41
en
7 Sensor Calibration
h
Note:Forbestprocesscalibrationresultsobservethefollowingpoints.Takegrabsampleas
closeaspossibletothemeasurementpointofthesensor.Measurethesampleatprocess
temperature.
Foradetaileddescriptionofthe“Process”,“1-Point”and“2-Point”calibrationmethodseeOp
-
erationManualoftheM800transmitter.ViaiSensesoftwareyoucancalibratethesensorwith
the“1-Point”or“2-Point”calibratemethod.SeeOperationManualoftheiSensesoftware.
Assoonasthecalibrationisinprogressnoothercalibrationcanbestarted.
Sensor Calibration Menu c
Aftereverysuccessfulcalibrationdifferentoptionsareavailable.If”Adjust”,“SaveCal”or”Cali-
brate”ischosen,themessage”Calibrationsavedsuccessfully!Reinstallsensor”isdisplayed.
Press”Done”.
Option Analog sensors
(1-channel version only)
ISM (digital) sensors
Analog sensors:
Save Cal
ISM sensors:
Adjust
Calibrationvaluesarestoredinthe
transmitterandusedforthemeasure-
ment.Additionally,thecalibrationvalues
arestoredinthecalibrationdata.
Calibrationvaluesarestoredinthesen-
sorandusedforthemeasurement.Ad-
ditionally,thecalibrationvaluesare
storedinthecalibrationhistory.
Calibrate Thefunction“Calibrate”isnotapplicable
foranalogsensors.
Calibrationvaluesarestoredinthecali-
brationhistoryfordocumentation,but
notbeusedforthemeasurement.The
calibrationvaluesfromthelastvalid
adjustmentarefurtherusedforthemea-
surement.
Cancel Calibrationvaluesarediscarded. Calibrationvaluesarediscarded.