Extech Instruments AN320 User manual

Category
Measuring, testing & control
Type
User manual
USERGUIDE
LargeVaneCFM/CMM
Anemometer/Psychrometer+CO
2

ModelAN320
AN320-en-GB_V1.3 11/15
2
Introduction
CongratulationsonyourpurchaseoftheExtechAN320VaneAirflowAnemometerPsychrometer
+CO
2
.ThisinstrumentmeasuresAirVelocity,AirFlow(volume),AirTemperature,Relative
Humidity,WetBulbTemperature,DewPointTemperature,andCO
2
(carbondioxide).Thelarge,
easytoreadbacklitLCDincludesprimaryandsecondarydisplaysplusnumerousstatusindicators.
Theoptionalroundandsquareairflowconeadaptorsallowforquickairflow(volume)
measurements.Thismeterisshippedfullytestedandcalibratedand,withproperuse,willprovide
yearsofreliableservice.Pleasevisitourwebsite(www.extech.com)tocheckforthelatestversion
ofthisUserGuide,ProductUpdates,andCustomerSupport.
CAUTIONS
Improperuseofthismetercancausedamagetothemeterandpersonalinjury.Readand
understandthisusermanualbeforeoperatingthemeter.
Inspecttheconditionoftheprobeandthemeterforanydamagebeforeoperatingthe
meter.Repairorreplacedamagebeforeuse.
Iftheequipmentisusedinamannernotspecifiedbythemanufacturer,theprotection
providedbytheequipmentmaybeimpaired.
Thisdeviceshouldnotbemadeavailabletochildren.Itcontainshazardousobjectsaswell
assmallpartsthatcanbeaccidentallyswallowed.Themeter’sbatteriesandpacking
material
canalsobedangeroustochildren.
Intheeventthatthemeteristobeunusedforanextendedperiodoftime,removethe
batteriestoprotectagainstbatteryleakage.
Expiredordamagedbatteriescanbehazardousifallowedtocomeincontactwithskin.Use
suitablehandprotectioninsuchcases.
Donotshortcircuitbatteriesorputbatteriesinfire.
AN320-en-GB_V1.3 11/15
3
MeterDescription
MeterDescription(front)
1. ToplineLCDdisplayreading
2. BottomlineLCDdisplayreading
3. U/modeAVGanduparrowbutton
4. L/modeanddownarrowbutton
5. ONOFF(Power)andSETbutton
6. HOLD,MAXMIN,andleftarrow(Enter)button
7. Unitofmeasuresymbolforlowerdisplayrow
8.
Unitofmeasuresymbolforupperdisplayrow
9. AirVelocityVane
MeterDescription(rear)
10. Airtemperature,humidity,andCO
2
sensors
11. Batterycompartment
AN320-en-GB_V1.3 11/15
4
SET
HOLD
MAX/MIN
U/mode
AVG
L/mode
U/mode
AVG
U/mode
AVG
L/mode
BriefKeypadDescription
PressmomentarilytoswitchthemeterONorOFF.Pressandholdthisbuttonfor2
secondstoenterorexittheSetupmode(refertoSetupmodesectionfordetails).
o Pressmomentarilytofreezethedisplayedreading(HOLDiconswitchesON);press
againmomentarilytoexittheHOLDmode(HOLDiconswitchesOFF)
o Pressandholdfor2secondstodisplaytheMIN(minimum)reading;pressandhold
againtodisplaytheMAX(maximum)reading;Pressandholdathirdtimetoexit
o ThisbuttonisalsousedintheAVERAGEmode(forrecordingreadingsoneatatime
orfora‘timed’recordingsession)andintheSETUPmode(forconfiguringthe
meter).PleasereviewtheAVERAGE
andSETUPmodesectionsfordetails
Pressmomentarilytotogglebetweenppm(CO
2
),humidity,andairvelocityreadings.
Pressandholdfor2secondstoaccesstheMultiPointAverageandTimedAverage
modes(theAVGiconswitchesON).Detailedinstructionsforthesemodesareprovided
laterinthisguide.IntheSetupmodepresstoselectacategoryortoincreasea
value
(refertoSetupmodelaterinthisguide)
Presstotogglethelowerdisplaybetweenairtemperature,airvolume,wetbulb
temperature,anddewpointtemperaturereadings;AlsousedintheMultiPoint
Average,TimedAverage,andSetupmodes;refertotheAVERAGEandSETUPmodes
sectionslaterinthisguide

+ WiththemeterswitchedOFF,pressandholdthesetwobuttonstodisable
AutoPowerOFF.DetailedAUTOPOWEROFF(Sleepmode)instructionsare
providedlaterinthisguide

++ Pressall3buttonssimultaneouslytoentertheCalibrationmode.
Calibrationinstructionsareprovidedindetaillaterinthisguide
AN320-en-GB_V1.3 11/15
5
DisplayLayout
1. LowerDisplaydigits(airvolume,temperature,timer,ormenuparameterdisplays)
2. UpperDisplaydigits(airvelocity,humidity,CO
2
,ormenudisplayparameters)
3. OptionalConeorFunnelattachmenticon
4. Unitsofmeasureforductareaprogramming
5. Vaneindicator(flashesverybrieflyonpowerup)
6. DisplayHOLDicon
7. MAXmodeicon
8. MINmodeicon
9. Recordingicon
10. ElapsedTimericon
11. AVGmode
icon
12. Batterystatusicon
13. Unitofmeasureicons*
*Fulllistshown.Availableunitsofmeasureiconsvarybymodel.
AN320-en-GB_V1.3 11/15
6
Operation
AirVelocityandAirVolumeMeasurements
1. SwitchthemeterONbypressingthePOWER buttonmomentarily.Themeterwill
countdownfrom30to0asitperformsaselftest.PressthePOWERbuttonagaintoswitch
themeterOFF.
2. Notethatthemeterwillautomaticallyenterthesleepmodeafter20minutesofinactivity.
Todisablethisfeature:withthemeterOFF,pressandholdthePOWERandtheU/mode
buttonsfor2seconds.TheLCDwilldisplay'n'asitswitchesON.Nowthesleepmodeis
disabledandtheusermustpressthePOWERbuttonmomentarilytoswitchthemeterOFF.
3. WhenthemeterisON,theupperdisplaycanindicate
airvelocity,relativehumidity,orCO
2
(usetheU/mode
buttontoscroll).Thelowerdisplaydigitscanindicate
airflow,dewpointtemperature,wetbulbtemperature,
orairtemperature(usetheL/modebuttontoscroll).
4. Placethesensorintheairstream.Theairmustenter
thevanefromtherearofthemeter.The
Air
Temperature,Relativehumidity,andCO
2
sensorsarelocatedatthecenterofthevaneon
therearofthemeter.
5. ReadthemeasurementdataontheLCD.Airvelocityreadingsareindicatedinmetersper
second(m/s)orfeetperminute(fpm).Temperaturereadings(air,dewpointDP,orwetbulb
WBT)areindicated
in°C/°F.Airflow(volume)isindicatedinCFM(cubicfeetperminute)or
CMM(cubicmetersperminute).UsetheL/modebutton(forlowerdisplay)andtheU/mode
button(forupperdisplay)toscrollthroughthedisplayedreadingoptions.
6. TochangebetweenImperialandMetricunitsofmeasurerefer
totheSETUPModesection.
7. TomeasuretheAirVolumeofaduct,firstmeasuretheduct’sarea(seetheAppendixfor
areacalculations)andthenentertheareavalueinthemeter’sSETUPmodeinin
2
orcm
2
.
Aftertheareavaluehasbeenentered,usetheL/modebuttontoscrolltotheairvolume
display(CMM/CFM)onthelowerdigits.Placethemeterinairstreamtomeasurevolume.
8. Optional:measureAirVolumewithanairflowConeorSquareadaptorbyfirstattachingone
ofthe
adaptorstothemeter(theoptionalconeadaptorkitincludesaroundandasquare
cone;kitpartnumberAN300C).Themeterwillautomaticallyrecognizethattheadaptoris
attachedtothemeterandaconeiconwillappearonthe display.Note:regardlessofthe
AREAsettingin
theSETUPmode,themeterwilldefaulttothedimensionsofthevane
opening.Ensurethattheairflowadaptoriswellmountedandlockedbeforetaking
measurements.
LCDbacklight
PresstheU/modebuttontoswitchONtheLCDbacklight.Thebacklightwillautomaticallyswitch
OFFafterapproximately10secondstoconservebatterypower.Notethatthebacklighting
switchesONautomaticallywhenthemeterisswitchedONbytheuser.
AN320-en-GB_V1.3 11/15
7
RelativeHumidity,WetBulb/DewPointTemperature,andCO
2
Readings
Sensorsarebuiltintothecenterofthe vane(rearofmeter)andareusedtomeasurerelative
humidity,airtemperature,andCO
2
.WetBulbandDewPointtemperaturesarecalculatedbased
ontemperatureandhumidityreadings.
1. PressU/modetoscrolltoRH(%)orCO
2
(ppm)modes.Themeasuredrelativehumidityor
carbondioxidereadingswillbedisplayedontheupperLCDdigitsasselected.
2. UsetheL/modebuttontoscrollthroughairtemperature,dewpointtemperature,wetbulb
temperature,andairflow(volume)readings.
3. Tomeasurethehumidityusinganoffsetreference
value,pleaseaccesstheSetupmodeand
inputtheoffsetvalueasdescribedintheSetupmodesection.TheLCDwilldisplaythe
measuredvalueminustheoffsetvalueoncetheoffsetisprogrammedintheSetupmode.
DataHoldMode
1. PresstheHOLDbuttonfromthenormaloperatingmodetofreezethecurrentmeasurement.
2. The‘HOLD’iconwillappearatthetopoftheLCDdisplay.
3. PressHOLDagaintoreturntonormaloperation.The‘HOLD’iconwillswitchOFFandthe
meterwillreturntodisplayingreadings
inrealtime.
MAX/MINMode
TheMAXMINfeaturerecordsthehighest(MAX)andlowest(MIN)readingscapturedsincethe
MAXMINmodewasaccessed.
1. PressandholdtheMAX/MINbuttonfor2secondstobegindisplayingthelowest(minimum)
readingsencounteredsincethismodewasaccessed.The‘MIN’iconwillappearatthetop
of
theLCDdisplayindicatingthatthereadingsshownarethelowestmeasured.
2. PressandholdtheMAX/MINbuttonagainfor2secondstodisplaythemaximumreadings
encountered.The‘MAX’iconwillappearontheLCD.
3. Pressandholdthisbuttonagainfor2secondstoreturntothenormalmodeofoperation
(the‘MIN’and‘MAX’iconsshouldswitchOFF).
4. WhileviewingtheMAXorMINdata,usetheL/modebuttontoswitchthedisplaysothatthe
AirVolumeMINandMAXreadingscanbeviewedalso.
5. PressandholdMAX/MINfor2seconds
toreturntothenormaloperationmode.
MultiPointAverageMode
1. FromthenormaloperatingmodepresstheU/modebuttonfor2secondstoenterMulti
PointAveragemode(theAVGdisplayiconandtheassociatedblackdotwillswitchON).
2. PresstheHOLDbuttonmomentarilytorecordonereading.Theupperdisplaydigitswill
showthereading’smemorylocationnumberandthelowerdisplaywillshowthemeasured
value.PresstheHOLDbuttonagaintotakeandstoreanotherreading;thememorylocation
numberwillincrementbyoneandthecurrentreadingwillbestoredinthisnewmemory
location.Themetercanstoreupto19,999pointsinmemory.
3. PressU/modemomentarilytoseethemultipointmean(average)forallofthereadings
recordedduringthesession(the'•AVG'displayiconwillflashduringthisprocess).
4. PressL/modetoviewtheaveragereadingfortheotheravailableparameters.
5. PressU/modemomentarilytoreturnto
thenormaloperatingmode.
AN320-en-GB_V1.3 11/15
8
TimedAverageMode
1. Fromthenormaloperatingmode,pressandholdtheU/modebuttonfor2secondstoenter
theMultiPointAveragemodefirstandthenpressU/modeonemoretimemomentarilyto
entertheTimedAverageMode(theAVGandtheelapsedtimer iconswillswitchON).
2. PressHOLDtobeginrecording.Theelapsedtimerstarts(asshownontheupperdisplay
digits)andtheclockiconflashesonandoff.Thelongestamountoftimethemetercan
operateinTimedAverageModeis19,999seconds.
3. PressU/modetostop
theclockandtocalculatethetimedaverage.TheupperLCDwill
displaytheelapsedtimeandthelowerLCDwillshowtheaveragedreadings.TheLCDwill
flashthe'AVG'andtheelapsedtimericons.
4. PressL/modetoviewtheaverageforotheravailablemeasurements.
5. PressU/modetoreturntothenormaloperatingmode.
SetupMode
EnteringSetupMode
PressandholdthePOWERSETbuttonfor2secondstoenterSetupmode.Twooptionsare
available:P1.0:UnitandP2.0:AREA.
Usetheup/downbuttonstotogglebetweenP1.0UNITandP2.0AREAoptions.
P1.0:UnitSelection
1. Onceinsetupmode,attheP1.0UNITdisplay,momentarilypress
theHOLDbuttontoenter
theP1.0UNITmenu.
2. Useeitherarrowbuttontoswitchbetweenmetricandimperialunits.Theavailableunits
are:airvelocity(m/s,fpm),temperature(C,F),airvolume(CMM,CFM),andareasize(cm²,
inch²).
3. PressHOLDmomentarilytoreturntotheP1.0
display.
4. PresseitherarrowbuttontomovetotheP2.0AREAoptionandfollowthestepsbelow.
P2.0:AreaSizeSelection(entersquareinchesorsquarecentimeters)
1. Insetupmode,fromtheP2.0AREAdisplay,pressHOLDtoentertheAREAsettingmenu.The
LCDwillshow‘99999’in
thelowerdisplaywiththefirstdigitflashing.
2. Usetheuparrowbuttonchangetheflashingdigit(from0through9).
3. Usethedownarrow buttontomovetothenextdigit.Usetheuparrowbuttonagainto
changetheflashingdigit.Continueinthiswayuntil
thedesiredAREAvalueisdisplayed.
4. PresstheHoldbuttonagainmomentarilytostorethe valueandtoreturntotheP2.0display.
5. PresseitherarrowbuttontomovetotheP3.0RHOFFSET optionandfollowthestepsbelow.
P3.0:RelativeHumidity(RH)OffsetSelection
1. WheninsetupmodeP3.0pressHOLDtoentertheRHoffsetsetting.TheLCDwillshow0.0
RHintheupperdisplay.
2. PressU/modetoincreaseoffsetvalueandL/modetodecreaseoffset.
3. PresHOLDoncemoretoreturntoP3.0.
4. PressandholdthePOWERSETbutton
for2secondstoreturntothenormaloperating
mode.
AN320-en-GB_V1.3 11/15
9
P4.0:AltitudePressureCompensation
5. WheninsetupmodeP4.0,pressHOLDtoenterthepressurecompensationparameter.The
LCDwillshowhPA(unitofmeasure:hectorPascalormillibar)onthelowerlineanda
flashing4digitnumber(from700to1990;thedefaultis1013hPAwhichis
standard
pressure)ontheupperline.
6. PressU/modetoincreasethepressurevalueandL/modetodecreaseit.
7. PressHOLDoncemoretoreturntoP4.0.
8. PressandholdthePOWERSETbuttonfor2secondstoreturntothenormaloperating
mode.
Calibration
CAUTION:DonotcalibratethemeterinairwithanunknownCO
2
concentration.Doingsowill
resultinthemeterbeingcalibratedtothestandard400ppmandcouldleadtoinaccurate
results.
1. Placethemeterinacalibratedenvironment.
2. WiththemeterOFF,pressSET,HOLDandL/modesimultaneouslytoenterCO
2
calibration
mode.
3. Themeterwillruna30secondcountdownand'CAL'willshowinthelowerdisplay.
4. DuringcalibrationaCO
2
readingbetween380420ppmwillflashintheupperdisplay.
5. Waitforapprox.10minutesforthereadingtostopflashing.
6. Thecalibrationisnowcompleteandthemeterwillreturntothenormaloperatingmode.
7. Toabortthecalibrationatanytime,turnOFFthemeter.
8. IfthecalibrationfailsforanyreasonpleasereturnthemetertoExtechInstrumentsforan
evaluationandpossiblerepair.
AN320-en-GB_V1.3 11/15
10
BatteryReplacement
WhenappearsontheLCD,theAAAbatteriesmustbereplaced.Thebatterycoverislocated
onthebackofthemeter.
1. Opentherearbatterycompartmentbycarefullypullingdownonthecompartment’slatch.
2. Thebatterycompartmentcovershouldbecompletelyremovedbeforeproceeding.
3. Replacethe
four(4)1.5V‘AAA’batteriesensuringproperpolarity.
4. Closethebatterycompartmentbeforeattemptingtousethemeter.
Neverdisposeofusedbatteriesorrechargeablebatteriesinhouseholdwaste.
Asconsumers,usersarelegallyrequiredtotakeusedbatteriestoappropriate
collectionsites,theretailstorewhere thebatteries
werepurchased,orwherever
batteriesaresold.
Disposal:Donotdisposeofthisinstrumentinhouseholdwaste.Theuseris
obligatedtotakeendoflifedevicestoadesignatedcollectionpointforthe
disposalofelectricalandelectronicequipment.
OtherBatterySafetyReminders
o Neverdisposeofbatteriesin
afire.Batteriesmayexplodeorleak.
o Nevermixbatterytypes.Alwaysinstallnewbatteriesofthesametype.
AN320-en-GB_V1.3 11/15
11
Specifications
AirVelocity Ranges Resolution Accuracy
m/s(meterspersec) 0.230m/s 0.01m/s ±(1.5%rdg+0.3m/s)
fpm(feetperminute) 405900fpm 1fpm ±(1.5%rdg+59fpm)
RelativeHumidity Range Resolution Accuracy
RH 0.1%99.9%RH 0.1RH ±3%RH(1090%RH)
±5%RH(<10%or>90%)
AirFlow Ranges Resolution AreaRange
CMM(cubicmeters/min) 099999m
3
/min
0.1upto9999.9
then1.0
0to99999cm
2
CFM(cubicft/min) 099999ft
3
/min
0.1upto9999.9
then1.0
0to99999in
2
Temperature Ranges Resolution Accuracy
o
F/
o
C
AirandWetBulb:
20to60
o
C(4to140
o
F)
DewPoint:
5to60
o
C(23to140
o
F)
0.1
o
C(
o
F)
±0.6°C(20to50°C)
±1.2°C(51to60°C)
±1.1°F(5to122°F)
±2.2°F(123to140°F)
CO
2
Range Resolution Accuracy
PPM(partspermillion)
09999ppm
(readingsover5000are
notspecified)
1ppm
±(30ppm+5%ofrdg)
from0to5000ppm
AN320-en-GB_V1.3 11/15
12
Circuit CustomLSImicroprocessor
Display Dualfunction13mm(0.5")4digitLCD
Samplingrate 1readingpersecondapprox.
Airvelocity/flowsensor Conventionalangledvanearmswithlowfrictionballbearing
TemperatureSensor NTCtypeprecisionthermistors
CO
2
Sensor NDIRCO
2
sensor
AutomaticPoweroff Autoshutoffafter20minutestopreservebatterylife(sleepmode)
OperatingTemperature 0°Cto50°C(32°Fto122°F)
StorageTemperature‐10°Cto50°C(14to122
o
F)
OperatingHumidity <80%RH
StorageHumidity <90%RH
OperatingAltitude 2000m(7000ft.)maximum
BatteryPower Four(4)‘AAA’1.5Vbatteries
BatteryLife >40hours
Batteryconsumption 8.3mADC(approx.)
Weight 725g(1.6lbs.)includingbatteryandprobe
Dimensions Maininstrument:269x106x51mm(10.6x
4.2x2”)
Vaneinnerdiameter:100mm/3.94inches
OptionalAirFlowAdaptors:
Round:210mm(8.3”)Diameter
Square:346x346mm(13.6x13.6”)
AN320-en-GB_V1.3 11/15
13
ErrorMessages
Errorcodesandsolutions
AirTemperature
E02:Temperaturereadingislowerthanthelowrangelimit.Placethemeterinroom
temperaturefor30minutes.IfE02continues,sendtheunittoExtechforrepair.
E03:Temperatureexceedsthehighrangelimit.Placethemeterinroomtemperaturefor30
minutes.IfE03continuestoappear,
sendtheunittoExtechforrepair.
E31:TheAtoDconverterrequiresreplacing.SendthemetertoExtechInstrumentsforrepair.
RelativeHumidity(RH)
E04:Temperatureerrorcorruptedthehumidityreading.Checkthetemperatureerrorsolution
section.
E11:RHcalibrationerror.SendtoExtechInstrumentsforrepair.
E33:RHmeasurementcircuiterror.SendtoExtechInstrumentsforrepair.
DewPoint(DP)andWetBulb(WBT)Temperature
E04:TemperatureorRHerror.ChecktemperatureorRHerrorsolutions.
Airvelocity
E03:Velocityexceedsthehighrangelimit.Testthemeterusinganairvelocitythatisknown
tobewithinthespecifiedrangelimits.SendtheunittoExtechforrepairiftheE03error
messagepersists.
Airvolume
E03:Readingexceedsthemeter’sdisplaylimit.CheckthattheAREAsettingiscorrectinthe
SetupMode.
E04:Airvelocityerror.ReturnmetertoExtechforrepair.
CO2
E01/E33:VoltagetooloworCO
2
moduleerror.Checkifthelowbatteryiconisdisplayedon
theLCD,ifyes,replacewithnewbatteries.Tryarecalibration,orsendtheunittoExtechfor
repairiftheproblempersists.
OtherErrors
E32:MemoryICerror.Rebootthemeterandcheckitagain;sendittoExtechInstrumentsfor
repairiftheerrormessagepersists.
Nodisplay
1. Checkthatthebatteriesaremakinggoodcontactandarealignedforcorrectpolarity.
2. Replacebatteries.
DisplayswitchesOFFautomatically
Thissymptomcouldbethenormal‘SleepMode’whichswitchesthemeterOFFafter20
minutesofinactivity.Ifthisisnotthe‘SleepMode’,checkwhetherthelowbatteryindicator
appearsbeforetheLCDattemptstoswitchON;ifyes,replacethebatteries.
AN320-en-GB_V1.3 11/15
14
Appendix:UsefulEquationsandConversions
Areaequationforrectangularorsquareducts

Areaequationforcircularducts
Cubicequations

NOTE:Measurementsmadeininchesorcentimetersmustbeconvertedtofeetormetersbeforeusingtheseformulae.
UnitofMeasureConversionTable
m/s ft/min knots km/h MPH
1m/s 1 196.87 1.944 3.6 2.24
1ft/min 0.00508 1 0.00987 0.01829 0.01138
1knot 0.5144 101.27 1 1.8519 1.1523
1km/h 0.2778 54.69 0.54 1 0.6222
1MPH 0.4464 87.89 0.8679 1.6071 1
Copyright©20142015FLIRSystems,Inc.
Allrightsreservedincludingtherightofreproductioninwholeorinpartinanyform
www.extech.com
Width(W)
Height(H) Area(A)=Width(W)xHeight(H)
CFM(ft
3
/min)=AirVelocity(ft/min)xArea(ft
2
)
CMM(m
3
/min)=AirVelocity(m/sec)xArea(m
2
)x60
Radius
Area(A)=pi xr
2
Wherepi=3.14andr
2
=radiusxradius
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14

Extech Instruments AN320 User manual

Category
Measuring, testing & control
Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI