Fritsch Particle Sizer ANALYSETTE 28 IS Operating instructions

  • Hello! I am a chat assistant and I have read the operating instructions for the FRITSCH Particle Sizer ANALYSETTE 28 IMAGESIZER. The document provides detailed information on how to setup, operate and maintain the device for both dry and wet measurements. I am ready to answer your questions regarding the features and capabilities of this particle sizer or any specific parts of operating instructions.
  • What is the purpose of the ANALYSETTE 28 ImageSizer?
    What is the measuring range of the ANALYSETTE 28 ImageSizer?
    What should be done before each measurement?
Operang instrucons
PARTICLE SIZER
ANALYSETTE 28 IMAGESIZER
Read the instrucons prior to performing any task!
Valid from: 28.2000/00131
Translaon of the original operang instrucons
Version 11/2023 Index 015
Fritsch GmbH
Milling and Sizing
Industriestraße 8
D - 55743 Idar-Oberstein
Telephone: +49 6784 70-0
Internet: www.fritsch.de
3
Cercaons and CE conformity
Cercaon
CE Conformity
Cercaons and CE conformity
Fritsch GmbH has been cered by the SGS-TÜV Saar GmbH.
An audit cered that Fritsch GmbH conforms to the requirements of the DIN EN
ISO 9001:2015.
The enclosed Conformity Declaraon lists the guidelines the FRITSCH instrument
conforms to, to be able to bear the CE mark and the UKCA mark!
4
Table of contents
Table of contents
1 Basic structure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2Safety informaon and use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.1 Requirements for the user. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.2 Scope of applicaon. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.3 Obligaons of the operator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.4 Informaon on hazards and symbols used in this manual. . . . . . . . . . . . 11
2.5 Device safety informaon. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.6 Hazardous points. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2.7 Electrical safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.7.1 General informaon. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.1 Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.2 Weight. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.3 Operang noise. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.4 Voltage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.5 Current consumpon. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.6 Protecon class. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.7 Power consumpon. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.8 Electrical fuses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4Installaon. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.1 Transport and storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.2 Unpacking. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.3 Scope of delivery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.3.1 Accessory case. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4.3.2 Accessories case 1 - dry. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
4.3.3 Accessory case 2 - wet dispersion unit. . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.3.4 Accessory case 3 - wet dispersion unit connecon kit. . . . . . . . . . . . . 21
4.3.5 Accessory set - US box. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.4 Seng up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.4.1 Transport securing devices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.5 Ambient condions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.6 Electrical connecon. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.6.1 Stability of the mains voltage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.7 Preparaon of the ANALYSETTE 28 ImageSizer. . . . . . . . . . . . . . . . . . . . . 24
4.7.1 Connecon of the camera to the lens and inseron. . . . . . . . . . . . . . . 24
5
Table of contents
4.8 Connecons for supply units. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.8.1 Mounng the pressure reducer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
4.9 Preparaon of the computer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
4.10 Checking the communicaon between the device and the PC. . . . . . . . 31
4.10.1 Checking the communicaon between the camera and the PC. . . . . 32
5 Using the device. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
5.1 Changing the lens. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
5.1.1 Adjusng the camera angle. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.2 Dry measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.2.1 Installing the feeder and funnel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.2.2 Funnel holder 250 ml. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.2.3 Adjusng the funnel height. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.3 Wet measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.3.1 Inserng the measuring cell. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.3.2 Selecng the liquids. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5.3.3 Filling the measuring circuit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.3.4 Sample addion. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5.3.5 Wet measurement EXTENDED. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
5.4 Fritsch test powder F16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
5.5 Fritsch test powder F70. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6.1 Lenses for dry measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6.2 Accessories for wet measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6.2.1 Immersion pump set. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6.3 Calibraon plate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
7 Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.1 Cleaning the device. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.1.1 Cleaning the falling chute. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.2 Cleaning the lens. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.3 Cleaning the windows. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.3.1 Removing the window holder. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
7.4 Cleaning the lens of the digital camera. . . . . . . . . . . . . . . . . . . . . . . . . . . 50
7.5 Cleaning the hoses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
7.6 Cleaning the wet measuring cell. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7.6.1 Design of wet measuring cell. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7.6.2 Preparaon. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
7.6.3 Emptying the system. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
7.6.4 Removing the wet measuring cell. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
6
Table of contents
7.6.5 Cleaning the measuring cell glasses. . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
7.6.6 Sealing rings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
7.6.7 Mounng the wet measuring cell. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
7.6.8 Check the wet measuring cell for impermeability. . . . . . . . . . . . . . . . . 54
8 Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
8.1 Replacing hoses in wet dispersion units. . . . . . . . . . . . . . . . . . . . . . . . . . . 56
9 Repairs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
9.1 Checklist for troubleshoong. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
10 Disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
11 Guarantee terms. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
12 Exclusion of liability. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
13 Safety logbook. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
14 Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
7
Basic structure
1 Basic structure
1 Funnel
2 Feeder
3 Rotary knob for funnel adjustment
4 Height-adjustable funnel holder
5 Funnel support column
6 Latches
7 Sample collecng vessel
8 Housing ap
8
Basic structure
9 5 MP digital camera
10 Lens holder
11 Lens
12 Lens holder end stop
13 Window holder
14 Falling chute (wet measuring cell holder)
15 Flash unit
16 Main switch
17 Mains connecon
18 Connecon, wet dispersion unit / ultrasonic box
19 USB port
9
Basic structure
20 Wet dispersion unit
21 ImageSizer measuring unit
22 Sample bath
23 Wet measuring cell
10
Safety informaon and use
2 Safety informaon and use
2.1 Requirements for the user
This operang manual is intended for persons assigned with operang and
monitoring the Fritsch ANALYSETTE 28 ImageSizer. The operang manual and
especially its safety instrucons are to be observed by all persons working on
or with this device. In addion, the applicable rules and regulaons for accident
prevenon at the installaon site are to be observed. Always keep the operang
manual at the installaon site of the ANALYSETTE 28 ImageSizer.
People with health problems or under the inuence of medicaon, drugs, alcohol
or exhauson must not operate this device.
The ANALYSETTE 28 ImageSizer may only be operated by authorised persons and
serviced or repaired by trained specialists. All commissioning, maintenance and
repair work may only be carried out by technically qualied personnel. Qualied
personnel are persons who, because of their educaon, experience and training
as well as their knowledge of relevant standards, regulaons, accident prevenon
guidelines and operang condions, are authorised by those responsible for the
safety of the machine to carry out the required work and are able to recognize
and avoid possible hazards as dened for skilled workers in IEC 364.
In order to prevent hazards to users, follow the instrucons in this manual.
Malfuncons that impair the safety of persons, the ANALYSETTE 28 ImageSizer or
other material property must be reced immediately. The following informaon
serves both the personal safety of operang personnel as well as the safety of
the products described and any devices connected to them: All maintenance and
repair work may only be performed by technically qualied personnel.
This operang manual is not a complete technical descripon. Only the details
required for operaon and maintaining usability are described.
Fritsch has prepared and reviewed this operang manual with the greatest care.
However, no guarantee is made for its completeness or accuracy.
Subject to technical modicaons.
2.2 Scope of applicaon
The ANALYSETTE 28 ImageSizer is used for the analysis of the parcle shape and
size of powders and bulk solids. The parcle sizer has an extra-large measuring
range of 20 µm to 20 mm! It is therefore an ideal device for uncomplicated
quality control in various areas of applicaon.
The parcle sizer described here is an operaons resource for use in industrial
environments. This operaons resource possesses hazardous live components
during operaon.
2.3 Obligaons of the operator
Before using the ANALYSETTE 28 ImageSizer, this manual is to be carefully read
and understood. The use of the ANALYSETTE 28 ImageSizer requires technical
knowledge; only commercial use is permied.
11
Safety informaon and use
The operang personnel must be familiar with the content of the operang
manual. For this reason, it is very important that these persons actually receive
the present operang manual. Ensure that the operang manual is always near
the device.
The ANALYSETTE 28 ImageSizer may exclusively be used within the scope of
applicaons set down in this manual and within the framework of guidelines put
forth in this manual. In case of non-compliance or improper use, the customer
assumes full liability for the funconal capability of the ANALYSETTE 28 Image-
Sizer and for any damage or injury arising from failure to full this obligaon.
By using the ANALYSETTE 28 ImageSizer the customer agrees with this and
recognizes that defects, malfuncons or errors cannot be completely excluded.
To prevent risk of damage to persons or property or of other direct or indirect
damage, resulng from this or other causes, the customer must implement
sucient and comprehensive safety measures for working with the ANALYSETTE
28 ImageSizer.
Neither compliance with this manual nor the condions and methods used
during installaon, operaon, use and maintenance of the ANALYSETTE 28
ImageSizer can be monitored by Fritsch GmbH. Improper execuon of the instal-
laon can result in property damage and thus endanger persons. Therefore,
we assume absolutely no responsibility or liability for loss, damage or costs
that result from errors at installaon, improper operaon or improper use or
improper maintenance or are in any way connected to these.
The applicable accident prevenon guidelines must be complied with.
Generally applicable legal and other obligatory regulaons regarding environ-
mental protecon must be observed.
2.4 Informaon on hazards and symbols used in this manual
Safety informaon
Safety informaon in this manual is designated by symbols. Safety informaon is
introduced by keywords that express the extent of the hazard.
DANGER
This symbol and keyword combinaon points out a directly
hazardous situaon that can result in death or serious injury if
not avoided.
WARNING
This symbol and keyword combinaon points out a possibly
hazardous situaon that can result in death or serious injury if
not avoided.
CAUTION
This symbol and keyword combinaon points out a possibly
hazardous situaon that can result in slight or minor injury if
not avoided.
12
Safety informaon and use
Special safety informaon
Safety informaon in the procedure instrucons
NOTICE
This symbol and keyword combinaon points out a possibly
hazardous situaon that can result in property damage if not
avoided.
To call aenon to specic hazards, the following symbols are used in the safety
informaon:
DANGER
This symbol and keyword combinaon points out a directly haz-
ardous situaon due to electrical current. Ignoring informaon
with this designaon will result in serious or fatal injury.
DANGER
This symbol and keyword combinaon designates contents and
instrucons for proper use of the machine in explosive areas
or with explosive substances. Ignoring informaon with this
designaon will result in serious or fatal injury.
DANGER
This symbol and keyword combinaon designates contents
and instrucons for proper use of the machine with combus-
ble substances. Ignoring informaon with this designaon will
result in serious or fatal injury.
WARNING
This symbol and keyword combinaon points out a directly
hazardous situaon due to movable parts. Ignoring informaon
with this designaon can result in hand injuries.
WARNING
This symbol and keyword combinaon points out a directly
hazardous situaon due to hot surfaces. Ignoring informaon
with this designaon can result in serious burn injuries due to
skin contact with hot surfaces.
Safety informaon can refer to specic, individual procedure instrucons. Such
safety informaon is embedded in the procedure instrucons so that the text can
be read without interrupon as the procedure is being carried out. The keywords
described above are used.
Example:
1. Loosen screw.
13
Safety informaon and use
Tips and recommendaons
Further designaons
2. CAUTION
Risk of entrapment at the lid.
Close the lid carefully.
3. Tighten screw.
This symbol emphasises useful ps and recommendaons as
wells as informaon for ecient operaon without malfunc-
on.
To emphasise procedure instrucons, results, lists, references and other ele-
ments, the following designaons are used in this manual:
Designaon Explanaon
1., 2., 3. ...
Step-by-step procedure instrucons
ðResults of steps in the procedure
References to secons in this manual and relevant
documentaon
Lists without a specic order
[Buon]
Operang elements (e.g. push buon, switch), dis-
play elements (e.g. signal lamps)
‘Display’
Screen elements (e.g. buons, funcon key assign-
ment)
2.5 Device safety informaon
The parcle sizer has been designed from the viewpoint of safety to users;
nevertheless, residual hazards cannot be ruled out.
The device may only be used for the purpose described under
Chapter 2.2
‘Scope of applicaon’ on page 10
.
Only use original accessories and original spare parts. Failure to observe this
instrucon can compromise the safety of the machine.
Safe conduct must be strictly observed during all work with the device.
All currently applicable naonal and internaonal accident prevenon guide-
lines must be complied with.
Do not remove the informaon signs.
14
Safety informaon and use
DANGER
Explosion hazard!
The device is not explosion-protected!
The device must not be used in electrically conducve,
dusty or damp environments.
DANGER
Mains voltage
Do not allow any liquids to run inside the device.
WARNING
The maximum accepted concentraon (MAC) levels of the valid
safety regulaons must be observed. If necessary, venlaon
must be provided or the machine must be operated under an
extractor hood.
NOTICE
Immediately replace damaged or illegible informaon signs.
Unauthorised alteraons will void Fritsch's declaraon of conformity to Euro-
pean direcves and void the guarantee.
If aer reading the operang manual you sll have quesons or problems,
please do not hesitate to contact our technical specialists.
2.6 Hazardous points
DANGER
There is a risk of fatal injuries from power surges when cleaning
the device. Disconnect the mains plug before cleaning the
device.
CAUTION
Crushing hazard!
When
Moving the funnel
Closing the device
Inserng the holders with the "measuring windows" 13 +
14
Inserng the lenses
Mounng the camera
Inserng the sample collecon container
15
Safety informaon and use
NOTICE
Do not leave any oxidizing parts on the channel or in the funnel
- risk of corrosion.
2.7 Electrical safety
2.7.1 General informaon
The main switch separates the device from the mains on two poles.
Switch o the main switch if the ANALYSETTE 28 ImageSizer is to stand idle
for a longer duraon (e.g. overnight).
In the switched-on state, the switch is illuminated red and the control lamp on
the front side blue.
16
Technical data
3 Technical data
3.1 Dimensions
ImageSizer:
84 x 20 x 28 cm (width x depth x height without funnel and feed channel)
Wet dispersion unit:
29 x 27.2 x 29 cm (W x D x H)
3.2 Weight
ImageSizer dry:
36.8 kg net
ImageSizer wet:
58.8 kg net
3.3 Operang noise
The noise level is up to approx. 78 dB(A).
3.4 Voltage
The voltage range is from 100 - 240 V.
3.5 Current consumpon
The maximum current consumpon is 0.2 A (230 V), 0.6 A (115 V) .
3.6 Protecon class
IP 22
3.7 Power consumpon
The maximum power consumpon is 50 W.
3.8 Electrical fuses
Micro-fuse 5 x 20 mm 2A slow-blow, 2 pieces The fuses are located inside the
mains power socket (19).
17
Installaon
4Installaon
4.1 Transport and storage
DANGER
Do not step under the transport pallet during transport.
The ANALYSETTE 28 is delivered in a wooden crate on a pallet. Do not lt or stack
this crate. The ImageSizer must be transported and stored in its packaging and
protected from weather and outside inuences.
Transport over larger distances only in the original Fritsch packaging.
We recommend using a forkli or pallet truck for transporng the packed device.
WARNING
Improper liing can lead to personal injury or property
damage. The machine is only to be lied with suitable equip-
ment and by qualied personnel.
The guarantee excludes all claims for damage due to improper transport.
4.2 Unpacking
Remove the transport packaging as shown in the following illustraons.
1. Cut open the straps of the package using side cuers.
2. Remove the staples using an appropriate tool.
3. Li the lid o.
The accessories case becomes visible.
4. Remove the surrounding packaging.
5. Remove the accessories case.
18
Installaon
6. Remove the foam mouldings and li the ImageSizer from the transport
cradle using two persons. If using the wet version, li out the wet disper-
sion unit as well from the transport cradle.
NOTICE
Do not use the funnel support column to li the ANALY-
SETTE 28 from the transport cradle.
For liing, reach under the base plate.
4.3 Scope of delivery
Please check the delivery against your order! If the delivery is incomplete or dam-
aged, inform the shipping agent immediately and inform Fritsch GmbH within 24
hours. Any claims received later cannot be considered.
4.3.1 Accessory case
Item Item number Quanty Designaon
1 81.1003.00
81.1015.00
1
1
USB cable. PC connecon
USB interface cable, camera connecon
2 28.2039.00 1 USB sck with soware and documentaon
3 86.0560.00 1 Mulple socket outlet
4 Round jar with small parts
19
Installaon
Item Item number Quanty Designaon
5 83.4305.00 1 Spirit level
6 81.6300.00
81.6310.00
1
1
Connecon cable (230V)
Connecon cable (110V)
7 28.2010.00 1 Digital camera
4.3.2 Accessories case 1 - dry
Item Item number Quanty Designaon
7 85.2300.00
85.2110.01
1
1
Standard test powder F16 and/or
Standard test powder F70
8 1 Lens (depending on order)
9 1 Funnel (depending on order)
10 90.0008.00 1 Lens brush
11 28.2210.10 1 Sha cover
12 83.4002.10 1 Chemical spatula
13 1 Vibraon feeder (depending on order)
14 85.6010.00 1 Flat brush
20
Installaon
Item Item number Quanty Designaon
15 90.0003.00 1 Anstac cloth
16 90.0007.00 1 Opcal ssue, 30 sheets
17 90.0009.00 1 Bellows
28.2215.00 1 When ordering lens 28.2061.00 and 28.2062.00:
Holder for 250 ml funnel
4.3.3 Accessory case 2 - wet dispersion unit
Item Item number Quanty Designaon
21 28.2605.00 1 Flow measuring cell
22 90.0012.00
90.0013.00
1 Polyurethane foam swabs, 12.7 x 130 mm, 50 pcs.
Polyurethane foam swabs, 2.7 x 80 mm, 500 pcs.
23 90.0003.00 1 Anstac cloth
24 22.8472.00 1 Dusazin 901 tenside
25 85.2302.00 1 Cleaning powder
26 85.2110.01 1 Standard test powder F70
/