CL500-en-GB_V1.3 1/16
3
GettingStarted
EquipmentRequired
Achlorinetestrequiresa10mlsamplebottle,reagentpowderfortestingFreeandTotal
chlorine,andtheCL500meter.
MeasurementProcedure
Calibrationconsiderations
1. Ensurethatthemeterhasbeencalibratedpriortouse.
2. Refertothecalibrationsectionforthisprocedure.
Testsample“Zero”measurement
Note:Priortotesting,setthemetermodetoFREEorTOTALbasedonthedesiredtest
method.SeeModeSelection.
1. ThemetershouldbeintheOFFposition.
2. Placethetestsample,withouttheDPDpowderadded,intothesamplebottleuptothe
10mLline(minimum).Place
thecaponthebottleandtighten.
3. Cleananddrytheoutsideofthesamplebottletoensureacleansurface.
4. Openthecoverofthetestchamberandinsertthebottlecompletelyintothechamber.
5. Lineupthewhitemarkonthebottlewiththewhite
markofthetestchamber.
6. Closethecoverandlockitinplace.
7. Pressthe“POWER”buttontoturnthemeterON.Thedisplaywillshow‘Free’or‘Total’
for1secondandthen‘CAL0’.Themeterisnowreadyforthe“zero”test.
8. Pressthe“Zero”button,
thedisplaywillflash“tEST”andthen“0.00”ppm.TheZeroTest
isnowcomplete.Donotpowerdownthemeter.
MeasurementofTestSolution
1. Openthetestchamberandremovethetestbottle.
2. AddtherequiredDPDpowder(FreeorTotal)intothesampleandthenshakefor20
secondsuntilcombined.Placethecaponthebottleandtighten.
FortheFreeChlorinetest,usetheFreeChlorineDPDpowder
Forthe
TotalChlorinetest,usetheTotalChlorineDPDpowder
3. Cleananddrytheoutsideofthesamplebottletoensureacleansurface.
4. Wait2minutebeforeplacingthetestbottleintothemetertoallowtimeforthesolution
tomixthoroughly.
5. Openthecoverof
thetestchamberandinsertthebottlecompletelyintothechamber.
6. Lineupthewhitemarkonthebottlewiththewhitemarkofthetestchamber.