Midmark Procenter Systems Installation guide

Type
Installation guide

Midmark Procenter Systems offer a range of capabilities to enhance dental procedures. These systems boast seamless integration with Elevance chairs, allowing for efficient control of chair functions, delivery systems, and assistant's modules. The consoles feature intuitive touchscreens for precise instrument selection and positioning, enhancing workflow and reducing treatment time. With customizable settings and user-friendly interfaces, Procenter Systems empower dentists to tailor their treatment setups for optimal patient care.

Midmark Procenter Systems offer a range of capabilities to enhance dental procedures. These systems boast seamless integration with Elevance chairs, allowing for efficient control of chair functions, delivery systems, and assistant's modules. The consoles feature intuitive touchscreens for precise instrument selection and positioning, enhancing workflow and reducing treatment time. With customizable settings and user-friendly interfaces, Procenter Systems empower dentists to tailor their treatment setups for optimal patient care.

1
© 2010 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Procenter Console on an Elevance Chair
Installation Guide
Console Arm
Language of origin: English
003-2036-00 Rev. D (8/13)
Mounting
Bolts (3)
Shipping
Bolt
Equipment Alert
When a console chair is ordered, the console mount arm is mounted
on the chair. Remove the shipping bolt before installing the mounting
bolts (in a bag attached to the arm).
Seat Shroud (RH)
Seat Frame
Seat
Upholstery
Step 2: Remove seat shrouds and seat frame.
A) Remove three screws / washers and RH seat shroud.
B) Repeat for LH seat shroud (not shown).
C) Remove two shoulder bolts / washers and lift seat frame.
Step 3: Install console mounting arm.
A) Remove shipping bolt.
B) Position mounting arm to the right or left as required.
C) Install and tighten three mounting bolts to secure the console
mounting arm.
WARNING
Always disconnect power before removing covers. Failure to do
so may result in personal injury.
Step 1: Remove the seat upholstery.
A) Rotate lock knob.
B) Lift seat upholstery to remove tabs from slots.
midmark.com
Style G
2
© 2010 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Install Console onto Chair
Vacuum
and Drain
Tubes
Console
Console Mounting
Arm
Wiring
Harness
& Tubing
Step 5: Fit the console hub into the console mounting arm while routing
the wiring harness, tubing and vacuum lines into the console.
Console
Hub
Outside cover lip ts under
console top, then push in
until ball studs snap into
ttings in frame.
Inside cover lip ts under
console frame and is secured
with screw on bottom.
Equipment Alert
When routing wiring harnesses be aware that all electrical
connections will need to be covered by the inside cover.
Leveling
Bolts (8)
Step 6: Level console by tightening, alternately, the eight leveling bolts on
console mounting arm.
Step 4: Remove the inside and outside covers.
midmark.com
3
© 2010 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
1/8 YL
3/4
1/8 BL
1/8 BL
1/4 BL
1/8
BL
1/4
BL
1/8
YL
1/8
BL
1/8
BL
1/8
YL
1/8
BL
1/8
BL
1/8
BR
1/4
YL
1/8
YL
1/8
YL
1/8
BR
3/4
1/8 YL
1/8 BL
Note:
There are two pieces of 1/8 inch blue tubing coming from the umbilical. Use
the piece that fits into the 1/8 inch blue tubing coming from the toggle switch to
connect city water to toggle switch.
The 2nd piece of 1/8 inch blue tubing, coming from the umbilical, and a 3rd
piece that is plugged, coming from the toggle switch, are to be used for an
assistant’s unit. The instructions for making this connection are included with
the Assistant’s Unit installation guide.
Tubing Adapter
30 PSI Air Supply
30 PSI Water Supply
80 PSI Air Supply
Vacuum
Umbilical to
J-box
To Cuspidor
80 PSI Air
80 PSI Air
30 PSI Water
City
Bottle
To Delivery
Unit
To Assistant’s
Unit
80 PSI Air
30 PSI Water
80 PSI Air
30 PSI Water
Step 7: Install Tubing Adapter and Connect Tubing.
A) Install tubing adapter (from hardware kit) onto 1/4” air supply
tube coming from umbilical.
B) Match colors and connect umbilical tubing to the installed
components.
Note: Tubing is provided to connect bottled water, a delivery unit, and an assistant’s
unit. Leave any unused tubing capped.
Toggle
Switch
midmark.com
Color Page
4
© 2010 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Jumper
Grounded to
Console Frame
J6
J10
J9
WHT
BLK
J4
J1 J2
Delivery
Power
Assistant’s
Power
Cuspidor
Connector
Communication
Connector
Elevance Procenter
Interface PCB
Delivery
Control
Note:
The chair umbilical provides connectors for a delivery unit, assistant’s unit and
cuspidor. Not all connectors will be used for all installations.
Elevance Procenter
Interface PCB
Ground
Wire
Chair Umbilical
Note:
If outlets on the PCB are identical, either one can be used to connect a mating plug.
Caution
Connect ground wire to ensure system ground continuity.
Failure to do so may result in electric shock.
Step 8: Connect electrical connections from umbilicals and
Elevance Procenter interface PC Board.
midmark.com
Color Page
5
© 2010 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Install Cover and Check Installation
Test Verify the following:
1 Console is level.
2 Electrical connections to PC Board are tight.
3
PC Board is secure inside the mounting box and is
properly grounded to the console frame.
4 Water bottle is screwed in tightly.
5 All console covers (4 total) are securely in place.
6
No wires or tubing are pinched by covers or
mountings.
7
Wipe console with a soft dry cloth to remove any dirt or
smudges.
Installation Checklist
Step 9: Install inside and outside covers.
Note: Cover all electrical connections.
DA2339i
Ball studs
Stud Fittings
Inside cover lip ts under
console frame and is secured
with screw on bottom.
Outside cover lip ts under
console top, then push in
until ball studs snap into
ttings in frame.
midmark.com
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5

Midmark Procenter Systems Installation guide

Type
Installation guide

Midmark Procenter Systems offer a range of capabilities to enhance dental procedures. These systems boast seamless integration with Elevance chairs, allowing for efficient control of chair functions, delivery systems, and assistant's modules. The consoles feature intuitive touchscreens for precise instrument selection and positioning, enhancing workflow and reducing treatment time. With customizable settings and user-friendly interfaces, Procenter Systems empower dentists to tailor their treatment setups for optimal patient care.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI