CALIBRATION PROCEDURE
for the
38094 ISOLATION AMPLIFIER
Section 1 - Voltmeter Calibration
A hand held digital voltmeter (FLUKE model 77 or equivalent) capable of reading 400V DC is
required to perform this test. This meter will be referred to as the DMM. Set the meter function switch to
Volts DC
Remove the two mounting screws that hold the ESP-200 front meter panel in place. This allows
access to the POT1 and POT2 potentiometers on the isolation amplifier PCB.
View from Front of Power Source Console
Viewed from Rear of Power Source Console
Note
Turning POT1 on the isolation amplifier PCB counter
clockwise increases voltage reading.
1.1 Disable the HF starting circuit to protect the DMM by disconnecting the wire to the HF transformer at
TB5-1. (right front bottom of console )
HF Wire
1.2 Connect the DMM to the console output (torch) connections, Positive (red) lead on torch electrode
terminal, Negative (black).lead on the work connection. The DMM will display a negative output
voltage reading when the main contactor is energized.
1.3 Turn the power switch on the console to ON and press the torch trigger switch with the torch in the
air AWAY FROM any metallic surfaces and the operator. The DMM should display approximately –
330V DC.
1.4 Adjust POT1 on PCB to adjust the front panel voltmeter (V) to agree with this reading on the DMM
within +/- 1 volt.
1.5 Turn the power switch on the console to OFF.
1.6 Remove DMM leads and reconnect the wire at TB5-1.