a) These steps will prime the syringe with water, isopropanol, water and then polymer.
During the polymer prime, monitor the syringe to determine if bubbles or leaks are
present in the syringe.
♦If bubbles are present refer to the Checking for Bubbles module.
♦If leaks are present refer to the Checking for Leaks module.
♦If necessary, replace the syringe.
Step 5:
♦If the array fill syringe does not move, remove the syringe from the pump.
a) Exit from the wizard and restart the system.
b) Start the change syringe wizard again and select the option to replace the syringe.
♦ If the pump moves, install a new syringe and continue with the wizard.
♦If the pump does not move, contact technical support.
Sample Transfer
This step of the wizard will prepare the syringes for removal by moving the autoloader tips to the
load bar to flush the lines and fill the syringes with water.
♦If the syringe is leaking, follow the wizard to remove it from the instrument.
♦If you are checking for bubbles or leakage go to step 4.
♦If you are verifying pump movement go to step 5.
Step 4:
♦If you are locating the source of bubbles or leaks, do not remove the syringe but forward to the
next section in the wizard.
Verifying Syringe Pumps Module pn4375504.doc
Rev A Page 7 of 9