12
9. Photometrical measuring
9.1 General notes for handling the DIT
9.1.1 Handling the photometer
• Keep the vial hole clean. No reagents should get into the hole.
Clean the vial hole carefully with a dry and soft cloth.
• The instrument must be horizontal during measuring.
• Protect from direct light (sunlight)!
• Condensation water in the vial hole can provoke incorrect
measurements! Clean the vial hole carefully with a dry and
soft cloth.
• Do not use aggressive solvents.
See section 12.3 Cleaning.
9.1.2 Handling the vials
• Keep the vials clean. They can be cleaned with normal
cleaning products.
• Lime scum can be removed with diluted hydrochloric acid
(HCl 10 %).
• Scratched or opaque vials should be replaced.
• It is not necessary to cover the vials, but direct sunlight should
be avoided.
• Observe the vial inscriptions! Vials especially dedicated for the
determination of one parameter must not be used for other
analysis to avoid the carry-over of chemicals.
• Take care that the vial is dry on the outside! Clean the vial with
a dry and soft cloth before measuring. Beware of moisture if
handling very cold sample water! Moisture on the vial or in the
vial hole provokes incorrect measurements!
• Observe the preference direction of the vial (black vertical
marking).
9.1.3 Handling and dosing the reagents
• Observe the best-before date of the reagents. It is printed on
the packaging and must not be exceeded.
• Store the reagents at temperatures between 15 °C and 25 °C
(unless otherwise indicated). Otherwise, the quality and
durability will suffer.
• To obtain a uniform size of drops, hold the dropping bottle
vertically.
• Powder reagents must not get moist.
• Protect all reagents from light.
• Close the reagent bottles immediately after use to avoid
chemical reactions for example with oxygen or carbon dioxide
in the air.
• Before measuring, the vials must be rinsed with water to be
tested.
• Never interchange the covering caps of the reagent bottles.
This could lead to chemical reactions, resulting in unusability
of the reagents.
9.2 Reference adjustment
9.2.1 Reference
Before doing photometrical measuring, the instrument must be
adjusted to zero. This is necessary after commissioning, after
long periods of storage or after temperature variation (10 °C).
The basic factory adjustment will not be influenced. Fill a vial with
sample water without adding any reagents. Select "photom." in
the main menu and then the menu item "reference". The
instrument then asks the user to insert the vial. After pressing
[OK], the DIT does the zero-measuring. This is finished after
70 seconds, and the instrument returns to the "photom." menu.
9.2.2 Zero check
Before each measuring, the zero adjustment can be checked.
When having selected a parameter, it is possible to select
"zerocheck". The instrument asks the user to fill a vial with 10 ml
of the sample water and insert it. Confirm both operations by
pressing [OK], and the DIT does a quick checking of the zero
value. After measuring, the result is displayed in extinction units
(dgt). Ideally, it should be 0.000 dgt. If it is more than ± 0.007 dgt,
the reference adjustment described above must be repeated.
If the result is acceptable, the display will show "zerocheck o.k.",
and in case of major difference "zerockeck error".
If different parameters shall be determined in succession, this
zero check has to be executed before each measurement.
9.3 Operator guidance and advice for photometrical
measuring
The basic operating mode is the same for all photometrical
measurings.
9.3.1 Carrying out the measuring
Select the menu item "photom." in the main menu and then the
parameter to be determined. Then the options "measure",
"zerocheck", "cancel" and, if necessary, "blanktest" will appear.
"measure" leads to the measurement itself. If it is selected by
pressing [OK] once, all analysis steps will be requested by the
instrument and have to be confirmed by the user. All steps and
waiting times listed in the following sections will be shown.
Especially in connection with waiting for the reaction time, this is
very useful, as no other clock will be needed. After the waiting
time, an acoustic signal indicates that the vial can be inserted,
which has to be confirmed by pressing [OK]. Now the DIT does
the photometrical measuring and shows the result or, if
necessary, an error message. The end of the measurement is
also indicated by an acoustic signal. After pressing [OK] again,
the result can be printed or stored, and the instrument returns to
the parameter menu.
9.3.2 Expert mode and professional mode
If "measure" is selected by pressing [OK] twice, the instrument
changes to the expert mode. The step-by-step guidance will then
not be shown. This is useful for proficient users, as it shortens the
operating time. However, the reaction times must be kept!
The expert mode cannot be switched on permanently, but has to
be selected for each measurement by pressing [OK] twice. If the
so-called professional mode ("profi mode") is switched on in the
configuration menu, the expert mode described above is
continuously switched on.
Caution
This photometer is a highly sensitive measuring
instrument.
For maximum performance and lifetime, careful
operation of the instrument and its accessories is
necessary.
Avoid vibrations, direct sunlight and humidity.
Caution
Due to conversion of production, the dosing
prescriptions may change.
In case of doubt, observe the prescriptions
printed on the reagent bottles.
Caution
Only an correctly made reference will give a
correct measurement!