  • Hello! I am an AI chatbot trained to assist you with the Midmark Procenter Systems Installation guide. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
1© Midmark Corporation 2003 SF-1667
DA137900i
003-1139-00 Rev. LA1 (9/17)
Assistant’s Unit Installation
[Console Mounted ]
Step 1: Move stop pin
into RH position.
Parts Bag Contents
T-Fitting w/tubing......... 1
Screw ......................... 3
Set Screw ................... 3
Sleeve Cap (red)......... 1
Toggle Cap (black) ...... 1
Water Q.D. Coupling ... 1
Special Tools
Level
Note
Step 1 is not necessary if installing this
instrument arm onto a LH console
(mounted on patient’s LH side).
Go to Step 2.
2
© Midmark Corporation 2003 SF-1667
Step 2: Rotate arm
adapter clockwise.
Step 3: Position arm as shown.
Install three mounting screws.
Do not tighten screws.
Step 4: Install three leveling screws.
Level arm.
Tighten three mounting screws.
Caution
Do not kink 5/8" tubing,
causing a restriction.
Note
The front console can be moved out of the
way (2 screws) for ease of installation. It is
not necessary to disconnect the tubing.
Note
If the 2" umbilical tubing prevents
routing 5/8" tubing at location of
umbilical, use alternate route (shown).
Step 5: Route tubing into console.
Alternate
Route for
5/8" Tubing
3© Midmark Corporation 2003 SF-1667
DA138100i
From
Assistant’s
Unit
Step 6: Remove plugged section of 1/8” blue tubing.
Connect 1/8” blue tubing from asst’s unit.
Step 7: Cut 1/8” yellow tubing.
Insert T-fitting w/tubing.
Connect to 1/8” yellow tubing from asst’s unit.
Step 8: Connect vacuum tubing from asst’s unit
to 5/8” ribbed tubing from umbilical.
In J-Box (not shown):
Connect 5/8”
ribbed tubing from umbilical to central
vacuum unit.
4
© Midmark Corporation 2003 SF-1667
Chair Position Touch Pad
Step 9: Connect wire harness
to J2 or J3 on PC Board.
Step 10: Connect harness to J1 and connect two wires
from wire harness to J6 terminals on PC Board
as shown.
Midmark Corporation
60 Vista Drive
P.O. Box 286
Versailles, OH 45380-0286
937-526-3662
Fax 937-526-5542
www.midmark.com
Assistant’s