1© Midmark Corporation 2003 SF-1828
003-1292-00 Rev. LA2 (4/2/19)
Side Delivery Unit Installation
Mounting Location
NOTE:
Verify desired position
(ex. RH / LH) with
proper personnel before
installation.
Minimum Cabinet Size Requirements
(Side Delivery)
Flex Arm
Fixed Arm
Short Flex Arm
28 in. ( 71.1 cm)
Long Flex Arm
36 in. ( 91.4 cm)
Long Flex Arm
27.5 in. ( 69.8 cm)
Short Flex Arm
17 in. ( 43.2 cm)
55 to 61 in.
(139.7 to 154.9 cm)
12 to 24 in.
(30.4 to 60.9 cm)
CONTENTS
Mounting Location ............................................ 1
Minimum Cabinet Size Requirements ............. 1
Mounting Requirements ................................... 1
Regulator Installation........................................ 2
Cabinet Mount Bracket Installation .................. 2
Wall Mount Bracket Installation ........................ 3
Bearing Bracket Installation ............................. 4
Unit Arm Installation ......................................... 5
Stop Pin Installation ......................................... 5
Flex Arm Lock Activation & Leveling ............... 6
Tubing Connections .......................................... 7
Dimensions, Range of Motion ......................... 8
Mounting Requirements
Mounting Structure
The mounting structure must be capable of supporting the weight / torque listed below without deflection:
Unit / Arm Assembly
Weight (Maximum) ................. 50 lbs (22.6 kgs.)
Torque (Maximum) ................. 170.8 ft-lbs. (231.5 N•M)
To prevent the unit from drifting, the mounting plate must deflect
no more than 1/16" (1.6 mm) over a 12" (30.5) distance.
Electrical Requirements: 24VAC, 3.2 amp
13VAC, 2.0 amp