Total Nucleic Acid Isolation
UPS Uninterrupted power supply
XML Extensible Markup Language
Abbreviation Definition
Table 3 Abbreviations
Roche Diagnostics
Application Manual · Version 1.0 11
AMPLILINK Software
Software Version 3.4 Series
General changes in the document o Any reference to the COBAS® AMPLICOR® analyzer removed.
o Any reference to the sample preparation methods (Generic and Probe) removed.
o The Microsoft Windows operating system version is no longer specified.
o Screenshots updated where necessary.
o Any reference to cobas Link (TeleService-Net) removed.
AMPLILINK Software
Software Version 3.4 Series
Roche Diagnostics
12 Application Manual · Version 1.0
System description
1 General safety information.............................................................................................................. 15
2 Introduction...................................................................................................................................... 23
3 Software............................................................................................................................................ 33
Roche Diagnostics
Application Manual · Version 1.0 15
AMPLILINK Software 1 General safety information
Table of contents Software Version 3.4 Series
General safety information 1
This chapter provides general safety informations for working with the AMPLILINK
Software. Specific safety informations are provided in the manual where required.
For specific safety informations refer also to the Instrument Manuals for
COBAS
®
AmpliPrep Instrument, COBAS
®
TaqMan
®
Analyzer, and
COBAS
®
TaqMan
®
48 Analyzer, the cobas p 630 instrument Operator’s Manual, and
the assay-specific package inserts and Method Manuals.
In this chapter
Chapter
1
Safety classifications.............................................................................................................17
Safety precautions.................................................................................................................18
Operator qualification...................................................................................................18
Safety summary ....................................................................................................................19
Warning messages..........................................................................................................19
Biohazardous materials ...........................................................................................19
Caution messages...........................................................................................................19
Reagents and other working solutions ..................................................................20
Samples and sample racks.......................................................................................20
Cross contamination of samples ............................................................................20
Notices .............................................................................................................................21
Software and data security......................................................................................21
Windows display settings warning ........................................................................22
Roche Diagnostics
16 Application Manual · Version 1.0
1 General safety information AMPLILINK Software
Table of contents Software Version 3.4 Series
Roche Diagnostics
Application Manual · Version 1.0 17
AMPLILINK Software 1 General safety information
Software Version 3.4 Series Safety classifications
Safety classifications
This section explains how precautionary information is presented in this manual.
The safety precautions and important user notes are classified according to the
ANSI Z535.6 Standard. Familiarize yourself with the following icons and their
meanings:
Generic safety precaution
r The safety alert symbol by itself without a signal word is used to promote awareness to
hazards which are generic or to direct the reader to safety information provided
elsewhere in the document.
These symbols and signal words are used for specific hazards:
WARNING
WARNING
r Indicates a hazardous situation which, if not avoided, could result in death or serious
injury.
CAUTION
CAUTION
r Indicates a hazardous situation which, if not avoided, could result in minor or moderate
injury.
NOTICE
NOTICE
r Indicates a hazardous situation which, if not avoided, could result in damage to
equipment.
Important information which is not safety relevant is indicated by the following
symbol:
q
Tip
Indicates additional information on correct use of the software, instruments, and analyzers
and gives useful tips.
Roche Diagnostics
18 Application Manual · Version 1.0
1 General safety information AMPLILINK Software
Safety precautions Software Version 3.4 Series
Safety precautions
Generic safety precaution
r Particular attention must be paid to the following safety precautions. If these safety
precautions are ignored, the operator may suffer serious or fatal injury. Each
precaution is important.
Operator qualification
Operators are required to have a sound knowledge of relevant guidelines and
standards as well as the information and procedures contained in all provided
manuals.
o Do not carry out operation and maintenance unless you have been trained by
Roche Diagnostics.
o Carefully follow the procedures specified in the provided manuals for the
operation of the software.
o Follow standard laboratory practices, especially when working with biohazardous
material.
Roche Diagnostics
Application Manual · Version 1.0 19
AMPLILINK Software 1 General safety information
Software Version 3.4 Series Safety summary
Safety summary
This safety summary contains the most important general warning and caution
messages for working with the AMPLILINK Software. Additionally, you will find
specific safety information at the beginning of Chapter Workflows and Chapter
Operation.
For specific safety informations refer also to the Instrument Manuals for
COBAS
®
AmpliPrep Instrument, COBAS
®
TaqMan
®
Analyzer, and
COBAS
®
TaqMan
®
48 Analyzer, the cobas p 630 instrument Operator’s Manual, and
the assay-specific package inserts and Method Manuals.
Warning messages
WARNING
List of warning messages
r Before operating the software, read the warning messages contained in this summary
carefully. Failure to observe them may result in death or serious injury.
Biohazardous materials
WARNING
Infection by samples and associated materials
r Contact with samples containing material of human origin may result in infection. All
materials and mechanical components associated with samples of human origin are
potentially biohazardous.
o Follow standard laboratory practices, especially when working with biohazardous
material.
o Be sure to wear appropriate protective equipment, including, but not limited to,
safety glasses with side shields, fluid resistant lab coat, and approved disposable
gloves.
o Wear a face shield if there is a chance of splash or splatter.
o If any biohazardous material is spilled, wipe it up immediately and apply
disinfectant.
o If sample or waste solution comes into contact with your skin, apply immediately a
disinfectant and then wash it off with soap and water. Consult a physician.
Caution messages
CAUTION
List of caution messages
r Before operating the instrument, read the caution messages contained in this summary
carefully. Failure to observe them may result in minor or moderate injury.
Roche Diagnostics
20 Application Manual · Version 1.0
1 General safety information AMPLILINK Software
Safety summary Software Version 3.4 Series
Reagents and other working solutions
WARNING
Incorrect results due to incorrect reagent volume
r Application faults may cause an undetectable loss of reagent.
o Always store reagents according to specified storage conditions.
o Reagents, once used, should not be used on other COBAS
®
AmpliPrep Instruments
which are not connected to the same AMPLILINK system (a system is an
AMPLILINK client-server installation and all connected instruments and analyzers).
WARNING
Incorrect results due to expired reagents or mixing kits
r Data obtained using expired reagents are not reliable. Reagents are supplied in a kit
package. Mixing reagents from different kits may also lead to incorrect results.
o Do not use reagents that have exceeded their expiration dates. Replace expired
reagent with unexpired reagents before sample processing.
o Do not mix kits. Reagent cassettes from the same kit package must all be loaded
together.
WARNING
Incorrect results due to reagent cassette interchange
r Reagent cassettes for the COBAS
®
AmpliPrep Instrument connected to the same
AMPLILINK system can be interchanged (a system is an AMPLILINK client-server
installation and all connected instruments and analyzers). But using a
COBAS
®
AmpliPrep Instrument cassette already used on a COBAS
®
AmpliPrep
Instrument that is connected to a different AMPLILINK system may lead to incorrect
results.
Samples and sample racks
WARNING
Incorrect results due to sample mismatch
r Manual mismatch of samples or sample IDs may lead to incorrect results. Whenever
possible use a barcode scanner or an LIS for input of sample IDs to minimize manual
input. Cross-check each manual input of sample IDs to minimize manual sample
mismatch.
WARNING
Incorrect results due to sample rack interchange
r Using a sample rack from a different AMPLILINK system may lead to incorrect results
(a system is an AMPLILINK client-server installation and all connected instruments and
analyzers).
Do not interchange sample racks among different AMPLILINK systems. Each sample
rack may be used on one individual AMPLILINK system only. Each sample rack has its
own individual rack ID, which the instrument or analyzer uses for identification
purposes.
If more than one AMPLILINK system is used in the same lab, be sure to dedicate each
sample rack to one single, individual AMPLILINK system (e.g. by using individual
sample rack ID ranges for each AMPLILINK system).
Cross contamination of samples
WARNING
Incorrect results due to carryover
r Traces of analytes or reagents may be carried over from one test to the next. Take
adequate measures to safeguard additional testing and to avoid potentially false
results.
/