Steris Amsco Rigid Sterilization Container System Operating instructions

Type
Operating instructions
i
A Word from STERIS Operator Manual 920003-690
This manual contains important information on the proper use and
maintenance of the Rigid MIS Instrument Rack. All personnel
involved in the use and maintenance of this rack must carefully
review and comply with the Safety Precautions and instructions
contained in this manual. These instructions are important to
protect the health and safety of personnel using this equipment and
should be retained in a conveniently accessible area for quick
reference.
Advisory IMPORTANT: A listing of the Safety Precautions to be observed
when using and servicing this equipment can be found in SECTION 1
of this manual. Do not use the equipment until you have become
familiar with this information.
STERIS does not intend, recommend nor represent in any way the
Rigid MIS Instrument Rack be used for the terminal disinfection or
sterilization of any regulated medical device. This rack is intended
only to perform an initial step in the processing of soiled, reusable
medical devices.
If medical devices will be contacting blood or compromised tissues,
such devices must be terminally processed in accordance with
Good Hospital Practices before each use in human patients.
STERIS does not recommend processing flexible endoscopes,
single-use instruments, rigid telescopes (optical lens), fiber-optic
cables, multi-part instruments that cannot be disassembled, heat-
sensitive devices and coagulating electrodes in the Rigid MIS
Instrument Rack. Processing these types of instruments may result in
instrument or equipment damage as well as improperly cleaned
items. All items treated in this rack must stand temperatures ranging
from 140 to 240°F [60 to 116°C].
The Rigid MIS Instrument Rack is specifically designed to only
process goods as outlined in this manual. If there is any doubt about
a specific material or product, contact the instrument manufacturer
for recommended washing techniques.
Indications For Use The Rigid MIS Instrument Rack is designed to be used in the washer/
disinfector for the efficient cleaning, rinsing, drying and low-level
thermal disinfection of rigid minimally invasive surgical instruments.
IMPORTANT: The Rigid MIS Instrument Rack is designed to be used
only in conjunction with the RIGID MIS cycle in the Reliance® 444
Single-Chamber Washer/Disinfector equipped with the Enzyme
treatment option or in conjunction with the Enzyme treatment option
in the Reliance® Synergy® Washer/Disinfector and Reliance® Genfore®
Washer/Disinfector.
©2022, STERIS Corporation. All rights reserved. Printed in Canada.
A WORD FROM STERIS
ii
920003-690 Operator Manual A Word from STERIS
Any alteration of this equipment not authorized or performed by
STERIS will void the warranty. Alteration of this equipment which
could adversely affect its operation and efficacy may violate national,
state and local laws or regulations.
Trademarks Please take note of the following registered trademarks that apply to
the wording used throughout the manual:
• Reliance® 444 Single-Chamber Washer/Disinfector
• Reliance® Synergy® Washer/Disinfector
• Reliance® Genfore® Washer/Disinfector
• Reliance® Load/Unload Modules,
• Reliance® ATS Automated Transport System
•Hydex
® is a registered trademark of A. L. Hyde Company.
•Teflon
® is a registered trademark of E. I. du Pont de Nemours
and Company.
Service Information A thorough preventive maintenance program is essential to help
ensure safe and proper washer/disinfector and rack operation. This
manual contains a routine maintenance procedure which should be
followed for satisfactory equipment performance.
Customers are encouraged to contact STERIS concerning extended
service maintenance agreements to give your washer/disinfector
and rack planned maintenance, which will help ensure equipment
performance according to factory specifications. STERIS maintains a
global staff of well-equipped, qualified service technicians to provide
this service, as well as professional repair services. Please contact
STERIS for details.
iii
A Word from STERIS Operator Manual 920003-690
Technical
Specifications
Overall Size: 24 x 24 x 14-3/4" (610 x 610 x
375 mm)
Weight: 38 lb (17 kg)
Total Load Capacity: 28" (730 mm)
Maximum Height: (Loaded):28" (730 mm)
Maximum Load Weight: 200 lb (91 kg), including the weight
of the accessory and/or basket(s).
Material:
- Stainless Steel 304,
- Rubber,
-Hydex
® Plastic,
- Silicone,
-Teflon
® Coated Stainless Steel
-Polypropylene
Certification The Reliance Synergy Washer/Disinfector, Reliance Genfore
Washer/Disinfector and Reliance 444 Single-Chamber Washer/
Disinfector, compatible with the Rigid MIS Instrument Rack, meet the
applicable requirements of certain standards. Refer to washer/
disinfector Operator Manual for the current list of standards.
iv
920003-690 Operator Manual A Word from STERIS
The base language of this
document is ENGLISH. Any
translation must be made from
the base language document.
Manufactured by:
STERIS Canada ULC
490 boulevard Armand-Paris
Qué bec (Qu éb e c)
Canada G1C 8A3
European Community Representative:
STERIS Ireland Limited
IDA Business and Technology Park
Tullamore
County Offaly
R35 X865
Ireland
Sales and Service -- Europe, Middle-East and Africa:
STERIS
116 Avenue de Magudas, 33185
Le Haillan
FRANCE
Support Service, by email:
Support_EMEA_IPT[email protected]
All other requests, by phone: +33 5 56 93 94 94
Website:
www.steris.com
Sales and Service -- Americas, Oceania and Asia:
STERIS Corporation
5960 Heisley Road
Mentor, Ohio 44060
440-354-2600 • 800-548-4873
v
920003-690 Operator Manual Table of Contents
A Word from STERIS ...................................................................................................................... i
Advisory ...................................................................................................................................................... i
Indications For Use .......................................................................................................................................... i
Trademarks ..................................................................................................................................................... ii
Service Information ......................................................................................................................................... ii
Technical Specifications..................................................................................................................................iii
Certification .....................................................................................................................................................iii
1 Safety Precautions .................................................................................................................... 1-1
2 Loading Instructions ................................................................................................................. 2-1
2.1 Rigid MIS Instrument Rack - Main Components ................................................................................2-1
2.2 Preparing Instruments ........................................................................................................................2-3
2.3 Loading Rack Lower-Level ................................................................................................................2-3
2.3.1 Lower-Level Manifolded Tray ...................................................................................................2-3
2.3.2 Lower-Level Compartmented Tray ...........................................................................................2-5
2.4 Loading Upper-Level Rack ................................................................................................................2-5
2.4.1 Instrument Holders for Cannulated Instruments ......................................................................2-5
2.4.2 Spindles ...................................................................................................................................2-9
3 Operating Instructions .............................................................................................................. 3-1
3.1 RIGID MIS Cycle Operation ...............................................................................................................3-1
3.2 How to program RIGID MIS Cycle .....................................................................................................3-3
3.2.1 Adding RIGID MIS Cycle to Cycle Menu ..................................................................................3-3
3.2.2 Configuring RIGID MIS Cycle ..................................................................................................3-6
4 Unloading Instructions .............................................................................................................. 4-1
4.1 Visual Inspection ................................................................................................................................4-1
4.2 Retractable Handle For Loading Cart ................................................................................................4-2
4.3 Unloading Instruments from Rigid MIS Instrument Rack ...................................................................4-2
5 Routine Maintenance and Replacement Parts ........................................................................ 5-1
5.1 Routine Maintenance .........................................................................................................................5-1
5.2 Replacement Parts ............................................................................................................................5-2
TABLE OF CONTENTS
Section
Number Description Page
vi
Table of Contents Operator Manual 920003-690
2 Loading Instructions
Figure 2-1. Rigid MIS Instrument Rack – General View ............................................................................. 2-1
Figure 2-2. Rigid MIS Instrument Rack - Main Components .......................................................................2-2
Figure 2-3. Lower-Level Manifold Tray ........................................................................................................2-4
Figure 2-4. Lower-Level Compartmented Tray ...........................................................................................2-4
Figure 2-5. Instrument Holders for Cannulated Instruments .......................................................................2-6
Figure 2-6. Spindles ....................................................................................................................................2-8
Figure 2-7. Fully Loaded Rigid MIS Instrument Rack (Sample Load – Side View) ...................................2-10
3 Operating Instructions
Figure 3-1. Sample Review Printout .........................................................................................................3-11
4 Unloading Instructions
Figure 4-1. Place Instrument Rack on Loading Cart ...................................................................................4-3
Figure 4-2. Fold-Down Cart Handle ............................................................................................................4-3
Figure 4-3. Slide Out Lower-Level Manifolded Tray ....................................................................................4-4
Figure 4-4. Slide Out Lower-Level Compartmented Tray ............................................................................4-4
5 Routine Maintenance and Replacement Parts
Figure 5-1. Rigid MIS Instrument Rack – Assembly ...................................................................................5-2
1 Safety Precautions
Table 1-1. Definition of Symbols ................................................................................................................1-2
3 Operating Instructions
Table 3-1. RIGID MIS Cycle Parameter Values for Reliance 444 Washer ................................................3-8
LIST OF ILLUSTRATIONS
Figure
Number Description Page
LIST OF TABLES
Table
Number Description Page
1-1
Safety Precautions Operator Manual 920003-690
1
The following Safety Precautions must be observed when operating or servicing this equipment. WARNING
indicates the potential for personal injury and CAUTION indicates the potential for damage to equipment. For
emphasis, certain Safety Precautions are repeated throughout the manual. It is important to review ALL Safety
Precautions before operating or servicing the equipment.
Strictly following these Safety Precautions enhances the Customer’s ability to safely and effectively utilize the unit
and helps avoid improper maintenance methods which may damage the unit or render it unsafe. It is important to
understand that these Safety Precautions are not exhaustive; Customers are encouraged to develop their own
safety policies and procedures to enhance and complement these Safety Precautions.
WARNING – PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD:
WARNING – CHEMICAL BURN AND/OR EYE INJURY HAZARD:
WARNING – BURN HAZARD:
Always load and unload items while facing rack. Trocar ends sticking out of rack can be extremely sharp
and can cause injuries such as cuts.
Door cannot be opened during a cycle. In case of an emergency requiring opening of door, press
STOP/RESET touch pad to interrupt cycle.
Regularly scheduled preventive maintenance is required for safe and reliable operation of this
equipment. Contact STERIS to schedule preventive maintenance or obtain the necessary washer/
disinfector Maintenance Manual if preventive maintenance is done by the Customer.
Repairs and adjustments to this equipment must be made only by STERIS or STERIS-trained service
personnel. Repairs and adjustments performed by unqualified personnel or installation of unauthorized
parts could cause personal injury, result in improper equipment performance, void the warranty or result
in costly damage. Contact STERIS regarding service options.
To avoid personal injuries, always load or unload rack on appropriate loading cart or surface.
Washer detergents can cause adverse effects to exposed tissues. Do not get in eyes, on skin or attempt
to swallow. Read and follow precautions and instructions on detergent label and in Safety Data Sheet
(SDS) prior to handling detergent containers or servicing detergent injection pumps. Wear appropriate
personal protective equipment (PPE) whenever handling detergents or servicing detergent injection
pumps and lines.
When cycle is completed, wait until water flow stops before opening door and unloading washer/
disinfector. Hot water/steam may be sprayed through door opening if door is opened too soon. Wear
appropriate Personal Protective Equipment (PPE) whenever reaching into wash chamber.
When cycle is complete, wash chamber, rack and clean items are extremely hot. Allow load to cool to
room temperature and always wear appropriate Personal Protective Equipment (PPE) before removing
rack from wash chamber and handling clean items.
SAFETY PRECAUTIONS
1-2
920003-690 Operator Manual Safety Precautions
WARNING – SLIPPING HAZARD:
CAUTION – POSSIBLE EQUIPMENT DAMAGE:
The table below contains symbols which may be on your equipment components:
To avoid slippery floor conditions, immediately wipe up any spilled liquids or condensation. If spills and
drippage are detergents or other chemicals, follow safety precautions and handling procedures set forth
on detergent or chemical label and/or Safety Data Sheet (SDS).
Before starting a cycle, ensure rotary spray arms in wash chamber and on manifolded accessory rack
are not obstructed and rotate freely. If rotary spray arms are obstructed, equipment damage may result
and washer/disinfector will be unable to effectively process load.
Instruments processed in this rack must be compatible with chemicals used. Always follow instrument
manufacturer's recommendations for chemicals.
To avoid equipment and product damage and maximize processing efficiency, always load soiled items
as instructed on rack.
To prevent voiding the warranty or damaging the equipment, use only authorized STERIS replacement
parts.
Use nonabrasive cleaners when cleaning rack. Follow directions on containers and rub in a back-and-
forth motion (in same direction as surface grain). Abrasive cleaners will damage stainless steel.
When choosing a chemical, select one with a low chloride content. Chemicals with a high chloride
content can be harmful to stainless steel.
Table 1-1. Definition of Symbols
Symbol Definition
Attention. Refer to Manual For Further Instructions.
2
2-1
Loading Instructions Operator Manual 920003-690
2.1 Rigid MIS
Instrument Rack - Main
Components
See Figure 2-1 and Figure 2-2 to identify the main components of the
Rigid MIS Instrument Rack.
IMPORTANT: A listing of the Safety Precautions to be observed when using and servicing this equipment
can be found in SECTION 1 of this manual. Do not use the equipment until you have become familiar with
this information.
WARNING – CHEMICAL BURN
AND/OR EYE INJURY HAZARD:
Washer detergents can cause
adverse effects to exposed
tissues. Do not get in eyes, on
skin or attempt to swallow.
Read and follow precautions
and instructions on detergent
label and in Safety Data Sheet
(SDS) prior to handling detergent
containers or servicing detergent
injection pumps. Wear
appropriate Personal Protective
Equipment (PPE) whenever
handling detergents or
servicing detergent injection
pumps and lines.
CAUTION – POSSIBLE EQUIPMENT
DAMAGE:
When choosing a chemical,
select one with a low
chloride content. Chemicals
with a high chloride content
can be harmful to stainless
steel.
Instruments processed in this
rack must be compatible with
chemicals used. Always follow
instrument manufacturer's
recommendations for chemicals.
Figure 2-1. Rigid MIS Instrument Rack –
General View
LOADING INSTRUCTIONS
2-2
920003-690 Operator Manual Loading Instructions
Figure 2-2. Rigid MIS Instrument Rack - Main Components
Spindles
UPPER LEVEL
LOWER LEVEL
Spring
Stopper
Spray Jet
Nozzles
Flexible
Tubes Instrument
Holders
Frame
Rotary Spray
Arm
Lower-Level
Manifold Tray
Lower-Level
Compartmented Tray
2-3
Loading Instructions Operator Manual 920003-690
2.2 Preparing
Instruments
1. Disassemble all instrument components prior to loading in rack.
2. Open stopcocks and instrument valves.
3. Fully open all handles, forceps jaws and graspers.
4. Remove seals and protective caps from each instrument.
2.3 Loading Rack
Lower-Level
Lower-level of Rigid MIS Instrument Rack is divided in two sections:
A large manifolded mesh tray to hold cannulated instruments.
Compartmented tray with a mesh cover, used to hold a variety of
small items such as end caps, compression springs, washers,
body caps as well as handles and small non-cannulated items.
2.3.1 Lower-Level Manifolded
Tray
See Figure 2-2, Figure 2-3 and Figure 2-4.
Lower-level manifolded tray can hold up to twelve cannulated
instruments (such as Verres needles, outer tubes, biopsy channels),
rods, as well as non-cannulated components.
1. Open hold-down frame.
2. Insert each cannulated instrument end into a spray jet nozzle.
Interior diameter: > 1 mm
Exterior diameter: < 4 mm
3. Manually clean any visible soil on port of cannulated instruments
where a flexible tubing connector will be installed.
4. Connect a flexible tubing to each cannulated instrument and
Verres needle.
5. Insert rods and non-cannulated instruments through end of tray
openings. Slide items under instrument guides.
6. Connect suction/irrigation port to flexible tubing (see Figure 2-3,
Detail B).
IMPORTANT: Do not connect single-port suction/irrigation
instruments to a flexible tube (see Figure 2-5, Detail A). On multi-port
instruments, one port must always be left open (see Figure 2-5,
Detail B). This allows cleaning solution to be evacuated out of
instrument.
7. Distal ends (forceps jaws, scissors cutting edge or graspers)
must be fully opened.
NOTE: Do not insert distal ends in instrument holders. Distal
ends are cleaned by water circulating through rotating spray arm.
8. Close hold-down frame. Push lower-level manifolded tray back
into position inside rack frame.
WARNING – PERSONAL INJURY
AND/OR EQUIPMENT DAMAGE
HAZARD:
To avoid personal injuries,
always load or unload rack
on an appropriate loading
cart or surface.
Always load and unload
items while facing rack.
Trocar ends sticking out of
rack can be extremely sharp
and can cause injuries such
as cuts.
CAUTION – POSSIBLE EQUIPMENT
DAMAGE: To avoid equipment
and product damage and
maximize processing efficiency,
always load soiled items as
instructed on rack.
2-4
920003-690 Operator Manual Loading Instructions
Figure 2-3. Lower-Level Manifold Tray
Hold-Down
Frame
Flexible
Tubing
Instrument
Guide
Instrument
Holder
DETAIL B
DETAIL A
DETAIL A
DETAIL B
Leave One Port Open
for Water Evacuation
Connect Flexible
Tubing to Ports of
Instruments
Lower-Level Manifolded Tray
Figure 2-4. Lower-Level Compartmented Tray
Instrument Jaws
Must Be Fully Open Mesh Cover
Put Light Items in
Covered
Compartments
Place Small Non-
Cannulated Items
Here
Instrument Jaws
Must Be Fully Open
DO NOT Insert Non-Cannulated
Instrument Ends in Instrument
2-5
Loading Instructions Operator Manual 920003-690
2.3.2 Lower-Level
Compartmented Tray
See Figure 2-4.
1. Pull out compartmented tray from rack frame.
2. Open mesh cover.
3. Place small light items such as end caps, compression springs,
washers and body caps into compartments with cover.
4. Place instrument handles and small non-cannulated items into
larger section of compartmented tray.
IMPORTANT: All instrument handles must be fully open.
5. Close compartmented tray cover. Push compartmented tray
back into position inside rack frame.
2.4 Loading Upper-
Level Rack
IMPORTANT: Hollow trocars (with air/gas channels) must be
manually pre-cleaned to ensure cleaning solution flow. Please refer to
instrument manufacturer’s recommended cleaning procedures.
IMPORTANT: Instrument channels must be clear of any debris
clogging channels prior to loading into rack. Cleaning solution must
be able to flow through instrument to clean interior surfaces.
2.4.1 Instrument Holders for
Cannulated Instruments
See Figure 2-5 and Figure 2-7.
Upper-level rack may hold up to four sleeves or reducer sleeves, or
sixteen cannulated instruments (such as outer tubes and biopsy
channels) using instrument holders.
Sleeves and Reducer Sleeves (Up to Four)
1. Sleeves and reducer sleeves specifications:
Interior diameter: > 5 mm
Exterior diameter: > 10 mm
Length: < 330 mm
2. Connect instrument holders to spray nozzle jets.
3. Insert sleeves and reducer sleeves into instrument holders.
4. Manually clean any visible soil on port of cannulated instruments
where a flexible tubing connector will be installed.
5. Connect flexible tubing to suction/irrigation port as required.
IMPORTANT: Do not connect single suction/irrigation port
instruments to a flexible tube (see Figure 2-5, Detail A). On multi-port
instruments, one port must always be left open (see Figure 2-5,
Detail B). This allows cleaning solution to be evacuated out of
instrument.
WARNING – PERSONAL INJURY
AND/OR EQUIPMENT DAMAGE
HAZARD:
To avoid personal injuries,
always load or unload rack
on an appropriate loading
cart or surface.
Always load and unload
items while facing rack.
Trocar ends sticking out of
rack can be extremely sharp
and can cause injuries such
as cuts.
CAUTION – POSSIBLE EQUIPMENT
DAMAGE: To avoid equipment
and product damage and
maximize processing efficiency,
always load soiled items as
instructed on rack.
2-6
920003-690 Operator Manual Loading Instructions
Figure 2-5. Instrument Holders for Cannulated Instruments
Spray Jet
Nozzles
Flexible
Tubes
Instrument
Holders
DETAIL A DETAIL B
DO NOT Insert
Non-Cannulated
Instrument Ends in
Instrument Holders
Connect Flexible
Tubing to Multi-Port
Instruments
Leave One Port Open
for Water Evacuation
DETAIL C
Leave One Port Open
for Water Evacuation
Connect Flexible
Tubing Here
2-7
Loading Instructions Operator Manual 920003-690
Cannulated Instruments (Up to Sixteen)
1. Cannulated instruments (such as outer tubes and biopsy
channels) specifications:
Interior diameter: > 3 mm
Exterior diameter: < 10 mm
Length: < 380 mm
2. Connect instrument holders to spray jets.
3. Insert cannulated instruments and reducers into instrument
holders.
4. Manually clean any visible soil on port of cannulated instruments
where a flexible tubing connector will be installed.
5. Connect flexible tubing to suction/irrigation port, as required.
IMPORTANT: Do not connect single suction/irrigation port
instruments to a flexible tube (see Figure 2-5, Detail A). On multi-port
instruments, one port must always be left open (see Figure 2-5,
Detail B). This allows cleaning solution to be evacuated out of
instrument.
2-8
920003-690 Operator Manual Loading Instructions
Figure 2-6. Spindles
Spring
Stoppers
Flexible
Tubes
Spindles
DO NOT Connect
Flexible Tubing to
Single-Port
Instruments
Connect Flexible
Tubing to Multi-Port
Instruments
Adjust Spring
Stopper
DETAIL B
Leave One Port Open
for Water Evacuation
DETAIL A
2-9
Loading Instructions Operator Manual 920003-690
2.4.2 Spindles See Figure 2-6 and Figure 2-7.
Upper-level rack also provides a total of twenty spindles in three
sizes for trocars, sleeves, bridges and automatic flap valves of
different sizes and lengths.
Trocars, Sleeves, Bridges and Automatic Flap Valves
(Up to Twenty)
1. SMALL SPINDLES (ten) – Trocars, sleeves, bridges, automatic
flap valves specifications:
Interior diameter: 3 to 7 mm
Length: < 185 mm
MEDIUM SPINDLES (six) – Trocars, sleeves, bridges, automatic
flap valves specifications:
Interior diameter: 5 to 11 mm
Length: < 260 mm
LARGE SPINDLES (four) – Trocars, sleeves, bridges, automatic
flap valves specifications:
Interior diameter: 10 to 15 mm
Length: < 235 mm
2. Slide spring stoppers onto spindles.
3. Adjust height of each spring stopper to instrument height.
4. Ensure each instrument is properly adjusted on spring stopper
to help ensure maximum coverage for proper processing.
5. Manually clean any visible soil on port of cannulated instruments
where a flexible tubing connector will be installed.
6. Connect flexible tubing to suction/irrigation port as required.
IMPORTANT: Do not connect single-port suction/irrigation
instruments to a flexible tube (see Figure 2-5, Detail A). On multi-port
instruments, one port must always be left open (see Figure 2-5,
Detail B). This allows cleaning solution to be evacuated out of
instrument.
2-10
920003-690 Operator Manual Loading Instructions
Figure 2-7. Fully Loaded Rigid MIS Instrument Rack
(Sample Load – Side View)
3
3-1
Operating Instructions Operator Manual 920003-690
3.1 RIGID MIS Cycle
Operation
Rigid MIS Instrument Rack must be used exclusively in conjunction
with any Reliance® 444 Single-Chamber Washer/Disinfector
equipped with Enzyme treatment option AND programmed with a
RIGID MIS Cycle, or in conjunction with any Reliance® Synergy®
Washer/Disinfector or Reliance® Genfore® Washer/Disinfector
equipped with Enzyme treatment option.
Following cycle description applies to pre-programmed RIGID MIS
cycle, which is the most extensive combination of pre-programmed
treatments.
For Reliance 444 washer only: If washer/disinfector is equipped
with Enzyme treatment option, but control panel is not already
programmed with a RIGID MIS Cycle, see SECTION 3.2, HOW TO
PROGRAM RIGID MIS CYCLE, for step-by-step instructions.
1. Ensure Main Power switch (located behind upper service access
panel) is in ON position.
2. Position POWER-OFF/STANDBY switch to POWER position.
Display shows (for example, depending on washer/disinfector):
... then:
__ indicates blinking position.
... and printer prints (for example, depending on washer/
disinfector):
IMPORTANT: A listing of the Safety Precautions to be observed when using and servicing this equipment
can be found in SECTION 1 of this manual. Do not use the equipment until you have become familiar with
this information.
RELIANCE
MODEL 444
INSTR UTENSILS
GLASSWARE PLASTIC
WARNING – PERSONAL INJURY
AND/OR EQUIPMENT DAMAGE
HAZARD: Door cannot be
opened during a cycle. In case of
an emergency requiring opening
of door, press STOP/RESET
touch pad to interrupt cycle.
WARNING – SLIPPING HAZARD:
To avoid slippery floor
conditions, immediately wipe
up any spilled liquids or
condensation. If spills and
drippage are detergents or
other chemicals, follow safety
precautions and handling
procedures set forth on
detergent or chemical label
and/or Safety Data Sheet (SDS).
CAUTION – POSSIBLE EQUIPMENT
DAMAGE: Before starting a cycle,
ensure rotary spray arms in
wash chamber and on
manifolded accessory rack are
not obstructed and rotate freely.
If rotary spray arms are
obstructed, equipment damage
may result and washer/
disinfector will be unable to
effectively process load.
CONTROL ON HH:MM:SS
YY/MM/DD
RELIANCE
MODEL 444
S/N 36XXXXXXXX
OPERATING INSTRUCTIONS
3-2
920003-690 Operator Manual Operating Instructions
3. If washer is equipped with Reliance® Load/Unload Modules,
Reliance® ATS Automated Transport System or SCS
Conveyor System: Cycle is started automatically when RIGID
MIS bar code is detected by bar code reader.
-- or --
Press CYCLE MENU touch pad to access second cycle menu.
Display shows:
__ indicates blinking position.
4. Press CURSOR arrows (left or right) to select RIGID MIS cycle.
__ indicates blinking position.
5. Press CYCLE START touch pad to confirm selection. Display
shows:
6. Press CYCLE START touch pad a second time. Cycle proceeds
automatically through selected treatments of RIGID MIS cycle.
For more information on typical cycle operation, see SECTION 4 of
OPERATOR MANUAL for
Reliance 444 Single-Chamber Washer/Disinfector (P122999-133),
Reliance Synergy Washer/Disinfector (P920013-971) or,
Reliance Genfore Washer/Disinfector (P920511-395).
WHEN CYCLE IS COMPLETED, UNLOAD RACK AS EXPLAINED
IN SECTION 4.
ANES RESP GENTLE
RIGID MIS CYCLE 8
ANES RESP GENTLE
RIGID MIS CYCLE 8
PRESS START TO
PROCESS – RIGID MIS
CURSOR
CYCLE
START
CYCLE
START
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38

Steris Amsco Rigid Sterilization Container System Operating instructions

Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI