10
© 2010 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Installation Checklist
Test Description of good installation criteria
1 Mount Arm and Cuspidor and Assistant’s Unit are level.
2 Mount Arm and Assistant’s Unit rotate and move easily.
3
Mount Arm stays where placed without drifting. Adjust drag
on pivot points if necessary.
4
Chair Lift Arm covers are in place and limit switches activate
when covers are tapped.
5
Chair Pivot covers are securely in place and the limit switch
activates when cover lowers into an obstruction.
7
Air and Water tools work on the Assistant’s Unit - no wires or
tubing are pinched by covers or mountings.
8
Wipe Cuspidor with a soft dry cloth to remove any dirt or
smudges.
9 Pressing Cup Fill button activates the Cup Fill function.
10
Pressing and holding the Cup Fill button runs water thru the
spout, until button is released.
11
Turning the Cup Fill Adjustment Knob allows you to modify
the amount of water the Cup Fill function delivers.
12 There are no leaks in the fill or drain connections or tubing.
DA2424i
Mount Arm
pivot points
Allen wrench
To Adjust drag on pivot points...
Tighten or loosen the set screw near the desired pivot point to
modify the pivot’s drag.