CAUTIONS FOR USE
Basic caution
When installing the instrument on a wall, use the wall-mounting metal fixture, which is one of the components,
and install it on the strong wall that can support its weight. (Do not install it on the plaster-board wall.)
[The instrument may fall off to cause a trouble or injure someone.]
Fix the wall-mounting metal fixture, which is one of the components, on the wall as taking care for its direction.
[The instrument may fall off to cause a trouble or injure someone.]
Set the test distance of the instrument correctly.
[The reliability of the measurement result may be lowered.]
Be careful not to damage the liquid crystal display or stain it with fingerprint, dust or others.
[The reliability of the measurement result may be lowered.]
Disposal
Dispose of the instrument according to local disposal and recycling laws.
ENVIRONMENTAL CONDITIONS FOR USE
Temperature : 10°C - 40°C
Humidity : 30% - 90% (non-condensing)
Air pressure : 700hPa - 1060hPa
STORING METHOD, USAGE PERIOD AND OTHERS
1. Storage (without wrapping (without package))
* Temperature: 10°C - 40°C
Humidity: 10% - 95% (without dew condensation)
Air pressure: 700hPa - 1060hPa
* THIS INSTRUMENT DOES NOT MEET THE TEMPERATURE REQUIREMENTS OF ISO 15004-
1 FOR STORAGE. DO NOT STORE THIS INSTRUMENT IN CONDITIONS WHERE THE TEM-
PERATURE MAY RISE ABOVE 40°C OR FALL BELOW 10°C.
2. Storage (with wrapping (with package))
Temperature: -20°C - 50°C
Humidity: 10% - 95%
3. Transportation (with wrapping (with package))
Temperature: -40°C - 70°C
Humidity: 10% - 95%
4. When storing the instrument, ensure that the following conditions are met:
(1) The instrument should not be splashed with water.
(2) Store the instrument away from the environment where air pressure, temperature, humidity,
ventilation, sunlight, dust, salty/sulfurous air, etc. could cause damage.
(3) Do not store or transport the instrument on a slope or uneven surface or in an area where it is
subject to vibrations or instability.
(4) Do not store the instrument where chemicals are stored or gas is generated.
5. Usage period
8 years from delivery providing regular maintenance is performed (according to the self-certification
[Topcon data])
CHECKPOINTS FOR MAINTENANCE
Maintenance by user
1. Regularly maintain and check the instrument and its parts.
2. Make sure that the test distance of the instrument is set correctly.
3. Make sure that the instrument is installed on the wall or on the Floorstand, which is an accessory, securely.
4. When using the instrument after a prolonged period of inactivity, confirm normal and safe operation before-
hand.
5. If the liquid crystal display is stained, clean it.
6. When this instrument is not in use, apply the dust cover to it.
For details, refer to "PREPARATION BEFORE USE" and "MAINTENANCE."