CV-5000 & KB-50 system realizes the refraction only with simple operations by automatic refraction program.
Programs 1-5 are customizable.
FLOW OF SIMPLE EXAMINATION
When customizing the defalut setting or programs, press [Shift] and [Menu] to access the [Settings] menu.
STANDARD course requires
only dial and [SET] button.
Smart Cross Cylinder
2
End of the test
1
Examination
This program proceeds automatic astigmatism test using cross cyinder.
Compare the first and the second Input the patient's reply.
Flow of Program
The flow of STANDARD program is as follows.
Look at the
title bar,
Dial Navigation,
Help menu,
then you perform the examination by
turning dial for "which is better 1 or 2"
and pressing [SET] button to go to the
next step.
Program Guided Examination
Press [Print] button to print out the data.
Press [Menu] button, if necessary, to acess [Eye
examination tool] for consultation.
Press [CL] button to reset. All for the test.
1
R/G Test
2
Cross Cylinder Test (axis)
3
Cross Cylinder Test (power)
4
R/G Test
5
Spherical / VA Test (for far vision)
RIGHT
6
R/G Test
7
Cross Cylinder Test (axis)
8
Cross Cylinder Test (power)
9
R/G Test
10
Spherical / VA Test (for far vision)
LEFT
BOTH
in case of STANDARD
Please use near vision chart for 13 and 14 which is the standard component.
11Binocular Balance Test
12
Spherical Test (for far vision)
13Addition Test
14VA Test (for near vision)
(from far to near)
C
Prepare CV-5000
Set the CV-5000 head in front of patient.
Press [PD] button and turn on the corneal
illumination.
Check the position of eye, by looking through
the examination windows.
Check the vertex position looking through the
corneal aligning windows.
Adjust forehead rest knob.
A Tu rn the power on.
Tu rn on the external units, such as Chart
Projector.
Tu rn on the CV-3000 & KB-50.
Auto-Refractometer Computerized
Lensmeter
RM CL
B
Input objective measurement data and lensmeter data.
Measure the eye with RM (Refractometer).
Press [PRINT] button on RM, then data is
output automatically.
Measure the glasses with CL (Computerized
Lensmeter).
Press [PRINT] button on CL, then data is
output automatically.
Press [IN] button on KB-50.
Data from RM and CL is displayed in data
reference area.
If in selectable mode, select the patient data
from each of the interfaced external devices.
D Select program course
Press [Prog] button.
Select the program using the dial.
Press [SET] button to start program course.
STANDARD is registered as default program
course.