Mettler Toledo pH transmitter pH 2800X Operating instructions

  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo pH transmitter pH 2800X Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
pH Transmitter 2800 X
Your Consultant:
02/97
52 120 694
Mettler-Toledo GmbH, Process, Postfach, 8902 Urdorf, Switzerland
Phone +41 (01) 736 22 11, Fax +41 (01) 736 26 36
pH Transmitter 2800X
Software Version: 2.x
TA–187.300–MTE01 020899
Warranty
Defects occurring within 3 years from delivery date
shall be remedied free of charge at our works
(carriage and insurance paid by sender).
Accessories and display backlighting: 1 year
Information III
Safety Information
Be sure to read and observe the following requirements
!
Before connecting the apparatus to mains, make sure that the mains voltage corre-
sponds to the voltage rating given on the nameplate.
Opening the apparatus exposes live parts. Therefore, the apparatus shall not be
opened. If repair should be required, return the apparatus to our factory.
If opening the apparatus is inevitable, it shall first be disconnected from all voltage
sources.
Make sure that the mains plug has been pulled out.
Repair or adjustment of an opened apparatus under voltage shall be carried out only
by a skilled person who is aware of the hazard involved.
Remember that the voltage across accessible parts of the open apparatus may be
dangerous to life.
Whenever it is likely that the protection has been impaired, the apparatus shall be
made inoperative and secured against any unintended operation.
The protection is likely to be impaired if, for example:
S the apparatus shows visible damage
S the apparatus fails to perform the intended measurements
S after prolonged storage at temperatures above 70 °C
S after severe transport stresses
Before recommissioning the apparatus, a professional routine test according to
EN 610101 shall be performed. This test should be carried out at our factory.
pH Transmitter 2800X
IV Information
Installation and Commissioning
Installation of the pH Transmitter 2800X must be
carried out only by specially trained personnel in
accordance with the relevant regulations and this
instruction manual. Be sure to observe the techni-
cal specifications and input ratings.
For information on installation, refer to chapter 1.
Commissioning of the pH Transmitter must be car-
ried out only by specially trained personnel in ac-
cordance with the relevant regulations and this
instruction manual.
Before first startup, a complete parameter setting
procedure must be performed by a system admin-
istrator.
At ambient temperatures below 0 °C LC display
readability might be restricted. This does not affect
instrument function.
Realtime clock, logbook, cal record and sensor
statistics are battery backed for approx. 1 year.
After longer power outages these data can be lost.
In that case the pH Transmitter displays Warn
Time/Date and the date is reset to 01011990.
Time and date must be updated.
Information V
Information on Electromagnetic Compatibility
Compliance with Interference Im-
munity Requirements
The pH Transmitter 2800X provides an immunity
to radiated interferences of 10 V/m (according to
NAMUR requirements). This field strength is gen-
erated for example by portable transceivers with
approx. 10 W output (radiotelephone) at a dis-
tance of 1 m.
All inputs and outputs of the pH Transmitter 2800X
are electrically isolated to each other.
The pH Transmitter 2800X meets the following ge-
neric standards:
S Electromagnetic Emission
EN 500811 01.92 Domestic, Commercial and
Light Industry
prEN 500812 08.92 Industry
S Immunity to Interference
EN 500821 01.92 Domestic, Commercial and
Light Industry
prEN 500822 07.93 Industry
and can therefore be used on residential, commer-
cial and light industrial premises and in industry.
pH Transmitter 2800X
VI Information
Package Contents and Unpacking
Unpack the instrument carefully.
Check the delivery for transport damages and for
completeness.
The package should contain:
S the pH Transmitter 2800X
S this instruction manual
S any accessories ordered
(for available accessories, see chapter 10)
Applications
The pH Transmitter 2800X and the pH Transmit-
ter 2800X-pH continuously measure pH and tem-
perature in liquids. The pH Transmitter 2800X-
COND also permits conductivity measurements.
The pH Transmitter is designed for industrial ap-
plications.
The weatherproof case is IP 65 protected and al-
lows direct wall mounting on site.
All instruments of the pH Transmitter 2800X Series
are approved for applications in hazardous areas.
Information VII
Description of this Manual
This manual describes
S what you can do with the pH Transmitter 2800X
S how to operate the pH Transmitter 2800X
S what you have to know for installation and
mounting
Warning
Warning means that ignoring the given instruc-
tions may lead to malfunction or damage of the
instrument or other equipment and to personal
injury.
Note
Notes call your attention to important information.
Remarks on Representation
The keys of the pH Transmitter 2800X are repre-
sented as follows:
meas , cal , maint , par , diag
A , " , Y , B , enter
A term printed in bold–faced text is explained in
chapter 15: Technical Terms.
Italics are used to emphasize certain information.
The representation of a menu in this manual can
slightly differ from the display of your pH Transmit-
ter 2800X. This depends on the options your
instrument is equipped with.
If the behavior of your instrument differs from the
description in this manual, check if the manual
corresponds to the software version of your instru-
ment: see page 64.
pH Transmitter 2800X
VIII Information
Blank page.
Contents IX
Contents
Safety Information III. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Installation and Commissioning IV. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Information on Electromagnetic Compatibility V. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Compliance with Interference Immunity Requirements V. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Package Contents and Unpacking VI. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Applications VI. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Description of this Manual VII. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Remarks on Representation VII. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1 Information on Mounting, Installation and Maintenance 11. . . . . . . . . . . . . . . . . . . . . . . .
Mounting 11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
How to mount the pH Transmitter 2800 in the ZU 0124 protective case 15. . . . . . . . . . . . . .
Installation 17. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Maintenance and Cleaning 18. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2 Capabilities of the pH Transmitter 2800X 21. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Overview on the pH Transmitter 2800X 21. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The pH Measurement 23. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SensoCheckr Electrode Monitoring 25. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ORP Measurement 26. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Simultaneous pH and ORP Measurement 27. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
rH Measurement 28. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Temperature Detection 28. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The Passive Outputs/Inputs 29. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The Interface 210. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Unical
79 (X) for Fully Automatic Calibration 210. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Typical Wirings 211. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Terminal Assignments 212. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
pH Transmitter 2800X
X Contents
3 Operating the pH Transmitter 2800X 31. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The User Interface 31. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The Measuring Mode 32. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The Measurement Recorder 34. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The Menu Structure 35. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Menu Operation 36. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4 Parameter Setting 41. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Language Selection 41. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The Three Levels of Parameter Setting 41. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The Factory Setting 43. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The Measurement Display 43. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The Input Filter 44. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Temperature Detection 44. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Temperature Compensation of the Measured Medium 45. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The Calimaticr Buffer Set 46. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The Tolerance Band Calibration 47. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ORP Check 48. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Nominal Electrode Zero and Nominal Electrode Slope 48. . . . . . . . . . . . . . . . . . . . . . . . . . . . .
rH Measurement 49. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The Delta Function 411. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Current Output 1 412. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Outputs/Inputs 2 and 3 412. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The Alarm Settings 419. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The NAMUR Contacts 422. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The V24 / 20 mA Interface/Unical
424. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Setting the Clock 430. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Point of Measurement/Note 430. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Device Diagnostics 430. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measurement Recorder 431. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Passcode Entry 432. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Release of Options 434. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contents XI
5 Calibration 51. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Why do you have to calibrate? 51. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Monitoring Functions for Calibration 51. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The Calibration Menu 52. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
What does First Calibration mean? 53. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Temperature Compensation during Calibration 54. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Single or DualPoint Calibration? 55. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Automatic Calibration with Calimatic
56. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Calibration with Manual Entry of Buffer Values 58. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Calibration by Entry of Premeasured Electrode Data 510. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Calibration by Sampling 510. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
ORP Check 511. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
pH Calibration with Unical
513. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6 The Diagnostics Menu 61. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The Message List 61. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The Electrode Records 61. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The Logbook 64. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The Device Description 64. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The Device Diagnostics 65. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measurement Recorder (Listing) 66. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7 The Maintenance Menu 71. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measurement Point Maintenance 71. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
The Current Source Function 72. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Temperature Probe Adjustment 72. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Manual Entry of Controller Output 73. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8 Error Messages 81. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
pH Transmitter 2800X
XII Contents
9 Interface Commands 91. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contents 91. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Transmission behavior 93. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
VALUE Commands: Query measured values 94. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
STATUS Commands: Query messages and states 95. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
PARAMETER Commands: Query settings and set parameters 99. . . . . . . . . . . . . . . . . . . . .
DEVICE Commands: Device Description 923. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
COMMAND Commands: Control Commands 923. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
10 Product Line and Accessories 101. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
11 Specifications 111. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
EMC Test Records 115. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Certificate of Conformity 116. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
12 Buffer Charts 121. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
13 Unical
Programs 131. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Unical
Program (dualpoint calibration) 131. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Unical
Program (singlepoint calibration) 132. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Unical
Program (ORP check) 133. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Unical
Program (original for copy) 134. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
14 Appendix 141. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
PROM Replacement 142. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
15 Technical Terms 151. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
16 Index 161. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Mounting 11
1 Information on Mounting, Installation and Maintenance
Mounting
S The weatherproof enclosure allows direct wall
mounting. For dimension drawing, see Fig. 11.
S With ZU 0126 mounting plate and ZU 0125
bracket kit, the instrument can also be post or
pipe mounted. For dimension drawing, see Fig.
12.
S ZU 0123 protective hood provides additional
protection against direct weather exposure and
mechanical damage. For dimension drawing,
see Fig. 12.
For mounting the protective hood, you require
ZU 0126 mounting plate.
S ZU 0124 protective case provides optimum
protection against dust, moisture and mechani-
cal damage. For dimension drawing, see Fig.
13.
With ZU 0128 bracket kit, the protective case
can also be post or pipe mounted.
pH Transmitter 2800X
12 Mounting
87 [3.43]
5 [0.2]
72 [2.84]
unit symbols
terminal compartment
control panel
304 [11.97]
295 [11.54]
14.2
6.3
124.5 [4.9]
223
nameplate
26
281 [11.06]
138 [5.43]
Pg 13.5 (9 pces)
[0.25]
[0.56]
Note: All dimensions in millimetres [inches].
Fig. 11 Dimension Drawing pH Transmitter 2800X
245 [9.65]
309
[
12.17
]
holes for wall mounting
for vertical or horizontal
posts or pipes
ZU 0125 bracket kit
ZU 0126 mounting plate
245 [9.65]
123 [4.84]
54 [2.13]
7
175 [6.89] 36 [1.42]
ZU 0123
85 [3.35] 139 [5.47]
3265
60 [2.36]
[1.262.56]
protective hood
Note: All dimensions in millimetres [inches].
Fig. 12 Dimension Drawing ZU 0126 Mounting Plate and ZU 0123 Protective Hood
Mounting 13
front removable for
instrument installation
polycarbonate
protective panel,
352 [13.86]
85.5 [3.37]
175 [6.89]
18 [0.71]
372 [14.65]
302 [11.89]
Pg 13.5
5 [0.2]
282 [11.1]
transparent
19 [0.75]
Note: All dimensions in millimetres [inches].
Observe ESD!
Fig. 13 Dimension Drawing ZU 0124 Protective Case
Pg11
35 [1.38]
35 [1.38]
148 [5.83]
160 [6.30]
45 [1.77]
75 [2.95]
4.4 [0.17]
57 [2.24]
approx. 120 [4.72]
Note: All dimensions in millimetres [inches].
Fig. 14 ZU 0074X Terminal Box
pH Transmitter 2800X
14 Mounting
90 [3.55] 175 [6.90]
31 [1.22]240 [9.46]
3265 [1.262.56]
ZU 0124 protective case
for vertical or horizontal
posts or pipes
Note: All dimensions in millimetres [inches].
Fig. 15 ZU 0128 Bracket Kit for ZU 0124 Protective Case
Mounting 15
How to mount the pH Transmitter
2800 in the ZU 0124 protective
case
Construction
The pH Transmitter 2800X is screwed to the cabi-
net of the protective case via two support rails.
The connection leads are passed through exten-
sion pieces to the bottom part of the protective
case where they are sealed by Pg threaded cable
glands.
Mounting Instructions
S Unscrew all Pg threaded cable glands and their
gaskets from the pH Transmitter 2800X and
save them for later mounting.
S Screw the included extensions (3) with their
corresponding gaskets (2) at the positions of
the Pg cable glands.
S Screw the two support rails (4), using two M4x8
screws and two 4.3 toothed lock washers each,
in the same direction into the cabinet of the pro-
tective case.
Do not tighten the screws before having aligned
the complete unit!
S Screw the pH Transmitter 2800X (using four
M5x16 screws and four 5.3 washers) onto the
two support rails. Do not tighten the screws be-
fore having aligned the complete unit!
S Push the four threaded inserts (5) evenly into
the free blind holes at the lid of the protective
case and spread them a bit apart.
S Screw the front cover (6) to the lid of the pro-
tective case using four screws and toothed lock
washers. Caution! The lid window must be hin-
ged at its bottom part!
S Place the lid onto the opening of the cabinet to
align the pH Transmitter 2800X in the protective
case.
S Screw the Pg cable glands with gaskets into
the extension pieces.
S Remove the lid and handtighten all fastening
screws.
S Make the electrical connections to the
pH Transmitter 2800X (see page 17).
S Fasten the lid to the cabinet using the four
assembly screws.
pH Transmitter 2800X
16 Mounting
cabinet
M4x8 screw M5x16 screw
M4x8 screw
threaded insert
nameplate
assembly screw
lid
Pg threaded cable gland
gasket
extension
gasket
pH Transmitter 2800X
front cover
plug
rail
4.3 toothed lock
washer (4x)
(4x) (4x)
5.3 washer
(4x) (4x)
4.3 toothed lock
washer (4x)
(4x)
(2x)
➃➄
(4x)
Fig. 16 Mounting the pH Transmitter 2800X in the Protective Case
Mounting 17
Installation
Installation of the pH Transmitter 2800X must be
carried out only by specially trained personnel in
accordance with the relevant regulations and this
instruction manual. Make sure that the technical
specifications and input ratings are observed.
Commissioning of the pH Transmitter 2800X must
be carried out only by specially trained personnel
in accordance with the relevant regulations and
this instruction manual.
Before first startup, a complete parameter setting
procedure must be performed by a system admini-
strator.
Observe the requirements of the Certificate of
Conformity and the relevant national installation
regulation.
Before connecting power supply, make sure that
your mains supply corresponds to the ratings on
the instruments nameplate:
S 230 Vac
S 115 Vac (option 363)
S 24 Vac/dc (option 298)
After connection of power supply current outputs
and contacts are frozen at the state before power
failure for approx. 10 s. This ensures that after
poweron no invalid messages are generated.
To connect the pH Transmitter 2800X, open the
cover of the terminal compartment (lower part of
the instrument) by removing the three screws.
The terminals are suitable for solid wires and
stranded wires up to 2.5 mm
2
.
At the left side of terminal 1 there is a connecting
strap with two clamping screws for connecting the
electrode cable shield (Fig. 213, Pg. 212).
This strap is electrically connected to terminal 3!
(See also wiring examples on pages 25 ff.)
As delivered, all terminals are open in order to al-
low easy insertion of the connecting wires.
If the terminals are only half open, it may occur
that the wire is pushed below the contacting ele-
ment and thus does not contact when the terminal
is closed.
Should your pH Transmitter be supplied with a
German assignment label, simply pull it out and
turn it over to read the English text.
pH Transmitter 2800X
18 Mounting
Maintenance and Cleaning
The pH Transmitter 2800X is maintenance free.
To remove dust, dirt and stains, the outer surfaces
of the instrument may be wiped using a soft, non
fluffing cloth moistened with water. If required, you
may also use a mild detergent.
When operating the pH Transmitter within a haz-
ardous area, prevent electrostatical charging!
Never wipe the unit with a dry cloth, for example.
/