  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo M300 Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Quick Setup Guide Transmitter M300
Download Manuals and Tools at
www
.mt.com/m300-downloads
Multilingual Quick Setup Guide
cs Průvodcerychlýmnastavením 3
da Lynvejledning 11
de Quick-Setup-Leitfaden 19
en Quick
SetupGuide 27
es Guía
deconguraciónrápida 35
 Pika-asetusopas 43
fr Guide
deparamétragerapide 51
hu Gyorsbeállítási
útmutató 59
it Guida
allacongurazionerapida 67
ja クイック セットアップ ガイド 75
ko 빠른 설정 가이드 83
nl Beknopte
handleiding 91
pl Podręcznik
szybkiejkonguracji 99
pt Guia
deConguraçãoRápida 107
ru Руководство по быстрой настройке 115
sv Snabbinstallationsguide 123
th คู่มือการตั้งค่าอย่างรวดเร็ว 131
zh 快速操作指南 139
©Mettler-ToledoGmbH10/2019.
Subjecttotechnicalchanges.PrintedinSwitzerland.30300498B
6
5 Struktura menu
Menu Screen
M300
S
Favorite
c
Calibration
C
Configuration
iMonitor
Messages
ISM Diagnostics
Calibration Data
Sensor Info
HW/SW Version
Set Favorite
Calibrate Sensor
Calibrate Electronics
Calibrate Meter
Calibrate
Analog Outputs
Calibrate
Analog Inputs
Maintenance
Pharma Waters
Verify
Measurement
Analog Outputs
Set Points
ISM Setup
General Alarm
ISM/Sensor Alarm
Clean
Display Setup
Digital Inputs
System
PID controller
Service
User Management
Reset
USB Output
14
5 Menustruktur
Menu Screen
M300
S
Favorite
c
Calibration
C
Configuration
iMonitor
Messages
ISM Diagnostics
Calibration Data
Sensor Info
HW/SW Version
Set Favorite
Calibrate Sensor
Calibrate Electronics
Calibrate Meter
Calibrate
Analog Outputs
Calibrate
Analog Inputs
Maintenance
Pharma Waters
Verify
Measurement
Analog Outputs
Set Points
ISM Setup
General Alarm
ISM/Sensor Alarm
Clean
Display Setup
Digital Inputs
System
PID controller
Service
User Management
Reset
USB Output
Content
1 SafetyandIntendedUse 28
2 TechnicalDataandFurtherInformation 29
3 Mounting 29
4 ElectricalConnection 29
5 MenuStructure 30
6 OperationOn-Site 31
7 SensorCalibration 33
8 Maintenance 34
9 Disposal 34
10 EC-DeclarationofConformity 34
English
Quick Setup Guide
Transmitter M300
28
1 Safety and Intended Use
h
Note:TheQuickSetupGuideisabriefoperatinginstruction.
TheM300transmittermustonlybeinstalled,connected,commissioned,andmaintainedby
qualifiedspecialistse.g.electricaltechniciansinfullcompliancewiththeinstructionsinthis
QuickSetupGuide,theapplicablenormsandlegalregulations.
ThespecialistmusthavereadandunderstoodthisQuickSetupGuideandmustfollowthe
instructionsitcontains.IfyouareunclearonanythinginthisQuickSetupGuide,youmust
readtheOperationManual(suppliedonCD-ROM).TheOperationManualprovidesdetailed
informationontheinstrument.
TheM300transmittershouldbeoperatedonlybypersonnelfamiliarwiththetransmitterand
whoarequalifiedforsuchwork.
Intended Use
TheM300isa4-wiretransmitterforanalyticalmeasurementswith4(0)to20mAoutput
signal.TheM300isamulti-parametertransmitterforpH/ORP,conductivity,dissolvedoxygen
anddissolvedozonemeasurementandavailableas1-channelor2-channelversion.Itis
compatiblewithanalogandISMsensors.
TheM300transmitterisdesignedforuseintheprocessindustries,innon-hazardousareas.
M300 parameter fit guide
Parameter M300 Process M300 Water
1)
M300 Water Cond/Res
pH/ORP
pH/pNa
UniCond2-e/4-e
Conductivity2-e/4-e Analogsensorsonly
Amp.DissolvedOxygenppm/ppb •/•
2)
–/•
2)
DissolvedOzone
1) Measuredtemperatureshigherthan100°Carenotdisplayed.
2) THORNTONHighperformancedissolvedoxygensensoronly
29
2 Technical Data and Further Information
Themostimportanttechnicaldatasuchassupplyvoltagearegivenonthenameplateeither
outsideorinsidethetransmitterhousing.Forfurthertechnicaldatasuchasaccuracyreferto
theOperationManual.Thisdocument,theOperationManualandsoftwareareonthesupplied
CD-ROM.YoucanalsodownloadthedocumentationviaInternet“www.mt.com/M300“.
3 Mounting
TheM300transmitterisavailableas½DINand¼DINversion.
ForinstallationdrawingsrefertotheOperationManual.
a
DANGER! Mortal danger by electric shock or risk of electrical shock:Themaximumscrew-
indepthofthemountingholesinthehousingis12mm(0.47inch).Donotexceedmaximum
screw-indepth.
1. ½DINonly:Mountthesuppliedcableglandsatthehousing.
2. Mountthetransmitter.Youhavethefollowingpossibilities:
Panelmounting:½DINand¼DINversion
Wallmounting:½DIN
Pipemounting:½DIN
4 Electrical Connection
a
DANGER! Mortal danger by electric shock:Poweroffinstrumentduringelectricalconnection.
1. Switchoffsupplyvoltage.
2. ConnectmainssupplytotheterminalsL,N,and
(Ground).
3. 1-channelversion:ConnectsensortoterminalblockTB3.
2-channelversion:ConnectsensoreithertoterminalblockTB3orTB4.
4. ConnectanalogoutputanddigitalinputsignalstoterminalblockTB2(TB2A,TB2B).
5. ConnectrelayoutputsignalstoterminalblockTB1.
ForterminaldefinitionsrefertotheOperationManual.
30
5 Menu Structure
Menu Screen
M300
S
Favorite
c
Calibration
C
Configuration
iMonitor
Messages
ISM Diagnostics
Calibration Data
Sensor Info
HW/SW Version
Set Favorite
Calibrate Sensor
Calibrate Electronics
Calibrate Meter
Calibrate
Analog Outputs
Calibrate
Analog Inputs
Maintenance
Pharma Waters
Verify
Measurement
Analog Outputs
Set Points
ISM Setup
General Alarm
ISM/Sensor Alarm
Clean
Display Setup
Digital Inputs
System
PID controller
Service
User Management
Reset
USB Output
31
6 Operation On-Site
1
2
3
4
5
A B
C
1
A Start screen (example)
1 Changingbetweenchannel1andchannel2,only2-channelversions
2 1stline,standardconfiguration
3 2ndline,standardconfiguration
4 3rdline,dependsonconfiguration
5 4thline,dependsonconfiguration
B Menu screen (example)
C ISM Menu screen
32
Operating element Description
*
EnterMessagesmenu
F
EnterMenuscreen
s
EnterStartscreen
EnterISMmenu
S
EnterFavoritemenu
c
EnterCalibrationmenu
C
EnterConfigurationmenu
H
ReturntoMenuscreen
c
Enternext-lowermenulevel,heree.g.iMonitor,
MessagesorISMDiagnostics
p
Returntonext-highermenulevel
< >
Changebetweenpageswithinonemenulevel
Changebetweenchannel1andchannel2,only
2-channelversions
33
7 Sensor Calibration
h
Note:Forbestprocesscalibrationresultsobservethefollowingpoints.Takegrabsampleas
closeaspossibletothemeasurementpointofthesensor.Measurethesampleatprocess
temperature.
Foradetaileddescriptionofthe“Process”,“1-Point”and“2-Point”calibrationmethodsee
operatinginstructionsoftheM300transmitter.Itisalsopossibletoconnectpre-calibrated
sensors,calibratedviaiSensesoftware.
Assoonasthecalibrationisinprogressnoothercalibrationcanbestarted.
Sensor Calibration Menu c
Aftereverysuccessfulcalibrationdifferentoptionsareavailable.If”Adjust”,“SaveCal”or”Cal
-
ibrate”ischosen,themessage”Calibrationsavedsuccessfully!Reinstallsensor”isdisplayed.
Press”Done”.
Option Analog sensors ISM (digital) sensors
Analog sensors:
Save Cal
ISM sensors:
Adjust
Calibrationvaluesarestoredinthe
transmitterandusedforthemeasure-
ment.Additionally,thecalibrationval-
uesarestoredinthecalibrationdata.
Calibrationvaluesarestoredinthesen-
sorandusedforthemeasurement.Ad-
ditionally,thecalibrationvaluesare
storedinthecalibrationhistory.
Calibrate Thefunction“Calibrate”isnotapplica-
bleforanalogsensors.
Calibrationvaluesarestoredinthecali-
brationhistoryfordocumentation,but
notbeusedforthemeasurement.The
calibrationvaluesfromthelastvalid
adjustmentarefurtherusedforthemea-
surement.
Cancel Calibrationvaluesarediscarded. Calibrationvaluesarediscarded.
34
8 Maintenance
Thetransmitterrequiresnomaintenance.
Cleanthesurfaceswithasoftdampclothanddrythesurfaceswithaclothcarefully.
9 Disposal
Observetheapplicablelocalornationalregulationsconcerningthedisposalof“Wasteelectri-
calandelectronicequipment”.
Disassemblethetransmitteraccordingtoresources.Sortresourcesandsupplythemtorecy-
cling.Non-recyclablematerialshavetobedisposedofinanenvironmental-friendlymanner.
10 EC-Declaration of Conformity
TheEC-Declarationofconformityispartofthedelivery.
54
5 Valikkorakenne
Menu Screen
M300
S
Favorite
c
Calibration
C
Configuration
iMonitor
Messages
ISM Diagnostics
Calibration Data
Sensor Info
HW/SW Version
Set Favorite
Calibrate Sensor
Calibrate Electronics
Calibrate Meter
Calibrate
Analog Outputs
Calibrate
Analog Inputs
Maintenance
Pharma Waters
Verify
Measurement
Analog Outputs
Set Points
ISM Setup
General Alarm
ISM/Sensor Alarm
Clean
Display Setup
Digital Inputs
System
PID controller
Service
User Management
Reset
USB Output
62
4. CsatlakoztassaazanalógkimenetiésdigitálisbemenetijeleketaTB2(TB2A,TB2B)sor-
kapocshoz.
5. CsatlakoztassaarelékimenetijeleitaTB1sorkapocshoz.
AkijelződefinícióitlásdaFelhasználóikézikönyvben.
5 Menüstruktúra
Menu Screen
M300
S
Favorite
c
Calibration
C
Configuration
iMonitor
Messages
ISM Diagnostics
Calibration Data
Sensor Info
HW/SW Version
Set Favorite
Calibrate Sensor
Calibrate Electronics
Calibrate Meter
Calibrate
Analog Outputs
Calibrate
Analog Inputs
Maintenance
Pharma Waters
Verify
Measurement
Analog Outputs
Set Points
ISM Setup
General Alarm
ISM/Sensor Alarm
Clean
Display Setup
Digital Inputs
System
PID controller
Service
User Management
Reset
USB Output
94
4. Sluitanalogeuitgangs-endigitaleingangssignalenaanopterminalblokTB2(TB2A,
TB2B).
5. SluitrelaisuitgangssignalenaanopterminalblokTB1.
Raadpleegvoorterminaldefinitiesdebedieningshandleiding.
5 Menustructuur
Menu Screen
M300
S
Favorite
c
Calibration
C
Configuration
iMonitor
Messages
ISM Diagnostics
Calibration Data
Sensor Info
HW/SW Version
Set Favorite
Calibrate Sensor
Calibrate Electronics
Calibrate Meter
Calibrate
Analog Outputs
Calibrate
Analog Inputs
Maintenance
Pharma Waters
Verify
Measurement
Analog Outputs
Set Points
ISM Setup
General Alarm
ISM/Sensor Alarm
Clean
Display Setup
Digital Inputs
System
PID controller
Service
User Management
Reset
USB Output
102
4. PodłączyćsygnaływyjściaanalogowegoiwejściacyfrowegodolistwyzaciskowejTB2
(TB2A,TB2B).
5. PodłączyćsygnaływyjściaprzekaźnikadolistwyzaciskowejTB1.
ObjaśnienieoznaczeńzaciskówznajdujesięwInstrukcjiobsługi.
5 Struktura menu
Menu Screen
M300
S
Favorite
c
Calibration
C
Configuration
iMonitor
Messages
ISM Diagnostics
Calibration Data
Sensor Info
HW/SW Version
Set Favorite
Calibrate Sensor
Calibrate Electronics
Calibrate Meter
Calibrate
Analog Outputs
Calibrate
Analog Inputs
Maintenance
Pharma Waters
Verify
Measurement
Analog Outputs
Set Points
ISM Setup
General Alarm
ISM/Sensor Alarm
Clean
Display Setup
Digital Inputs
System
PID controller
Service
User Management
Reset
USB Output
126
5 Menystruktur
Menu Screen
M300
S
Favorite
c
Calibration
C
Configuration
iMonitor
Messages
ISM Diagnostics
Calibration Data
Sensor Info
HW/SW Version
Set Favorite
Calibrate Sensor
Calibrate Electronics
Calibrate Meter
Calibrate
Analog Outputs
Calibrate
Analog Inputs
Maintenance
Pharma Waters
Verify
Measurement
Analog Outputs
Set Points
ISM Setup
General Alarm
ISM/Sensor Alarm
Clean
Display Setup
Digital Inputs
System
PID controller
Service
User Management
Reset
USB Output
134
 
Menu Screen
M300
S
Favorite
c
Calibration
C
Configuration
iMonitor
Messages
ISM Diagnostics
Calibration Data
Sensor Info
HW/SW Version
Set Favorite
Calibrate Sensor
Calibrate Electronics
Calibrate Meter
Calibrate
Analog Outputs
Calibrate
Analog Inputs
Maintenance
Pharma Waters
Verify
Measurement
Analog Outputs
Set Points
ISM Setup
General Alarm
ISM/Sensor Alarm
Clean
Display Setup
Digital Inputs
System
PID controller
Service
User Management
Reset
USB Output
Mettler-Toledo GmbH
ProcessAnalytics
ImHackacker15
CH-8902Urdorf,Switzerland
Phone +41447296211
Fax +41447296636
www.mt.com/pro
Mettler-Toledo Thornton, Inc.
900MiddlesexTurnpike,Bldg.8,Billerica,MA01821,USA
Phone +17813018600,Tollfree+18005107873(inUSonly)
Fax +17812710214
www.mt.com/thornton
Subjecttotechnicalchanges.
©Mettler-ToledoGmbH10/2019.
PrintedinSwitzerland.30300498B
*30300498B*
ForaddressesofMETTLERTOLEDO
MarketOrganizationspleasegoto:
www.mt.com/pro-MOs
/