Quick start guide
si792 E/T 2-wire conductivity transmitter DOC016.53.00796
4–20 mA signal values
Default settings:
0 mS at 4 mA
100 mS at 20 mA
1. Press
CONF, enter 1200 for the “out. 1MNU” menu.
2. Press
ENTER to go to the "o1.4mA" or "o1.20mA" sub-menu.
3. Set the values for 4 mA and 20 mA.
4. Press
CONF, then ENTER to exit the configuration mode.
Calibration
Zero calibration in air
1. Press
CAL, enter passcode: 1001, ENTER. CAL ZRO will be displayed for 3 seconds.
2. Remove and clean the sensor.
Note: Make sure that the sensor is dry.
3. Use the UP ARROW and RIGHT ARROW to change the value until zero is displayed as the
conductivity value.
Note: If the entry of the zero point value takes longer than 6 seconds, the display will alternate
between showing the conductivity value and the temperature value.
4. Press ENTER to confirm.
5. The display will show the new cell constant value and zero point. Press ENTER.
6. Immerse the sensor in the process water.
7. The transmitter remains in the HoLD mode. Press
ENTER again.
The transmitter remains in HoLD mode for approximately 20 seconds after calibration and
returns to the measurement mode.
Note: Repeat the calibration when “Err” is displayed.
Calibration with a calibration solution
Use a calibration solution with a known conductivity value to adjust the transmitter to show
the same value as the solution.
Note: Stabilize the temperature during calibration.
1. Press CAL, enter passcode: 0110, ENTER. CAL SOL will be displayed for 3 seconds. The
transmitter is ready for calibration.
2. Remove and clean the sensor.
3. Immerse the sensor in the calibration solution.
4. Use the
UP ARROW and RIGHT ARROW to enter the known value of the calibration solution.
Note: If the entry of the calibration solution value takes longer than 6 seconds, the display will
alternate between showing the cell constant and the temperature value.
5. Press ENTER to save the value.
6. The display will show the new cell constant. Press
ENTER.
7. The transmitter remains in HoLD mode. Press
ENTER again.
The transmitter remains in HoLD mode for approximately 20 seconds after calibration and
returns to the measurement mode.
The display will show the temperature-corrected value of the calibration solution (unless
temperature compensation is set to off in the configuration menu).
Note: Repeat the calibration when “Err” is displayed.
DANGER
Explosion hazard. This Quick Start Guide is intended for setup of the si792
transmitter in a non-hazardous location. It is not a substitute for the instructions in
the user manual (DOC026.53.00796). Installation in a hazardous location must follow
the applicable control drawing or ATEX certificate (refer to the documentation CD).
DANGER: Electric shock hazard. Disconnect power to all wiring before making
connection to the si792 E/T series transmitters.
Connect 24 VDC power and communications
Remove the power source before making any electrical connections to the si792 transmitter.
Connect the sensor—Hach GLI 3700 series inductive conductivity sensors
For sensors other than Hach 3700 series, refer to the user manual.
Basic configuration—4–20 mA
1. Press CONF and enter 1200 to enter the configuration mode. The following menus can
be changed:
4–20 mA current output—out.1MNU
Temperature compensation—tc MNU
Alarms—ALrt MNU
2. To move to a different menu, press the
RIGHT ARROW ">", then press enter.
3. To change a setting within a menu, use the
UP ARROW "^" or RIGHT ARROW ">" buttons
to change the setting.
4. Press
ENTER to move to the next item in the menu.
To exit the configuration mode at any time, press
CONF and then ENTER. Refer to the user
manual for additional configuration options.
Terminal No.
si792 transmitter version
4–20 mA/HART Profibus PA Foundation Fieldbus
10
(+) (–) (–)
11
(–) (+) (+)
Jumper Terminal No. 3700 series color 7MA2200 series color
1 white green
Jumper between terminals 2
and 3
2blue yellow
3 yellow white
4 green brown
5 shield violet/black
7 blue/grey
—8redred/pink
Calibration by comparison/grab sample
The sensor can be calibrated by measuring the process water with a portable instrument or
taking a grab sample and measuring the value in the lab (one-point calibration). The sensor
does not need to be removed from the process during calibration. The sample temperature
should correspond to the measured process temperature for accurate results.
Note: The calibration by comparison option is not available when the range and units are set
to 000.0%.
1. Press CAL, enter passcode: 1105, ENTER. The display will show CAL PRD and then StorE.
2. Measure the process water with a portable instrument or collect a grab sample for mea-
surement in the lab. Press
ENTER to proceed.
The cal mode indicator will flash to indicate that calibration has not been completed. The
transmitter will continue to measure and display the current sample values.
3. Measure the grab sample with a laboratory or portable meter.
4. Press
CAL, enter passcode: 1105, ENTER to access the product calibration once more.
The display will briefly show CAL PRD and then CALC.
5. Edit the displayed value to match the value measured by the laboratory or portable meter
and press
ENTER. The display will show the new cell constant.
6. Press
ENTER to end the calibration. The display will alternate between the measured value
and Hold.
Press
ENTER to return to the measuring mode. The outputs will remain in the hold mode for
approximately 20 seconds.
Note: For comparison calibration via Foundation Fieldbus, see the User Manual.
Table 1 Default passcode settings
Display Function Default setting
Cal INFO Calibration information 0000
CAL 0-CAL Calibration zero 1001
CAL CELL Calibration—cell factor adjustment 1100
CAL SOL Calibration with standard solution 0110
1
CAL PROD Product calibration 1105
1
CAL RTD Temperature sensor adjustment 1015
CFG ERR Error information 0000
CFG CONF Configuration mode 1200
CFG SNSR MO Sensor monitor 2222
CFG OUT Current source 5555
CFG SPCL ST
Administrator passcode. Use
ARROW keys to
select NO or YES
1989
NO SPCL ST
Do not change
administrator
passcode
Press
ENTER to return to the default
Press
CONF to exit the editor
YES SPCL ST
Change administrator
passcode
Press
ENTER to accept the new passcode
Press
CONF to exit the editor
1
Not available with si792x E/T-PA
  • Page 1 1
  • Page 2 2

Hach si792 E Quick start guide

Category
Measuring, testing & control
Type
Quick start guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI