YSI 900 Chlorine Colorimeter User manual

Type
User manual
900 Chlorine Colorimeter
OPERATIONS MANUAL
INSTRUCTION MANUAL: PAGES 2  21
MODE D‘EMPLOI: PAGES 22  41
INSTRUCCIONES: PÁGINA 42 61
Contact Information
ORDERING & TECHNICAL SUPPORT
Telephone: (800) 897-4151
(937) 767-7241
Monday through Friday, 8:00 AM to 5:00 PM ET
Fax: (937) 767-9353 (orders)
(937) 767-1058 (technical support)
Mail: YSI Incorporated
1725 Brannum Lane
Yellow Springs, OH 45387
USA
Internet: www.ysi.com
When placing an order please have the following information available:






SERVICE INFORMATION
YSI has authorized service centers throughout the United States and Internationally. For the nearest

Support directly at 800-897-4151.



3
Table of Contents
CAUTION



• General notes .............................................................4
 .......................................4
Method notes ..............................................................4
 .....................................................5
• Functional description .....................................................6
Operation .................................................................6
 ...........................................................7
 ........................................................7
 ...............................................................7
• Methods .................................................................8
 ...................................................8
 ..............................................10
 .............................................12
 .............................................14
• Menu options ............................................................16
Menu selections ...........................................................16
 .......................................................16
Transmitting stored data ....................................................16
Setting date and time .......................................................17
• Calibration Mode .........................................................17
 ...........................................................17
 ....................................................19
• Technical data ...........................................................20
Operating messages .......................................................21
Error codes ..............................................................21
Battery disposal ...........................................................21

4
Guidelines for photometric measurements
 

 

 

 

vial aligned with the 
 
shielding ring.
 

 
 


 


 
Method notes
 

 

 
 
General Notes
5
Replacement of batteries:
Correct position of the vial (Ø 24 mm):
CAUTION:

 
 


AAA/LR 03
AAA/LR 03
AAA/LR03
AAA/LR03

(E) seal ring



(A) screws
Correctllingofthevial:
correct
10 ml
5 ml
10 ml
5 ml
wrong
correct correct wrongwrong

6
Operation



Scroll Memory (SM)
-
















The result appears in the display.
The result is saved automatically.
Repeating the test:

Repeating the zero:
Press the 
METHOD
METHOD
METHOD
RESULT
On
Off
Mode
Zero
Test
Zero
Test
Zero
Test
0.0.0
Zero
Test
METHOD
Functional Description
7

Display backlight
-

Recall of stored data

4 seconds to access the recall menu.
Countdown / reaction period





-
surement starts immediately.
Caution:
An incomplete reaction period can lead to incorrect test
results.
!
!
!
Zero
Test
8
Methods
!
Zero
Test
Zero
Test
CL 6
Tablet Powder
CL 6
Tablet Powder
CL 6
Tablet Powder
0.0.0
Tablet Powder
RESULT
Tablet Powder
RESULT
Tablet Powder
Chlorine with Tablet: 0.01 – 6.0 mg/l
a) Free Chlorine
Fill a clean vial (24 mm Ø) with 10 ml of the water sample  

empty it, leaving a few drops
remaining in the vial.
Add one DPD No. 1 tablet 



dissolved.





b) Total Chlorine
Add one DPD No. 3 tablet 


dissolved.


Wait for a reaction period of 2 minutes.



c) Combined Chlorine

Tolerances:





!
Zero
Test
Zero
Test
CL 6
Tablet Powder
CL 6
Tablet Powder
CL 6
Tablet Powder
0.0.0
Tablet Powder
RESULT
Tablet Powder
RESULT
Tablet Powder
9

Notes:
1. Vial cleaning:



Preparation:
then rinse all glassware thoroughly with deionized water.


3. Preparing the sample:






5. Exceeding the measuring range:


6. Turbidity (can lead to errors):









10
Methods
!
Zero
Test
Zero
Test
CL 10
Tablet Powder
CL 10
Tablet Powder
0.0.0
Tablet Powder
CL 10
Tablet Powder
RESULT
Tablet Powder
RESULT
Tablet Powder
Chlorine HR with DPD Tablet: 0.1 – 10 mg/l
a) Free Chlorine
Fill a clean vial (24 mm Ø) with 10 ml of the water sample  

empty it, leaving a few drops
remaining in the vial.
Add one DPD No. 1 HR tablet 



dissolved.


.
.

b) Total Chlorine
Add one DPD No. 3 HR tablet
.

dissolved.


Wait for a reaction period of 2 minutes.

.

c) Combined Chlorine

Tolerances:
 
 
 

11

Notes:
1. Vial cleaning:



Preparation:
then rinse all glassware thoroughly with deionized water.
2. Preparing the sample:


 



4. Turbidity (can lead to errors):

3


 

12
Methods
Chlorine with Powder Pack: 0.02 – 2.0 mg/l
a) Free Chlorine
Fill a clean vial (24 mm Ø) with 10 ml of the water sample


one Chlorine Free-DPD/F10 Powder Pack


sec.).





b) Total Chlorine



one Chlorine Total-DPD/F10 Powder Pack


sec.).


Wait for a reaction period of 3 minutes.



c) Combined Chlorine

Tolerances:


!
Zero
Test
Zero
Test
CL 2
Tablet Powder
CL 2
Tablet Powder
CL 2
Tablet Powder
0.0.0
Tablet Powder
RESULT
Tablet Powder
RESULT
Tablet Powder
13

Notes:
1. Vial cleaning:



Preparation:
then rinse all glassware thoroughly with deionized water.
 

 

4. Preparing the sample:


 



6. Exceeding the measuring range:


 

14
Methods
Chlorine with Powder Pack: 0.1 – 8.0 mg/l
a) Free Chlorine
Fill a clean vial (10 mm
) with 5 ml of the water sample   

two Chlorine Free-DPD/F10 Powder Pack







b) Total Chlorine
Fill a clean vial (10 mm
) with 5 ml of the water sample   

two Chlorine Total-DPD/F10 Powder Pack

(20 sec.).


Wait for a reaction period of 3 – 6 minutes.



c) Combined Chlorine

Tolerances:



Zero
Test
Zero
Test
CL 8
Tablet Powder
CL 8
Tablet Powder
CL 8
Tablet Powder
RESULT
Tablet Powder
RESULT
Tablet Powder
0.0.0
Tablet Powder
0.0.0
Tablet Powder
15

Notes:
1. Vial cleaning:



Preparation
then rinse all glassware thoroughly with deionized water.
 

 

4. Preparing the sample:


 



 

7. 

 

16
Menu Options
Menu selections




1
diS: recall stored data
1
Prt: printing stored data
2
3
setting the date and time
 
1
diS – Recall of stored data


result is displayed):
 
Year YYYY (e.g. 2010)
 
Time hh:mm (hourhour:minuteminute)
Test Method
 



1
Prt – Transmitting stored data (to Printer or PC)
900 Series Data Hub-

The 900 Series Data Hub
       


instrument returns to test mode.

On
Off
On
Off
Mode
!
!
Mode
Mode
Zero
Test
PrtG
Store Date
Print Time
2
1
Cal
diS
1
Store Date
Print Time
Prt
1
Store Date
Print Time
1
1
2
17

Mode
!
SET
DATE
YYYY
(2 sec.)
Mode
Zero
Test
E 132


the instrument returns to test mode (see also 900 Series Data Hub manual).
2
Setting date and time (24-hour-format)









the instrument returns to the measurement mode.
2
Store Date
Print Time
18
Calibration Mode
USER CALIBRATION (Cal)
CAL
METHOD
0.0.0
METHOD
METHOD
CAL


















1 digit.

standard.



CAL
cAL
Zero
Test
Zero
Test
Cal
Store Date
Print Time
CAL
RESULT
+ x
CAL
Mode
Zero
Test
RESULT


19

FACTORY CALIBRATION RESET
cAL
SEL
SEL
CAL
CAL
SEL
-
ods and ranges.

displayed.

and hold.





or:







Mode
On
Off
Mode
On
Off
Zero
Test
RESULT
Cal
Store Date
Print Time
20
Technical Data
Instrument: single wavelength, direct reading colorimeter
Light source: 

530 nm

Wavelength accuracy: 
Photometric accuracy*:   
Photometric resolution: 0.01 A
Power supply:  

Auto-OFF: 
Display: 
Storage: 
Interface: 
Time: 
Calibration: 

Dimensions: 
Weight: 
Ambient conditions: 
rel. humidity: 3090% (non-condensing)
Waterproof: 
CE: 
*measured with standard solutions


  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64