Hach si792 D series User manual

Category
Carbon monoxide (CO) detectors
Type
User manual

This manual is also suitable for

DOC026.52.00797
si792 D
si792x D
si792x D-FF
si792x D-PA
Dissolved Oxygen
2-Wire Transmitters
USER MANUAL
03/2014, Edition 8
2
1
Table of Contents
Section 1 Specifications ......................................................... 5
Section 2 General Information........................................... 11
2.1 Safety information .................................................................... 11
2.1.1 Use of hazard information ................................................. 11
2.1.2 Precautionary labels.......................................................... 11
2.2 General product information..................................................... 13
2.2.1 Product overview............................................................... 13
2.2.2 FDA 21 CFR part 11 Compliance (HART only)................. 13
2.2.2.1 Electronic signatures for si792(x) D transmitters ....... 14
2.2.2.2 Audit trail for si792(x) D transmitters.......................... 14
Section 3 Installation.............................................................. 15
3.1 Hazardous location................................................................... 16
3.2 Unpacking the transmitter ........................................................ 16
3.3 Mechanical installation ............................................................. 18
3.3.1 Transmitter assembly ........................................................ 18
3.3.2 Mounting ........................................................................... 18
3.3.2.1 Wall mount ................................................................. 18
3.3.2.2 Panel or pipe mount (optional) ................................... 18
3.4 Wiring Safety Information ......................................................... 20
3.5 Electrical installation................................................................. 21
3.5.1 Wire preparation................................................................ 21
3.5.2 Power and communication connections............................ 23
3.5.2.1 si792(x) D (4–20 mA/HART) wiring............................ 24
3.5.2.2 si792x D-FF and si792x D-PA wiring ......................... 25
3.5.3 Sensor wire connection ..................................................... 25
3.5.3.1 Oxygold G/Oxyferm sensor—si792(x) D
(4–20 mA/HART)................................................................. 26
3.5.3.2 Oxygold G/Oxyferm sensor—si792x D-FF or
si792x D-PA ........................................................................ 27
3.5.3.3 Oxysens sensor—si792(x) D (4–20 mA/HART)......... 28
3.5.3.4 Oxysens sensor—si792x D-FF and si792x D-PA ...... 29
3.6 HART communication connection ............................................ 30
Table of Contents
2
Section 4 Interface and navigation.................................. 31
4.1 si792(x) D (4–20 mA/HART) interface...................................... 31
4.2 si792x D-FF and si792x D-PA interface ................................... 32
4.3 Display...................................................................................... 33
Section 5 Operation—si792(x) D
(4–20 mA/HART)......................................................................... 35
5.1 Measure mode ......................................................................... 35
5.2 Configuration ............................................................................ 35
5.2.1 Output configuration .......................................................... 36
5.2.1.1 Output range .............................................................. 38
5.2.1.2 Time averaging filter................................................... 38
5.2.1.3 Output signal during errors......................................... 38
5.2.1.4 Output signal during HOLD ........................................ 39
5.2.2 Temperature sensor configuration ....................................40
5.2.3 Correction menu configuration .......................................... 40
5.2.4 Calibration mode configuration.......................................... 41
5.2.5 Alarm settings configuration .............................................. 41
Section 6 Operation—Foundation Fieldbus................ 43
6.1 Configuration ............................................................................ 43
6.1.1 Configuration steps ........................................................... 43
6.1.2 Configuration menu ........................................................... 44
6.2 Foundation Fieldbus communication........................................ 46
6.2.1 Standard resource block (RB) ........................................... 46
6.2.2 Standard analog input block (AI) .......................................49
6.2.2.1 Operating modes........................................................ 49
6.2.2.2 Set the parameter and units ....................................... 49
6.2.2.3 Data processing ......................................................... 50
6.2.2.4 Alarms ........................................................................ 50
6.2.2.5 Bus parameters for the analog input block................. 51
6.2.2.6 Cyclic measured value status..................................... 54
6.2.2.7 Measured value limits—limit bits................................ 54
6.2.3 Transducer block............................................................... 55
6.2.4 Calibration via Foundation Fieldbus ..................................63
Table of Contents
3
Section 7 Operation—Profibus PA................................... 65
7.1 Configuration ............................................................................ 65
7.1.1 Configuration steps ........................................................... 65
7.1.2 Configuration menu........................................................... 66
7.2 Profibus PA communication ..................................................... 67
Section 8 Calibration.............................................................. 71
8.1 Calibration ................................................................................ 71
8.1.1 Zero calibration ................................................................. 72
8.1.2 DO calibration (default) ..................................................... 73
8.1.2.1 Calibration in air (recommended)............................... 73
8.1.2.2 Calibration in water .................................................... 74
8.1.3 Gas calibration (volume concentration)............................. 74
8.1.4 Calibration by comparison................................................. 75
8.2 Temperature sensor adjustment .............................................. 76
Section 9 Maintenance........................................................... 77
9.1 Cleaning the instrument ........................................................... 77
9.2 Sensor maintenance ................................................................ 77
Section 10 Troubleshooting................................................ 79
10.1 Sensoface .............................................................................. 79
10.2 Sensocheck............................................................................ 79
10.3 Error codes............................................................................. 80
10.4 Calibration errors.................................................................... 82
10.5 Diagnostic functions ............................................................... 83
Section 11 Parts and accessories.................................... 85
11.1 si792 D series versions .......................................................... 85
11.2 Accessories............................................................................ 85
Appendix A Passcode editor.............................................. 87
Table of Contents
4
5
Section 1 Specifications
Specifications are subject to change without notice.
Transmitter
Composition PBT (polybutylene terephthalate)
Display LCD
Fittings
3 knockouts for M20 x 1.5 strain reliefs
2 knockouts for ½ inch NPT or rigid
metallic conduit
Power Requirements—HART
12–30 VDC (30 VDC maximum)
100 mA maximum
0.8 W maximum (Ex)
Power requirements—FF and
Profibus PA
FISCO bus supply: 9 to 17.5 VDC
Linear barrier: 9 to 24 VDC
Loop current—HART
4–20 mA floating;
3.80–22.00 mA specifiable
Current consumption—FF and
Profibus PA
<13.2mA
Maximum current in case of fault
(FDE)—FF and Profibus PA
<17.6mA
Measurement error
1
<0.3% current value + 0.05 mA
Specifications
6
Certification (may not apply to all
sensors. Refer to the control drawing or
listing for certification information for
the sensor that is used)
US:
si792 D
FM listed for:
Class I, Division 2
si792x D; si792x D-FF; si792x D-PA
FM listed for:
Class I, Division 1, Groups A, B, C, D
Class II, Division 1, Groups E, F,
Class III, Division 1
Class I, Zone 0, AEx ia, Group IIC T4
Enclosure: Type 2
Canada:
si792 D CSA certified to:
Class I, Division 2
si792x D CSA certified and
si792x C-FF; si792x D-PA cFMus
certified to:
Class I, Division 1, Groups A, B, C, D
Class I, Division 2, Groups A, B, C, D
Sensor: Class I, Zone 0, Group IIC
Transmitter: Class I, Zone 1, Group IIC
Enclosure: Type 2
EU:
si792x P;
ATEX Certification:
II 2 (1) G Ex ib [ia Ga] IIC T6 Gb
si792x D-FF; si792x D-PA
II 2 (1) G Ex ia IIC T4 Ga
CE Marked to:
EMC Directive 2004/108/EC
ATEX Directive 94/9/EC
Enclosure: IP65
Lightning protection EN 61000-4-5
Output averaging time constant (HART) 0–120 seconds
Output span allowed
2%–500%
200–9999 µg/L (ppb)
0.5–50 mg/L (ppm)
Specifications
7
Storage Temperature –20 to 70 °C (–4 to 158 °F)
Operating Temperature –20 to 55 °C (–4 to 131°F)
Weight Approximately 1 kg
Dimensions H 144 mm, W 144 mm, D 105 mm
Data retention
Parameters and calibration data
>10 years (EEPROM)
Passcodes
Modifiable according to FDA 21 CFR
Part 11 “Electronic Signatures”
Sensocheck
Automatic monitoring for short circuits
or open circuits (can be disabled).
Delay: 30 seconds.
Probe monitor
Direct display of uncorrected sensor
signal (sensor current/temperature)
Communications
HART communication
Digital communication by FSK
modulation of loop current, reading of
device identification, measured values,
status and messages, reading and
writing of parameters, start of product
calibration, signaling of configuration
changes according to FDA 21 CFR
Part 11.
Foundation Fieldbus (FF_H1)
Bus-powered device with constant
current consumption. Cyclic and acyclic
data exchange. 1 resource block,
1 transducer block, 3 analog input
function blocks (switchable: O
2
saturation, O
2
concentration,
temperature, zero, slope, volume
concentration)
Execution time: 50 ms
Certified to ITK 4.6
Physical interface: to IEC 1158-2
Address range: 017 to 246
Specifications
8
Profibus-PA (DPV1)
Bus-powered device with constant
current consumption. Cyclic and acyclic
data exchange. Physical block,
2 analog input function blocks,
2 discrete input blocks, logbook block,
alarm block.
PNO directive: PROFIBUS-PA,
Profile for Process Control Devices,
Version 3.0
Physical interface: MBP-IS
(Manchester Bus Powered-Intrinsically
Safe) to IEC 1158-2 (DIN-EN 61158-2)
Connection: via segment coupler to
SPC, PC, PCS
Address range: 1 to 126
Temperature Input
Probe
NTC 30 k/NTC 22 k (selectable),
2-wire connection
Range –20.0 to 150.0 °C (–4 to 302 °F)
Adjustment range 10 K
Resolution 0.1 °C; 0.1 °F
Measurement error
1
,
2
<0.5 K
(<1 K at T >100 °C)
Dissolved oxygen input
Type A sensor
3
Oxyferm, Oxysens (Hamilton)
Type B sensor
3
Oxygold G (Hamilton)
Measuring current 0–1200 nA; resolution: 20 pA
Measurement error
1
,
2
0.5% measured value + 0.05 nA
Temperature coefficient 0.005 nA/K
Permitted guard current
20 µA
Polarization voltage
400–1000 mV; resolution:
approximately 3 mV
Range, percent saturation 0–199.9%/200–500% (–10 to 80 °C)
Specifications
9
Range, concentration (–10 to 80 °C)
0–50.00 mg/L; 0–50.00 ppm
0–9999 µg/L; 0–9999 ppb
Range, volume concentration in gas
4
0–9999 ppm; 0–120.0% (–10 to 80 °C)
Process pressure 0–9.999 bars (0–999.9 kPa/0–145 PSI)
Pressure correction 0–9.999 bars (0–999.9 kPa/0–145 PSI)
Salinity correction 0–45 g/kg
Probe calibration
Type A sensor
Slope 25–130 nA (25 °C, 1013 mbars)
Zero point ± 2 nA
Range, Type B sensor
Slope 200–550 nA (25 °C, 1013 mbars)
Zero point ± 2 nA
Calibration timer 0 to 9999 hours
1
IEC 746 Part 1, at nominal operating conditions
2
(± 1 count plus sensor error)
3
Sensor specifications given in Table 1.
4
si792(x) D and si792x D-FF only
Table 1 Sensor specifications
Sensor type Screw cap
Sensor
current
in air
(25 °C)
Detection
limit
Temperature
sensor
Type A
sensor
Oxysens
5 meter,
fixed cable
with stripped
ends
40–80 nA 0.04 ppm 22 NTC
Oxyferm VP 40–80 nA 0.01 ppm 22 NTC
Type B
sensor
Oxygold G VP 180–400 nA 0.001 ppm 22 NTC
Specifications
10
11
Section 2 General Information
2.1 Safety information
Please read this entire manual before unpacking, setting up, or
operating this equipment. Pay attention to all danger and caution
statements. Failure to do so could result in serious injury to the operator
or damage to the equipment.
To ensure that the protection provided by this equipment is not
impaired, do not use or install this equipment in any manner other than
that specified in this manual.
2.1.1 Use of hazard information
DANGER
Indicates a potentially or imminently hazardous situation which, if
not avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation that may result in
minor or moderate injury.
Important Note: Information that requires special emphasis.
Note: Information that supplements points in the main text.
2.1.2 Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or
damage to the instrument could occur if not observed. A symbol, if
noted on the instrument, will be included with a danger or caution
statement in the manual.
This symbol, if noted on the instrument, references the instruction
manual for operation and/or safety information.
General Information
12
Electrical equipment marked with this symbol may not be disposed of
in European public disposal systems after 12 August of 2005. In
conformity with European local and national regulations (EU
Directive 2002/96/EC), European electrical equipment users must
now return old or end-of life equipment to the Producer for disposal
at no charge to the user.
Note: For return for recycling, please contact the equipment
producer or supplier for instructions on how to return end-of-life
equipment, producer-supplied electrical accessories, and all auxiliary
items for proper disposal.
This symbol, when noted on a product enclosure or barrier, indicates
that a risk of electrical shock and/or electrocution exists.
This symbol, when noted on the product, identifies the location of the
connection for Protective Earth (ground).
This symbol, when noted on the product, identifies the location of a
fuse or current limiting device.
This symbol, when noted on the product, identifies a risk of chemical
harm and indicates that only individuals qualified and trained to work
with chemicals should handle chemicals or perform maintenance on
chemical delivery systems associated with the equipment.
This symbol, when noted on the product, identifies the presence of a
strong corrosive or other hazardous substance and a risk of chemical
harm. Only individuals qualified and trained to work with chemicals
should handle chemicals or perform maintenance on chemical deliv-
ery systems associated with the equipment.
This symbol, when noted on the product, indicated the presence of
devices sensitive to Electro-static Discharge (ESD) and indicated
that care must be taken to prevent damage with the equipment.
This symbol, when noted on the product, identifies the presence of a
noxious substance and a risk of chemical harm. Only individuals
qualified and trained to work with chemicals should handle chemicals
or perform maintenance on chemical delivery systems associated
with the equipment.
General Information
13
2.2 General product information
2.2.1 Product overview
The si792 D and si792x D transmitters are used for continuous
dissolved oxygen and temperature measurement. Applications include
biotechnology, chemical, pharmaceutical and food processing as well
as environmental monitoring and sewage treatment. This manual
describes the installation, operation and maintenance for standard and
EU models of the si792 D series transmitters.
The transmitter accepts various amperometric sensors, for example:
Hamilton Oxysens, Oxyferm, Oxygold G
The molded transmitter enclosure can be attached to a panel, wall, post
or pipe railing. The optional hood (see Accessories on page 85)
provides protection against direct weather exposure and mechanical
damage.
Three communication options are available for the si792 transmitter:
4–20 mA/HART—si792 D and si792x D (si792(x) D)
Foundation Fieldbus—si792x D-FF
Profibus PA—si792x PD-PA
This user manual includes instructions for all three protocols.
Important Note: An ‘x’ in the model number is an indication of an
instrinsically safe (IS) instrument.
2.2.2 FDA 21 CFR part 11 Compliance (HART only)
In the directive “Title 21 Code of Federal Regulations, 21 CFR Part 11,
Electronic Records; Electronic Signatures“ the US American health
agency FDA (Food and Drug Administration) regulates the production
and processing of electronic documents for pharmaceutical
development and production. The features described in section 2.2.2.1
and section 2.2.2.2 make the transmitter compliant with the
requirements of FDA 21 CFR Part 11.
General Information
14
2.2.2.1 Electronic signatures for si792(x) D transmitters
Device functions are protected by passcode access, which prevents
unauthorized modification of device settings or manipulation of
measurement results. Passcodes may be used as electronic
signatures. Passcodes can be edited with the passcode editor
(Appendix A on page 87).
2.2.2.2 Audit trail for si792(x) D transmitters
The si792(x) D can automatically track all changes to the device
settings. Each change is tagged with a Configuration Change flag,
which is documented using HART communication. Altered device
settings or parameters can be retrieved from the transmitter using
HART communication.
15
Section 3 Installation
DANGER
Explosion hazard. Trained personnel only must install or
commission the equipment.
DANGER
Explosion hazard. Never connect items to the transmitter that are
not specified on the control drawing. Do not connect or
disconnect any equipment unless power has been switched off or
the area is known to be non-hazardous.
DANGER
Explosion hazard. The safety of the transmitter may be impaired if
any of the following conditions have occurred:
visible damage
storage above 70 °C for prolonged periods
exposure to severe transport stresses
previous installation
failure to operate properly
If any of these conditions have occurred, return the device to the
manufacturer for recertification.
The si792 and si792x transmitters differ in hazard classification. The
si792 transmitter is designed for non-hazardous or Class I, Division 2,
Groups A, B, C, D hazardous locations. The si792x transmitter is
designed for Class I, Division 1, Groups A, B, C, D hazardous locations.
Refer to the control drawing or listing applicable to the site location.
For outdoor installation, use of a protective hood or sunshield is
recommended (section 11.2 on page 85).
Installation of the si792x in an outdoor hazardous location per FM or
CSA control drawings requires a suitable enclosure and must follow
NEC guidelines. Refer to NEMA 250 to determine enclosure needs.
Installation
16
3.1 Hazardous location
Before installation, review the applicable Hazardous Location Control
Drawings or ATEX EC-type Examination certificate included with the
instrument and on the provided documentation CD. Follow all
regulations specified for the installation location.
Refer to the documentation CD for manuals provided in
other languages.
3.2 Unpacking the transmitter
Check the shipment for transport damage and make sure all
components have been shipped complete. The package includes:
Display module Bag of hardware and fasteners
Back enclosure Test report and user manual
Installation
17
Figure 1 Instrument Components
1 Back enclosure 8 Strain relief (3x)
2 Optional conduit hardware 9 Filler plug (3x)
3 Conduit washer 10 Enclosure screw (4x)
4 Jumper (2x) 11 Display module
5 Cable tie (3x) 12 Hinge pin
6 Sealing insert 13 Hex nuts (5x)
7 Rubber reducer 14 Sealing plug (2x)
Installation
18
3.3 Mechanical installation
3.3.1 Transmitter assembly
Refer to Figure 1 and the following instructions to assemble the
transmitter.
1. Insert the strain relief fittings in the holes of the back enclosure and
secure with the hex nuts (Figure 2).
2. Insert the conduit hardware or plugs in the back enclosure and
secure with the hex nuts.
3. Attach the display module to the back enclosure using the hinge pin.
3.3.2 Mounting
Refer to the following sections to mount the transmitter on a wall, panel
or pipe.
3.3.2.1 Wall mount
1. Use a punch to open the two wall-mount holes in the back enclosure
(Figure 2).
2. Drill holes in the wall suitable for the user-supplied mounting bolts.
3. Attach the back enclosure to the wall using two customer-supplied
bolts.
4. Insert the clear plastic plugs into the mounting holes.
3.3.2.2 Panel or pipe mount (optional)
Refer to Figure 2 and the instructions supplied with the panel and pipe
mounting kits (see Accessories on page 85).
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96