Mettler Toledo InPro 5000 (i) Sensors Operating instructions

Type
Operating instructions
14 InPro 5000 (i) CO
2
Sensor Series
© 06/2019Mettler-ToledoGmbH InPro5000(i)SensorSeries
30 313 393 B PrintedinSwitzerland
1 Introduction ................................................................... 15
2 Safetyinstructions .......................................................... 15
3 Productdescription ........................................................ 15
4 Sensorinstallation .......................................................... 16
4.1 Connection .................................................................... 16
5 Operation ..................................................................... 16
5.1 CalibrationofthepHsensor ............................................ 16
5.2 Mountingthemembranebody ......................................... 17
5.3 CalibrationoftheCO
2
sensor .......................................... 17
6 Maintenance .................................................................. 17
7 Disposal ....................................................................... 17
8 Warranty ....................................................................... 17
9 Storagecondition ........................................................... 17
10 ECDeclarationofConformity ........................................... 17
Contents
InPro 5000 (i)
CO
2
Sensors
Quick Setup Guide
InProandISMareregisteredtrademarksof
Mettler-ToledoGmbHinSwitzerland,theUSA,
theEuropeanUnionandafurthervecountries.
en
en QuickSetupGuide
InPro 5000 (i) CO
2
Sensor Series 15
InPro 5000 (i) Sensor Series © 06/2019Mettler-ToledoGmbH
PrintedinSwitzerland 30 313 393 B
1 Introduction
Thank you for buying this InPro 5000(i) sensor from
METTLER TOLEDO. InPro 5000(i) sensors are intended
solelyfor in-line measurementof dissolved CO
2
(carbon
dioxide).
2 Safety instructions
h
Notice:TheInPro5000(i)sensorshouldbeinstalled,oper-
ated,maintainedonlybypersonnelfamiliarwiththesensor
andwhoarequaliedforsuchwork.
h
Notice:Adefectivesensormustneitherbeinstallednorput
intoservice.
a
Caution: Only the maintenance work described in these
operatinginstructionsmaybeperformedonthesensors.
h
Notice:Whenchangingfaultycomponents,useonlyorig-
inal spare parts obtainable from your METTLER TOLEDO
supplier.
h
Notice:Nomodicationstothesensorortheaccessories
areallowed.Themanufacturer/supplieracceptsnorespon-
sibilityfordamagecausedbyunauthorizedmodications.
Theriskisborneentirelybytheuser.
3 Product description
Eachsensorissuppliedtogetherwith:
1bottleCO
2
-electrolyte25ml
1qualitycontrolcerticate
–1inspectioncerticate3.1(inaccordancewithEN
10 2 0 4. 3 / 3.1)
1quicksetupguide
1ECDeclarationofConformity
Membranebodyandcapsleevearenotmounted.Theinte-
riorbodyisprotectedagainstdryingoutwithawettingcap,
lledwithCO
2
-electrolyte.
Pleasecheckcompleteness.
en
16 InPro 5000 (i) CO
2
Sensor Series
© 06/2019Mettler-ToledoGmbH InPro5000(i)SensorSeries
30 313 393 B PrintedinSwitzerland
1 VPconnector
2 ISMconnector
3 Hexagonalscrew
4 ThreadedsleevePg13.5
5 WasherPTFE
6 O-Ring(10.77×2.62mm,
SiliconeFDA/USPVI)
7 Retainernut
8 Membranebody
9 Interiorbody
10 Capsleeve
11 Wettingcap
4 Sensor installation
4.1 Connection
Theinteriorbody(pHelectrode)isreadyforcalibration.
–Onunpacking,checkthepHelectrodeformechanicalda-
mage.ReportanydamageimmediatelytoyourMETTLER
TOLEDOsupplier.
–Removethewateringcap,capsleeve,membranebody
andbrieyrinsethesensorwithde-ionizedwater.
–Afterrinsing,thesensorshouldonlybedabbeddrywitha
tissue.DonotrubthepH-sensitiveglassasthiscanlead
toelectrostaticchargingandsluggishresponsetimes.
–CheckinsidethepH-sensitiveglassforthepresenceof
any air bubbles and remove same by gently swinging
theelectrodeinaverticalplane(shakingdownaswitha
thermometer).
–CalibratethepHsensorasdescribedinSection5.
–Placetheelectrodeintothehousingasdescribedinthe
instructionsforthehousing.
h
Notice:Mountingupside-downorhorizontallyisnotpos-
sibleduetotheinternalpHelectrode.Anangleof15°or
greatertohorizontalisneeded.
–Connect the electrode and the pH transmitter with the
correspondingcable(AK9fortheInPro5000(i),VPfor
the InPro 5000). Observe the connection scheme that
accompaniesthecableorthetransmitter.
ISM
®
sensorssuchasInPro5000ienablePlugandMea-
surestartupandexpandeddiagnostics.Forinstallation,
commissioningandoperationoftheISMsystemplease
refertoinstructionsforthetransmitter,moduleandcable.
5 Operation
5.1 Calibration of the pH sensor
(NotneededforInPro5000i)
a:Min.immersiondepth>25mm/1"
For the analog CO
2
sensor InPro 5000,we recommend
a2-pointcalibrationinpHbuffers7.00and9.21.Before
calibration,rstremovewettingcap.StartwithbufferpH
7.00fordeterminingzeropointandthenbufferpH9.21for
determiningtheslope.Pleaserefertoinstructionmanualof
thetransmitterforfurtherdetails.
h
Notice:Onthetransmitter,selectPt1000asthetempera-
turesensorfortheanalogInPro5000.TheInPro 5000i
doesnotrequireatemperaturesetting.
en
InPro 5000 (i) CO
2
Sensor Series 17
InPro 5000 (i) Sensor Series © 06/2019Mettler-ToledoGmbH
PrintedinSwitzerland 30 313 393 B
h
Notice:Pleasedonotleavethesensorinthecalibration
bufferforlongerthan1hour.Toolonganexposuremay
resultinhighsensordriftintheprocess.
5.2 Mounting the membrane body
–AftercalibrationofthepHelectroderinsethepHelectrode
withde-ionizedwaterandcarefullydabitwithapaper
tissue.
FillthenewmembranebodyhalfwaywithCO
2
electroly te.
–Makesurethatallairbubblesareremovedfromthemem-
brane body. Air bubbles can be removed by carefully
tappingonthemembranebody.
–Slip the membrane body over the pH electrode while
holding the sensor in a vertical position. The excess
electrolytethatwillbedisplacedshouldberemovedwith
apapertissue.
h
Notice: No electrolyte, sample media or contamination
shouldbepresentbetweenthemembranebodyandthe
capsleeve.Pleasecheckcarefully.
–Carefully slip the cap sleeve over the tted membrane
body,holdingthesensorinaverticalposition,andscrew
ittight.Thecapsleevemustbecleanandtight.
5.3 Calibration of the CO
2
sensor
TheInPro5000isinsitusterilizableorinanautoclave.After
sterilizationaprocesscalibrationstepisneededtoachieve
accuratemeasurements.
The process calibration can be performed by using the
transmitter’s calibration menu. For further details please
refertothemainmanual.
6 Maintenance
Werecommendusinganewmembranebodyaftereach
batch.AftereachcalibrationofthepHelectrodenewCO
2
electrolyteshouldbelledinthemembranebody.
IfthepHelectrodeshowsasluggishresponsetimeand/or
insufcientsensitivity,thepH-sensitiveglasscanbereac-
tivatedusinganHFsolution(p/n51319053).Forfurther
detailspleaserefertothemainmanual.
Typically,after10–20sterilizationcyclesthepHelectrode
shouldbereplaced.Pleaserefertotheinstructionmanual
ofthereplacementinteriorbodyInPro5000(i).
7 Disposal
Itistheuser’sresponsibilitytodisposeofthespentspare
partsandifnecessarythesensorinaprofessionalmanner.
Sensorscontainelectroniccomponentsthatrequirespecial
disposalwithoutanyhazardforhumans,andwithnorisk
ofharmtotheenvironment.Pleaseobservetheapplicable
localor national regulationsconcerning the disposal of
"wasteelectricalandelectronicequipment".
8 Warranty
Warrantedfreeofmanufacturingdefectsforatermof12
monthsafterdelivery.
9 Storage condition
Forstoragewerecommendllingthemembranebodywith
freshCO
2
electrolyteandreconnectingittothesensor.The
wetting cap, lled with CO
2
-electrolyte, will prevent dry-
ing-outofthemembrane.
10 EC Declaration of Conformity
Thecompletedeclarationisavailableinthedeclarationof
conformitycerticate.
en
74 InPro 5000 (i) CO
2
Sensor Series
© 06/2019Mettler-ToledoGmbH InPro5000(i)SensorSeries
30 313 393 B PrintedinSwitzerland
Notes
InPro 5000 (i) CO
2
Sensor Series 75
InPro 5000 (i) Sensor Series © 06/2019Mettler-ToledoGmbH
PrintedinSwitzerland 30 313 393 B
Notes
Mettler-ToledoGmbH,ProcessAnalytics
ImHackacker15,CH-8902Urdorf,Switzerland
Phone+41447296211,Fax+41447296636
Subjecttotechnicalchanges.
©06/2019METTLERTOLEDO.Allrightsreserved.
PrintedinSwitzerland.30313393B
www.mt.com/pro
ForaddressesofMETTLERTOLEDO
MarketOrganizationspleasegoto:
www.mt.com/pro-MOs
Management System
certified according to
ISO 9001 / ISO 14001
9001
certified
ISO
14001
certified
ISO
1258
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76

Mettler Toledo InPro 5000 (i) Sensors Operating instructions

Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI