  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo Transmitter M100 Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Quick Setup Guide Transmitter M100 Head mount
Multilingual Quick Setup Guide
cs Průvodcerychlýmnastavením 3
da Lynvejledning 13
de Quick-Setup-Leitfaden 23
en QuickSetupGuide 33
es Guíadeconguraciónrápida 43
 Pika-asetusopas 53
fr Guidedeparamétragerapide 63
hu Gyorsbeállításiútmutató 73
it Guidaallacongurazionerapida 83
ja
93
ko
103
nl Beknoptehandleiding 113
pl Podręcznikszybkiejkonguracji 123
pt GuiadeConguraçãoRápida 133
ru Руководство по быстрой 143
sv Snabbinstallationsguide 153
th
 163
zh
快速操作指南 173
©Mettler-ToledoGmbH08/2016.
Subjecttotechnicalchanges.PrintedinSwitzerland.30130606
7
5 Struktura menu
Device Setup
Sensor
Calibration *
Diagnostics &
Service
Detailed Setup
• Load
Configuration **
• Measurement
• Output Condition
• HART Info
• ISM Setup
• System
• Alarm Setup
• Device Info
• Test device
• Loop test
• D/A Trim
• HW Diagnostics
Review
• Tag
• Long Tag
• Sensor type
• Dynamic variable
• Range
• pH Process
• ORP Process
• Verify
pH/ORP
pH/pNa
Oxygen Conductivity
• O2 Slope
• O2 Offset
• Ain calibration
• Verify
• Conductivity
Process
• Resistivity
Process
• Verify
Process variables
Process variable Measurement chart Analog output chart
* Sensor Calibration menu dependent on the connected sensor
** Load Configuration only for HART handheld terminal
Set Date/Time
cs
17
5 Menustruktur
Device Setup
Sensor
Calibration *
Diagnostics &
Service
Detailed Setup
• Load
Configuration **
• Measurement
• Output Condition
• HART Info
• ISM Setup
• System
• Alarm Setup
• Device Info
• Test device
• Loop test
• D/A Trim
• HW Diagnostics
Review
• Tag
• Long Tag
• Sensor type
• Dynamic variable
• Range
• pH Process
• ORP Process
• Verify
pH/ORP
pH/pNa
Oxygen Conductivity
• O2 Slope
• O2 Offset
• Ain calibration
• Verify
• Conductivity
Process
• Resistivity
Process
• Verify
Process variables
Process variable Measurement chart Analog output chart
* Sensor Calibration menu dependent on the connected sensor
** Load Configuration only for HART handheld terminal
Set Date/Time
da
27
5 Menüstruktur
Device Setup
Sensor
Calibration *
Diagnostics &
Service
Detailed Setup
• Load
Configuration **
• Measurement
• Output Condition
• HART Info
• ISM Setup
• System
• Alarm Setup
• Device Info
• Test device
• Loop test
• D/A Trim
• HW Diagnostics
Review
• Tag
• Long Tag
• Sensor type
• Dynamic variable
• Range
• pH Process
• ORP Process
• Verify
pH/ORP
pH/pNa
Oxygen Conductivity
• O2 Slope
• O2 Offset
• Ain calibration
• Verify
• Conductivity
Process
• Resistivity
Process
• Verify
Process variables
Process variable Measurement chart Analog output chart
* Sensor Calibration menu dependent on the connected sensor
** Load Configuration only for HART handheld terminal
Set Date/Time
de
33
Quick Setup Guide
Transmitter M100 Head mount
Content
1 SafetyandIntendedUse 34
2 TechnicalDataandFurtherInformation 35
3 Mounting 35
4 ElectricalConnection 36
5 MenuStructure 37
6 GeneralSetup 38
7 SensorCalibration 41
8 Maintenance 42
9 Disposal 42
10 EC-DeclarationofConformity 42
en QuickSetupGuide
en
34
1 Safety and Intended Use
h
Note:TheQuickSetupGuideisabriefoperatinginstruction.
TheM100transmittermustonlybeinstalled,connected,commissioned,andmaintainedby
qualifiedspecialistse.g.electricaltechniciansinfullcompliancewiththeinstructionsinthis
QuickSetupGuide,theapplicablenormsandlegalregulations.
ThespecialistmusthavereadandunderstoodthisQuickSetupGuideandmustfollowthe
instructionsitcontains.IfyouareunclearonanythinginthisQuickSetupGuide,youmust
readtheOperatingInstructions(suppliedonCD-ROM).TheOperatingInstructionsprovides
detailedinformationontheinstrument.
TheM100transmittershouldbeoperatedonlybypersonnelfamiliarwiththetransmitterand
whoarequalifiedforsuchwork.
Intended Use
TheM100isa2-wireheadmounttransmitterforanalyticalmeasurementswithHARTcom
-
municationcapabilities.TheM100isa1-channel,multi-parameterunitforpH/ORP,conduc-
tivityanddissolvedoxygenmeasurement.ItisonlycompatiblewithISM
®
sensors.
TheM100transmitterisdesignedforuseintheprocessindustriesandisintrinsicallysafe
certifiedforinstallationsinhazardousareas.Forapplicationsinhazardousareasrefertothe
OperatingInstructions.
M100 parameter fit guide
Parameter M100 XH
ISM
pH/ORP
pH/pNa
Conductivity4-e
Amp.DissolvedOxygenppm/ppb/trace
1)
/ •
2)
/•
2)
1) INGOLDandTHORNTONsensors
2) THORNTONsensors
en
35
2 Technical Data and Further Information
Themostimportanttechnicaldatasuchassupplyvoltagearegivenonthenameplateonthe
transmitter.ForfurthertechnicaldatasuchasaccuracyrefertotheOperatingInstructions.This
document,theOperatingInstructionsandsoftwareareonthesuppliedCD-ROM.Youcanalso
downloadthedocumentationviaInternet“www.mt.com/M100“.
3 Mounting
1
2
3
4
h
Note:Poweroffdeviceduringinstallation.
1. InstallsensorhousingasdescribedintheOperat
-
ingInstructionsofthesensorhousing.
2. Carefullyinsertsensorintothesensorhousing.
Screwinsensorhand-tight.
3. Screwsensorhousingandextensiontubehand-
tight.DonotturntheM100transmitter.Turnex
-
tensiontubeonly.
4. UnscrewcoveroftheM100transmitter.
5. Performwiring.See“ElectricalConnection”
chapteronPage36.
6. ScrewcoveroftheM100transmitter.
1 M100transmitter
2 Extensiontube
3 Sensororelectrode
4 Sensorhousing
en
36
4 Electrical Connection
h
Note:Poweroffdeviceduringelectricalconnection.
1. Switchoffsupplyvoltage.
2. Connectmainssupply(14to30VDC)totheterminalsAout+, HART+andAout–.
HART–.Noticepolarity.
3. ConnectanalogoutputandHARTsignaltoterminalsHART+andHART–.
4. ConnectdigitalinputforswitchingtransmitterinHoldstatetoterminalsDI+andDI–.
5. Connectanaloginputsignal4to20mAforpressurecompensationtoterminalsAin+
andAin–.
14 30 V DC
Aout–, HART–
DI+
DI–
ISM-Data
Ain–
Ain+
Aout+, HART+
ISM-GND
Port
Terminal Definition
Port Interfaceforservicee.g.firmwareupdate.
Aout+,HART+
Aout–,HART–
Notice polarity.
Powerconnection14to30VDC,
AnalogoutputandHARTsignal.
DI+andDI– Digitalinput(forswitchingtransmitterin
Holdstate).
ISM-DATA
ISM-GND
Sensorinput,see“SafetyandIntended
Use”chapteronPage34“.Wiredby
thefactory.Donotdisconnect.
Ain+andAin– Analoginput4to20mA(forpressure
compensation).
en
37
5 Menu Structure
Device Setup
Sensor
Calibration *
Diagnostics &
Service
Detailed Setup
• Load
Configuration **
• Measurement
• Output Condition
• HART Info
• ISM Setup
• System
• Alarm Setup
• Device Info
• Test device
• Loop test
• D/A Trim
• HW Diagnostics
Review
• Tag
• Long Tag
• Sensor type
• Dynamic variable
• Range
• pH Process
• ORP Process
• Verify
pH/ORP
pH/pNa
Oxygen Conductivity
• O2 Slope
• O2 Offset
• Ain calibration
• Verify
• Conductivity
Process
• Resistivity
Process
• Verify
Process variables
Process variable Measurement chart Analog output chart
* Sensor Calibration menu dependent on the connected sensor
** Load Configuration only for HART handheld terminal
Set Date/Time
en
38
6 General Setup
YouconfiguretheM100transmittereitherviaaconfigurationtool,anasset-management-tool
orviaaHARThandheldterminal.
Prerequisite:TheM100transmitterandthesensoraremountedandelectricallyconnected.
Configuration tool or asset-management tool
h
Note:TheDTMandtheconfigurationtoolPACTWare™areonthesuppliedCD-ROM.Youcan
alsodownloadtheDTMviaInternet“www.mt.com/M100“.
Forsteps1to5andstep13refertothedocumentationofthetool.
1. Installconfigurationtoole.g.PACTWare™orasset-management-tool.
2. InstallDTMforHARTinterfaceandDTMforM100transmitter.
3. Updatedevicecatalog.
4. Buildupconnection.CheckCOMportsettingsifnecessary.
5. SelectSensor Type.
Menupath:DetailedSetup>Measurements>ChannelSetup>SensorSetup
6. Loadconfigurationfromdevice.
7. SetTagor/andLong Tag.
Menupath:DeviceSetup>DetailedSetup>HARTInfo
8. SetDateandTime.Setthetimeinthe24hourformat.Thetimeformatcannotbe
changed.Menupath:DeviceSetup>SetDate/Time
9. Setrangeoftheanalogoutputsignal.
Menupath:Detailedsetup>OutputCondition>Analogoutput>Range
URV(UpperRangeValue)andLRV(LowerRangeValue):Thevaluescanbechanged
tothecurrentmeasurementrange.Thevalueshavetobewithinthemeasurementlimits
ofthesensor.
USL(UpperSensorLimit)andLSL(LowerSensorLimit):Thelimitsaredefinedbythe
sensorandcannotbechanged.
en
39
10.DefinetheprocessvariablesPV,SV,TVandQV.
Menupath:DeviceSetup>DetailedSetup>Measurements>ChannelSetup
11.Calibratesensor.Menupath:DeviceSetup>SensorCalibration
12.Performfurthersettings.SeeOperatingInstructionsoftheM100transmitter.
13.Storeconfigurationtodevice.
HART handheld terminal
h
Note:TheDD“008E8E7D0101.hhd”isonthesuppliedCD-ROM.Youcanalsodownloadthe
DDviaInternet“www.mt.com/M100“.
For“Step1”refertothedocumentationoftheHARThandheldterminal.
1. CheckiftheDDoftheM100transmitterhasalreadybeeninstalledontheHARThandheld
terminal.InstalltheDDifnecessary.
2. Thecommunicationisbuiltupautomatically.
3. SelectSensor Type.
Menupath:DetailedSetup>Measurements>ChannelSetup>SensorSetup
4. Loadconfigurationfromdevice.Menupath:DeviceSetup>DetailedSetup
5. SetTagor/andLong Tag.
Menupath:DeviceSetup>DetailedSetup>HARTInfo
6. SetDateandTime.Setthetimeinthe24hourformat.Thetimeformatcannotbe
changed.Menupath:DeviceSetup>SetDate/Time
7. Setrangeoftheanalogoutputsignal.
Menupath:Detailedsetup>OutputCondition>Analogoutput>Range
URV(UpperRangeValue)andLRV(LowerRangeValue):Thevaluescanbechanged
tothecurrentmeasurementrange.Thevalueshavetobewithinthemeasurementlimits
ofthesensor.
USL(UpperSensorLimit)andLSL(LowerSensorLimit):Thelimitsaredefinedbythe
sensorandcannotbechanged.
8. DefinetheprocessvariablesPV,SV,TVandQV.
Menupath:DeviceSetup>DetailedSetup>Measurements>ChannelSetup
en
40
9. Calibratesensor.Menupath:DeviceSetup>SensorCalibration
10.Performfurthersettings.SeeOperatingInstructionsoftheM100transmitter.
Change Sensor Type
Prerequisite:TheM100transmitterandanothersensortypearemountedandelectricallycon
-
nected.
Ifyouchangethesensortypee.g.aConductivitysensorwithapHsensor,performthefollow-
ingprocedure:
1. Builtupconnection.
2. SelectSensor Setup.
Menupath:DetailedSetup>Measurements>ChannelSetup>SensorSetup
3. Start“SensorSetup”.
4. SelectthenewsensortypefortheSensor Typeparameter.
5. Loadconfigurationfromdevice.
6. TheconfigurationintheconfigurationtoolorintheHARThandheldisupdated.Ifyou
haveselectedthecorrectsensortype,the“Verify”menuwillbeshown.
en
41
7 Sensor Calibration
(PATH:Devicesetup/SensorCalibration)
h
Note:Forbestprocesscalibrationresultsobservethefollowingpoints.Takegrabsampleas
closeaspossibletothemeasurementpointofthesensor.Measurethesampleatprocess
temperature.
Viaconfigurationtool,asset-management-toolorHARThandheldterminalyoucancalibrate
thesensorwiththe“Process”,“1-Point”or“2-Point”calibrationmethod.Forthecalibration
methods see Operating Instructionsof the M100transmitter. Via iSense software you can
calibratethesensorwiththe“1-Point”or“2-Point”calibratemethod.SeeOperatingInstruc
-
tionsoftheiSensesoftware.
Assoonasthecalibrationisinprogressnoothercalibrationcanbestarted.
Sensor Calibration Menu
Forthecalibrationmethodssee“MenuStructure”chapteronPage37.
Aftereverysuccessfulcalibrationthefollowingoptionsareavailable:
Adjust
Calibrationvaluesareadoptedandusedforthemeasurement.Additionally,thecalibration
valuesarestoredinthecalibrationhistory.
Calibrate
Calibrationvaluesarestoredinthecalibrationhistoryfordocumentation,butnotbeused
forthemeasurement.Thecalibrationvaluesfromthelastvalidadjustmentarefurtherused
forthemeasurement.
Abort
Calibrationvaluesarediscarded.
en
42
8 Maintenance
Thetransmitterrequiresnomaintenance.
Cleanthesurfaceswithasoftdampclothanddrythesurfaceswithaclothcarefully.
9 Disposal
Observetheapplicablelocalornationalregulationsconcerningthedisposalof“Wasteelectri-
calandelectronicequipment”.
Disassemblethetransmitteraccordingtoresources.Sortresourcesandsupplythemtorecy-
cling.Non-recyclablematerialshavetobedisposedofinanenvironmental-friendlymanner.
10 EC-Declaration of Conformity
TheEC-Declarationofconformityispartofthedelivery.
ISMisaregisteredtrademarkoftheMettler-ToledoGroupinSwitzerland,Brazil,USA,China,
EuropeanUnion,SouthKorea,RussiaandSingapore.
en
47
5 Estructura de menús
Device Setup
Sensor
Calibration *
Diagnostics &
Service
Detailed Setup
• Load
Configuration **
• Measurement
• Output Condition
• HART Info
• ISM Setup
• System
• Alarm Setup
• Device Info
• Test device
• Loop test
• D/A Trim
• HW Diagnostics
Review
• Tag
• Long Tag
• Sensor type
• Dynamic variable
• Range
• pH Process
• ORP Process
• Verify
pH/ORP
pH/pNa
Oxygen Conductivity
• O2 Slope
• O2 Offset
• Ain calibration
• Verify
• Conductivity
Process
• Resistivity
Process
• Verify
Process variables
Process variable Measurement chart Analog output chart
* Sensor Calibration menu dependent on the connected sensor
** Load Configuration only for HART handheld terminal
Set Date/Time
es
57
5 Valikkorakenne
Device Setup
Sensor
Calibration *
Diagnostics &
Service
Detailed Setup
• Load
Configuration **
• Measurement
• Output Condition
• HART Info
• ISM Setup
• System
• Alarm Setup
• Device Info
• Test device
• Loop test
• D/A Trim
• HW Diagnostics
Review
• Tag
• Long Tag
• Sensor type
• Dynamic variable
• Range
• pH Process
• ORP Process
• Verify
pH/ORP
pH/pNa
Oxygen Conductivity
• O2 Slope
• O2 Offset
• Ain calibration
• Verify
• Conductivity
Process
• Resistivity
Process
• Verify
Process variables
Process variable Measurement chart Analog output chart
* Sensor Calibration menu dependent on the connected sensor
** Load Configuration only for HART handheld terminal
Set Date/Time
67
5 Structure du menu
Device Setup
Sensor
Calibration *
Diagnostics &
Service
Detailed Setup
• Load
Configuration **
• Measurement
• Output Condition
• HART Info
• ISM Setup
• System
• Alarm Setup
• Device Info
• Test device
• Loop test
• D/A Trim
• HW Diagnostics
Review
• Tag
• Long Tag
• Sensor type
• Dynamic variable
• Range
• pH Process
• ORP Process
• Verify
pH/ORP
pH/pNa
Oxygen Conductivity
• O2 Slope
• O2 Offset
• Ain calibration
• Verify
• Conductivity
Process
• Resistivity
Process
• Verify
Process variables
Process variable Measurement chart Analog output chart
* Sensor Calibration menu dependent on the connected sensor
** Load Configuration only for HART handheld terminal
Set Date/Time
fr
77
5 Menüstruktúra
Device Setup
Sensor
Calibration *
Diagnostics &
Service
Detailed Setup
• Load
Configuration **
• Measurement
• Output Condition
• HART Info
• ISM Setup
• System
• Alarm Setup
• Device Info
• Test device
• Loop test
• D/A Trim
• HW Diagnostics
Review
• Tag
• Long Tag
• Sensor type
• Dynamic variable
• Range
• pH Process
• ORP Process
• Verify
pH/ORP
pH/pNa
Oxygen Conductivity
• O2 Slope
• O2 Offset
• Ain calibration
• Verify
• Conductivity
Process
• Resistivity
Process
• Verify
Process variables
Process variable Measurement chart Analog output chart
* Sensor Calibration menu dependent on the connected sensor
** Load Configuration only for HART handheld terminal
Set Date/Time
hu
87
5 Struttura del menu
Device Setup
Sensor
Calibration *
Diagnostics &
Service
Detailed Setup
• Load
Configuration **
• Measurement
• Output Condition
• HART Info
• ISM Setup
• System
• Alarm Setup
• Device Info
• Test device
• Loop test
• D/A Trim
• HW Diagnostics
Review
• Tag
• Long Tag
• Sensor type
• Dynamic variable
• Range
• pH Process
• ORP Process
• Verify
pH/ORP
pH/pNa
Oxygen Conductivity
• O2 Slope
• O2 Offset
• Ain calibration
• Verify
• Conductivity
Process
• Resistivity
Process
• Verify
Process variables
Process variable Measurement chart Analog output chart
* Sensor Calibration menu dependent on the connected sensor
** Load Configuration only for HART handheld terminal
Set Date/Time