Extech Instruments WQ500 User manual

Category
Measuring, testing & control
Type
User manual

This manual is also suitable for

USERGUIDE
WQSeriesWaterQualityBenchtopMeters
ModelWQ500pH,ORPandTemperature
ModelWQ510pH,ORP,Conductivity,TDS,Salinity,&Temperature
ModelWQ530pH,ORP,Conductivity,TDS,Salinity,DO,&Temp.
WQ500 Series-en-GB_V1.3 9/15
2
Introduction
CongratulationsonyourpurchaseoftheWQseriesmeter.Thisbenchtopdeviceemploys
leadingedgetechnologywithintegratedmicroprocessorformeasuringpH,ORP,Conductivity,
DO,Salinity,TDS,andTemperature(dependingonmodel).Thesedevicescanstore150labeled
readingsforrecallatalatertimeandincludesaMAXMINmemoryfeature.Inaddition,theWQ
seriesmetersofferaPCinterfacefordatatransferandotherremotefunctionality.Thismeteris
shippedfullytestedandcalibratedand,withproperuse,willprovideyearsofreliableservice.
Pleasevisitourwebsite(www.extech.com)tocheckforthelatestversionofthis
UserGuide,
ProductUpdates,andCustomerSupport.
FEATURES
Microprocessorbaseddevicewithruggedhousingandsplashproofkeyboard
LargebacklitLCDdisplay
AutomaticTemperatureCompensation(ATC)
ManualSalinityandAltitudeCompensation
Memoryfunctionstoresandrecallsupto150datapoints
MAXMINstoreandrecall
DataLock
Uniquemeasurementplatform
andelectrodeholder
9VDCuniversalACadapter(UA100240)
PCinterfaceforcapturingmeasurementdata
SUPPLIEDACCESSORIESANDEQUIPMENT
WQ500:
pHglasselectrode,Temp.probe,Buffer7.00x100ml,Buffer4.01x100ml,Electrodeclamp
&rod,SoftwareandUSBcable,AC/DCadaptor,Instructionmanual,Giftbox,Stirrer
WQ510
pHglasselectrode,Conductivitycell,Temp.probe,Buffer7.00x100ml,Buffer4.01x
100ml,1413μSx100ml,
12.88mSx100ml,Electrodeclamp&rod,SoftwareandUSBcable,
AC/DCadaptor,Instructionmanual,Giftbox,Stirrer
WQ530
pHglasselectrode,Conductivitycell,DOprobe,Temp.probe,Membranecapx4pcs,Buffer
7.00x100ml,Buffer4.01x100ml,1413μSx100ml,12.88mSx100ml,Electrolyte
x50ml,
Plasticburette,Sandpaper,Electrodeclamp&rod,SoftwareandUSBcable,AC/DCadaptor,
Instructionmanual,Giftbox,Stirrer
WQ500 Series-en-GB_V1.3 9/15
3
Descriptions
METERDESCRIPTION
1. LCDDisplay
2. PrimaryKeypad
3. SecondaryKeypad
4. SecondaryKeypadlegend
5. BeakerTestPlate
6. RearConnectionaccess
7. Electrodeholderpost
8. Electrodeholder
LCDDESCRIPTION
1. Measurementtype
2. PrimaryReadingDisplayarea
3. CalibrationMode
4. Calibrationerrorindicator
5. Readingstoredindicator
6. MAX/MINmodeicons
7. RecallMode
8. Lockmodeicon
9. Temperaturereading
10. Autotemperaturecondensation
11. Unitsofmeasureicons

WQ500 Series-en-GB_V1.3 9/15
4
KEYPADDESCRIPTION
LockthecurrentreadingontheLCD.
Pressandholdfor3secondstoenterorexitMAX/MINmode.
InMAX/MINmode,presstobrowseMAXandMINreadings.
Storethecurrentreading.
Pressandholdfor3secondstoenterRecallmode.
InRecallmode,usetheupanddownarrowstobrowserecords.
Press and hold simultaneously for 3 seconds to enter the advanced
settingmode(seeAdvancedSettingssection).
Selectameasurementfunction.
Press andholdfor3 seconds to switchC/ForswitchpHmVor ORP
mV(inORPmode),orswitchmg/Ltoppm(inDOmode)
SwitchPowerONorOFF.
Pressandholdfor3secondstoentertheCalibrationmode
1. Increasestirrerspeed
2. Decreasestirrerspeed
3. Presstoturnonthestirrerorsetstir
mode(30,60,120minutesor
continuously);Press3sec.toturnoff
thestirrer
4. StirmodeLEDindicator:30,60,120
minutesorstircontinuously.
WQ500 Series-en-GB_V1.3 9/15
5
Preparation
pH,ORP,Conductivity,TDS(TotalDissolvedSolids),andSalinity
1. ConnectACadaptertoapowersource.
2. Removetheprotectioncapfromtheelectrodeandconnecttotheinputconnector.
3. Connectthetemperatureprobetothemeterandpoweron.
4. Rinsetheelectrodewithcleanwaterandwipedry.
DO(DissolvedOxygen)
1. ConnecttheDOprobeandtemperatureprobetothemeterandremovetheprotectivecap.
2. Carefullyremovethemembranecap.
3. Fillmembranecapwiththeelectrodesolutiontothebottomofthethreads insidethecap.
4. PressthePowerbuttontoswitchthemeterON.
CalibrationandOperation
pHCalibration(NOTE:ORPdoesnotrequirecalibration)
Beforeaccuratemeasurementscanbeobtained,itisnecessarytocalibratethemeterwiththe
electrode.Typically,pH7iscalibratedfirstandthenoneormoreoftheotherbuffers.
1. ConnectthepHelectrodetothemeter’sBNCconnectorinput.
2. DiptheelectrodeandtemperatureprobeintothepH7.00buffersolution.Stirgentlyand
waitforastablereading.
3. PressandholdthePowerbuttonfor3secondstoentercalibrationmode.
4. TheLCDwilldisplayCALanda 7.00willflash.
5. WhencalibrationiscompletethedisplaywillstopflashingandshowSA
thenEND.
6. Themeterwillnowreturntonormaloperatingmode.
7. Rinsetheelectrodeandwipedry.DiptheelectrodeinpH4.01orpH10.01.TheLCDwill
indicateapercentageofslope(PTS)toshowthestatusoftheelectrode.
8. IfPTSisbelow70%
orabove130%theelectrodemustbereplaced.Aslopeof100%isideal.
Notes:
1. Calibrationerrorindicatoriconwillappear(ErrinsteadofSA)ifcalibrationfails.
2. Whendoinga2or3pointcalibration,calibratewithbufferpH7first,andthenfollowwith
bufferpH4orpH10.
3. pHcalibrationtypeUSAorNISTcanbechangedintheAdvancedSettingmode(seethe
AdvancedSettingssection).
4. ThecalibrationpointsforUSAare1.68,4.01,7.00,10.01and12.45.
5. ThecalibrationpointsforNISTare1.68,4.01,
6.86,9.18and12.45.
WQ500 Series-en-GB_V1.3 9/15
6
ORPCalibration
CalibrationisnotnecessaryforORP.However,ORPshouldbeverifiedwithORPstandardsolution
tocheckwhethertheelectrodeisinproperworkingorderornot.
Conductivity,TDS,andSalinityCalibration
1. DiptheConductivitycellandthetemperatureprobeintothestandardsolution1413μS/cm.
2. Stirgentlyandwaituntilthereadingisstable.
3. Press and hold the Power button for 3 sec. to enter the calibration mode. The display will
showCALand1413μS/cmwillflash.
4.
When calibration is complete the display will stop flashing and indicate SA, then End. The
meterwillthenreturntothenormaloperatingmode.
Notes:
1. Calibratingwith12.88mS/cmstandardsolutionisbetterformeasuringwithhighconductivity
solutions.
2. TheiconCONDwilldisplayautomaticallyduringcalibrationmode.
3. Calibrationerrorindicatoriconwillappear(ErrinsteadofSA)ifcalibrationfails.
4. Ifthereadingisnot0μS/cmwhilethecellisinair(notdippedintoanysolution),calibratein
airtoobtainareading
of0μS/cm.
5. The Conductivity calibration points are: 0, 84μS/cm, 1413μS/cm, 12.88mS/cm and
80.0mS/cm.
DOCalibration
1. WiththeDOprobeattached,pressPowerbuttontoturnonmeterandthenpressMODEto
selectO2mode.
2. Wait10to30minutesfortheprobetopolarize.Thereadingshouldbeapprox.101.7%
(saturation)aftertheprobehascompletelypolarized
3. Removetheprobe.PressandholdPowerfor3sec.toentercalibrationmode.Thedisplaywill
showCALand101.7%willflash.
ThedisplaywillstopflashingandindicateSA,thenEndwhen
calibrationiscomplete.Themeterwillreturntothenormaloperatingmode.
4. Optional’zerooxygen’calibration:(improvesmeasurementaccuracyforveryloworveryhigh
DOmeasurements).Placetheprobeintoazerooxygencalibrationsolution,suchas
5%
sodiumsulfite,waitforstabilityandpressandholdPowertoentercalibration.Stabilityina
zerosolutionmaytakemanyminutes,dependingonprobehistory.
Note:
1. TheiconO2willdisplayautomaticallyduringcalibrationmode.
2. Calibrationerrorindicatoriconwillappear(ErrinsteadofSA)ifcalibrationfails.
3. Ifthe reading is not0%when the probe is disconnected,calibrateitin air without probe to
obtainareadingof0%.
WQ500 Series-en-GB_V1.3 9/15
7
pHMeasurement
1. PressMODEtochoosepHmode.
2. Aftercalibration,rinsethepHelectrodewithcleanwaterandwipeitdry.
3. Dip the electrode and the temperature probe into the sample solution that is going to be
measured.
4. Stirgentlyandwaituntilastablereadingcanbeobtained.
ORPMeasurement
1. InserttheORPelectrode,andpressMODEtochooseORPmode.
2. RinsetheORPelectrodewithcleanwaterandwipeitdry.
3. Diptheelectrodeintothesamplesolutionthatisgoingtobemeasured.
4. Stirgentlyandwaituntilastablereadingcanbeobtained.
Notes:
1. Thedisplaywillshow‐‐‐‐inanoverrangecondition.
2. Aftermeasurement,rinsetheelectrodewithcleanwater.
3. Replacethesoakingbottle.Thebottleshouldalways be filledwithsoakingsolution(4MKCL).
COND,TDS,SalinityMeasurement
1. InsertConductivitycell,andpressMODEtochooseCOND,TDSorSaltmode.
2. Aftercalibration,rinsetheconductivitycellwithcleanwaterandwipeitdry.
3. Dip the electrode and the temperature probe into the sample solution that is going to be
measured.
4. Stirgentlyandwait
untilastablereadingcanbeobtained.
Notes:
1. Thedisplaywillshow‐‐‐‐inanoverrangecondition.
2. ThemeterautorangestoμS/cmormS/cm,orppmorppt.
3. Aftermeasurement,rinsethecellwithcleanwaterandreplacetheprotectivecap.
4. Donot
touchorwipethesurfaceoftheinnerblackplateoftheconductivitycell.
DOMeasurement
1. RemovetheprotectioncapandturnonthepowerandpressMODEtochooseO2mode.Wait
10to30minutesfortheprobetopolarize.Thereadingshouldbeapprox.101.7%(saturation)
aftertheprobehascompletelypolarized.
2. SelectthedesiredunitsofmeasurebypressingMODEuntil
theproperunitsareshowninthe
display.
3. Placetheprobeinthesampletobemeasured.Stirtheprobeinthesampletoremoveany
trappedairbubblesfromthemembranesurface.
4. Allowthemetertimetosettletothefinalmeasurementvalue.
WQ500 Series-en-GB_V1.3 9/15
8
Notes:
1. Thelargerthedifferenceintemperaturebetweentheprobeandthesolution,thelongerit
willtakeforthereadingtostabilize.Stabilizationtimecanvaryfromten(10)secondstofive
(5)minutes.
2. Covertheprobewiththeprobecap.Thespongecontainedinthecapshould
bemoistened
(notsoaked)withDI(distilledwater)orcleantapwater.
150ReadingMeasurementStoreandRecall
1. PresstheSTORE/RECALLbuttontostorethecurrentreading.Presseachtimeareadingisto
bestored.TheSTOREiconappearsasareadingisstored.
2. Ifanattemptismadetostoremorethan150readings,thestoredreadings(startingwiththe
firstreading)willbeoverwritten.
3. Torecallreadings,pressandholdtheSTORE/RECALLbuttonuntilthe‘RECALL’iconappears
ontheLCD.Usetheup/downarrowbuttonstoscrollthroughthestoredreadings.
4. Toexitthismode,pressSTORE/RECALLtoreturntonormaloperatingmode.
5. Pressandholdtheupanddownbuttonssimultaneouslyfor3secondstoclearthememory.
MAX/MINMemoryDisplay
1. Themetermemorizesthehighest(MAX)andlowest(MIN)readingswhenprompted.
2. PressandholdtheMAX/MINbuttonuntilboththeMAXandMINiconsappearflashingon
theLCD.ThemeterwillnowbeginkeepingtrackoftheMINandMAXreadings.
3. ToviewMAX/MINvalues,presstheMAX/MINbuttonagain;themeterautomatically
displaysthehighestandlowestreadingsinsuccessionandthenreturnstotheMAX/MIN
recordingmode.
4. ToexittheMAX/MINmode,pressandholdtheMAX/MINbuttonuntiltheMAXandMIN
iconsdisappear.
WQ500 Series-en-GB_V1.3 9/15
9
AdvancedSettings
InpH,TDS,orDOmodepresstheupanddownarrowssimultaneouslyfor3secondstoenterthe
AdvancedSetupmode.
pHAdvancedSettings
1. InpHsettingmode,pressSTORE/RECALLbuttontosetthepHcalibrationtype.Usetheup
anddownarrowstoselectUSAorNIST.USAandNISTcalibrationpointsdiffer,refertothe
Calibrationsectionofthisguideforcomparisonofcalibrationpoints.
2. PressMODEtosavesettingandreturntonormaloperatingmode.
3. PressLOCK/MAXMINtoturnAUTOLOCKONorOFF.Usetheupanddownarrowstoselect
ONorOFF.InAutoLockmodethekeypadislockedandcannotbetampered.
4. PressMODEtosavesettingandreturntonormaloperatingmode.
TDSAdvancedSettings
1. PressSTORE/RECALLtosettheratiobetweenconductivityandTDS.
2. Usetheupanddownarrowstoadjusttheratiofrom0.5to1.0andthenpressMODEtosave
settingandreturntonormaloperatingmode.
DOAdvancedSettings
1. PressSTORE/RECALLtosetthesaltcompensation.Usetheupanddownarrowstoadjustthe
valuefrom0to50ppt.
2. PressMODEtosavesettingandreturntonormaloperatingmode.
3. PressLOCKtosetthealtitudecompensation.Usetheupanddownarrowstoadjustthe
valuefrom0to7000m(20000ft.).
4. PressMODEtosavesettingandreturntonormaloperatingmode.
Note:Inanyadvancedsettingmode,pressthepowerbuttontoresetallsettingsbacktofactory
setup(exceptpHcalibrationandAutolock)
PCInterface
ThisWQseriesmetercanbeconnectedtoaPCfordatatransferandotherremotefunctionality.
PleaserefertothePCInterface SoftwareInstructions suppliedontheincludedprogramCDROM
and/ortheinstructionsprovidedintheHELPUtilityoftheSoftwareProgramitselfonceitis
opened.Pleasecontact
ourCustomerCareTechnicalSupportdepartmentforfurtherPC
interfacerelatedassistance.Checkthewebsitewww.extech.comsoftwaredownloadpagefor
thelatestversionofprogramandfortheoperatingsystemcompatibility.
WQ500 Series-en-GB_V1.3 9/15
10
MembraneCapReplacement
IMPORTANTNOTE:OnlytheelectrolytesolutionsuppliedbyExtechwillworkcorrectlyforthis
probe.Ifathirdpartysolutionisused,theprobewillnotworknormally.
1. Donottouchthemembraneasskinoilswillinterferewiththeoxygenpermeabilityrateofthe
membrane.Replacethecapcarefully.
2. Itisrecommendedthattheproberemainattachedtothemeterduringthisreplacement
process.
3. Unscrewthecapfirmlyandcarefullyfromtheprobe.
4. RinsetheoldelectrolytesolutionfromtheCathodeandAnode.
5. UsethesuppliedPolishingStripstoclean,polish,shine,and/orremovescratches
fromthe
cathode.Besuretomoistentheclothbeforepolishingthecathode.Donotoverpolishthe
sensitivegoldcathode.
6. Setthenewreplacementmembranecaponaflatsurface.Leavethecapinthisposition
duringthereplacementprocess.
7. Fillthemembranecapwiththeelectrolytesolution
uptothebottomofthethreadsonthe
insideofthecap.
8. Tapthemembranecaptoreleaseandpreventairbubbleinelectrolytesolution.
9. Keepingthecapinafixedpositiononaflatsurface,carefullyinserttheprobeintothenew
capbyfirstdipping
andremovingtheprobeseveraltimesfromthecap.Witheachdip,push
theprobeprogressivelydeeperintothebondedcap.Finally,screwtheprobeslowlyontothe
capuntilfullytightened.Thedippingandremovaltechniqueminimizestheintroductionofair
bubblesintotheelectrolytesolution.Airbubblesinthe
electrolytecanaffectmeasurements.
10. Itisnormalthatexcesselectrolytesolutionwillleakfromthecapduringthisreplacement
sinceitminimizestheintroductionofairpockets.Cleanofftheexcesselectrolytebeforeuse.
1) MembraneCap
2) Cathode
3) Anode
4) DOProbe
CleaningtheDOProbe
WhentheDOreadingisunstableorincorrect,theprobeneedstobecleaned:
1. Unscrewthemembranecapfromtheprobe.
2. Placesandpaperonatablewiththecoarsesidefaceup,andaddsomewater.
3. Rubthecathodepartagainstthesandpaperabout10times.
4. Usethesandpapertocleantheanodepart.
5. Aftercleaningtheprobe,pleaserefillmembranecapwithnewelectrolyte.
WQ500 Series-en-GB_V1.3 9/15
11
ProbeStorage
pHprobe(WQPH)
Afteruse,cleantheprobetipwithdistilledwater.Theprobetipneedstobestoredwetin
pH4buffersolution.PlacetheprotectivecapwithpH4solutionovertheendoftheprobe.

ORPprobe(67500B)
Afteruse,cleantheprobetipwithdistilledwater.Theprobetipneedstobestoredwetin
pH4buffersolutionor4.0MKCLsolution.Placetheprotectivecapwiththesolution overthe
endoftheprobe.
ConductivityProbe(WQEC)
Afteruse,cleantheprobetipwithdistilledwater.Theprobeshouldbestoredwetin
deionizedordistilledwater.Placetheprotectivecapwiththewaterovertheendofthe
probe.
DissolvedOxygenprobe(WQDO)
Afteruse,cleantheprobetipwithdistilledwater.
Shorttermstorage‐Lessthan2weeks
Theprobeshouldbestoredwetindeionizedordistilledwater.Placetheprotectivecapwith
thewaterovertheendoftheprobe.
Longtermstoragemorethan2weeks
The
probeshouldbestoreddry.Removethe membranecapandrinseofftheelectrodewith
deionizedordistilledwaterandthenallowtheelectrodetodry.Placethemembranecapback
overtheelectrodeforprotectionandplacetheprobetipintotheprotectivecap.
WQ500 Series-en-GB_V1.3 9/15
12
Specifications
pH ORP
Range‐2.00to16.00pH‐1999to‐200mV
199.9to499.9mV
500to1999mV
Accuracy ±(0.02%FS+1d) ±(2%FS+1digit)
Resolution 0.01pH 0.1/1mV
Compensation ATC:0to100
o
C(32to212
o
F) N/A
Conductivity TDS Salt
Range 0.0to199.9μS
200to1999μS
2.00to19.99mS
20.0to100.0mS
0.0to131.9ppm
132to1319ppm
1.32to13.19ppt
13.2to66.0ppt
0.0to99.9ppm
100to999ppm
1.00to9.99ppt
10.0to50.0ppt
Accuracy ±2%FS ±2%FS ±2%FS
Resolution 0.1/1μS/0.01/0.1mS 0.1/1ppm/0.01/0.1ppt 0.1/1ppm/0.01/0.1ppt
Compensation ATC:0to60
o
C(32to140
o
F) ATC:0to60
o
C(32to140
o
F) ATC:0to60
o
C(32to
140
o
F)
DO O2 Temperature
Range 0to20.00mg/L
0to20.00ppm
0to200.0% 0to110
o
C(32to230
o
F)
Accuracy ±(0.2%FS+1digit) ±2%FS ±1
o
C(±1.8
o
F)Note:Accuracy
isratedbetween10and65
o
C
(50and149
o
F)
Resolution 0.01mg/L 0.10% 0.1
o
C/F
Compensation AutomaticTemperatureCompensation(ATC):0to60
o
C(32to140
o
F)
ManualSalinityCompensation(MSC):0to50ppt
ManualAltitudeCompensation(MAC):0to7000m(20000ft.)
OperatingTemperature 0to50
o
C(32to122
o
F)
OperatingHumidity 85%maximum(noncondensing)
Power 9Vdc,universalACadapter
Dimensions 24cm(L)x17cm(W)x5.7cm(H)or31.2cm(H)withrod
 9.5”(L)x6.7”(W)x2.2”(H)or12.3” (H)withrod
Weight 740g(1.6lbs.)
Copyright©20132015FLIRSystems,Inc.
Allrightsreservedincludingtherightofreproductioninwholeorinpartinanyform
www.extech.com
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12

Extech Instruments WQ500 User manual

Category
Measuring, testing & control
Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI