Thermo Fisher Scientific Orion 2111XP Sodium Analyzer User guide

Category
Oxygen Equipment
Type
User guide
User Manual
Thermo Scientific
Orion 2111XP
Sodium Analyzer
UM-263639-005 Revision B January 2017
© 2013-2017 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the
property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and
pricing are subject to change. Not all products are available in all countries. Please
consult your local sales representative for details.
No part of this publication may be reproduced, transmitted, transcribed, stored in a
retrieval system, or translated into any language or any computer language, in any
form or by any third party, without the prior written permission of Thermo Fisher
Scientific.
Disclaimer
Thermo Fisher Scientific does not accept any responsibility for any damage caused to its
products by unauthorized personnel. Use of non-Thermo Scientific™ reagents and/or
replacement parts will void all warranties.
Contents
Section 1 General Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Features and Benefits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Application Packages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Principles of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Principles of Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Double Known Addition (DKA) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Offline Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Fluidics Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Two Channel Analyzer Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Section 2 Analyzer Preparation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Unpacking the Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Mounting and Plumbing Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Sample Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Electrical Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Safety Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Warning Labels and Locations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Wiring the Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Terminal Assignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Electrode Wiring Assignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Installation of DIPA Reagent and Diffusion Tubing . . . . . . . . . . . . . . . . . . . . . . . . . 17
Installation of Ammonia Reagent and Diffusion Tubing . . . . . . . . . . . . . . . . . . . . . . 20
Installation of New Electrode Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Installation of the ATC Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Installation of a New Reference Electrode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Section 3 Analyzer Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
Description of Basic Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Description of Keypad Icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Use of the Setup Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Setup Mode Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Shutdown and Start-Up Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Section 4 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
Calibration Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Flow Cell Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Rinsing the Flow Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Air Regulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Before Performing a DKA Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Performing a DKA Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Calibration Abort Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .62
Calibration Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63
Calibration At Custom Concentrations Using DKA . . . . . . . . . . . . . . . . . . . . . . . . . .64
Span Check Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .65
Offline Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .65
Section 5 Analyzer Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67
Weekly Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67
Monthly Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68
Yearly Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72
Section 6 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Diagnostics Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75
Slope Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .80
Troubleshooting Matrix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .82
Error/Event Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .83
Resetting the Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .86
Serial Number and Software Revision . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .87
Service and Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .87
Section 7 Customer Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
Notice of Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .90
Statement of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .91
WEEE Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .92
Terms and Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .92
Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .95
Mounting Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .95
ISE Default Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .96
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .96
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .100
Accessory Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .102
Field Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .102
Recommended Consumables for Annual Operation . . . . . . . . . . . . . . . . . . . . . . . .103
Pipet Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .104
Pipet Techniques . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .105
Thermo Scientific Orion 2111XP Sodium Analyzer | 1
S E C T I O N
1General Information
This user guide covers the operation, maintenance and troubleshooting for the Thermo
Scientific Orion 2111XP sodium analyzer, which offers unmatched reliability in monitoring
critical sample streams throughout the power/steam generation and industrial water industry.
Introduction
Monitoring the sodium ion content of steam and water circuits to produce accurate and
reproducible results requires a very well designed and maintained system. The system must
optimize the fluidic design with the sensing technology to enable low level (ppb) measurement
of the contaminants as well as measuring across the linear range of the analyzer.
The 2111XP sodium analyzer meets all of the criteria for accurate and dependable sodium
monitoring and more. The 2111XP analyzer incorporates innovative technologies that include:
 Premium electrodes
 Accurate and precise flow cell design
 Marquee help screen
 Pump-less reagent addition and DKA calibration system
1 | General Information
2 | Orion 2111XP Sodium Analyzer Thermo Scientific
Section
Thermo Scientific Orion 2111XP Sodium Analyzer
Markets
 Power
 Semiconductor
 Chemical and petrochemical
 Pulp and paper
Applications
 Feedwater / make-up water
 Boiler feedwater
 Drum boilers
 Demineralized water
 Steam condensate
 Cation exchange breakthrough
 High acid samples
Features and Benefits
The Thermo Scientific Orion 2111XP sodium analyzer features an expandable platform design
that incorporates over 30 years of expertise in sodium measurements combined with innovative
Thermo Scientific Orion technologies for superior performance.
 Choice of application/reagent packages:
Flexible configurations for applications ranging from low-level detection to high acid
samples.
 Accurate and precise measurements in the range of 0.30 ppb to 200 ppm:
Reliable measurements and a wide measurement range with selectable resolution.
 Premium reference and sensing electrodes:
Superior accuracy and stability over a wide temperature range.
 Advanced flow cell design with air stirring:
Automatic sample handling and contamination control with no moving parts.
 Patented scrolling marquee:
Intuitive menu-driven, digital user interface.
 Data log of previous measurements and calibration:
View measurement, calibration and error history.
 Self diagnostics:
Ease of maintainability.
 Password protection:
Security and peace of mind for your operation.
 Auto-ranging electronics with an easy to read backlit LCD display:
Analyzer determines the best range.
1 | General Information
Section
Thermo Scientific Orion 2111XP Sodium Analyzer | 3
Application Packages
The Thermo Scientific Orion 2111XP sodium analyzer is ideally suited to meet the demanding
needs of high purity water measurements and high acid-cation exchange applications, all in one
system and all from one of the most trusted names in sodium monitoring.
The 2111XP analyzer is offered in three application packages that are uniquely designed to
accommodate the changing requirements for successful sodium monitoring. Our flexible reagent
kits are available in prepackaged bottles for safe and convenient replacement – save your
valuable time and money!
 Ammonia Application Package – for general purpose sodium measurements, provides up
to 60 days of continuous operation.
 Diisopropylamine (DIPA) Application Package – for low range sodium detection,
provides up to 60 days of continuous operation.
 Cation/High Acid Application Package for cation exchange breakthrough and high acid
samples, provides up to 45 days of continuous operation.
Application Packages
Ammonia Application Package
(Cat. No. 2111XA)
Reagent: Ammonia
Range: 0.30 ppb to 200 ppm
Accuracy: ± 5% or 0.3 ppb (with DKA calibration)
DIPA Application
Package
(Cat. No. 2111XD)
Reagent:
Diisopropylamine (DIPA)
Range: 0.10 ppb to 10 ppm
Accuracy: ± 5% or 0.1 ppb (with DKA calibration)
Cation / High Acid
Application Package
(Cat. No. 2111XC)
Reagent:
Ammonia
Range: 1.0 ppb to 200 ppm
Accuracy: ± 5% or 2 ppb (with DKA calibration)
1 | General Information
4 | Orion 2111XP Sodium Analyzer Thermo Scientific
Section
Principles of Operation
The sample enters the Thermo Scientific Orion 2111XP sodium analyzer and passes through
the inlet valve, bypass/needle valve, inlet filter, pressure regulator, flow meter and into the
restrictor tubing. The sample then passes through the flow cell manifold into a reagent bottle and
through a diffusion tubing assembly where pH adjustment takes place. The pH-adjusted sample
then flows back through the manifold into the flow cell as air is introduced from the air pump to
ensure proper mixing and fast response. The sample then flows into an atmospheric drain via
the diverter valve.
Figure 1: 2111XP Schematic
The sensing electrode responds logarithmically to changes in the sodium ion concentration. This
response is described by the Nernst equation:
E = E0 + 2.3 (RT/nF) log (C/Ciso)
Where:
E = measured electrode potential, mV
E0= potential, when C equals Ciso, mV
Safety Drain for
Reagent Overflow
Restrictor Tubing
Pressurizes Flow Cell
Calibration Mode - PUSH
to Initiate Siphoning
Valve Positions
Sample/Measure Mode - PULL
Analyzer
12345
Power
4-20mA Output
Air Pump
FFllooww MMeetteerr
Bypass / Needle
Inlet Valve
Valve and Filter
Regulator
Valve
Reference Solution
Port
Calibration
REF
TEMP
Sense
Flow Cell Block
Calibration Level
Air Filter
Reagent
Reagent Bottle
Diffusion Tube
Fluid Path
Reagent Manifold
Electric
Fluid Restrictor Tubing
Air Path
Reference Solution Path
Diffusion Tubing
KEY
Reagent Overflow Path
Main Feed
Air Inlet
Measure Level
Drain
Check
Valve
Sample
VVVaaaalllvlvveee
Cal
1 | General Information
Section
Thermo Scientific Orion 2111XP Sodium Analyzer | 5
R = ideal gas constant
T = temperature of sample, degrees K
n = valence of ionic species (+1 for sodium ion)
F = Faraday’s constant
C = effective sodium ion concentration (activity)
Ciso = concentration (activity) of sodium ion where potential E is temperature
independent (isopotential point)
The above equation indicates that the measured potential varies with both temperature and the
concentration of the ion of the interest. In order to eliminate error caused by fluctuations in
sample temperature, the 2111XP microprocessor constantly updates temperature corrections
from data supplied by the ATC probe.
From the Nernst equation, the theoretical response of a sodium ion selective electrode to a
ten-fold change in concentration at 25 °C is 59.16 mV. This is referred to as the electrode slope
(S). Most electrodes, however, do not exhibit a theoretical slope. Therefore, the analyzer is
calibrated to determine its actual value. Two standards are used to provide information
necessary for the microprocessor to compute the actual slope and E0 for use during sample
analysis.
In order to eliminate interference from hydrogen ions, which can become significant when
measuring low levels of sodium, the 2111XP analyzer raises the sample pH. This pH adjustment
is accomplished by the patented passive-diffusion process wherein the sample passes through
a length of tubing contained in the reagent bottle. The reagent diffuses through the tube wall and
mixes with the sample, which raises the sample pH.
Principles of Calibration
Calibration procedures for analytical instruments are important and must be performed carefully.
The calibration procedure used in the Thermo Scientific Orion 2111XP analyzer is a variation of
Double Known Addition (DKA) using advanced electrode and flow cell technology in
combination with the passive diffusion system. This method has the distinct advantages of being
fast, easy, and accurate.
Double Known Addition (DKA)
Before calibration begins, the diverter valve is pushed in to divert flow from the measure drain,
allowing the flow cell to fill.
At the beginning of the DKA calibration the actual concentration in the sample is unknown. The
analyzer measures the potential (Es) and stores this value in the microprocessor. A known
amount of standard 1 solution is added to the flow cell, which increases the concentration (Cs)
with a corresponding known amount (dc1). During this process, air is pumped into the flow cell,
1 | General Information
6 | Orion 2111XP Sodium Analyzer Thermo Scientific
Section
thoroughly mixing sample and standard in a closed- loop system. The new potential (E1) is
measured and stored automatically when stability is reached. Adding standard 2, preferably 10
times more concentrated than standard 1, increases the concentration (dc2) in the sample
reservoir. Again, the new potential (E2) is measured and stored when the reading is stable. Now,
we have the following three unknowns:
Es = E0 + S(Ts/298.15) log (Cs/Ciso)
E1 =E
0 + S(T1/298.15) log [(Cs + dC1)/Ciso]
E2 =E
0 + S(T2/298.15) log [(Cs+ dC1+ dC2)/Ciso]
S is the Slope at 25 °C (298.15 K)
T is the temperature in Kelvin, measured when the potential E is measured.
Es, E1, E2 have been determined during the calibration procedure. The microprocessor solves
these three equations, to obtain the values of S and E0. The calibration result is stored for use
during online monitoring to convert the measured potential and temperature in the sample into
concentration values in either ppm or ppb.
When the calibration is complete the flow cell drains as the sample flow returns. The flow cell
volume returns to the measurement level. After allowing approximately 30 minutes for
concentrated calibration solution to be flushed from the system, the 2111XP analyzer can begin
sample measurement again.
In addition to Double Known Addition (DKA), the 2111XP analyzer also allows the operator the
ability to perform an offline calibration.
Figure 2: Flow Cell Volume for DKA
1 | General Information
Section
Thermo Scientific Orion 2111XP Sodium Analyzer | 7
Offline Calibration
The offline calibration feature of the 2111XP analyzer allows the operator to adjust the analyzer
to values determined by alternate methods used in their laboratory such as elemental
spectroscopy and ion chromatography.
The offline calibration is essentially a one point calibration. To perform an offline calibration, a
sample is taken from the bypass of the analyzer; the sample concentration value is stored in
memory; the sample is analyzed by an alternate method of choice; the previously stored reading
is adjusted to the lab method result; and the analyzer is then returned to the analysis mode. The
term “offline calibration” refers only to the fact that a sample from 2111XP analyzer bypass is
taken “offline” for laboratory analysis; in fact, no downtime is experienced during the procedure
and the analyzer remains online throughout.
Fluidics Diagram
Figure 3: Fluidics Diagram
Scrolling Marquee
LCD Display
Warning Icon
Mode Indicator
Mounting
Hole
Keypad
Flow Cell
Reservoir
Siphon Tube
Flow Cell Block
Inlet Valve
Reagent
Bottle
Adapter
Assembly
Pressure
Regulator
Status Indicators
Flow Meter
Calibration Port
Reference
Filling
Solution
Restructor
Tube
Assembly
Air Pump
Bypass/Needle
Valve
Inlet Filter
Reagent
Manifold
Diverter Valve
Diffusion Tubing
Assembly
Reagent Bottle
1 | General Information
8 | Orion 2111XP Sodium Analyzer Thermo Scientific
Section
Glossary
Refer to Figure 3.
Inlet Valve – Accepts the sample stream via 1/4 inch NPTF connector. The operator must
supply the sample with a pressure between 8 and 100 psig.
Inlet Filter – 60 micron stainless steel filter traps particulate matter in the sample stream.
Bypass/Needle Valve – Used to redirect flow in the bypass system.
Pressure Regulator – Adjusts flow on the incoming sample stream.
Flow Meter – Measures sample flow rate.
Restrictor Tube Assembly – Used in conjunction with the pressure regulator to lower
downstream pressure.
Reagent Manifold – Directs sample flow in and out of the reagent bottle assembly.
Reagent Bottle Adapter Assembly – Connects the reagent bottle assembly to the manifold.
Diffusion Tubing Assembly Semi-permeable tubing through which reagent diffuses into the
sample.
Reagent Bottle – Contains water-soluble amine or ammonia reagent that raise the sample pH.
Flow Cell – Contains sodium sensing electrode, reference electrode and ATC probe.
Diverter Valve – Allows the flow cell reservoir to fill during calibration by forming a closed-loop
system.
Sodium Sensing Electrode – Senses sodium ions in sample stream and produce an electrical
potential dependent on sample concentration.
Reference Electrode – Provides a constant reference potential and completes the
measurement circuit.
Reference Electrode Filling Solution Bottle – Provides constant flow of electrolyte solution
through reference electrode for maximum stability.
ATC Probe – Measures sample temperature and inputs data to microprocessor for automatic
temperature compensation (ATC).
1 | General Information
Section
Thermo Scientific Orion 2111XP Sodium Analyzer | 9
Calibration Port – Allows introduction of standards to the sample reservoir during calibration.
LCD Display – Provides digital readouts of concentration, temperature, millivolts and error
codes.
Air Pump – Used to mix the sample during both measurement and calibration.
Keypad – Consists of five mode keys, four prompt indicator lights, two scroll keys and one key
for entering data. Mode and error indicators are also incorporated on the keypad.
Status Indicator – Two LED lights that illuminate according to current status of the analyzer.
Note: When either the yellow or red LED is lit, there may be an entry in the diagnostics mode
that indicates the error. The logging feature must be initiated in the setup mode. Refer to
Section 3, Use of the Setup Mode for instructions.
Two Channel Analyzer Configurations
A pH/ORP module or conductivity module can be added by the operator to the second channel
of the Thermo Scientific Orion 2111XP sodium analyzer for the ultimate flexibility in
measurement capabilities. The 2100 series pH/ORP and conductivity analyzers provide
accurate and reliable measurements in ultra pure water as well as the harshest industrial
environments. Combined with decades of superior Thermo Scientific Orion sensor technology,
our systems provide rapid results with complete stability.
When a pH/ORP module or conductivity module is installed on the second channel of the
2111XP sodium analyzer, refer to the Thermo Scientific Orion 2100 Series pH/ORP Analyzer
and Conductivity Analyzer User Guide for detailed instructions on operating the pH/ORP or
conductivity analyzer. Visit www.thermofisher.com/water to download any of the 2100 series
analyzer user guides.
Green Light: Indicates that system is in correct working condition.
Yellow Light: Indicates a warning, system in hold or that maintenance is required.
Red Light: Indicates that something is seriously wrong.
Cat.No Description
2100PH2 Second channel module for pH/ORP
2100CD2 Second channel module for conductivity
10 | Orion 2111XP Sodium Analyzer Thermo Scientific
S E C T I O N
2Analyzer Preparation
WARNING: The instructions provided in this user guide are recommendations from the
manufacturer to ensure safe and correct operation of the analyzer. If the analyzer is not used as
recommended by the manufacturer this can lead to incorrect operation or injury.
Unpacking the Analyzer
Thermo Scientific Orion analyzers are assembled, tested and packaged with great care. Refer
to Figure 4.
Report any obvious damage of shipping container to carrier and hold for inspection. The carrier
(not Thermo Fisher Scientific) is responsible for any damage incurred during shipment.
1. Open the outer box. Remove the top two foam corner support pieces.
2. Open the inner box. This box should contain the analyzer and ATC temperature probe,
sodium electrode box, reference electrode box, low level standards kit (two 2 oz. bottles of
standard 1, 19.1 ppm sodium; two 2 oz. bottles of standard 2, 192 ppm sodium; and
one 2 oz. bottle of etching solution), high level standards kit (two 2 oz. bottles of standard
1, 1910 ppm sodium; two 2 oz. bottles of standard 2, 19200 ppm sodium; and one 2 oz.
bottle of etching solution), the options kit and user guide CD.
3. Remove the cardboard retaining shell by sliding it over the entire mounting board and the
analyzer.
4. Carefully remove the entire mounting board with analyzer from the inner box.
Note: Do not lift or pull the analyzer by the fluidics or the electronic components. Use the
back panel to lift the analyzer system.
2 | Analyzer Preparation
Section
Thermo Scientific Orion 2111XP Sodium Analyzer | 11
Figure 4: Unpacking the Analyzer
5. Unbolt the analyzer from the mounting board by removing the four mounting bolts with a
9/16'' wrench. These bolts may be discarded.
6. Carefully place the analyzer at a convenient location until proper installation can be
completed.
Mounting and Plumbing Instructions
Refer to the Appendix, Mounting Dimensions section.
WARNING: Do not connect power prior to the mounting and plumbing of the analyzer.
Recommendations
 Select a site for the analyzer that allows it to be permanently bolted with ample height for
atmospheric drain operation. Be sure that there is ready access to the electronic controls,
calibration port and electrodes.
 A clearance of 15 inch (about 40 cm) must be allowed above the flow cell calibration port.
Insert the pipet vertically (not angled) during the calibration.
 The analyzer location must permit connections to a sample line, drain and AC power supply
and any connections for output devices.
 The analyzer should be mounted as close to the sampling point as possible. This ensures
the fastest possible response to a changing sample condition. Refer to the Appendix,
Sample Conditions section.
 For proper flow cell operation, the analyzer must be installed straight and level upon its
mounting location. Failure to level the analyzer may cause poor siphoning in the flow cell.
2 | Analyzer Preparation
12 | Orion 2111XP Sodium Analyzer Thermo Scientific
Section
Instructions
1. Prepare the mounting holes. Carefully lift the analyzer and bolt it into place. Do not lift the
analyzer by holding on to any of the plumbing or fluid handling components.
2. Connect a waste line to the outlet of the analyzer, which is 34'' NPT male. The waste line
should be connected to a drain of sufficient capacity, 0.5 inch (1.27 cm) OD is
recommended.
3. Connect a sample line to the inlet of the analyzer, which is 14'' NPT female. It is
recommended that a shutoff valve be installed at the sampling point.
4. The analyzer must be mounted and leveled vertically for proper operation.
Sample Requirements
Additional information is listed in the Appendix, Specifications section.
Sample inlet connection – 1/4'' NPTF. If particulate matter is present in the sample,
pre-filtration is necessary. The 60 micron stainless steel filter located after inlet valve will remove
moderate amounts of particulates.
Flow rate – 40 mL/minute (nominal) for the Ammonia and DIPA Application Packages and 25
mL/minute (nominal) for the Cation/High Acid Application Package.
Pressure – 8 to 100 psig. Consult Technical Support for details on sample handling if the
pressure is outside of this range.
Temperature – Temperature must be between 5 and 45°C.
Sodium level – Sodium levels are read directly in ppb or ppm, when calibrated with Thermo
Scientific Orion sodium standards 1 and 2.
Sample alkalinity – Sample alkalinity should be less than 250 ppm CaCO3 equivalent. For
higher sample alkalinity, contact Technical Support.
Electrical Wiring
The warning icon highlights important information that should be strictly followed when using the
analyzer for your own safety. Failure to follow these instructions may result in injuries.
WARNING: Read and observe the following safety recommendations.
Safety Requirements
 Prior to wiring, a switch or circuit breaker for disconnecting the analyzer from power supply
should be installed.
The switch should be in close proximity to the analyzer and with easy reach of the user.
The switch should be marked as the disconnecting device for the analyzer.
2 | Analyzer Preparation
Section
Thermo Scientific Orion 2111XP Sodium Analyzer | 13
 To reduce the risk of shock hazard, disconnect the power prior to opening the analyzer.
 Before connecting the analyzer to the main, make sure that the voltage lies within either
range: 100-120V 200mA / 200-240V 100mA; 50-60 Hz AC.
 Cutting off the power by disconnecting power source will not reset the analyzer. This
analyzer incorporates a non-volatile memory and will maintain calibration and settings after
power failure. Battery power is supplied to the display for the date and time functions.
 If a repair is required, or to arrange Return Material Authorization, call Technical Support or
contact your local authorized dealer.
 Installation and wiring of the analyzer may only be carried out in accordance with applicable
local and national codes per this user guide.
 Be sure to observe the technical specifications and input ratings.
Warning Labels and Locations
WARNING: The following section provides important information that should be strictly followed
when using the analyzer for your own safety. Failure to follow these instructions may result in
injuries.
The safety warning icons are used in two locations on the analyzer.
 Faceplate – Refer to Figure 5.
Figure 5: Faceplate
 Power supply – Refer to Figure 6.
Figure 6: Power Supply
.
Note: Replace the fuse only with a fuse of same rating.
