7 Sensor Calibration
h
Note:Forbestprocesscalibrationresultsobservethefollowingpoints.Takegrabsampleas
closeaspossibletothemeasurementpointofthesensor.Measurethesampleatprocess
temperature.
Foradetaileddescriptionofthe“Process”,“1-Point”and“2-Point”calibrationmethodsee
operatinginstructionsoftheM300transmitter.Itisalsopossibletoconnectpre-calibrated
sensors,calibratedviaiSensesoftware.
Assoonasthecalibrationisinprogressnoothercalibrationcanbestarted.
Sensor Calibration Menu c
Aftereverysuccessfulcalibrationdifferentoptionsareavailable.If”Adjust”,“SaveCal”or”Cal
-
ibrate”ischosen,themessage”Calibrationsavedsuccessfully!Reinstallsensor”isdisplayed.
Press”Done”.
Option Analog sensors ISM (digital) sensors
Analog sensors:
Save Cal
ISM sensors:
Adjust
Calibrationvaluesarestoredinthe
transmitterandusedforthemeasure-
ment.Additionally,thecalibrationval-
uesarestoredinthecalibrationdata.
Calibrationvaluesarestoredinthesen-
sorandusedforthemeasurement.Ad-
ditionally,thecalibrationvaluesare
storedinthecalibrationhistory.
Calibrate Thefunction“Calibrate”isnotapplica-
bleforanalogsensors.
Calibrationvaluesarestoredinthecali-
brationhistoryfordocumentation,but
notbeusedforthemeasurement.The
calibrationvaluesfromthelastvalid
adjustmentarefurtherusedforthemea-
surement.
Cancel Calibrationvaluesarediscarded. Calibrationvaluesarediscarded.