Abbott CELL-DYN 3000 Troubleshooting Manual

Type
Troubleshooting Manual

This manual is also suitable for

24 Sep 1997
16:11
CELL-DYN
®
3000/3500 SAMPLE LOADER
TROUBLESHOOTING GUIDE
CATALOG NO. 8921133701
Abbott Laboratories © 1997 Abbott Laboratories
Abbott Park, IL 60064
Click on item below:
Troubleshooting
Parts Lists
Verification Procedures
Sensors, LEDs, POTs
Error Messages
EXIT
CD-ROM
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
Search
Book TOC
Go Back
28 Oct
1997 20:44
PROPRIETARY INFORMATION
i
PROPRIETARY INFORMATION
This CELL-DYN 3000/3500 Sample Loader Troubleshooting Guide was developed solely for use with
Abbott Laboratories equipment as specified in the operating instructions. The Guide is copyrighted by
Abbott. No part of this document may be reproduced, stored, or transmitted in any form or by any
means electronic, mechanical, photocopied, recorded, or otherwise without the prior written
permission of Abbott Laboratories.
This CELL-DYN 3000/3500 Sample Loader Troubleshooting Guide is published by Abbott Diagnostics
Division, Abbott Laboratories, Abbott Park, IL 60064-3500, U.S.A. Please direct all inquiries
concerning information in this guide to the foregoing address.
Vacutainer is a registered trademark of Becton-Dickinson and Company.
This troubleshooting guide was developed to be used, in the field, by trained Abbott Laboratories Field
Service Representatives and by in-house support personnel. The revision status of this guide is the
responsibility of the guide holder.
In no event shall Abbott Laboratories or its subsidiaries be liable for any injury, loss or damages
incurred in connection with or arising from the use of this guide by unauthorized personnel or persons
not fully trained by Abbott Laboratories.
© 1997 by Abbott Laboratories. All rights reserved.
END
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
Search
Book TOC
Go Back
28 Oct
1997 20:44
REVISION LOG
ii
REVISION LOG
The Revision Status of this guide is indicated below. Be sure that the guide contains the latest revision
number of all pages. Additional copies of this guide may be ordered using Catalog Number
8921133701.
END
REVISION DATE
PAGES
REVISED/ADDED
TSBs
INCORPORATED
ISAs
INCORPORATED
Original Issue 9/95 Not Applicable 95-001 through 95-053
97-001 through 97-012
95-001 through 95-029
97-001 through 97-006
Rev B 6/97 Add Chapter 3 Parts List 95-030 through 95-042
97-013 through 97-022 97-006 through 97-020
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
Search
Book TOC
Go Back
28 Oct
1997 20:44
HAZARDS
iii
HAZARDS
General Biosafety
WARNING: Potential Biohazard. Consider all clinical specimens and reagent controls that contain
human blood or serum (calibrators, etc.) and contaminated instruments as potentially infectious. Wear
gloves, lab coats, and safety glasses, and follow other biosafety practices as specified in the OSHA
Bloodborne Pathogen Rule (29 CFR Part 1910.1030) or other equivalent biosafety procedures.
Needles
Needles are sharp and potentially contaminated with infectious materials. Avoid any contact with
needles.
Handling spills
Consider all samples, reagents, calibrators, and controls that contain human blood or serum as
potentially infectious. Clean up spills of potentially infectious materials in accordance with established
biosafety practices. A generally accepted procedure for cleaning such spills is to absorb the spill with
toweling or other absorbent material, wipe the area with a detergent solution, and then wipe the area
with an appropriate tuberculocidal disinfectant such as 10% chlorine bleach solution.
Handling Waste and Waste Containers
Dispose of all clinical specimens, reagents, controls, calibrators, standards, cuvettes, and other
disposables that may be contaminated, in accordance with local, state, and federal regulations
governing the treatment of regulated medical waste. Sharps, such as contaminated probes, must be
placed in an appropriately marked, puncture-resistant container prior to treatment and disposal.
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
Search
Book TOC
Go Back
28 Oct
1997 20:44
HAZARDS
iv
Decontamination Procedures
The OSHA Bloodborne Pathogen Rule (20 CFR 1910.1030) requires the decontamination of
laboratory equipment prior to servicing or shipment. (Note: see Operations Manual, Section 9, Service
and Maintenance)
Needles
Flush the needles with a disinfectant solution, if possible, or a buffer or water solution used in the
assay system.
Fluid Pathway
Cycle the pathway with a disinfectant solution, if possible, or a buffer or water solution. If during
maintenance or repair the lines will be disconnected or changed, purge the waste reagents out of the
lines.
Instrument Surfaces
Wipe the surface of the instrument with a detergent solution to remove any soiling. Then wipe the unit
down with a hospital disinfectant such as 10% chlorine bleach solution.
Waste Container
Empty and rinse all waste containers with disinfectant or water prior to shipment.
Disposables
Remove all samples, reagents, controls and disposables from the unit prior to shipment.
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
Search
Book TOC
Go Back
28 Oct
1997 20:44
HAZARDS
v
Electrical Shock Hazard
WARNING: Electrical Shock Hazard. Basic electrical hazard awareness is essential to the safe
operation of any electrical device. Personnel must practice good habits of electrical safety such as the
following:
• Periodically inspect electrical cabling into and on the system for signs of wear or damage.
• Do not disconnect any electrical connection while the power is on.
• In the event of a blown fuse or thrown circuit breaker, determine and correct the cause before
attempting to resume operation of the equipment.
• Keep liquids away from all connectors of electrical or communication components.
• Keep the floor and countertop dry and clean under and around the system.
Laser Hazard
DANGER: Class 3B Laser System. Avoid direct exposure to beam. Wear approved laser eye
protection and observe procedural controls.
Electrostatic Discharge
Many of the electronic components on the CELL-DYN circuit boards are susceptible to electrostatic
discharge (ESD). To prevent ESD damage, follow these guidelines:
1. Place the static protective work mat on a solid surface close to the instrument. Allow the ground
strap to reach the instrument.
2. Attach the ground clip to the instrument chassis.
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
Search
Book TOC
Go Back
28 Oct
1997 20:44
HAZARDS
vi
3. Attach the other end of the ground clip cable and the connector from the wrist strap to the work
mat. (Exception: Some wrist strap cables provide a clip to connect to the same ground source as
the mat cable.)
4. Attach wrist strap to your wrist, make sure the metallic button on the inside of the wrist strap is in
direct contact with your skin.
5. Place PC Boards, removed from the instrument, on the work mat.
6. Replacement PC Boards should be placed on the work mat before removing from the protective
bag and remain on the mat until installation.
Defective PC Boards should be placed in the static-protective bag before removal from the work mat
area.
END
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
Search
Book TOC
Go Back
28 Oct
1997 20:44
How to Identify a CE MARK Loader
vii
How to Identify a CE MARK Loader
The previous version of the Guide (ISA 95-049 Attachment) used
[NEW]
and
[OLD]
references. These
have been replaced with
[CE MARK]
and
[NON-CE MARK]
.
To determine if the Loader is CE MARK Certified, look for the CE Label affixed to the front cover near
the Power Input Panel (See ).
If the Loader is CE MARK Certified, follow the
[CE MARK]
instructions throughout the Guide. Follow
the
[NON-CE MARK]
instructions for all other configurations.
NOTE:The CE MARK Loader is also referred to as Sample Loader 96 in some ISA/TSB Documenta-
tion.
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
Search
Book TOC
Go Back
28 Oct
1997 20:44
How to Identify a CE MARK Loader
viii
Figure i - 1 CE MARK Label Location
END
Power Input
Panel
SL-276
CE MARK
Label
28 Oct
1997 20:45
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
TROUBLESHOOTING 1 - 1
1
Search
Book TOC
Go Back
CHAPTER 1 TABLE OF CONTENTS
1.1 OVERVIEW
1.2 ERROR MESSAGES
1.3 ISOLATION PROCEDURES
IP - 1 Blood Detected In Sample Line
IP - 2 Erratic Results
IP - 3 Mixin
g
Fault
IP - 4 Incomplete Aspiration
IP - 5 No Li
g
hts, Some Other Functions Work
IP - 6 Sample Loader Dead, No Li
g
hts, No
sounds
IP - 7 Alarm 4 (a) Needle Home & Limit Sensors
IP - 8 Alarm 4 (b) X-Axis Home And Limit
Sensors
IP - 9 Alarm 4 (c) Mixin
g
Home And Contact
Sensors
IP - 10 Alarm 4 (d) Mixin
g
Home And Limit
Sensors
IP - 11 Alarm 4 (e and j) Time-out Movin
g
X-Axis
IP - 12 Alarm 4 (f) Time-out Movin
g
Pusher
Y-Axis
IP - 13 Alarm 4 (
g
and k) Time-out Movin
g
Needles
IP - 14 Alarm 4 (h) Time-out Movin
g
Mixin
g
Paddle
IP - 15 Alarm 4 (i) Time-out Movin
g
X-Axis
IP - 16 Alarm 4 (l) Time-out Movin
g
Mixin
g
Paddle
IP - 17 Rack Fault or Alarm 4 (m, n, o, p, q, r)
Rack Sensors
IP - 18 Alarm 4 (s) Sample Tube Not Sensed
Properly
IP - 19 Covers Open or Alarm 4 (t) Safety Cover
Removed
IP - 20 Alarm 4 (u) Rack Number Misread
IP - 21 Alarm 1 (v, w, y, z) Communications
Problem
IP - 22 Alarm 2, Mixin
g
Failure
--- END ---
28 Oct
1997 20:45
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
TROUBLESHOOTING
1.1 OVERVIEW
1 - 2
1
Search
Book TOC
Go Back
1.1 OVERVIEW
This chapter contains the followin
g
information
for Type E, F, & G Sample Loaders:
• 1.2 Error Messa
g
es
• 1.3 Isolation Procedures (IPs)
Troubleshooting Approach
Troubleshootin
g
is required whenever an
Observed Error or a Displayed Error occurs.
Sample Loader Displayed Errors can be
viewed on the Bulletin Line of the Data Station
screen. (The Bulletin Line is down about 3/4
way on the screen.)
When a malfunction occurs, follow these
recommended procedures:
1. Decontaminate the Sample Loader and the
Analyzer accordin
g
to General Biosafety
Procedure.
2. Use this Troubleshootin
g
Guide as follows:
a. If possible, be
g
in by verifyin
g
the
reported error. Perform the exact pro-
cedure that produced the error condi-
tion.
b. When an error occurs, either observed
or displayed, look up the error in Sec-
tion 1.2, Error Messa
es table.
c. Perform the Corrective Action.
d. If directed to an IP, flip pa
es while look-
in
on ed
e of sheets for the desired IP
Number.
e. Turn to pa
e 1 of x, and be
in at the
START box.
f. Follow the directions in the IP.
g
. If directed to perform a Verification Pro-
cedure (VP), flip pa
es while lookin
on
ed
e of sheets for the desired VP Num-
ber.
h. If a part is adjusted or replaced,
o to
Section 3.2, Interrelationship Matrix.
i. Find part in left column.
j. Perform VPs in order
iven (i.e. 1, 2, 3,
4, etc.)
3. Perform Total Call.
--- END ---
28 Oct
1997 20:45
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
TROUBLESHOOTING
1.2 ERROR MESSAGES
1 - 3
1
Search
Book TOC
Go Back
1.2 ERROR MESSAGES
Table 1 - 1 Error Messages
MESSAGE
CORREC-
TIVE
ACTION
Alarms Fault Messa
g
eNone
Alarm 1 (v) Last messa
g
e sent to Analyzer was acknowled
g
ed ne
g
atively
(improper format or improper information.
IP - 21
Alarm 1 (w) Time-out - CELL-DYN did not Acknowled
g
e last messa
g
e from
Sample Loader.
IP - 21
Alarm 1 (x) Analyzer sent F command to Sample Loader, and Sample Loader
responded.
None
Alarm 1 (y) Time-out durin
g
aspiration. Sample Loader and Analyzer maybe
out of Sync.
IP - 21
Alarm 1 (z) Time-out durin
g
mixin
g
. Sample Loader and Analyzer maybe out of
Sync.
IP - 21
Alarm 2 Mixin
g
Failure IP - 22
Alarm 3 Not Used None
Alarm 4 (a) Z Ventin
g
/Samplin
g
Needle’s Ori
g
in and Limit Sensors BOTH ON
at Power-On.
IP - 7
28 Oct
1997 20:45
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
TROUBLESHOOTING
1.2 ERROR MESSAGES
1 - 4
1
Search
Book TOC
Go Back
Alarm 4 (b) X Ori
g
in and Limit Sensors BOTH ON at Power On. IP - 8
(X Transport Home and Limit Sensors both active durin
g
initialization of Sample Loader.)
Alarm 4 (c) Mixin
g
Paddle’s Ori
g
in and Position Sensors BOTH ON at Power
On.
IP - 9
Alarm 4 (d) Mixin
g
Paddle’s Ori
g
in and Limit Sensors BOTH ON at Power On. IP - 10
Alarm 4 (e) X Home Sensor Failure. IP - 11
Time-out in an attempt to drive X-axis mechanism to/from its ori
g
in.
A test performed durin
g
initialization and Run where the particular
mechanism is moved sli
g
htly past a particular sensor then back
and forth several times. The Sensor operation is verified each
time.
Alarm 4 (f) Y Home Sensor Failure. IP - 12
Time-out in an attempt to drive Y-axis mechanism to/from its ori
g
in.
A test performed durin
g
initialization and Run where the particular
mechanism is moved sli
g
htly past a particular sensor then back
and forth several times. The Sensor operation is verified each
time.
Table 1 - 1 Error Messages(continued)
MESSAGE
CORREC-
TIVE
ACTION
28 Oct
1997 20:45
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
TROUBLESHOOTING
1.2 ERROR MESSAGES
1 - 5
1
Search
Book TOC
Go Back
Alarm 4 (
g
) Time-out in an attempt to drive the Ventin
g
/Samplin
g
needle’s axis
mechanism to/from its ori
g
in.
IP - 13
Alarm 4 (h) Time-out in an attempt to drive the Mixin
g
Z2-axis mechanism
to/from its ori
g
in.
IP - 14
Alarm 4 (i) Time-out in an attempt to drive the X-axis mechanism to its
Slowdown Sensor.
IP - 15
Alarm 4 (j) X Limit Sensor Time-out in an attempt to drive the X-axis
mechanism to/from its limit sensor.
IP - 11
Alarm 4 (k) Z1 Limit Sensor Time-out in an attempt to drive the
Ventin
g
/Samplin
g
needle’s axis mechanism to/from its ori
g
in.
IP - 13
Alarm 4 (l) Time-out in an attempt to drive the Mixin
g
Paddle’s axis
mechanism to its Position Sensor.
IP - 16
Alarm 4 (m) Rack-B Sensor did not detect a Rack. IP - 17
Alarm 4 (n) At least 2 Rack Sensors A throu
g
h D not in correct state durin
g
phases of rack rotation.
IP - 17
Alarm 4 (o) Rack Sensor A not in correct state durin
g
rotation phase. IP - 17
Alarm 4 (p) Rack Sensor B not in correct state durin
g
rotation phase. IP - 17
Alarm 4 (q) Rack Sensor C not in correct state durin
g
rotation phase. IP - 17
Alarm 4 (r) Rack Sensor D not in correct state durin
g
rotation phase. IP - 17
Table 1 - 1 Error Messages(continued)
MESSAGE
CORREC-
TIVE
ACTION
28 Oct
1997 20:45
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
TROUBLESHOOTING
1.2 ERROR MESSAGES
1 - 6
1
Search
Book TOC
Go Back
Alarm 4 (s) Sample Tube not sensed correctly. IP - 18
Alarm 4 (t) Safety Cover removed durin
g
sample processin
g
. IP - 19
Alarm 4 (u) Rack Number Misread. Barcode reader did not properly read a
rack barcode.
IP - 20
Autosampler Command Ne
g
atively Acknowled
g
ed Reseat
RS232
Cable At
Both
Ends
And
Restart
Auto Stand-by Mode None
Blood Detected
in Sample Line
If system automatically rinses, and error does not recur. None
If "Blood Detected in Sample Line" error displays on bulletin line. IP - 1
If FATAL error displays on bulletin line. Cycle
Power,
then
IP - 1
Check Rack Alarm None
Cover Open IP - 19
Table 1 - 1 Error Messages(continued)
MESSAGE
CORREC-
TIVE
ACTION
28 Oct
1997 20:45
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
TROUBLESHOOTING
1.2 ERROR MESSAGES
1 - 7
1
Search
Book TOC
Go Back
Emer
g
ency STOP Switch Pressed None
End Ma
g
azine (Rack) Detected None
Erratic Results IP - 2
Incomplete Aspiration (or Samplin
g
Error) IP - 4
Laser OFF None
Laser ON None
Manual Mode None
Mixin
g
Fault for 1st pre-mixer on E-Loader, or F-Loader. Mixin
g
1 Overloaded. IP - 3
Mixin
g
Fault for 2nd pre-mixer on E-Loader. Mixin
g
2 Overloaded. IP - 3
Mixin
g
Faults IP - 3
No Li
g
hts on Control Panel - Other Functions Work IP - 5
Pause Executed None
Re-Mixin
g
Off None
Repeat Auto-Run None
Table 1 - 1 Error Messages(continued)
MESSAGE
CORREC-
TIVE
ACTION
28 Oct
1997 20:45
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
TROUBLESHOOTING
1.2 ERROR MESSAGES
1 - 8
1
Search
Book TOC
Go Back
--- END --
Request to Clean Ventin
g
Needle Clean
Ventin
g
Needle
Per
Operator
Manual
Reset Started None
Sample is at Mixin
g
Position None
Sample Loader Dead - No Li
g
hts -No Sound IP - 6
Samplin
g
Error (or Incomplete Aspiration) IP - 4
Start Key Pressed or Initialize started. None
Start: Send Barcode, CKSUM, RACK, Tube Numbers None
Vertical Barcode Number for Dia
g
nostics. DA=Ali
g
nment, DV=Verification None
Table 1 - 1 Error Messages(continued)
MESSAGE
CORREC-
TIVE
ACTION
28 Oct
1997 20:45
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
TROUBLESHOOTING
1.3 ISOLATION PROCEDURES
1 - 9
1
Search
Book TOC
Go Back
1.3 ISOLATION PROCEDURES
28 Oct
1997 20:45
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
TROUBLESHOOTING
IP - 1 Blood Detected In Sample Line, 1 of 3
1 - 10
1
Search
Book TOC
Go Back
IP - 1 Blood Detected In Sample Line, 1 of 3
Blood in
Line
?
Check for Blood in
Sample Line Near
Blood In-Line
Sensor
1
No
START
From the Data Station Main Menu
- Go to Special Protocols and Press
CLEAN NEEDLE Key
- Check All Tubing, T-Connectors, and
Y-Fittings leading Through SOL #96
and #92 Ending at Waste Chamber
#2
- Check 12" (82.7 kPa) Vacuum Used
for Evacuating Leftover Sample
- Check Diluent Line leading Through
SOL #91 along With 4 PSI (27.6 kPa)
Used to Flush the Line
- Check Vacuum Connection from
Waste Chamber #2 to Sample
Loader Wash Block
YES
28 Oct
1997 20:45
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
TROUBLESHOOTING
IP - 1 Blood Detected In Sample Line, 2 of 3
1 - 11
1
Search
Book TOC
Go Back
IP - 1 Blood Detected In Sample Line, 2 of 3
Problem
Corrected
?
Perform Total
Call
Yes
1
- Turn OFF
Sample Loader
- Reseat CN5 on
CPU Main Bd
- Reseat
Connector on
Blood In-Line
Sensor Bd
- Perform Blood
In-Line Sensor
Check (VP-12.1)
Replace Sample
Line (PL-7.8)
Problem in
Line
?
Pull Sample Line Through
Sensor About 1/2 "
(1.27cm), and Inspect Line
For Dark Spots and Kinks
That Might Cause Error
Yes
No
No
2
Blood
In-Line
Sensor
SL-267
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227
  • Page 228 228
  • Page 229 229
  • Page 230 230
  • Page 231 231
  • Page 232 232
  • Page 233 233
  • Page 234 234
  • Page 235 235
  • Page 236 236
  • Page 237 237
  • Page 238 238
  • Page 239 239
  • Page 240 240
  • Page 241 241
  • Page 242 242
  • Page 243 243
  • Page 244 244
  • Page 245 245
  • Page 246 246
  • Page 247 247
  • Page 248 248
  • Page 249 249
  • Page 250 250
  • Page 251 251
  • Page 252 252
  • Page 253 253
  • Page 254 254
  • Page 255 255
  • Page 256 256
  • Page 257 257
  • Page 258 258
  • Page 259 259
  • Page 260 260
  • Page 261 261
  • Page 262 262
  • Page 263 263
  • Page 264 264
  • Page 265 265
  • Page 266 266
  • Page 267 267
  • Page 268 268
  • Page 269 269
  • Page 270 270
  • Page 271 271
  • Page 272 272
  • Page 273 273
  • Page 274 274
  • Page 275 275
  • Page 276 276
  • Page 277 277
  • Page 278 278
  • Page 279 279
  • Page 280 280
  • Page 281 281
  • Page 282 282
  • Page 283 283
  • Page 284 284
  • Page 285 285
  • Page 286 286
  • Page 287 287
  • Page 288 288
  • Page 289 289
  • Page 290 290
  • Page 291 291
  • Page 292 292
  • Page 293 293
  • Page 294 294
  • Page 295 295
  • Page 296 296

Abbott CELL-DYN 3000 Troubleshooting Manual

Type
Troubleshooting Manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI