2.2 02/2008, version 1.0
Installation
Transportation and Storage Conditions
The Docking Station is transported in a palletized container.
The Docking Station weight, including accessories, pallet and packaging
materials, is approximately 35 kg (77 lb).
Docking Station Dimensions and Weight
Unpacking and Installation
The original shipping container must be transferred unopened to the
installation site. The Docking Station should be unpacked and installed by
trained Roche Service personnel. No attempt should be made by the user
to remove, set up, or operate the instrument without Roche Service
personnel being present.
The packing list that accompanies the Docking Station contains the
description, part number, and quantity of all component parts,
accessories, and consumables delivered with the instrument. This packing
list should be verified against the delivered items and retained for future
reference.
Electrical Power Requirements
The Docking Station is designed in accordance with EN/IEC 61010-1,
EN/IEC 61010-2-101, UL 61010A-1, and CAN/CSA C22.2 No. 1010.1.
Environmental Requirements
The Docking Station is designed for indoor use at an altitude below 2,000
meters (6,500 feet). Room temperature should be between 15° C to 32° C
(59° F to 89° F), and relative humidity should not exceed 80% at 32° C
(89°
F) without condensation. Atmospheric conditions should conform to
Pollution Degree 2 (according to EN/IEC 61010-1).
Width 35 cm (13.75 in)
Depth 43 cm (17 in)
Height 76.5 cm (30 in)
Weight 25 kg (55 lb)
Docking Station
power consumption
typically 30 VA
maximum 90 VA