Abbott i-STAT 1 System Manual

Type
System Manual
i-STAT
Important!
i-STAT
®
1 System Manual for Waived Tests Update
Art: 732080-00F Page 1 of 1 Rev. Date: 19-Oct-17
As of November 2017, the current i-STAT
®
1 System Manual for Waived Tests has been updated. Please ADD and DELETE the
sheets as listed below. Once the updates have been completed these instructions may be discarded. If you have any questions
about these instructions, please contact your i-STAT support services provider.
A ADD SHEET D DESTROY SHEET
Item Art#
A Quick Reference Guide 716942-00L
D Quick Reference Guide 716942-00K (or lower)
A Section 1 – Start-up Guide 720672-00J
D Section 1 – Start-up Guide 720672-00I (or lower)
A Section 2 – Testing Procedures for i-STAT Chemistry Cartridges 725592-00K
D Section 2 – Testing Procedures for i-STAT Chemistry Cartridges 725592-00J (or lower)
A Section 3 – System Resources 720660-00K
D Section 3 – System Resources 720660-00J (or lower)
END
© 2017
A
bbott Point of Care Inc., 100 and 200 Abbott Park Road,
A
bbott Park, Il 60064 • USA
i-STAT is a trademark of the Abbott Group of Companies in various jurisdictions.
Quick Reference Guide
for the i-STAT
®
1
Handheld and
i-STAT Chemistry Cartridges
For use with a CLIA Certificate of Waiver
CHEM8+
Crea
G
E3+
EC4+
6+
2
Quick Reference Guide for the i-STAT
®
1
Handheld and i-STAT Chemistry Cartridges
For use with a CLIA Certificate of Waiver
This Quick Reference Guide is for use with the following CLIA-waived i-STAT
cartridges: CHEM8+, 6+, EC4+, E3+, Crea, and G. These cartridges include
various subsets of the following tests: sodium, potassium, chloride, total carbon
dioxide, ionized calcium, glucose, urea nitrogen, creatinine and hematocrit.
CLIA-waived status applies to testing performed on venous whole blood collected in
tubes with lithium heparin anticoagulant only.
Your facility must have a CLIA Certificate of Waiver in order to perform waived tests,
must meet all applicable state and local laboratory testing laws, and must follow the
manufacturer’s instructions.
You should read the complete test procedure before performing the test.
See the
i-STAT System Manual for Waived Tests for complete information.
Individual Cartridge: Must be at room temperature for at least 5 minutes before using it.
Box of Cartridges: Must be at room temperature for at least 1 hour before using the cartridges.
Do not remove the cartridge from the pouch until instructed to in the procedures.
Prepare the Cartridge
To scan the barcoded cartridge lot number, position the barcode 3 - 9 inches from laser barcode scanne
r
window on the handheld. Press and hold to activate the scanner. Align the red laser light so it covers
the entire barcode. The handheld will beep when it reads the barcode successfully.
SCAN
Tear open pouch at notch.
Remove cartridge.
Always hold by sides.
Place on level surface.
Scan the barcode.
Fill Mark
Tab for Snap Closure
Sample Well
(Blood or control fluid
goes here)
Cartridge Features
3
Control Testing Frequency
• Test one cartridge from each lot in each shipment upon receipt.
Test a single cartridge from the refrigerator monthly. Select cartridge in the following order:
CHEM8+, 6+, Crea, EC4+, Glucose, E3+
Prepare Control
Control Ampule: Must be at room temperature for at least
30 minutes before using it.
CHEM8+ Cartridges: use the TriControl Level 1 control.
6+ & Crea Cartridges: use the i-STAT or TriControl Level 1
control.
EC4+, E3+, Glucose Cartridges: use the i-STAT or TriControl
Level 3 control.
Test Control
Control Testing Procedure
Shake ampule Break ampule
MENU
1
3
1. Press to turn on handheld.
2. Press to change screen to Administration Menu.
3. Press for the Quality Tests menu.
4. Press for Control.
5. As prompted by the handheld, scan or manually enter the
Operator ID and Control Lot number.
6.
Scan the barcoded cartridge lot number.
7. Remove cartridge from pouch.
8. Shake the ampule hard for 10 seconds.
9. Break the ampule open. Use an ampule breaker or
protect fingers by covering the ampule with gauze.
Continued
4
10. Withdraw the control fluid slowly into a 1 or 3 cc plain
syringe until the syringe is half filled. Dispense 3 drops
of control fluid onto a gauze pad.
