Mettler Toledo pH electrode 405-DPAS-SC Operating instructions

  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo pH electrode 405-DPAS-SC Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
405-DPAS-SC pH Electrodes
53 500 085
Instruction Manual
Type 405-DPAS-SC
pre-pressurized
gel-reference
pH Electrodes
cs Návod k obsluze 2
da Brugsanvisning 4
de Betriebsanleitung 6
en Instruction manual 8
es Manual de instrucciones 10
fr Instructions dutilisation 12
 Käyttöohje 14
hu Használatiutasítás 16
it Istruzioni per l‘uso 18
ja 20
ko 22
nl Gebruikershandleiding 24
pl Instrukcjaobsługi 26
pt Manual de instruções 28
ru Инструкция по эксплуатации 30
sv Bruksanvisning 32
th ำ�ำ� 34
zh 36
8 405-DPAS-SC pH Electrodes
© 04 / 16 Mettler-Toledo GmbH 405-DPAS-SC pH Electrodes
53 500 085 Printed in Switzerland
1 Introduction
The METTLER TOLEDO Type 405-DPAS-SC combination
pH electrode is a low-maintenance, pre-pressurized,
autoclavable electrode with a gel-type reference. The
reference element incorporates a silver-ion trap to prevent
silversuldecontaminationof the reference junction. The
gel-typereferenceisalsoavailableinacombinationORP
and as a separate reference electrode
2 Preparation
Afterunpacking,examinetheelectrodeforanymechanical
damage.
METTLER TOLEDO electrodes are factory tested and are
shippedready-to-use.(Tomaintainfactorypressurization,
the Type 405-DPAS-SC electrodes have a silicone seal
coveringtheceramicdiaphragm;thismustberemovedwith
theenclosedbladepriortouse.)Removethehydrationcap
andremoveanyelectrolytecrystalsbyrinsingtheelectrode
withdistilledwater.Patdry;donotrubasthismaybuildup
astaticchargeandlengthentheresponsetime.
Eliminateanyairbubblesaroundtheinternalelementinthe
pH-sensitivemembrane.
3 Operation
METTLERTOLEDOpHelectrodescanbeusedwithmostpH/
mVmeters.
Followmanufacturer’sinstructionsforuse.
Atwo-pointcalibrationisrecommended,usingfreshbuffers
ofknownpHvalueswhichbracketthesamplepH;onebuffer
shouldbeclosetopH7tosetthezeropoint.Forfastroutine
analysis,astandardizationusingonebufferatavaluenear
thesamplepHcanbedone.
4 Sterilization
Type 405-DPAS-SC electrodes are steam-sterilizable in-
situandautoclavable.Followoperatinginstructionsforthe
autoclave. It is advisable(but notnecessary) toput the
closurecap(p/n002011096)ontheconnectorpriorto
autoclaving.
Repeatedsterilizationmaycausetheelectrolytetodiscolor
slightlyduetotheelevatedtemperatures.Thisdoesnotaffect
theelectrodeperformance.
5 Maintenance
When properly cared for, METTLER TOLEDO electrodes
shouldprovideaccurateresultsoverthousandsofuses.
Storage:Electrodesshouldberinsedthoroughlyafteruse
andstoredin3MKClorpH4buffersolutionwithaddedKCl.
DONOTstoretheelectrodedryorindistilledortapwater;
thisdehydratesthepHmembraneandtheelectrodemust
berehydratedbeforeusebyimmersingforseveralhoursin
3MKCl.
Relling:Type405-DPAS-SCgel-reference electrodes are
notrellable.
en
en Instruction manual
Instructions for Type 405-DPAS-SC
pre-pressurized gel-reference pH Electrodes
405-DPAS-SCpHElectrodes 9
405-DPAS-SC pH Electrodes © 04 / 16 Mettler-Toledo GmbH
Printed in Switzerland 53 500 085
en
ProteinContamination:CleantheelectrodewithpHElectrode
CleanerforProteinsfollowinginstructionsonthebottle.
Reactivating pH membrane: Use Reactivating Solution
followingdirectionsCAREFULLY.
Servicingacloggedordryreferencejunction: Immerse
theelectrodeinabeakerof3MKCl.Heatthesolutionto
approximately70°C(160°F)andallowelectrodetosoak
inthissolutionasitcoolstoroomtemperature.
6 Troubleshooting
Noresponse(symptom)
– Poor meter connection
 Checkconnectorsatthemeterandat
theelectrode
– Cracked glass
Replaceelectrode
Slowresponse(symptom)
–DehydratedpHmembrane
Rehydrate(seeMaintenance)
–Cloggedreferencejunction
SoakinwarmKCl(seeMaintenance)
– Protein contamination
UsepHElectrodeCleaner(see
Maintenance)
–Depletedgellayer
UseReactivatingSolution(see
Maintenance)
Noise(symptom)
– Poor meter connection
Checkconnector
–Cloggedreferencejunction
(Seeabove)
Lowslope(symptom)
–DehydratedpHmembrane
Rehydrate(seeMaintenance)
– Contaminated membrane
(Seeabove)
Forfurtherassistance,callourCustomerServiceCenterat
+1-800-352-8763.
38 405-DPAS-SC pH Electrodes
© 04 / 16 Mettler-Toledo GmbH 405-DPAS-SC pH Electrodes
53 500 085 Printed in Switzerland
Notes
405-DPAS-SCpHElectrodes 39
405-DPAS-SC pH Electrodes © 04 / 16 Mettler-Toledo GmbH
Printed in Switzerland 53 500 085
Notes
Brazil Mettler-ToledoInd.eCom.Ltda.
AvenidaTamboré,418–Tamboré,
BR-06460-000Barueri/SP,Brazil
Phone+551141667400
e-mailmettler@mettler.com.br;service@mettler.com.br
France Mettler-ToledoAnalyseIndustrielleS.A.S.
30,Bld.deDouaumont,FR-75017Paris,France
Phone+33147370600
e-mailmtpro-f@mt.com
Germany Mettler-ToledoGmbH
Prozeßanalytik,Ockerweg3,DE-35396Gießen
Phone+49641507-444
e-mailprozess@mt.com
Switzerland Mettler-Toledo(Schweiz)GmbH
ImLangacher,CH-8606Greifensee
Phone+41449444760
e-mailProSupport.ch@mt.com
UnitedStates METTLERTOLEDO
ProcessAnalytics
900MiddlesexTurnpike,Bld.8,Billerica,MA01821,USA
Phone+17813018800
Freephone+18003528763(onlyUSA)
e-mailmtprous@mt.com
FormoreaddressesofMETTLERTOLEDOMarketOrganizationspleasegoto:
www.mt.com/pro-MOs
Mettler-ToledoGmbH,ProcessAnalytics,ImHackacker15,CH-8902Urdorf
Phone+41447296211,Fax+41447296636
Subjecttotechnicalchanges
04/2016©Mettler-ToledoGmbH
PrintedinSwitzerland.53500085
www.mt.com/pro