  • Hello! I am an AI chatbot trained to assist you with the Roche cobas s 201 system User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
  • What is the Docking Station used for in the Roche cobas s 201 system?
    How many K-tubes can a K-carrier hold?
    What is the purpose of a K-tube?
    What is a multicarrier?
0
0
cobas s 201 System 0
Operator’s Manual Supplement
for Use of the Docking Station
0
P/N: 05216486190-01
0.2 05/2008, version 1.0
This page intentionally left blank.
Preface
05/2008, version 1.0 0.3
Table of Contents
Preface
Table of Contents .......................................................................... 0.3
Revision History ............................................................................0.4
How to Use this Supplement ........................................................ 0.6
Conventions Used in This Supplement .......................................0.7
Glossary ........................................................................................ 0.10
Operational Precautions and Limitations .................................. 0.12
Warnings ...................................................................................... 0.13
Intended Use ................................................................................0.13
1. Workflow
Overview ........................................................................................ 1.1
Perform Startup Procedures ......................................................... 1.2
Load Reagents ................................................................................1.4
Load Consumables ........................................................................1.6
Create Orders ................................................................................. 1.8
Load SK24 Racks ...........................................................................1.9
Start the Run ................................................................................1.10
Remove Used Consumables ........................................................ 1.12
Review Run Messages .................................................................. 1.12
Accept Results .............................................................................. 1.12
2. Docking Station
Overview ........................................................................................ 2.1
Installation .....................................................................................2.2
Methodology .................................................................................. 2.3
Loading Empty K-carriers ............................................................. 2.5
Docking Station Maintenance ...................................................... 2.6
3. Index
0.4 05/2008, version 1.0
Revision History
The Docking Station is manufactured by:
Manual Revision Revision Date
1.0 May 2008
Manufacturer Roche Instrument Center AG
CH-6343 Rotkreuz
Switzerland
Distributed by
Roche Diagnostics
Indianapolis, IN 46256 USA
(For Technical Assistance call the
Roche Response Center
toll-free 1-800 526 1247)
Roche Diagnostics
H7V 4A2 Laval, Quebec
(For Technical Assistance call:
Pour toute assistance technique,
appeler le: 1-877 273 3433)
Roche Diagnostics (Schweiz) AG
CH-6343 Rotkreuz
Roche Diagnostics
F-38240 Meylan
Roche Diagnostics GmbH
D-68298 Mannheim, Germany
Distributore in Italia:
Roche Diagnostics SpA
Piazza Durante 11
I-20131 Milano
Roche Diagnostics S.L.
E-08006 Barcelona
Distribuidor em Portugal:
Roche Farmacêutica Química, Lda
P-2700 Amadora
l
Declaration of Conformity
The cobas s 201 system Docking Station complies with the requirements of
the Directive 98/79/EC of the European Parliament and of the Council of
27 October 1998 on In-vitro Diagnostic Medical Devices.
Preface
05/2008, version 1.0 0.5
Questions or comments regarding the contents of this manual can be
directed to the address below or to your Roche representative.
Roche Molecular Systems, Inc.
4300 Hacienda Drive
Pleasanton, California 94588-2722
ROCHE, AMPERASE, AMPLILINK, COBAS, COBAS S, AMPLIPREP and
TAQMAN are trademarks of Roche.
ROCHE RESPONSE CENTER is a servicemark of Roche.
Microlab is a registered trademark of Hamilton Company.
Microsoft, Windows, and Windows XP are either registered trademarks or
trademarks of Microsoft Corporation in the United States and/or other
countries.
AMPLILINK Software:
The Source Code version of the executable code (the "Software") is
available under the terms of the Interbase Public License Version 1.0. The
user may obtain a copy of the license at:
http://interbase.com/IPL.html.
The Software is offered solely under the terms of the Interbase Public
License Version 1.0.
Copyright © 2008, Roche Molecular Systems, Inc. All rights reserved.
0.6 05/2008, version 1.0
How to Use this Supplement
This supplement is to be used in conjunction with the cobas s 201 system
Operator’s Manual.
Before using the Docking Station, it is essential that the Warnings,
Cautions, and Safety Requirements that are contained in that manual and
in this supplement are read and understood by the user.
Refer to the section titled Symbols on the following page for a description
of the warning symbols and caution symbols that appear in this
supplement.
This supplement contains the following sections:
Preface (Preface) summarizes safety precautions necessary when
operating the system.
Section 1 (Workflow) outlines the procedures that are different
during routine operation of the cobas
s 201 system with
the Docking Station.
Section 2 (Docking Station) describes the major components of the
Docking Station.
Index
Preface
05/2008, version 1.0 0.7
Conventions Used in This Supplement
Text Conventions To convey information readily and consistently, certain text conventions
are used throughout this supplement. These text conventions are as
follows:
Text convention Used for
Numbered lists Numbered procedures that should be
implemented in order during an operation:
1 Turn power switch on.
2 Click on File.
Bold typeface, lower case Emphasis of any term or component that is
being described:
A profile is a user-defined series of tests to be
performed.
Italics Designating another section of this
Operator’s Manual that should be consulted:
Details concerning order entry are outlined
in the Orders section of this Manual.
0.8 05/2008, version 1.0
Symbols Certain symbols are used throughout this supplement to provide a ready
visual reference. These symbols are as follows:
Symbol Used for
Information note. Designates a note that
provides additional information concerning
the current topic or procedure.
Important note. Designates an important
note that must be reviewed and understood.
Danger. Indicates a direct danger that, if not
avoided, may result in death or serious
injury.
War ning. Indicates a possible hazardous
situation that, if not avoided, may result in
death or serious injury.
Caution. Indicates a possible hazardous
situation which, if not avoided, may result in
personal injury and/or in damage to the
system.
Warning Potential Biohazardous Material.
Indicates a possible hazardous situation due
to biohazardous material that may result in
death or serious injury.
Warning Laser Transmitter. Designates the
presence of a laser transmitter. Do not stare
into the laser transmitter.
Warning Moving Parts. Indicates a possible
hazardous situation. Keep hands away from
moving parts.
Preface
05/2008, version 1.0 0.9
The following symbols appear on the instrument type plate.
Electrical and electronic equipment marked
with this symbol are covered by the European
directive WEEE. This symbol denotes that
the equipment must not be disposed of in the
municipal waste system.
(WEEE) Directive 2002/96/EC of the
European Parliament and the Council of 27
January 2003 on waste electrical and
electronic equipment
Danger Area. The triangular warning symbol
on the back of the analyzer is a general
reminder that users read the safety
information contained in this manual. Users
must be able to identify specific hazards and
take appropriate action to avoid them.
Symbol Used for
Caution: Refer to accompanying documents.
Manufacturer of device.
0.10 05/2008, version 1.0
Glossary
The terminology that is used in this supplement is generally the same as
that used in clinical laboratories. Certain specific terms, however, are used
for description of an operation or a component.
Refer to the cobas s 201 system Operators Manual for a summary of
terms that are used with this system and their definitions.
A summary of the terms used in this Supplement and their definitions is
provided below:
K-carrier Reusable metallic holder that accommodates
up to 24 K-tubes that contain PCR-ready
samples for analysis on the COBAS
®
TaqMa n
®
Analyzer.
K-tube Kinetic tube. The disposable sample tube that
is used to contain a prepared sample. K-tubes
come prepackaged in a covered K-tube rack.
Multicarrier Holds a K-carrier for transport from the
COBAS
®
AmpliPrep Instrument to the
COBAS
®
TaqMa n
®
Analyzer by the Docking
Station and vice versa.
SK24 rack Reusable rack that holds S-tubes or K-tubes
in barcode clips. The SK24 rack loads and
unloads samples throughout the pipetting,
extraction, and amplification/detection
process.
Thermal Cycler Programmable temperature block in the
COBAS
®
TaqM an
®
Analyzer. The thermal
cycler rapidly changes temperature according
to the amplification profile defined in the test
definition files.
Preface
05/2008, version 1.0 0.11
Units
The following measurement abbreviations are used:
°C Degrees Centigrade
°F Degrees Fahrenheit
CAN Canada.
CE Conformité Européenne
cm Centimeters
CSA Canadian Standards Association
EN European Normative
FCC Federal Communications Commission
IEC International Electrotechnical Commission
in Inches
kg Kilogram
lb Pounds (weight)
LED Light Emitting Diode
μL microliters(10
-6
liters)
UL Underwriters Laboratories (USA)
VA Volt-Ampères (power consumption)
VA C Volts (alternating current)
0.12 05/2008, version 1.0
Operational Precautions and Limitations
Safety Requirements
The safety precautions that are necessary when installing, operating and
servicing the Docking Station are summarized in the following sections
of this manual. It is important that you carefully read and understand
the statements regarding safety contained in these sections. This
information should also be made available for new employees and for
future reference.
Disposal Recommendations
All electrical and electronic products should be disposed of separately
from the municipal waste system. Proper disposal of your old appliance
prevents potential negative consequences for the environment and
human health.
Disposal of the Instrument
The instrument must be treated as biologically contaminated-
hazardous waste. Decontamination (i.e., a combination of processes,
including cleaning, disinfection and/or sterilization) is required before
reuse, recycling or disposal.
Dispose the instrument according to local and/or labor regulations. For
more information, contact your local Roche Support personnel.
Disposal of Control Unit Components
Components of your control unit such as the computer, monitor,
keyboard, etc. which are marked with the crossed-out wheeled bin
symbol are covered by the European Directive 2002/96/EC (WEEE).
These items must be disposed of via designated collection facilities
appointed by government or local authorities.
For more information about disposal of your old product, please
contact your city office, waste disposal service or your local Roche
Support personnel.
Constraint
It is left to the responsible laboratory organization to determine
whether control unit components are contaminated or not. If
contaminated, treat in the same way as the instrument.
Preface
05/2008, version 1.0 0.13
Warnings
Intended Use
The Docking Station connects the COBAS
®
AmpliPrep Instrument to the
COBAS
®
TaqMa n
®
Analyzer and fully automates sample preparation,
amplification, and detection.
This supplement contains instructions for using the Docking Station with
the cobas s 201 system. This supplement should be used in conjunction
with the cobas s 201 system Operator’s Manual.
WARNING
For service, contact your Roche Service representative, who is trained
and qualified to perform the necessary servicing.
WARNING
Never open the Docking Station while a run is in progress. Failure to
follow this procedure can result in potential injury to the operator.
WARNING
After the Docking Station is installed by a Roche Service representative,
the COBAS
®
AmpliPrep Instrument and the COBAS
®
TaqMa n
®
Analyzer must not be moved. Movement of the docked system can
cause the Docking Station to become non-operational.
Electrical safety
The Docking Station is designed in accordance with safety standard EN/
IEC 61010-1. Grounding of the instrument and those surfaces the user
can come into contact with is provided by a grounded cable in
accordance with Protection Class I (IEC).
Users should not perform any servicing.
0.14 05/2008, version 1.0