Roche cobas 8000 core unit User manual

Type
User manual
cobas
®
8000 modular analyzer
series
Safety Guide – Version 5.1
Software version 06-03
Roche Diagnostics
cobas® 8000 modular analyzer series · Software version 06-03 · Safety Guide · 5.1
2
Publication information
Edition notice This publication is intended for operators of the
cobas
®
8000 modular analyzer series.
The cobas
®
8000 modular analyzer series is composed of
2 main components:
The instrument includes the analytical modules, the
control unit, and other core components.
The cobas
®
8000 data manager consists of a PC
which coordinates data in real time between the
instrument and the host.
Every effort has been made to ensure that all the
information contained in this publication is correct at the
time of publishing. However, the manufacturer of this
product may need to update the publication information
as output of product surveillance activities, leading to a
new version of this publication.
Where to find information The Online Help contains all information about the
product, including the following:
Routine operation
Maintenance
Safety
Troubleshooting information
A software reference
Configuration information
Background information
The Safety Guide contains important safety information.
You must read the Safety Guide before operating the
instrument.
The Quick Reference Guide provides you with an
overview of pre-routine and routine tasks on the data
manager and instrument.
The Operator’s Manual focuses on routine operation
and maintenance. The chapters are organized according
to the normal operation workflow.
Publication
version
Software version Revision date Change description
5.0 06-02 2016-05 Publication redesigned according to the new Roche
documentation concept.
First version of the separated Safety Guide.
5.1 06-03 2016-10 Minor changes throughout the document.
y Revision history
Roche Diagnostics
cobas® 8000 modular analyzer series · Software version 06-03 · Safety Guide · 5.1
3
A separate Operator’s Manual is available for the
cobas
®
8000 data manager.
The PC manufacturer’s manual contains all information
about the control unit hardware.
!
General attention
To avoid serious or fatal injury, ensure that you are
familiar with the system and safety information before you
use the system.
r Pay particular attention to all safety precautions.
r Always follow the instructions in this publication.
r Do not use the instrument in a way that is not
described in this publication.
r Store all publications in a safe and easily retrievable
place.
!
General attention
To avoid serious or fatal injury, ensure that you are
familiar with the system and safety information before you
use the system.
r Pay particular attention to all safety precautions.
r Always follow the instructions in this publication.
r Do not use the instrument in a way that is not
described in this publication.
r Store all publications in a safe and easily retrievable
place.
Training Do not carry out operation tasks or maintenance actions
unless you have received training from Roche
Diagnostics. Leave tasks that are not described in the
user documentation to trained Roche Service
representatives.
Screenshots The screenshots in this publication have been added
exclusively for illustration purposes. Configurable and
variable data, such as tests, results, or path names visible
therein must not be used for laboratory purposes.
Warranty Any customer modification to the system renders the
warranty or service agreement null and void.
For conditions of warranty, contact your local sales
representative or refer to your warranty contract partner.
Always leave software updates to a Roche Service
representative or perform such updates with their
assistance.
Copyright © 2007–2016, Roche Diagnostics GmbH.
All rights reserved.
Roche Diagnostics
cobas® 8000 modular analyzer series · Software version 06-03 · Safety Guide · 5.1
4
Trademarks The following trademarks are acknowledged:
COBAS, COBAS C, COBAS E, COBAS INTEGRA, and LIFE
NEEDS ANSWERS are trademarks of Roche.
All other trademarks are the property of their respective
owners.
Feedback Every effort has been made to ensure that this publication
fulfills the intended purpose. All feedback on any aspect
of this publication is welcome and is considered during
updates. Contact your Roche representative, should you
have any such feedback.
Approvals
The cobas
®
8000 modular analyzer series meets the
requirements laid down in:
Directive 98/79/EC of the European Parliament and of the
Council of 27 October 1998 on in vitro diagnostic medical
devices.
Directive 2014/53/EU of the European Parliament and of
the Council of 16 April 2014 on the harmonization of the
laws of the Member States relating to the making
available on the market of radio equipment and repealing
Directive 1999/5/EC.
Directive 2011/65/EU of the European Parliament and of
the Council of 8 June 2011 on the restriction of the use of
certain hazardous substances in electrical and electronic
equipment.
Compliance is provided by means of the Declaration of
Conformity.
The following marks demonstrate compliance:
For in vitro diagnostic use.
Complies with the IVD directive 98/79/EC on in vitro
diagnostic medical devices
Issued by Intertek for Canada and the US.
Roche Diagnostics
cobas® 8000 modular analyzer series · Software version 06-03 · Safety Guide · 5.1
5
Instrument approvals Furthermore, the instrument is manufactured and tested
according to the following international safety standards:
IEC 61010-1
IEC 61010-2-081 and A1
IEC 61010-2-101
UL 61010-1
CAN/CSA C22.2 No. 61010-1
CAN/CSA C22.2 No. 61010-2-101
CAN/CSA C22.2 No. 61010-2-081
The instrument complies with the emission and immunity
requirements described in standard IEC 61326-2-
6/EN 61326-2-6.
Roche Diagnostics
cobas® 8000 modular analyzer series · Software version 06-03 · Safety Guide · 5.1
6
Contact addresses
Inside the European Union and
EFTA member states
Outside the European Union and
EFTA member states
Manufacturer of the
instrument
Hitachi High-Technologies
Corporation
1-24-14 Nishi-Shimbashi
Minato-ku Tokyo 105-8717
Japan
Manufacturer of
cobas
®
8000 data
manager
Roche Diagnostics GmbH
Sandhofer Strasse 116
D-68305 Mannheim
Germany
Authorized
representative
Roche Diagnostics GmbH
Sandhofer Strasse 116
D-68305 Mannheim
Germany
Manufactured by: Hitachi High-Technologies
Corporation
Manufactured for: Roche Diagnostics GmbH
Sandhofer Strasse 116
68305 Mannheim
Germany
Roche Diagnostics
cobas® 8000 modular analyzer series · Software version 06-03 · Safety Guide · 5.1
7
T
able of contents
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Symbols and abbreviations . . . . . . . . . . . . . . . . . . . 9
Safety classifications . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Safety precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
About operator qualification . . . . . . . . . . . . . . . . . . 13
About safe and proper use of the system . . . . . . . 14
About installation and deinstallation . . . . . . . . . . . 15
About environmental conditions . . . . . . . . . . . . . . 16
Warning messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Electrical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Biohazardous materials . . . . . . . . . . . . . . . . . . . . . . 18
Waste . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Caution messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Mechanical safety. . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Reagents and working solutions . . . . . . . . . . . . . . 28
Software and data security . . . . . . . . . . . . . . . . . . . 29
Notices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Circuit breakers and fuses . . . . . . . . . . . . . . . . . . . 30
Spillage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Magazine drawer (e 801 and e 602 module) . . . . 31
Preparation table (c 502 module). . . . . . . . . . . . . . 31
Transfer mechanism (c 702 module) . . . . . . . . . . . 31
Data security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Product safety labels . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
List of product safety labels . . . . . . . . . . . . . . . . . . 33
Product safety labels on the core unit . . . . . . . . . . 35
Product safety labels on the ISE module. . . . . . . . 37
Product safety labels on the c 702 module . . . . . . 40
Product safety labels on the c 701 module . . . . . . 44
Product safety labels on the c 502 module . . . . . . 48
Product safety labels on the e 801 module. . . . . . 52
Product safety labels on the e 602 module. . . . . . 60
Safety information for barcode readers. . . . . . . . . . . . 65
About the laser barcode reader . . . . . . . . . . . . . . . 65
About the LED barcode readers. . . . . . . . . . . . . . . 66
Safety interlock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
About electronic safety interlock for various
covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Resuming operation after a stop by safety
interlock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
About mechanical safety interlock of top cover
on e 801 and c 702 modules. . . . . . . . . . . . . . . . . . 70
Safety information for disposal. . . . . . . . . . . . . . . . . . . 72
Roche Diagnostics
cobas® 8000 modular analyzer series · Software version 06-03 · Safety Guide · 5.1
8
Disposal information . . . . . . . . . . . . . . . . . . . . . . . . 72
Roche Diagnostics
cobas® 8000 modular analyzer series · Software version 06-03 · Safety Guide · 5.1
Preface 9
Preface
Use this publication together with the Operator’s Manuals
of the cobas
®
8000 modular analyzer series.
Operation and maintenance actions are described in the
Operator’s Manual and the Online Help.
In this section
Intended use (9)
Symbols and abbreviations (9)
Intended use
The cobas
®
8000 modular analyzer series system is a
fully automated random-access software-controlled
system for immunoassays and photometric analysis
intended for qualitative and quantitative in vitro
determinations using a wide variety of tests.
The cobas
®
8000 modular analyzer series is an in vitro
diagnostic (IVD) medical equipment.
Symbols and abbreviations
Product names Except where the context clearly indicated otherwise, the
following product names and descriptors are used.
Product name Descriptor
cobas
®
8000 modular analyzer
series
System
cobas
®
8000 data manager
Data manager
cobas
®
ISE module
ISE module
cobas c 702 module c 702 module
cobas c 701 module c 701 module
cobas c 502 module c 502 module
cobas e 801 module e 801 module
cobas e 602 module e 602 module
cobas c pack reagent cassette
cobas e pack reagent cassette
cobas e pack green reagent cassette
cobas e flow cobas e flow test
Self-Contained ISE Support Box ISE support box
Sample Cup standard cup
y Product names
Roche Diagnostics
cobas® 8000 modular analyzer series · Software version 06-03 · Safety Guide · 5.1
10 Preface
Symbols used in the publication
Symbols used on products
Micro-Sample Cup micro cup
Sarstedt micro tube cryo cup
AssayTip/AssayCup tray magazine
WasteLiner solid waste box
Product name Descriptor
y Product names
Symbol Explanation
o List item
u Related topics containing further information
q
Tip. Extra information on correct use or useful
hints
r Start of a task
I Extra information within a task
f Result of an action within a task
c Frequency of a task
n Duration of a task
d Materials that are required for a task
j Prerequisites of a task
u Topic. Used in cross-references to topics
p Task. Used in cross-references to tasks
w
Figure. Used in figure titles and cross-
references to figures
y
Table. Used in table titles and cross-references
to tables
z
Equation. Used in cross-references to
equations
y Symbols used in the publication
Symbol Explanation
Global Trade Item Number
Quantity contained in the package
y Symbols used on products
Roche Diagnostics
cobas® 8000 modular analyzer series · Software version 06-03 · Safety Guide · 5.1
11
Abbreviations The following abbreviations are used:
Quantity contained in the package
Orientation of the package during
transportation
Complies with the directive 2011/65/EU on
RoHS
Symbol Explanation
y Symbols used on products
Abbreviation Definition
ANSI American National Standards Institute
EC European Community
ECL Electrochemiluminescence
EFTA European Free Trade Association
EN European standard
IEC International Electrical Commission
ISE Ion-selective electrode
ISO International Organization for
Standardization
IT Information technology
IVD In vitro diagnostic
LED Light-emitting diode
LIS Laboratory information system
NaOHD Detergent for reagent probe wash and
reaction cell wash (D1)
QC Quality control
STAT Short Turn-Around Time
UPS Uninterruptible power supply
WEEE Waste Electrical and Electronic
Equipment
y Abbreviations
Roche Diagnostics
cobas® 8000 modular analyzer series · Software version 06-03 · Safety Guide · 5.1
12 Safety classifications
Safety classifications
The safety precautions and important user notes are
classified according to the ANSI Z535.6 standard.
Familiarize yourself with the following meanings and
icons:
!
Safety alert
r The safety alert symbol is used to alert you to potential
physical injury hazards. Obey all safety messages that
follow this symbol to avoid possible damage to the
system, injury, or death.
These symbols and signal words are used for specific
hazards:
WARNING
!
Warning…
r …indicates a hazardous situation that, if not avoided,
could result in death or serious injury.
CAUTION
!
Caution…
r …indicates a hazardous situation that, if not avoided,
could result in minor or moderate injury.
NOTICE
Notice...
...indicates a hazardous situation which, if not avoided,
may result in damage to the system.
Important information that is not safety relevant is
indicated with the following icon:
q Tip...
...indicates additional information on correct use or useful
tips.
Roche Diagnostics
cobas® 8000 modular analyzer series · Software version 06-03 · Safety Guide · 5.1
13
Safety precautions
!
To avoid serious or fatal injury, read and
comply with the following safety precautions.
The hazard warnings in the Safety Guide, in the
Operator’s Manual, in the OLH, and on the instrument
cannot cover every possible case, as it is impossible to
predict and evaluate all circumstances beforehand.
Just following the given directions may, therefore, be
inadequate for operation. Always be alert and use your
common sense.
In this section
About operator qualification (13)
About safe and proper use of the system (14)
About installation and deinstallation (15)
About environmental conditions (16)
About operator qualification
Insufficient knowledge and skills As an operator, ensure that you know the relevant safety
precaution guidelines and standards, and the information
and procedures contained in these instructions.
r Do not carry out operation and maintenance unless
you have been trained by Roche Diagnostics.
r Leave maintenance, installation, or service that is not
described to trained Roche Service representatives.
r Carefully follow the procedures specified in the
instructions for operation and maintenance.
r Follow Good Laboratory Practices, especially when
you work with biohazardous material.
Roche Diagnostics
cobas® 8000 modular analyzer series · Software version 06-03 · Safety Guide · 5.1
14 Safety precautions
About safe and proper use of the system
Missing personal protective equipment Working without personal protective equipment means
danger to life or health.
r Wear appropriate personal protective equipment,
including, but not limited to, the following items:
o Eye protection with side shields
o Fluid-resistant laboratory coat
o Approved lab gloves
o Face shield if there is a chance of splashing or
splattering
Fatigue due to long hours of operation Looking at the monitor over an extended time may lead to
eye strain or body fatigue.
r Take a break to relax, in accordance with your local
regulations.
System not used for an extended period r Follow the decommissioning procedure.
r Set the power switch to OFF if you do not use the
system for an extended period.
r Remove and refrigerate any remaining reagents.
r For further information, call your Roche Service
representative.
Abnormal condition During operation, always check for any abnormal sound,
water leakages, or other abnormal conditions.
r If a trouble occurs, take suitable safety measures
according to the condition and contact your Roche
Service representative.
Non-approved parts Use of non-approved parts or devices may result in
malfunction of the system and may render the warranty
null and void.
r Use only parts and devices approved by Roche
Diagnostics.
Roche Diagnostics
cobas® 8000 modular analyzer series · Software version 06-03 · Safety Guide · 5.1
15
About installation and deinstallation
Errors in installation Only trained Roche Service representatives may install
the system.
r Leave installation that is not described to trained
Roche Service representatives.
Damage in transit r Do not attempt to relocate or transport the system.
r Leave relocation and transportation to Roche Service
representatives.
Disposal An infectious system may lead to infection.
r If you must dispose of the system, read the following
information.
u Disposal information (72)
Roche Diagnostics
cobas® 8000 modular analyzer series · Software version 06-03 · Safety Guide · 5.1
16 Safety precautions
About environmental conditions
Unsuitable environmental conditions Operation outside of the specified ranges may lead to
incorrect results or malfunction of the system.
r Use the system indoors only, and avoid heat and
humidity outside of the specified range.
r Make sure that the system's ventilation openings
remain unobstructed always.
r To maintain the environmental conditions of the
system, perform maintenance in accordance with the
specified intervals.
r Keep the operating instructions undamaged and
available for use. Operating instructions must be
easily accessible for all users.
Power interruption A power failure or momentary drop in voltage may
damage the system or lead to data loss.
r Operate only with an uninterruptible power supply
(UPS).
r Ensure periodic maintenance of the UPS.
r Perform regular backups of results.
r Do not switch off power while the control unit
accesses the hard disk or a storage medium.
Roche Diagnostics
cobas® 8000 modular analyzer series · Software version 06-03 · Safety Guide · 5.1
17
Warning messages
!
List of warning messages
Failure to observe warning messages may result in death
or serious injury.
r Before operating the system, read the warning
messages carefully.
In this section
Electrical safety (17)
Biohazardous materials (18)
Waste (20)
Results (21)
Electrical safety
Electric shock Removing the covers of electronic equipment can cause
electric shock because there are high-voltage parts
inside.
r Do not attempt to work in any electronic equipment.
r Do not remove any cover of the system except those
covers specified in the Operators Manual.
r Do not open the top cover and touch the ultrasonic
mixing unit during operation or when the analyzer
performs maintenance.
r Only Roche Service representatives may install,
service, and repair the system.
Roche Diagnostics
cobas® 8000 modular analyzer series · Software version 06-03 · Safety Guide · 5.1
18 Warning messages
Biohazardous materials
Infectious samples Contact with samples containing material of human origin
may result in infection. All materials and mechanical
components associated with samples containing material
of human origin are potentially infectious.
r Follow Good Laboratory Practices, especially when
working with infectious material.
r Keep all covers closed while the system is operating.
r Before you work with an opened cover, always switch
off the instrument or select a mode appropriate for the
particular maintenance action.
r Wear appropriate personal protective equipment.
r If any infectious material is spilled, wipe it up
immediately and apply a disinfectant.
r If sample or waste comes into contact with your skin,
wash the affected area immediately with soap and
water and apply a disinfectant.
Consult a physician.
Spilled sample The sample containers may vibrate during rack transport
and rotation of sample tubes. Sample may be spilled on
the transport mechanisms.
r Avoid overfilling sample tubes and cups.
r Only use specified sample tubes and cups.
r Leave approximately 10 mm clearance between the
liquid and the top of the sample tube or cup.
r If sample is spilled on the instrument, wipe it up
immediately and apply disinfectant. Be sure to wear
protective equipment.
Ultrasonic mixers Contact with splashes of reaction mixture or by inhaling
reaction mixture aerosol from the ultrasonic mixer may
result in infection.
r Keep all covers closed while the system is operating.
Sharp objects Contact with probes or needles may result in infection.
r When you wipe probes or needles, use several layers
of gauze and wipe from the top down.
r Take care not to puncture yourself.
r Wear appropriate personal protective equipment.
Take extra care when working with lab gloves, which
can easily be pierced or cut, leading to infection.
Roche Diagnostics
cobas® 8000 modular analyzer series · Software version 06-03 · Safety Guide · 5.1
19
Moving parts Contact with moving parts may result in infection or
personal injury.
r Keep all covers closed and in place while the system is
operating.
r Before you work with an opened cover, always switch
off the instrument or select a mode appropriate for the
particular maintenance action.
r Only trained personnel should have access to the keys
to the covers of the instrument.
r Do not touch any parts of the system except those
parts specified. Keep away from moving parts during
operation.
r Only load or unload racks when the green status
indicators at the rack input buffer are lighting.
r During operation and maintenance, carefully follow
the Operator’s Manual.
Roche Diagnostics
cobas® 8000 modular analyzer series · Software version 06-03 · Safety Guide · 5.1
20 Warning messages
Waste
Infectious waste Contact with waste (liquid and/or solid) may result in
infection. All materials and mechanical components
associated with the waste systems are potentially
infectious.
r Wear appropriate personal protective equipment.
Take extra care when working with lab gloves. They
can easily be pierced or cut, leading to infection.
r If any biohazardous material is spilled, wipe it up
immediately and apply a disinfectant.
r If waste comes into contact with your skin, wash the
affected area immediately with soap and water and
apply a disinfectant.
Consult a physician.
Waste must be treated in accordance with the relevant
laws and regulations. Any substances contained in
reagents, calibrators, and QC material, which are legally
regulated for environmental protection, must be disposed
of according to the relevant water discharge facility
regulations. For the legal regulations on water discharge,
contact local authorities.
Two kinds of liquid waste are discharged by the analyzer:
Concentrated liquid waste that contains highly
concentrated reaction mixture. This waste must be
treated as infectious waste as specified by the
relevant regulations.
Dilute waste: A non-concentrated liquid waste diluted
with rinsing water from reaction cell wash or water
from the incubation bath. When using NaOHD for
washing the reaction cells, alkaline concentration is
0.1 mmol/L to 1.0 mmol/L.
Harmful fumes The combination of bleach with the contents of the liquid
waste container can cause potentially harmful fumes.
r Do not use bleach to clean the liquid waste container.
Environmental harm The system generates liquid and/or solid waste. This
waste contains concentrated reaction mixtures and is
potentially infectious. Improper disposal may contaminate
the environment.
r Treat this waste as infectious waste.
r Dispose of waste in accordance with the local
regulations.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72

Roche cobas 8000 core unit User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI