Electronic Calibration
If electronic scaling is the accessory medium for calibration, see the accessory technical brief for instructions.
Servicing
The following instructions are for removal, inspection and installation of meter parts and assemblies. Refer to applicable
illustrated parts list for part numbers of components and ordering information. Accessory service and repair procedures are
provided in literature specific to that device.
Meter Disassembly Method
These procedures are for disassembly of the meter:
1. Shut o the uid ow to meter. Place a container under the meter and relieve uid pressure in the meter by uniformly
loosening the bolts (bronze meter) or the retaining ring (plastic meter). Fluid will run out into the container.
2. Remove the bolts or the retaining ring while holding the housing bottom in place. Remove the housing bottom. The
chamber assembly may drop from the housing as the bottom is removed.
3. If the chamber assembly is out, set it aside. If not, rst remove the chamber strap and then remove the chamber assembly.
4. Remove the meter screen. Inspect the screen for dirt and corrosion. Clean or replace the screen as necessary.
5. Inspect the chamber assembly:
a. Check the disc in the chamber assembly for warpage, cracks or wear. A severely worn disc can cause
over-delivery of the fluid being metered.
b. Check the thrust roller and thrust roller insert for excessive wear.
After inspection, clean or replace the chamber assembly as necessary. If the chamber is worn or corroded, replace the entire
chamber and disc assembly.
6. Re-install the housing chamber assembly, screen, chamber strap and bottom as follows:
a. Assemble the screen in the proper location.
b. Position the chamber assembly in the proper orientation. Add the chamber strap.
c. Position the housing bottom with the correct seal. (Replacement of seal is always recommended.)
d. Assemble and tighten the bolts or retaining ring.
Alternative Disassembly Method for Service
Loosen all the connection pieces slightly. Turn the meter upside-down, putting gravity on your side. This will help eliminate
the possibility of the chamber assembly and or the screen component dropping during disassembly and causing damage.
Recalibration
After repairing or replacing a meter component or assembly, perform the calibration check and adjustment procedure to
ensure that the meter is properly calibrated and will operate in accordance with published specifications.
www.badgermeter.com
Trademarks appearing in this document are the property of their respective entities. Due to continuous research, product improvements and enhancements, Badger Meter reserves
the right to change product or system specications without notice, except to the extent an outstanding contractual obligation exists. © 2014 Badger Meter, Inc. All rights reserved.
The Americas | Badger Meter | 4545 West Brown Deer Rd | PO Box 245036 | Milwaukee, WI 53224-9536 | 800-876-3837 | 414-355-0400
México | Badger Meter de las Americas, S.A. de C.V. | Pedro Luis Ogazón N°32 | Esq. Angelina N°24 | Colonia Guadalupe Inn | CP 01050 | México, DF | México | +52-55-5662-0882
Europe, Middle East and Africa | Badger Meter Europa GmbH | Nurtinger Str 76 | 72639 Neuen | Germany | +49-7025-9208-0
Europe, Middle East Branch Oce | Badger Meter Europe | PO Box 341442 | Dubai Silicon Oasis, Head Quarter Building, Wing C, Oce #C209 | Dubai / UAE | +971-4-371 2503
Czech Republic | Badger Meter Czech Republic s.r.o. | Maříkova 2082/26 | 621 00 Brno, Czech Republic | +420-5-41420411
Slovakia | Badger Meter Slovakia s.r.o. | Racianska 109/B | 831 02 Bratislava, Slovakia | +421-2-44 63 83 01
Asia Pacic | Badger Meter | 80 Marine Parade Rd | 21-06 Parkway Parade | Singapore 449269 | +65-63464836
China | Badger Meter | 7-1202 | 99 Hangzhong Road | Minhang District | Shanghai | China 201101 | +86-21-5763 5412 Legacy Document: IOM-067-05-EN 53400-067
Control. Manage. Optimize.