39
10.DefinetheprocessvariablesPV,SV,TVandQV.
Menupath:DeviceSetup>DetailedSetup>Measurements>ChannelSetup
11.Calibratesensor.Menupath:DeviceSetup>SensorCalibration
12.Performfurthersettings.SeeOperatingInstructionsoftheM100transmitter.
13.Storeconfigurationtodevice.
HART handheld terminal
h
Note:TheDD“008E8E7D0101.hhd”isonthesuppliedCD-ROM.Youcanalsodownloadthe
DDviaInternet“www.mt.com/M100“.
For“Step1”refertothedocumentationoftheHARThandheldterminal.
1. CheckiftheDDoftheM100transmitterhasalreadybeeninstalledontheHARThandheld
terminal.InstalltheDDifnecessary.
2. Thecommunicationisbuiltupautomatically.
3. SelectSensor Type.
Menupath:DetailedSetup>Measurements>ChannelSetup>SensorSetup
4. Loadconfigurationfromdevice.Menupath:DeviceSetup>DetailedSetup
5. SetTagor/andLong Tag.
Menupath:DeviceSetup>DetailedSetup>HARTInfo
6. SetDateandTime.Setthetimeinthe24hourformat.Thetimeformatcannotbe
changed.Menupath:DeviceSetup>SetDate/Time
7. Setrangeoftheanalogoutputsignal.
Menupath:Detailedsetup>OutputCondition>Analogoutput>Range
– URV(UpperRangeValue)andLRV(LowerRangeValue):Thevaluescanbechanged
tothecurrentmeasurementrange.Thevalueshavetobewithinthemeasurementlimits
ofthesensor.
– USL(UpperSensorLimit)andLSL(LowerSensorLimit):Thelimitsaredefinedbythe
sensorandcannotbechanged.
8. DefinetheprocessvariablesPV,SV,TVandQV.
Menupath:DeviceSetup>DetailedSetup>Measurements>ChannelSetup
en