ABI PRISM® 3700 Instrument-Hardware-Array View Troubleshooting Module
Please contact technical support if you have any questions regarding the document.
Text in blue indicates a link to another section or another document.
Observations
Possible Causes
Recommended Actions
Refer to the Checking the CCD calibration window Module to determine if bubbles
have been removed from the cuvette.
Verify that within the polymer bottle, the tubing is immersed in polymer.
Check for leaking (dried polymer) at the in-line filter or syringes and if necessary
replace components.
Verify that the fittings on the port above the sheath flow syringe are finger tight.
Turquoise color
Go to larger picture
Presence of bubbles in the
cuvette. The source of the
bubbles will need to be
identified, refer to the Checking
for Bubbles module.
Run the cuvette flush module, this may require several flushes until bubbles are
removed.
Leak in sheath flow syringe
Replace the sheath flow syringe using the change syringe wizard.
Impaired movement of the
syringe pump.
Check the syringe pump, refer to the Verifying Syringe Pumps module.
Leak at the inline filter
Clean dried polymer from the area and replace the inline filter.
Verify that the fittings on the port above the sheath flow syringe are finger tight.
Presence of bubbles in the
sheath flow syringe.
Verify that the tubing in the polymer bottle is immersed in polymer.
Yellow Color
Go to larger picture
Sheath flow syringe or pump
may not be functioning
Verify the syringe pump movement with and without the sheath flow syringe, refer to
the Verifying the Syringe Pumps module.
Array View Colors Module pn 4375401 .doc
Rev A Page 1 of 19