  • Hello! I am an AI chatbot trained to assist you with the Midmark 626 Barrier-Free® Exam Chair (-001 thru -006) Installation guide. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
1
© 2018 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
MA10555-2i
003-10173-00 Rev. AA1 4/9/18
Step 1: Remove Top Cover
A) Raise chair backrest to the full upright position.
B) Unplug chair.
C) Remove two screws from back of Top Cover.
D) Pull cover toward head end to disengage tabs,
then lift up to remove cover.
Screws
Note
For parts list, refer to the 626 Parts manual.
Caution
Always disconnect power to chair before removing
covers or performing any service procedures.
Tabs
Diagnostic Interface PCB installation (002-10267-00)
Style A
midmark.com
2
© 2018 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
MA10837i
Diagnostic
Interface PCB
Step 2: Replace the Diagnostic Interface PCB.
A) Remove wire connections to Diagnostic Interface PCB.
B) Push four standoff tabs while pulling up on Diagnostic
Interface PCB to release and remove the board.
C) Position the new Diagnostic Interface PCB as shown.
D) Press board onto four standoffs.
Step 3: Install connectors to Diagnostic Interface PCB.
A) Install wire harness as shown.
B) Plug in chair to verify power to Diagnostic Interface PCB.
C) Display on the Diagnostic Interface PCB should illuminate.
D) Unplug chair.
E) Reinstall top rear cover.
Stando
Stando