2 | Analyzer Preparation
14 | Orion 2111XP Sodium Analyzer Thermo Scientific
Section
Wiring the Analyzer
WARNING: Read and observe the following requirements. If you install the wrong fuse for your
system, you could damage the analyzer. Make sure that you select the correct fuse rating and
discard the additional fuses supplied in the fuse kit.
Required Tools
 Options kit – includes fuses, cable glands, conduit fitting and green screw terminal.
 Phillips head screwdriver.
 2 mm blade flat-head screwdriver.
Figure 7: Electronic Enclosure with Cable
Glands
1. Open the faceplate – loosen the four screws using a Phillips head screwdriver. The
electronics faceplate will open via the hinge pin connection.
2. Remove one or two of the two unused cable glands as required for wiring power cable or
auxiliary connections. Power cable optional hole locations are shown in Figure 7.
3. Select and install the appropriate size cable gland or conduit fitting as required.
Figure 8: Terminal Connector Location
Terminal Connector
2 | Analyzer Preparation
Section
Thermo Scientific Orion 2111XP Sodium Analyzer | 15
4. Feed the power cable through the conduit or cable glands as required.
5. Wire the power cable to the green screw terminal connector from the options kit. Select
correct terminal for hot conductor depending on line voltage, insert ground wire into the
correct terminal and connect the other cable to the neutral terminal. Refer to Figure 8 for
terminal connector location.
6. Plug the terminal connector into the power supply. Refer to Figure 6.
7. The universal power supply uses both fuses in the fuse kit. Install by inserting the fuse in the
fuse holder and secure it using the twist and lock method. The fuses are clearly labeled with
the appropriate voltages for your system. Refer to Figure 6 and Figure 8 for the correct fuse
holder positions. Refer to the table below for fuses required.
Terminal Assignments
AC Voltage Fuse rating
115V 200mA, 250V, Fast Acting
230V 100mA, 250V Fast Acting
Terminal Layout Terminal Layout Terminal Layout Terminal Layout
1 mA1 output 9Relay 1 26 Sensing electrode 28 Do not connect
2GND common ground 10 Relay 1 27 Do not connect 29 Do not connect
3 mA2 output 11 Relay 2 30 Preamp power
4Air pump (ISE only) 12 Relay 2 31 Preamp ground
5Air pump (ISE only) 13 Relay 3 32 Shield
6Shield ground for conductivity 14 Relay 3 33 Shield
7Do not connect 15 Do not connect 34 Jumper to pin 26 when
using preamp
8Do not connect 16 Temperature ground
17 Temperature drive
18 Temperature sense
19 Solution ground
20 Conductivity drive +
21 Conductivity sense +
22 Conductivity sense -
23 Conductivity drive -
24 Reference electrode
25 Jumper to pin 24
when using preamp
2 | Analyzer Preparation
16 | Orion 2111XP Sodium Analyzer Thermo Scientific
Section
Figure 9: Terminal Assignments
Electrode Wiring Assignments
Sodium Electrode
26 Sensing Electrode Connect clear wire
33 Shield Connect black wire
Reference Electrode
24 Reference electrode Connect clear wire
32 Shield Connect black wire
2100TP Temperature Probe
16 Temperature ground /thermistor Connect white wire
17 Temperature drive /thermistor Connect green wire
19 Solution ground Connect red wire
2001TM Temperature Probe
16 Temperature ground /thermistor Connect white wire
17 Temperature drive /thermistor Connect green wire, jumper 17 and 18
18 Temperature sense Jumper to 17
19 Solution ground Connect red wire
2001SC pH Electrode
24 Reference electrode Connect black wire
26 Sensing electrode Connect clear wire
110250 ORP Electrode
19 Solution ground Connect black wire
24 Reference electrode Connect purple wire
26 Sensing electrode Connect coax center wire
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110

Thermo Fisher Scientific Orion 2111XP Sodium Analyzer User guide

Category
Oxygen Equipment
Type
User guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI