Steris Amsco 7053L Single-Chamber Washer/Disinfector Operating instructions

  • Hello! I've reviewed the operator manual for the AMSCO 7000 Series 7053L Washer/Disinfector. This manual provides detailed information on operating, maintaining, and troubleshooting the device. The washer/disinfector is designed for cleaning and disinfecting reusable medical devices and has features like thermal disinfection, touchscreen controls, and various safety mechanisms. I'm ready to answer any specific questions you might have about this equipment.
  • What is the intended use of the AMSCO 7053L Washer/Disinfector?
    What kind of waste materials are associated with the washer/disinfector?
    Does the AMSCO 7053L have any safety features?
10085881
Revision K
AMSCO® 7000 Series
Washer/Disinfector Model # 7053L
Operator Manual
10085881_K AMSCO® 7000 Series Operator Manual
i
Operator Manual AMSCO® 7000 Series 10085881_K
A Word from STERIS
Introduction
This manual contains important information on the proper use and routine maintenance of the AMSCO 7053L Single-Chamber
Washer/Disinfector. All personnel involved in the use and maintenance of this equipment must carefully review and comply
with the Safety Precautions and instructions contained in this manual. These instructions are important to protect the health
and safety of personnel operating this washer/disinfector and should be retained in a conveniently accessible area for quick
reference. This equipment is specifically designed only for the uses outlined in this manual.
Complete instructions for uncrating and connecting utilities, as well as the equipment drawing, have been provided. If they are
missing, contact STERIS for replacement copies, providing the serial number and model of the washer/disinfector.
STERIS provides a line of accessories for use with this washer/disinfector. A STERIS representative will gladly review these with
you.
Please familiarize all users with this manual and store in an accessible location.
Also, refer to GENERAL NOTES – APPLICABLE TO STERILIZER AND WASHER EQUIPMENT DRAWINGS (P062941-091) for more
information about utilities requirements.
Advisory
IMPORTANT: A listing of the Safety Precautions to be observed when operating and servicing this washer/disinfector can be found in
Section 1.1, Safety Precautions — Introduction of this manual. Do not operate the equipment until you have become familiar with this
information.
STERIS does not intend, recommend or represent in any way that the AMSCO 7053L Single-Chamber Washer/Disinfector be
used for the terminal disinfection or sterilization of any regulated medical device. The AMSCO 7053L Single-Chamber
Washer/Disinfector is intended only to perform an initial step in the processing of soiled, reusable medical devices. If medical
devices will be contacting blood or compromised tissues, such devices must be terminally processed in accordance with device
manufacturer's instructions and/or Good Hospital Practices before each use in human patients.
Any alteration of this equipment not authorized or performed by STERIS will void the warranty. Alteration of this equipment
which could adversely affect its operation and efficacy may violate national, state and local laws or regulations.
The protection provided by the equipment may be impaired if the equipment is used in a manner not specified by STERIS.
To help ensure operators are adequately trained in the safe use of the equipment, STERIS recommends:
all personnel who operate or maintain the equipment are trained in its operation and in its safe use;
personnel working with toxic chemicals and vapors (if applicable) have comprehensive instructions in the
washer/disinfector process, relevant health hazards and methods to detect the escape of toxic materials;
there is regular training of all personnel concerned with the operation and maintenance of the equipment; attendance
records are maintained; and that evidence of understanding is demonstrated;
current Safety Data Sheets (SDS) should be available to all users in the department. For current SDS, contact STERIS.
Serious incidents that have occurred in relation to this medical device should be reported to the manufacturer and competent
authority in the country where the incident occurred.
Waste Products Disposal
The following are waste materials associated with the washer/disinfector. When disposing of waste materials, ensure disposal
is in compliance with federal, state and local regulations.
Printer paper – Recyclable.
Printer ribbon – Not recyclable.
ii
10085881_K AMSCO® 7000 Series Operator Manual
Optional bar code reader battery – Lithium-Ion battery. Refer to local Regulatory and other laws and regulations about the
environment.
HEPA Filter – Place HEPA filter in a large, sealable bag. Discard the bagged HEPA filter in an outdoor waste container.
Chemicals – Dispose in a safe manner in accordance with local/national regulations and current Safety Data Sheets (SDS).
Chemical containers – Dispose of empty containers in a safe manner in accordance with local/national regulations.
Entire washer/disinfector (end-of-life) – Contact STERIS Corporation for disposal or recycling recommendations.
Indications For Use
The AMSCO 7053L Single-Chamber Washer/Disinfector with touch screen control is intended for use in the cleaning and
intermediate-level disinfection of soiled reusable simple hard-surfaced rigid surgical instruments (such as forceps and clamps),
utensils (such as bedpans, urinals, trays, bowl, basin and kidney dishes), rubber and plastic goods, theatre shoes, and other
similar and related articles found in healthcare facilities.
Service Information
A thorough preventive maintenance program is essential to help ensure safe and proper equipment operation. This manual
contains maintenance schedules and procedures which should be followed for satisfactory equipment performance.
Customers are encouraged to contact STERIS concerning our comprehensive annual maintenance program. Under the terms of
the program, preventive maintenance, adjustments and replacement of worn parts are performed on a scheduled basis to help
ensure equipment performance at peak capability and to help avoid untimely and costly interruptions. STERIS maintains a global
staff of well-equipped, qualified service technicians to provide this service, as well as professional repair services. Please
contact STERIS for details.
Addresses
The base language of this document is
ENGLISH. Any translation must be
made from the base language
document.
Manufactured by:
STERIS Canada ULC
490 boulevard Armand-Paris
Québec (Québec)
CANADA G1C 8A3
Sales and Service –America, Oceania and Asia:
STERIS Corporation
5960 Heisley Road
Mentor, Ohio 44060
800-548-4873
Sales and Service – Europe, Middle-East and Africa:
STERIS
116 Avenue de Magudas, 33185
Le Haillan
FRANCE
Support Service, by email:
All other requests, by phone:
+33 5 56 93 94 94
European Community Representative:
STERIS Ireland Limited
IDA Business and Technology Park
Tullamore
County Offaly
R35 X865
iii
Operator Manual AMSCO® 7000 Series 10085881_K
Ireland
Website: www.steris.com
© 2022, STERIS Corporation. All rights reserved. Printed in Canada.
Permissible Environmental Conditions
The AMSCO 7053L Single-Chamber Washer/Disinfector is designed to operate under the following conditions:
Indoor use only;
Altitude of operation up to 6560 ft (2000 m);
Temperature of 41 to 104°F (5 to 40°C);
Relative humidity of 30% to 80% non-condensing;
Pollution Degree 2.
Pollution degree 2: Equipment must be installed in an environment where normally only non-conductive pollution occurs but
where occasional, temporary conductivity caused by condensation can be expected (as determined according to International
Standard EN/IEC 61010-1, Second Edition).
Seismic Anchorage System (Accessory)
A seismic anchorage system (based on California requirements) is available for high-risk seismic zones (CBC 2013 report
#10094972 and CBC 2019 report #11863506).
Noise Level
Equivalent Sound Pressure Level at work station (measured 39" [1.0 m] away from equipment and at 63" [1.6 m] from ground):
60.2 dB (A). (Results determined according to ISO-3746: 2010 Third Edition Standard: Acoustics – Determination of sound power
levels and sound energy levels of noise sources using sound pressure – Survey method using an enveloping measurement surface over
a reflecting plane.)
Certification
This AMSCO 7053L Single-Chamber Washer/Disinfector meets the applicable Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use:
General Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use:
Underwriters Laboratories (UL) Standard UL: 61010-1, Second Edition (2004).
EN/IEC 61010-1, Second Edition (2001).
Canadian Standards Association (CSA): CAN/CSA-C22.2 No. 61010-1, Second Edition (2004).
Particular Requirements for Sterilizers and Washer/Disinfectors used to Treat Medical Materials:
Canadian Standards Association (CSA): CAN/CSA-C22.2 No. 61010-2-040, First Edition (2007).
EN/IEC 61010-2-040, First Edition (2005).
This AMSCO 7053L Single-Chamber Washer/Disinfector meets the following performance requirements:
General requirements, terms, definitions and tests:
EN ISO 15883-1:2006.
ANSI/AAMI ST15883-1:2009.
CAN/CSA Z15883-1:2009.
iv
10085881_K AMSCO® 7000 Series Operator Manual
Requirements and tests for washer/disinfectors employing thermal disinfection for surgical instruments, anesthetic
equipment, bowls, dishes, receivers, utensils, glassware, etc:
EN ISO 15883-2:2006.
ANSI/AAMI ST15883-2:2013.
CAN/CSA Z15883-2:2009.
This AMSCO 7053L Single-Chamber Washer/Disinfector meets the following EMC/EMI regulations:
Code of Federal Regulation, CFR 47 (2012) – FCC PART 15, SUBPART B Class A – Unintentional Radiators for Use in
Commercial and Industrial Areas.
Industry Canada, ICES-001 – Industrial, Scientific and Medical (ISM) Radio Frequency Generators, Issue 4, June 2006.
European Community, EMC directive 2004/108/EC – Class A, Group 1 ISM Equipment.
EN 61326-1 (2013)/IEC 61326-1, Second Edition (July 2012)
This AMSCO 7053L Single-Chamber Washer/Disinfector
meets the following Governing Directives for affixing of the CE
mark:
Medical Device Directive 93/42/EEC as amended by
2007/47/EC.
Machinery Directive 2006/42/EC.
EMC Directive 2004/108/EC - Class A, Group 1 ISM
Equipment.
RoHS Directive 2011/65/EU
v
Operator Manual AMSCO® 7000 Series 10085881_K
TABLE OF CONTENTS
Section Number Description Page
1 Safety Precautions ........................................................................................................................................ 1
1.1 Safety Precautions Introduction .............................................................................................. 1
1.2 Safety Precautions ......................................................................................................................... 1
2 Symbols and Labels....................................................................................................................................... 5
2.1 Equipment Symbols........................................................................................................................ 5
2.2 Nameplate Symbols ....................................................................................................................... 5
2.3 Transportation and Storage........................................................................................................... 6
3 Component Identification ............................................................................................................................. 9
3.1 Component Identification Introduction.................................................................................... 9
3.2 Washer/Disinfector General..................................................................................................... 9
3.3 Control System ............................................................................................................................. 13
3.4 Emergency Safety Features......................................................................................................... 18
3.5 SCS Conveyor System (Option) .................................................................................................... 19
3.6 ATLAS Workflow Automation Vehicle System ........................................................................... 20
3.7 Accessories ................................................................................................................................... 21
3.8 Flexible Hoses for Cannulated Instruments (Accessory) ........................................................... 37
3.9 Chemicals and Cycles ................................................................................................................... 43
4 Operating Instructions ................................................................................................................................ 51
4.1 Operating Instructions Introduction....................................................................................... 51
4.2 Before Operating Equipment ....................................................................................................... 51
4.3 Loading Washer/Disinfector Manually........................................................................................ 53
4.4 Loading Washer/Disinfector With a Washer Automation System............................................ 55
4.5 Typical Cycle Operation ................................................................................................................ 60
4.6 Unloading Washer/Disinfector .................................................................................................... 64
4.7 Visual Inspection........................................................................................................................... 65
4.8 Responding to an Alarm............................................................................................................... 65
4.9 Aborting Cycle Operation ............................................................................................................. 67
4.10 Shutdown ...................................................................................................................................... 67
4.11 Standby ......................................................................................................................................... 68
4.12 Power Door Operation ................................................................................................................. 70
5 Cycle and Control Value Programming ..................................................................................................... 73
5.1 Cycle and Control Value Programming Introduction ............................................................ 73
5.2 General .......................................................................................................................................... 73
5.3 Supervisor Mode Menus .............................................................................................................. 82
5.4 Cycles Menu .................................................................................................................................. 83
5.5 Transfer Cycle Reports Menu ...................................................................................................... 87
5.6 Miscellaneous Menu..................................................................................................................... 88
5.7 Password Menu............................................................................................................................ 91
5.8 Chemical Pumps Name Menu...................................................................................................... 92
5.9 Prime Chemical Pumps Menu...................................................................................................... 94
5.10 Units Menu.................................................................................................................................... 95
5.11 Print Menu (Accessory) ................................................................................................................ 96
5.12 Date and Time Menus .................................................................................................................. 98
5.13 Language Menu ............................................................................................................................ 99
vi
10085881_K AMSCO® 7000 Series Operator Manual
5.14 Software Version Menu ............................................................................................................... 99
5.15 Test Menu ...................................................................................................................................100
5.16 Cycle History Menu.....................................................................................................................102
5.17 Formatting USB Key ...................................................................................................................108
6 Routine Maintenance ................................................................................................................................115
6.1 Routine Maintenance Introduction ......................................................................................115
6.2 General ........................................................................................................................................115
6.3 Routine Maintenance .................................................................................................................116
6.4 Debris Screen..............................................................................................................................118
6.5 Water Outlet Connection............................................................................................................121
6.6 Cleaning Door Gasket .................................................................................................................121
6.7 Clean Washer/Disinfector Exterior............................................................................................122
6.8 Decontamination Cycle...............................................................................................................123
6.9 Chemical Container Replacement..............................................................................................126
6.10 Changing Thermal Printer Paper Roll (Accessory)....................................................................127
6.11 Changing Impact Printer Paper Roll and Printer Ink Cartridge (Accessory)............................129
6.12 Inspect and Clean Rotary Spray Arm Assemblies and Accessory Corner Spray
Nozzles ........................................................................................................................................132
6.13 Door Safety Test.........................................................................................................................135
6.14 Sump Safety Level Test..............................................................................................................137
6.15 Water Inlet Pressure Test ..........................................................................................................138
6.16 Verifying Emergency Stop Pushbuttons ..................................................................................142
7 Troubleshooting.........................................................................................................................................145
7.1 Troubleshooting — Introduction...............................................................................................145
7.2 Troubleshooting General ......................................................................................................145
7.3 Alarm# 1 LOAD DOOR OPEN (CRITICAL)...................................................................................146
7.4 ALARM# 2 UNLOAD DOOR OPEN (CRITICAL)...........................................................................146
7.5 System Error# 3 CYCLES RETURNED TO DEFAULT VALUES (CRITICAL) ................................146
7.6 Alarm# 4 SUMP TOO LONG TO FILL (CRITICAL) .......................................................................147
7.7 Alarm# 5 SUMP NOT DRAINED (NON-CRITICAL).....................................................................147
7.8 Alarm# 6 WATER REMAINED IN CHAMBER (CRITICAL) ..........................................................148
7.9 Alarm# 7 SUMP TOO LONG TO HEAT (NON-CRITICAL)...........................................................148
7.10 Alarm# 8 SUMP SAFETY WATER LEVEL REACHED (CRITICAL)...............................................149
7.11 Alarm# 9 RECIRCULATION TEMPERATURE PROBE #1 DEFECTIVE (CRITICAL).....................150
7.12 Alarm# 10 SUMP TEMPERATURE TOO HIGH (CRITICAL) ........................................................150
7.13 Alarm# 11 DRYING TEMPERATURE TOO HIGH (CRITICAL) .....................................................151
7.14 Alarm# 12 SUMP OVER TEMPERATURE SWITCH TRIPPED (CRITICAL).................................151
7.15 Alarm# 13 PUMP DID NOT START (CRITICAL)..........................................................................151
7.16 Alarm# 14 DRYING OVER TEMPERATURE SWITCH TRIPPED (CRITICAL) ..............................153
7.17 Alarm# 15 RECIRCULATION TEMPERATURE PROBE #2 DEFECTIVE (CRITICAL) ..................153
7.18 Alarm# 16 CHEMICAL INJECTED VOLUME TOO LOW (CRITICAL)............................................153
7.19 Alarm# 17 RECIRCULATION TEMP. PROBES VARIANCE EXCEEDS RANGE
(CRITICAL) ....................................................................................................................................154
7.20 Alarm# 18 MOTOR OVERLOAD DETECTED (CRITICAL)............................................................155
7.21 Alarm# 19 POWER FAILURE DURING CYCLE (Abort in Progress...) ........................................155
7.22 System Error# 20 I/O BOARDS COMM. FAILURE (CRITICAL) ..................................................155
7.23 System Error# 21 CYCLE AND UNIT CONFIG. RESET TO DEFAULT (CRITICAL)......................156
7.24 Alarm# 22 EMERGENCY STOP PRESSED (CRITICAL)...............................................................156
7.25 Alarm# 23 PRESSURE TRANSMITTER DEFECTIVE (CRITICAL)................................................156
vii
Operator Manual AMSCO® 7000 Series 10085881_K
7.26 Alarm# 24 LOAD DOOR DID NOT OPEN (NON-CRITICAL) .......................................................157
7.27 Alarm# 25 LOAD DOOR DID NOT CLOSE (NON-CRITICAL)......................................................157
7.28 Alarm# 26 LOAD DOOR OBSTRUCTION (NON-CRITICAL)........................................................158
7.29 Alarm# 27 UNLOAD DOOR DID NOT OPEN (NON-CRITICAL)..................................................158
7.30 Alarm# 28 UNLOAD DOOR DID NOT CLOSE (NON-CRITICAL).................................................159
7.31 Alarm# 29 UNLOAD DOOR OBSTRUCTION (NON-CRITICAL) ..................................................159
7.32 Alarm# 30 WATER LEVEL DEFECTIVE (CRITICAL) ....................................................................160
7.33 Alarm# 31 LOW PRESSURE DETECTED (NON-CRITICAL)........................................................160
7.34 Alarm# 32 DRYING SYSTEM NOT HEATING (CRITICAL)...........................................................162
7.35 Alarm# 34 CONDUCTIVITY TRANSMITTER DEFECTIVE (CRITICAL).........................................162
7.36 Alarm# 35 DRYING TEMPERATURE PROBES VARIANCE EXCEEDS RANGE
(CRITICAL) ....................................................................................................................................163
7.37 Alarm# 36 DRYING TEMPERATURE PROBE #2 DEFECTIVE (CRITICAL).................................163
7.38 Alarm# 37 DRYING TEMPERATURE PROBE #1 DEFECTIVE (CRITICAL).................................164
7.39 System Error# 38 SYSTEM READ/WRITE ERROR (CRITICAL) .................................................164
7.40 Alarm# 39 WATER CONDUCTIVITY OUT OF RANGE (NON-CRITICAL) ....................................164
7.41 System Error# 40 SOFTWARE APPLICATION ERROR..............................................................164
7.42 Alarm# 41 to 45 CHEMICAL CONTAINER [X] EMPTY ..............................................................165
7.43 Abnormal Conductivity Readings for Thermal Rinse Draining ................................................165
7.44 Abnormal Noise Generated by Washer/Disinfector.................................................................166
7.45 AIR PRESSURE TOO LOW ..........................................................................................................167
7.46 Auto-Docking Long Chamber Transfer Cart does not Remain Docked to FD209 Transfer
Cart Adaptor................................................................................................................................167
7.47 CHAMBER DOES NOT DRAIN PROPERLY.................................................................................168
7.48 Chamber Rotary Spray Arm Assembly does not Rotate During Cycle....................................168
7.49 CLEAN SUMP FILTER..................................................................................................................168
7.50 CYCLE HISTORY DOWNLOAD DUE OR CYCLE IS ALMOST FULL; PLEASE BACK UP CYCLE
DATA............................................................................................................................................169
7.51 CYCLE HISTORY IS FULL.............................................................................................................169
7.52 DECONTAMINATION CYCLE DUE...............................................................................................169
7.53 Display blank; no acknowledgment when touch pads pressed ..............................................170
7.54 Display Shows Red Screen with an Error Message .................................................................170
7.55 DOOR SAFETY TEST DUE...........................................................................................................170
7.56 Excessive Foam in Chamber ......................................................................................................171
7.57 EXCESSIVE FOAMING DURING PRE-WASH, PHASE REPEATED ............................................171
7.58 FATAL ERROR: MEMORY READ/ WRITE ERROR (CRITICAL) ...................................................172
7.59 Hand-Held Bar Code Reader (EMEA Option) does not Scan ...................................................172
7.60 If Washer/Disinfector is Equipped with a SCS Conveyor System, Bar Code Reader does
not Operate Properly..................................................................................................................172
7.61 If Washer/Disinfector is Equipped with a SCS Conveyor System, Transfer does not
Occur Between Conveyors and Chamber .................................................................................173
7.62 If washer/disinfector is Equipped with Accessory Printer, Generated Printout is Light or
Blank on Generated Printout .....................................................................................................173
7.63 If Washer/Disinfector is Equipped with Accessory Printer, Paper is not Taken Up Into
Power Take-Up...........................................................................................................................174
7.64 If Washer/Disinfector is Equipped with Accessory Printer, Printer does not Work ..............174
7.65 If Washer/Disinfector is Equipped with Accessory Printer, Printer Printout is
Erratic ..........................................................................................................................................174
7.66 INVALID DIRECTORY NAME .......................................................................................................175
viii
10085881_K AMSCO® 7000 Series Operator Manual
7.67 Load Comes Out Dirty At the End of a Cycle ............................................................................175
7.68 Load Comes Out Wet at the End of a Cycle..............................................................................176
7.69 NO AUDIBLE ALARM SIGNAL WHEN ALARM OCCURS, WHEN CYCLE IS COMPLETED,
ETC...............................................................................................................................................177
7.70 NO CODE DETECTED OR CYCLE LOCKED OR INVALID CODE ..................................................177
7.71 PHASE STOPS FOR NO APPARENT REASON...........................................................................178
7.72 PLEASE CLOSE DOOR.................................................................................................................178
7.73 PREVENTIVE MAINTENANCE DUE ............................................................................................178
7.74 PRINTER NOT CONNECTED .......................................................................................................179
7.75 Pump Starts Before Appropriate Chamber Water Level is Reached ......................................179
7.76 RETRIEVAL FAILED, NETWORK CONNECTION ERROR............................................................179
7.77 RETRIEVAL FAILED, NO LOAD NUMBER READY......................................................................180
7.78 SAVE COMPLETE SOME CORRUPT FILES WERE DISCARDED ................................................180
7.79 Interior Light not Turned On or Defective.................................................................................180
7.80 SUMP SAFETY LEVEL TEST DUE ...............................................................................................180
7.81 TEST FAILED: CALL STERIS ........................................................................................................181
7.82 Touch Screen does not Work Properly......................................................................................181
7.83 UNLOAD CONVEYOR NOT READY.............................................................................................182
7.84 UNLOAD DOOR DOES NOT OPEN .............................................................................................182
7.85 WASHER NOT CALIBRATED.......................................................................................................182
7.86 Water Leaks From Washer/Disinfector ....................................................................................183
8 Replacement Parts and Consumable Products......................................................................................185
9 Software Code Copyright..........................................................................................................................187
ix
Operator Manual AMSCO® 7000 Series 10085881_K
LIST OF FIGURES
Description Page
Figure 3-1 AMSCO 7053L Single-Chamber Washer/Disinfector Typical Configuration ........................... 10
Figure 3-2 Chamber and Service Access Interior ............................................................................................ 11
Figure 3-3 Chemical Injection System ............................................................................................................. 12
Figure 3-4 Debris Screen .................................................................................................................................. 13
Figure 3-5 Control System Location ................................................................................................................ 14
Figure 3-6 Power (ON/OFF) Toggle Switch ..................................................................................................... 17
Figure 3-7 Handheld Bar Code Reader ............................................................................................................ 18
Figure 3-8 Emergency Stop Pushbutton ......................................................................................................... 19
Figure 3-9 SCS Conveyor System (Option) ...................................................................................................... 20
Figure 3-10 ATLAS Workflow Automation Vehicle ......................................................................................... 21
Figure 3-11 Flexible Hold-Down Screen.......................................................................................................... 22
Figure 3-12 Mesh Instrument Tray.................................................................................................................. 23
Figure 3-13 Instrument Stringer ...................................................................................................................... 23
Figure 3-14 Multi-Function Rack for Small Items Long ............................................................................. 24
Figure 3-15 Multi-Function Rack for Large Items – Long ............................................................................. 24
Figure 3-16 GPB Accessory Long ................................................................................................................. 25
Figure 3-17 GPB Accessory Long with Length Divider ............................................................................... 26
Figure 3-18 Rack for Five Sterile Goods Containers 150x300x600 .............................................................. 26
Figure 3-19 Insert for OR Shoes ...................................................................................................................... 27
Figure 3-20 Two-Level Manifold Rack — Long and Multi-Function Racks ................................................. 28
Figure 3-21 Two-Level Manifold Rack with Equal Level Space — Long ...................................................... 29
Figure 3-22 Three-Level Manifold Rack Long and Multi-Function Rack................................................. 29
Figure 3-23 Four-Level Manifold Rack — Long ............................................................................................. 30
Figure 3-24 Four-Level Orthopedic Manifold Rack Long............................................................................ 31
Figure 3-25 Four-Level Manifold Rack for US Trays– Long .......................................................................... 31
Figure 3-26 Five-Level Manifold Rack — Long .............................................................................................. 32
Figure 3-27 Three-Level MIS Rack with Removable Upper Insert Long ................................................... 33
Figure 3-28 Three-Level Ophthalmic Rack With Removable Upper Insert Long ...................................... 34
Figure 3-29 Three-Level Base Rack For MIS, Ophthalmic or Dental Application Long ............................ 34
Figure 3-30 Upper Insert for MIS Instruments 68 Jets Connection / 24 Tube Connections ....................... 35
Figure 3-31 Upper Insert for Ophthalmic Instruments / 4x12 Connections Large................................... 35
Figure 3-32 Upper Universal Insert Large.................................................................................................... 36
Figure 3-33 Upper Insert for Two Sterile Goods Containers.......................................................................... 36
Figure 3-34 Anesthesia/Respiratory Rack Long ......................................................................................... 37
Figure 3-35 Flexible Hoses Connected to Irrigation Ports ............................................................................. 38
Figure 3-36 Examples of Cannulated Instruments......................................................................................... 38
Figure 3-37 Correct Placement of Flexible Hoses for Cannulated Instruments........................................... 41
Figure 4-1 Chemical Container ......................................................................................................................... 53
Figure 4-2 Cycle Bar Code Tag Insertion ......................................................................................................... 56
Figure 6-1 Debris Screen ................................................................................................................................119
Figure 6-2 Brushing Debris Screen ...............................................................................................................120
Figure 6-3 Rinsing Debris Screen...................................................................................................................120
Figure 6-4 Water Outlet Connection and Door Gasket.................................................................................121
Figure 6-5 Cleaning Washer/Disinfector .......................................................................................................123
x
10085881_K AMSCO® 7000 Series Operator Manual
Figure 6-6 Chemical Container .......................................................................................................................127
Figure 6-7 Cleaning Spray Arms and End Spray Jets....................................................................................133
Figure 6-8 Top and Accessory Rotary Spray Arm Assemblies.....................................................................134
Figure 6-9 Bottom Rotary Spray Arm Assembly ..........................................................................................135
Figure 6-10 Detection Bar and Holes ............................................................................................................137
Figure 6-11 Emergency Stop Pushbutton.....................................................................................................142
Figure 6-12 Emergency Stop Pushbutton.....................................................................................................143
xi
Operator Manual AMSCO® 7000 Series 10085881_K
LIST OF TABLES
Description Page
Table 2-1 Definition of Symbols......................................................................................................................... 6
Table 2-2 Definition of Symbols on Crates........................................................................................................ 7
Table 3-1 Flexible Hose and Cannulated Instrument Connection ................................................................. 39
Table 3-2 Incorrect Placement of Flexible Hoses and Cannulated Instruments .......................................... 42
Table 3-3 Cycles Using Prolystica Ultra Concentrate Chemicals ................................................................... 44
Table 3-4 Prolystica Ultra Concentrate Chemicals Specifications ................................................................. 45
Table 3-5 Cycles Using Prolystica Ultra Concentrate HP Chemicals ............................................................ 46
Table 3-6 Prolystica Ultra Concentrate HP Chemicals Specifications ........................................................... 47
Table 3-7 Other STERIS Chemicals Specifications .......................................................................................... 48
Table 3-8 Other Chemicals Specifications ....................................................................................................... 49
Table 4-1 Cycle Names and Cycle Bar Code Tag Numbers ............................................................................ 56
Table 5-1 AMSCO 7053L – Cycle Description Chart (60 Hz – With Prolystica Ultra Concentrate
Chemicals) ............................................................................................................................................. 75
Table 5-2 AMSCO 7053L – Cycle Description Chart (50 Hz – With Prolystica Ultra Concentrate
Chemicals) ............................................................................................................................................. 77
Table 5-3 AMSCO 7053L – Cycle Description Chart (50 Hz and 60 Hz – With Other STERIS
Chemicals) ............................................................................................................................................. 79
Table 5-4 AMSCO 7053L Pump Designation, Chemical Type and Concentrations Available................... 81
Table 5-5 AMSCO 7053L A0 Values Selection ............................................................................................ 81
Table 6-1 Routine Maintenance - Check List ................................................................................................116
Table 8-1 Replacement Parts and Consumables..........................................................................................185
xii
10085881_K AMSCO® 7000 Series Operator Manual
1 — Safety Precautions
1
Operator Manual AMSCO® 7000 Series 10085881_K
Safety Precautions
1.1 Safety Precautions — Introduction
The following Safety Precautions must be observed when uncrating/installing, operating or servicing this equipment.
WARNING indicates the potential for personal injury and CAUTION indicates the potential for damage to equipment. It
is important to review ALL Safety Precautions before operating or servicing the equipment.
Strictly following these Safety Precautions enhances the Customer’s ability to safely and effectively utilize the unit and
helps avoid improper maintenance methods which may damage the unit or render it unsafe. It is important to
understand that these Safety Precautions are not exhaustive; Customers are encouraged to develop their own safety
policies and procedures to enhance and complement these Safety Precautions.
1.2 Safety Precautions
WARNING
PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD
Always load empty baskets on appropriate loading cart or surface.
Always use a hold-down screen to secure small or light items. While loading instrument
racks, do not place miscellaneous articles on the sides of instrument trays. Incorrect
rack loading/overloading could lead to injury and/or damage to the equipment or
instruments.
Always wear Personal Protective Equipment (PPE) appropriate to the task being
performed.
Do not place empty baskets on the conveyor modules to load soiled articles. If you do,
the proximity sensor will detect the baskets and indexing will automatically start.
If an obstruction is present in the chamber door, obstruction sensor will detect
obstruction and door will not close. Wait until water flow has stopped before removing
an obstruction.
Regularly scheduled routine maintenance, as well as preventive maintenance, are
required for safe and reliable operation of this equipment. Routine maintenance
procedures are explained in this manual. For preventive maintenance, contact STERIS to
schedule preventive maintenance or obtain the necessary Maintenance Manual
(P764335497) if preventive maintenance is done by the Customer.
1 — Safety Precautions
2
10085881_K AMSCO® 7000 Series Operator Manual
Maximum operating weight, including weight of basket and load, is 160 lb (73 kg) for
AMSCO 7053L Single-Chamber Washer/Disinfector.
Repairs and adjustments to this equipment must be made only by STERIS or STERIS-
trained service personnel. Repairs and adjustments performed by unqualified personnel
or installation of unauthorized parts could cause personal injury, result in improper
equipment performance, void the warranty or result in costly damage. Contact STERIS
regarding service options.
Only use AMSCO 7053L long racks in the washer/disinfector. DO NOT use Vision Single
Chamber racks and accessories.
WARNING
PERSONAL INJURY HAZARD
Risk of pinch point between door and threshold when the door opens. Keep fingers
away from threshold.
Risk of pinch point between door and upper panel. Do not push on top portion of doors;
do not push on door when door is rising; do not push on door when door is jammed.
When baskets are present on the conveyor modules, keep fingers and hands away from
wash chamber doors and moving baskets.
If it ever occurs that the sensor does not detect the obstruction, never use your hand to
push on the obstruction trying to dislodge it. Call STERIS for this abnormal situation.
WARNING
ELECTRIC SHOCK AND/OR BURN HAZARD
Fasteners and star washers are used to ensure protective bonding continuity. Always
reinstall any star washer which may have been removed during installation or servicing.
Pressing bottom portion of POWER (ON/OFF) toggle switch to turn off
washer/disinfector DOES NOT cut off electrical power. Lockout/Tagout building
electrical disconnect switch to turn off power to washer/disinfector for maintenance or
repairs.
1 — Safety Precautions
3
Operator Manual AMSCO® 7000 Series 10085881_K
Repairs and adjustments to this equipment must be made only by STERIS or STERIS-
trained service personnel. Service personnel must disconnect all utilities to unit before
servicing. No one should service unit unless all utilities have been properly locked out.
Always follow local electrical codes and safety-related work practices.
WARNING
CHEMICAL BURN AND/OR EYE INJURY HAZARD
Chemicals are caustic and can cause adverse effects to exposed tissues. Do not get in
eyes, on skin or do attempt to swallow. Read and follow the precautions and
instructions on the chemical label and in the Safety Data Sheet (SDS) prior to handling
the chemical, refilling the chemical container or servicing the chemical injection pump.
Wear appropriate Personal Protective Equipment (PPE) whenever handling the chemical
or servicing the chemical injection pump and lines.
Wear gloves and eye protection when using a chemical. Avoid contact with the eyes or
skin. If a chemical spills or splashes on you, flush the affected area with water for 15
minutes. If swallowed, DO NOT induce vomiting. Seek medical attention immediately.
WARNING
BURN HAZARD
In case of an emergency, always stop cycle by pressing Emergency Stop pushbutton
and wait for water flow to stop prior to trying to open door. Wear appropriate Personal
Protective Equipment (PPE) whenever reaching into the chamber.
Inner surfaces of washer/disinfector are very hot after cycle is completed. Allow
washer/disinfector to cool down before touching chamber. Always wear appropriate
Personal Protective Equipment (PPE), including gloves as well as a face shield, and avoid
all contact with inner walls when reaching into chamber.
Rack and items are extremely hot when cycle is completed. Allow rack and items to cool
to room temperature and always wear appropriate Personal Protective Equipment (PPE)
before reaching into chamber.
1 — Safety Precautions
4
10085881_K AMSCO® 7000 Series Operator Manual
WARNING
SLIPPING HAZARD
To prevent slips, keep floors dry. Promptly clean up any spills or drippage. For spills or
drippage of detergents or other chemicals, follow safety precautions and handling
procedures set forth on detergent or chemical label and/or Safety Data Sheet (SDS).
CAUTION
POSSIBLE EQUIPMENT DAMAGE
Always position manifold rack over the manifold connector before operating
washer/disinfector. If manifold rack and/or bottom rotary spray arm assembly are not
positioned correctly, damage may result and washer/disinfector will be unable to
effectively wash load.
Always select a cycle appropriate for the items being processed. Failure to do so may
result in product damage.
Suction tips (as well as other items) must be fully enclosed within the accessory (curved
suction tips facing inward) to prevent obstruction in conveyor system.
Use nonabrasive cleaners when cleaning unit. Follow directions on containers and rub in
a back-and-forth motion (in same direction as surface grain). Abrasive cleaners will
damage stainless steel. Cleaners rubbed in a circular motion or applied with a wire
brush or steel wool on door and chamber assemblies can damage stainless steel. Do
not use these cleaners on painted surfaces.
When choosing a detergent, select one with a low-chloride content. Detergents with a
high-chloride content must not be used, as such detergents may damage stainless
steel.
Wipe all plastic surfaces or painted surfaces with 70% isopropyl alcohol cleaner (such as
glass or countertop cleaners).
2 — Symbols and Labels
5
Operator Manual AMSCO® 7000 Series 10085881_K
Symbols and Labels
2.1 Equipment Symbols
The table below contains symbols which may be on your washer/disinfector components and/or throughout this
manual:
Symbol Definition
Protective Earth (Ground)
Warning! Risk of Electrical Shock
Attention. Refer to Manual For Further Instructions
ON (For Part of Equipment)
OFF (For Part of Equipment)
Transfer of Heat, Hot Surface
Risk of Crushing Fingers or Hands
2.2 Nameplate Symbols
The table below contains symbols which may be on the identification nameplate of your equipment:
2 — Symbols and Labels
6
10085881_K AMSCO® 7000 Series Operator Manual
Table 2-1. Definition of Symbols
Symbol Definition
Model of Equipment.
Serial Number of Equipment.
Unique Device Identification.
Year of Manufacture of the Equipment.
Safe Disposal of the Device — Refer to Instruction for Use.
Medical Device.
Power Rating of Processor: Voltage, Frequency (Hertz), Type of Supply.
Address of Manufacturing Facility.
Address of European Community Representative.
2.3 Transportation and Storage
Symbols on the shipping crates stipulate equipment transport and storage environmental conditions for temperature
and relative humidity as well as storage conditions and handling information.
/