Draw fluid
Fill cartridge
Fold closure
Control Testing Procedure - continued
Review Results
Compare results on handheld to the Value Assignment
Sheet.
If any result is out of
range, repeat the control
testing with
a fresh ampule and syringe.
If any result is still out of range, do not use the cartridge
lot, and call Technical Support at 800-284-0702,
Option 1.
11. Fill the cartridge with the control fluid. Fold the
closure over the sample well until it snaps into place.
12.
Push cartridge into the handheld port until it clicks into place, and do not remove until the “DO
NOT REMOVE CARTRIDGE – CARTRIDGE LOCKED” message is gone
.
5
Prepare the Sample
Mix the sample tube by gently inverting it 2-3 times before filling a
cartridge. Do not use a partially filled collection tube.
CHEM8+: Test the sample within 10 minutes after collecting it.
6+, EC4+, E3+, Crea, and G: Test the sample within 30 minutes
after collecting it.
Patient Testing Procedure
1. Press to turn on handheld.
2. Press (i-STAT Cartridge).
3. As prompted by the handheld, use number
keys to enter the Operator ID and Patient ID.
4. Scan the barcoded cartridge lot number.
5. Remove cartridge from pouch.
6. Use a sample transfer device to remove some
sample from
the tube. Dispense 3 drops of
sample onto a gauze pad.
7.
Hold the sample transfer device over the
cartridge sample well.
Test Sample
Get sample
Fill cartridge
2
8. Dispense sample until blood reaches the
cartridge fill mark. Some blood must be left in
the cartridge well.
Continued
6
9. Fold the snap closure over the sample well until it clicks into place.
Press on the tab. Do not press directly over the sample well.
Pt: 12345
16:10 25JanYY
i-STAT CHEM8+
Na mmol/L 141
K mmol/L 4.2
Cl mmol/L 104
iCa mmol/L 1.22
TCO2 mmol/L 26
Page
Pt: 12345
16:10 25JanYY
i-STAT CHEM8+
Glu mg/dL 90
BUN mg/dL 17
Crea mg/dL 1.0
Hct %PCV 42
H b g/dL 14.3
AnGap mmol/L 16
Page
If all results are not displayed on the first
screen,
press to page through all
screens.
Fold closure
Insert cartridge
Review Results
View results on the handheld.
Report results as per your
facility policy.
Patient Testing Procedure - continued
10.
Push the cartridge into the handheld port until it clicks into place.
Do not remove the cartridge from the handheld until the
“DO NOT REMOVE CARTRIDGE - CARTRIDGE LOCKED” message
is gone. For Hct testing, the handheld should remain level until a
result is obtained. A level surface includes running the handheld in the
downloader/recharger.
See i-STAT 1 System
Manual for Waived Tests
if the results are not
displayed or if results are
questioned.
11. After reviewing results, discard gloves, tube, transfer device and cartridge in
biohazard waste container.
7
Abbott Point of Care Inc.
400 College Road East
Princeton, NJ 08540
www.pointofcare.abbott
i-STAT is a registered trademark of the
Abbott Group of Companies in various jurisdictions.
06F20-06
Art: 716942-00L
Rev. Date: 11-Oct-17
Cartridge Receipt
Check the temperature strip included with each shipment.
Do not use the cartridges if the 3rd or 4th windows on the strip are colored.
Contact Technical Support at 800.284.0702, option 1.
Cartridge Storage
Do not use the cartridges after the expiration date on the cartridge pouch and
box.
Refrigerated Storage: 2-8 ˚C (35 to 46 ˚F)
Room Temperature Storage: Follow instructions on the cartridge pouch
and box.
Control Storage
Do not use controls after the expiration date on the labeling.
Refrigerated Storage: 2-8 ˚C (35 to 46 ˚F)
Room Temperature Storage: 18-30 ˚C (64 to 86 ˚F) for up to 5 days.
Cartridge and Control Storage
Precautions
To protect yourself and others from infection, use standard precautions
as defined by your organization or by the Occupational Safety and
Health Administration (OSHA).
A falling or dropped handheld:
place handheld and peripherals on a stable
surface.
Barcode scanner: do not look into laser beam coming from scanner, or
point into eyes of someone else
.
Potential Sources of
Harm to the Operator
Art: 720672-00J, Rev. Date: 11-Oct-17 1
Start-Up
Overview of the i-STAT System ........................................................................3–4
Preparing the Handheld ...................................................................................5–6
Install/Change Batteries
Check/Change Date and Time
Check Software and Status
Entering Information into the Handheld ................................................................7
Reviewing Stored Results ...................................................................................8
Printing Results....................................................................................................9
Quality Control...................................................................................................10
Checking Handheld with the Electronic Simulator ..............................................11
Cleaning and Decontaminating the Handheld ....................................................12
Customizing the Handheld Settings ...................................................................12
Updating the Handheld Software.......................................................................13
Thermal Probe Check........................................................................................13
Warranty.......................................................................................................13-14
Testing Procedures
i-STAT Chemistry Cartridges
System Resources
This section has its own table of contents about a wide range of information that may be useful
as a reference, although only applicable or needed in waived test settings on an occasional basis.
Regulatory Guide
This section has its own table of contents and provides regulatory information regarding
waived testing with the i-STAT
®
1 System.
Contact Information
To place an order for cartridges, supplies or equipment, contact Abbott Point of Care Inc.
at
800-284-0702 – Option 2, or order through your distributor. Cartridges and controls
are shipped Monday through Thursday and delivered within 24 hours.
For Technical Support: Phone: 800.284.0702, Option 1
Fax: 609.419.9371
Table of Contents
2 Art: 720672-00J, Rev. Date: 11-Oct-17
Infrared
Communication
Window
Laser Barcode
Scanner Window
Battery
Compartment
Display
ON/OFF Key
Print Key
Backlight
and 0 Key
Menu Key
Cartridge Port
Numeric
Keys (0-9)
Page Key
Scan Key
Clear Key
Enter Key
Art: 720672-00J, Rev. Date: 11-Oct-17 3
The i-STAT System incorporates components needed to perform blood analysis at the
patient’s side. A portable handheld, a cartridge with the required tests, and 1 to 3
drops of blood will allow the caregiver to view quantitative results for tests commonly
needed in physician offices, nursing homes, and clinics. A portable printer allows test
records to be printed.
T
o perform a test, the operator fills a cartridge with sample, seals the cartridge with its
closure, and inserts the cartridge into the handheld. The unit-use cartridge contains all
components needed to perform the tests. The handheld automatically controls all steps in
the testing cycle including: fluid movement, reagent mixing, calibration and temperature
control. Quality checks are performed continuously throughout the testing cycle. Operator
and patient identification and other patient information can be entered into the test record.
When the test cycle is complete, results are displayed and the test record is stored. This
degree of automation, along with the ability to test fresh whole blood, eliminates many
sources of error as well as time-consuming and costly steps inherent in other methods.
Overview of the i-STAT System
Fill and seal cartridge
with patient sample.
Insert into handheld.
Review, report and record results in minutes.
4
4
4 Art: 720672-00J, Rev. Date: 11-Oct-17
Note Regarding System Reliability: The i-STAT System automatically runs a
comprehensive set of quality checks of handheld and cartridge performance each time
a sample is tested. This internal quality system will suppress results if the handheld or
cartridge does not meet certain internal specifications. To minimize the probability of
delivering a result with medically significant error, the internal specifications are very
stringent. It is typical for the system to suppress a very small percentage of results in
normal operation given the stringency of these specifications. If however the handheld
or cartridges have been compromised, results may be persistently suppressed, and
one or the other must be replaced to restore normal operating conditions. Where
unavailability of results while awaiting replacement of handhelds or cartridges is
unacceptable, Abbott Point of Care Inc. recommends maintaining both a backup
i-STAT System handheld and cartridges from an alternate lot number.
Art: 720672-00J, Rev. Date: 11-Oct-17 5
+
Preparing the Handheld
Install/Change Batteries
The handheld requires two 9-volt lithium batteries. Before using
a new handheld, or when battery replacement is indicated, install
batteries as follows:
1. Slide the battery compartment door off.
2. Tilt the handheld to slide out battery carrier.
3. Locate two (2) 9 volt lithium batteries. Insert a battery into
each side of battery carrier, matching + and - marks on
batteries to the + and - marks on battery carrier.
4.
Slide battery carrier into handheld as shown (label facing up).
5. Slide battery door back into place.
Note:
Two 9 volt lithium batteries will provide power for approximately
400 chemistry cartridges. The handheld will display a “BATTERY
LOW” message when the On/Off key is pressed, and a flashing
battery icon will appear when battery replacement is needed.
Stored results are not lost when batteries are fully discharged or
removed. Stored results and clock settings are maintained by an
internal lithium battery that should last for seven years.
Remove the batteries if the handheld is not to be used for an
extended period of time.
4
4
4
Before using a new or replacement handheld, complete these simple procedures:
Install batteries
Check the handheld’s date and time
Check the handheld’s software and status
Customize handheld
6 Art: 720672-00J, Rev. Date: 11-Oct-17
Check/Change Date and Time
1. Press to turn on handheld.
2. Press
to change screen to Administration Menu.
3. Press (Set Clock).
4. Press (No password is required).
5. Use or to move the cursor to a digit if it
needs to be changed. Use number keys to change digit.
Note: Handheld uses a 24 hour clock.
Therefore, 1:00 PM is displayed as 13:00.
6.
Press to accept changes, or to cancel changes.
MENU
MENU
5
ENT
3
ENT
3
4
4
13:26 18MAY10
A
dministration Menu
1 - Analyzer Status
2 - Data Review
3 - Quality Tests
4 - Customization
5 - Set Clock
6 - Transmit Data
7 - Utility
Enter Current
Time And Date
13:36
01/18/10
mm/dd/yy
ENTER - Set And Exit
MENU - Cancel
Check Software and Status
1. From the
Administration Menu, press
2. Check that software (CLEW & Version) in a newly
received handheld is the same as software in other
handhelds in your facility (if applicable).
1
15:26 18JUN07
Administration Menu
1 - Analyzer Status
2 - Data Review
3 - Quality Tests
4 - Customization
5 - Set Clock
6 - Transmit Data
7 - Utility
Analyzer Status
Temp: 27.1˚C
Pressure: 761mmHg
Battery: 8.54V
Uses: 100
Serial: 30144-B
CLEW: A12
Version: JAMS121B
Custom: Default0
Stored Records
Total: 100
Unsent: 100
Temp: Handheld’s reading of room temperature.
Handheld will function between 16-30 ˚C.
Pressure: Handheld’s reading of barometric pressure.
Handheld will function between
300-850 mmHg.
Battery: Handheld’s reading of battery voltage.
Uses:
Total number of cartridge
runs on handheld.
Serial: Serial number of handheld.
CLEW: Standardization software installed on handheld.
Maintains accuracy of tests.
Version: Application software installed on handheld.
Custom: Name of customization profile in handheld.
Default0 indicates default settings. Handheld can be
customized for certain site-specific and cartridge
testing characteristics.
Total: Total number of test records stored in memory.
Unsent: Number of records that have not been sent to a
data management system.
Art: 720672-00J, Rev. Date: 11-Oct-17 7
Entering Information into the Handheld
Instructions for performing a patient and control test can be found in the
Testing Procedures section for each cartridge type. Information entered into the handheld
during the testing process is stored with results in a patient record that can be accessed
and printed at any time.
After Operator and Patient ID are entered (up to 15 characters each), additional
information can be entered on the Chart Page, which is accessed by pressing the
key. Press to move from field to field on the Chart Page.
Sample Type:
1. ART (arterial) 4. CAP (capillary)
2. VEN (venous) 5. CORD (umbilical cord)
3. MXVN (mixed venous) 6. Other
Note: In waived testing environments,
• Venous whole blood collected in lithium or
sodium heparin tubes is the only acceptable
sample type.
The CPB sample type (if displayed) should be “NO”.
Fields 1, 2 and 3:
up to 9 characters
can be entered
Note: To enter letters, press the key. The letter A is automatically
entered. Use the
arrow keys to move up and down the alphabet until the desired letter
appears. Press the
key again to advance to the next position (To erase a letter,
press the key
to move to the next position, then press to erase the letter). Once all desired characters
have
been input, press to advance to the next field.
Operator ID numbers, Patient ID numbers and information for Fields 1, 2 and 3 on the
Chart Page can be also be entered by scanning a barcode.
Caution: DO NOT stare into the laser aperture or the laser beam,
or point the laser beam at other persons.
Press and hold to activate laser barcode scanner.
Align the red laser light so it covers the entire barcode
being scanned. The handheld will beep when it reads the
barcode successfully. Try again if the “Scan or Enter”
message is still displayed.
Pt:12345
Scan or Enter Data
Sample Type _
Field 1 _ _ _ _ _ _ _ _ _
Field 2 _ _ _ _ _ _ _ _ _
Field 3 _ _ _ _ _ _ _ _ _
1 - ART 4 - CAP
2 - VEN 5 - CORD
3 - MXVN 6 - OTHR
E
NT
3
4
ABC
SCAN
ABC
ABC
ENT
3
4
Reviewing Stored Results
A maximum of 1000 test result records are stored by the handheld, and can be reviewed
by accessing the Data Review Function:
1. Press to turn on handheld.
2. Press
to change screen to Administration Menu.
3. Press (Data Review).
4. At the Data Review screen, choose the category of
results for review. Use to move from the most
recent record to additional records.
Data Review Definitions
Patient: Records for patients are recalled by scanning or manually entering a
patient ID number. If no patient ID is entered, all patient test records
are recalled when is pressed. Not all tests may be displayed on
first screen. Press handheld’s arrow keys to page through all screens.
Control: All quality control test records.
Proficiency: Proficiency testing is a type of Quality Test. Not required for testing
under a Certificate of Waiver.
Cal Ver: Calibration Verification is a type of Quality Test. Not required for testing
under a Certificate of Waiver.
Simulator: All external and internal Electronic Simulator records.
All: All test records in the handheld’s memory.
List: Records are listed with cartridge type, date and time of test, and
patient or control ID (lot number). Records can be selected for viewing
or printing using the numbered keys. Pressing the number key
corresponding to a record selects the record; pressing the number key
a second time deselects the record. To view one or more records,
select the records and press the key.
M
ENU
2
2
13:26 18MAY10
Administration Menu
1 - Analyzer Status
2 - Data Review
3 - Quality Tests
4 - Customization
5 - Set Clock
6 - Transmit Data
7 - Utility
Data Review
1 – Patient
2 – Control
3 – Proficiency
4 – Cal Ver
5 – Simulator
6 – All
7 – List
ENT
3
ENT
3
8 Art: 720672-00J, Rev. Date: 11-Oct-17
A small thermal printer is available (purchased separately through Abbott Point of Care
Inc.) for obtaining printouts of test records.
To print a test record shown on the display, point the handheld’s Infrared Communication
Window at the printer’s IR LED window (on the left side of the front) and press the key
on the handheld. The printer must be within 1 to 5 in (2.5 to 12.7 cm) of the handheld.
Do not move handheld or printer until printing is complete.To print a stored result, recall
the result to the display using the procedure in “Reviewing Stored Results”.
To print a selection of different stored records and for information on set-up, use and
maintenance of the printer, see the System Resources section of this Manual.
Printing Results
Art: 720672-00J, Rev. Date: 11-Oct-17 9
10 Art: 720672-00J, Rev. Date: 11-Oct-17
Quality Control
Quality control procedures are used to ensure the continued accuracy of a test system.
The quality control program for the i-STAT System includes:
Automatic quality checks: A series of automatic quality checks are performed
during each test cycle. When there is a quality check failure, a message is displayed
with the cause and corrective action. A complete list of quality check messages can
be found in the System Resources section of this manual. The quality checks detect
improper environmental conditions, handheld function, cartridge filling, cartridge
function and sensor function.
Electronic Simulator check: An independent check of the handheld's ability to take
accurate and precise readings from the sensors is performed automatically every 24
hours when cartridges are being tested. An external Electronic Simulator is used to
verify an internal Electronic Simulator failure and perform the twice yearly Thermal Probe
check described in the System Resources section. Both the internal and external
simulator results are stored in the handheld’s memory.
Both the internal and external Electronic Simulator send signals that simulate
those of a cartridge to the handheld's signal detection system. The signals
are below and above the measurement ranges of the tests and the
acceptance limits are tighter than those for liquid control samples. Therefore,
the simulator test is more sensitive to an out-of-specification condition than
liquid control samples.
The internal simulator check is triggered by the insertion of a cartridge once
every 24 hours. If the check passes, the cartridge test cycle continues. If the
check fails, "FAIL" and a failure code are displayed. A cartridge test cannot be
performed until the handheld passes the simulator check. If the FAIL
message is observed when it occurs, the cartridge can be re-inserted. If FAIL
is displayed a second time, the external simulator can be used to verify that
the failure is being caused by the handheld and not by a faulty cartridge.
Note that if there is a delay between the time the cartridge is inserted and the
time the display is read, use a fresh cartridge and sample or the external
simulator rather than re-inserting the original cartridge.
Liquid control samples: Used to perform independent checks of system
performance. Their use is an accepted way of verifying performance with traditional
quantitative tests. Although this is a unit use test system, the waived status
categorization for this product requires laboratories to test controls. Control testing
frequency: test one cartridge from each lot in each shipment upon receipt and test a
single cartridge from the refrigerator monthly. Select cartridges in the order as
specified in the Testing Procedures section.
Cartridge storage: Proper cartridge storage conditions as described in the cartridge
Testing Procedures section are required for reliable results.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84

Abbott i-STAT 1 System Manual

Type
System Manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI