Pathatrix® 10-Pooling E. coli O157:H7 Kit Linked to Selective Agar Plates
Chapter1 Pathatrix® 10-Pooling E. coli O157:H7 Kit – Same-Day–Enrichment Format
Detection – direct plating
Detection – direct plating
Note: Werecommendstreakingthesampleouttogenerateindividualcolonies,as
opposedtospreadplating.
1. PipetallofthePathatrix®paramagneticbeadsuspensionontotheedgeofwell‐
dried,selectiveagarplates(forexample,CefiximePotassiumTelluriteSorbitol‐
MacConkeyAgar[CT‐SMAC],CHROMagarO157)forisolationofthetarget.
Note: Retainanamountofbeadsuspensionyouwishtokeep.Dividethe
remainderofthebeadsuspensionintoequalamountsforstreakingonyour
selectiveagarplates.
2. Allowtheplatestodryforapproximately10minutestheninvertandincubateat
attherequiredtemperaturefor18–24hours.
Ifanegativeresultisobtainedfromthepooledsample,theindividualenrichmentscan
bediscarded,asfurthertestingisnotrequired.
Apresumptivepositiveresultisdefinedastheisolationoftypical,suspicious,or
atypicalE.coliO157:H7coloniesontheagarplatesused.Apresumptivepositive
isolateshouldbesubsequentlyconfirmedbytheuseofsubculture,appropriate
biochemicalandserologicaltests(forexample,asdetailedinISO16654:2001orUSDA
MicrobiologyLaboratoryGuidebook[MLG]5.04asusedintheAOACResearch
Institutevalidationstudy[See“References”onpage30]).
Ifapositiveresultisobtainedfromthepooledsample,theenrichmentscanberetested
toallowidentificationofwhichindividualsamplesinthepoolproducedthepositive
result(seenextsection,“Retestingofindividualsamplesafter“PositivePooled
Sample””)
Retesting of individual samples after “Positive Pooled Sample”
1. Individualsamplesthatrequireretesting,shouldberewarmedto42±1°Cbefore
having10mLremovedandtransferredintoaSampleVessel.
2. Individualsamplesshouldberetestedimmediatelybyrepeatingallthestepsin
SamplesandconsumableloadingthroughDetection–directplatingdescribed
above.
Test result interpretation and classification
ThePathatrix®10‐PoolingE.coliO157:H7Kitisdesignedasasamplepreparation
methodforpresence/absencedetectionofE.coliO157:H7infoodmatrices.
UsingthePathatrix®10‐PoolingE.coliO157:H7Kit(same‐day–enrichmentformat)
linkedtoselectiveagarplates,presumptiveresultscanbeobtained,priorto
confirmation,within26hours.
Onceconfirmed,theresultsarereportedas:
•E.coliO157:H7Detectedin25–375g(samplematrices)
•E.coliO157:H7Notdetectedin25–375g(samplematrices)