Kanomax 3718-A User manual

Type
User manual

This manual is also suitable for

Component List
Standard
ITEM
MODEL
QTY
Main Unit (3718-A3719-A)
Contained Analog Output
terminal connector
3718-A
3719-A
1
AC Adapter
1
Filter
1
Adapter 1
1
Adapter 2
1
Tygon Tube 1
1
User’s Manual(English)
User’s Manual(Japanese)
Settings Software (DVD
3718/19-40
1
Optional Extras
ITEM
MODEL
QTY
USB-RS485 Converter
with 3716A-21
3716A-20
1
Converter Connecting Cable
3716A-21
1
RS232C-RS485 Converter
with 3716A-21
3716A-22
1
Isokinetic Suction Probe
3718-01
1
Isokinetic Suction Probe
3905-07
1
Tygon Tube 1
1
Tygon Tube 2
1
Consumables
ITEM
MODEL
QTY
Zero Filter
(w/ joint and tube (70cm))
3910-04
1
For more information on optional extras, please contact your distributor or
your KANOMAX service center.
Important Safety Information
In this manual, warning types and classifications are defined as follows:
[Classification]
WARNING: To Prevent Serious Injury or Death
Indicates a potentially hazardous situation which, if not avoided,
may result in serious injury or death.
CAUTION: To Prevent Damage to the Product
Indicates a potentially hazardous situation which, if not avoided,
may result in damage to the product that may void the product
warranty.
[Description of Symbols]
Indicates a condition (including danger) that requires caution. The subject
of each caution is illustrated inside the triangle (e.g., the symbol shown to
the left is the high temperature caution).
Indicates a prohibition. Do not take the prohibited action shown inside or
near this symbol (e.g., The symbol shown to the left prohibits disassembly).
Indicates a mandatory action. A specific action is described near the
symbol.
WARNING
Do not disassemble, modify or repair the instrument.
…… Misuse of the instrument may result in electric shock, fire, or
damage to the instrument, etc.
Use this instrument properly by carefully following this User's
Manual.
…… Misuse of the instrument may result in electric shock, fire, or
damage to the instrument. etc.
Handle Properly.
Do not
modify/disassemble
ii
If any abnormal noises, unusual odors or smoke are observed, or
any liquid enters the instrument, turn the power off immediately,
remove the battery and disconnect the power cable.
…… It may result in electric shock, fire or damage to the instrument.
Contact your distributor or your KANOMAX service center for repair.
Prohibition
Do not install the instrument where it will be
exposed to rain and/or water drops.
…… It may result in electric shock, fire, serious injury
or death.
Caution
Unplug the power cord when the instrument is not in use.
……Failure to observe the above may result in electrical shock, fire
or damage to internal circuitry.
Handle Properly
When using a power cord or an AC adapter, make sure to use the one
provided with this instrument.
……Other commercially available cords may have different voltage
specifications and polarity, which could result in a short circuit, fire
or damage to the instrument.
Prohibition
Do not perform measurements in environments exceeding or falling
below the specified temperatures and RH levels of the instrument.
……The instrument may not function properly outside the specified
environment (10-35 , 2085RH, with no condensation).
Prohibition
Do not subject the instrument to strong
shock. Do not place heavy objects on the
instrument or sensor.
……Failure to observe the above may cause
damage to the instrument.
Water
drops
STATUS
PARTICLE
MODEL 3719-A
LASER PARTICLE SENSOR
STATUS
PARTICLE
MODEL 3719-A
LASER PARTICLE SENSOR
Table of Contents
1. Part Names and Functions………………………………………………………………………1
1.1 General Outline…………………………………………………………………………………1
1.2 Part Names and Functions……………………………………………………………………1
1.3 USB-RS485 Converter…………2
1.4 Connection with the Distributor for the CRMS………………………………………………4
2. Getting Started………………………………………………………………………………………5
2.1 Power Supply………………………………………………………………………………5
2.1.1 AC Adapter…………………………………………………………………………………5
2.1.2 DC Power Supply…………………………………………………………………………5
2.2 Parameter Settings……………………………………………………………………………6
2.3 Connecting the Vacuum Source……………………………………………………………8
2.4 Connecting the Analog Output Terminal………………………………………………8
2.5 Connecting RS485 Connector………………………………………………………………11
2.6 Connecting the Zero Filter…………………………………………………………………11
3. Measurements………………………………………………………………………………………12
4. Main Specifications…………………………………………………………………………………13
5. Troubleshooting……………………………………………………………14
6. Warranty and After-sales Service……………………………………………………………15
- 1 -
1. Part Names and Functions
1.1 General Outline
This instrument (Model 3718-A/3719-A is a light-scattering laser particle sensor
using a laser diode as the light source. The output of the measurement data is from
4-20mA, based on the current system.
This instrument is connectable to the Clean Room Monitoring System CRMS ,
allowing it to perform measurements in a clean room environment for a prolonged
time.
The status lamp allows this instrument to display LD errors and flow rate errors.
When a particle concentration higher than the configured one in the alarm setting, is
detected the device will output the alarm signal.
1.2 Part Names and Functions
(Unit: mm)
67
Approx.133
Approx.
32
STATUS
PARTICLE
MODEL 3719-A
LASER PARTICLE SENSOR
8
7
6
5
4
3
2
1
Power(9- 28V)
0.5μm
5.0μm
A-GND
P-GND
ALM.c h1
ALM.c h2
ALM.C OM
1
2
3 5
4
RS-485
DC-IN
Approx.
8 0
Approx.
1 6
Ap pro x.
95
Adapter 1
Adapter 2
- 2 -
Number
Name
Function
Inlet
This is where sample air is taken in.
It can also be used with the provided Tygon tubing.
Analog
Output
Terminal
This consists of a DC power input, an analog output terminal,
and an alarm output terminal. For detailed functions, refer to
2.4 Connecting the Analog Output Terminal.
RS485
Connector
This is a connector to communicate with a PC. To connect it,
a USB-RS485 Converter (Converter + Converter Connecting
Cable) or a distributor (distributor for the CRMSwill be
required.
The address of the main unit, sampling time, measurement
range and alarm settings can be configured by a PC. For more
details, refer to 2.5 Connecting RS485 Connector.
Outlet
Connect a vacuum source to the outlet.
For information about the performance of the vacuum source,
refer to 2.3 Connecting the Vacuum Source.
Power Supply
Inlet
This is an inlet to supply DC power to the main unit.
Connect the provided AC adapter to this inlet.
Refer to 2.1.1 AC Adapter.
PARTICLE
Lamp
This is the orange lamp. When a particle is detected,
the lamp blinks once. When no particles are detected, the
lamp is not lit.
STATUS
Lamp
When the conditions of the flow rate and laser diode are
normal, this lamp is green. When a flow error or a
laser diode error occurs, this lamp is red.
1.3 USB-RS485 Converter
To set the parameters of this instrument, use the provided setting software (MODEL
3718/19-40). To configure the settings, a USB-RS485 converter (including a
converter connecting cable) is required. A USB-RS485 Converter, MODEL 3716A-20
(with a Converter Connecting Cable MODEL, 3716A-21) is available as an optional
extra.
For more details, please contact your distributor or your KANOMAX service center.
When preparing a USB-RS485 Converter, please refer to the instruction on the next
page. Converter connecting cables (MODEL 3716A-21) are also sold as optional
extras.
- 3 -
Connection Method
The following Figures show how to connect the instrument, a USB-RS485 Converter
and a PC. For more details, refer to the information accompanied with USB-RS485
Converter MODEL, 3716A-20 (with a Converter Connecting Cable, MODEL 3716A-21).
USB-RS485 Converter ModelUSB-4W485-RJ45-T9P
Manufacturer: System Sacom Industry Corp.
Connection Method of the Converter Connecting Cable and USB-RS485 Converter
Number
Signal name
(this instrument)
Number
RS485
Converter
1
TXD+
4
RXD+
2
TXD-
5
RXD-
3
RXD+
1
TXD+
4
RXD-
2
TXD-
5
GND
3
GND
6
GND
6
GND
7
+24V input
7
NC
8
+24V input
8
NC
Case
Shield
9
NC
USB-4W485-RJ45-T9P
Converter Cable 3716A-21
Note: To supply power through the converter connecting cable, supply 9-28V
to +24V and GND.
When power is supplied through the provided AC adapter and analog output
terminal, do not connect to the converter. In this case, the terminal not in
use should be insulated.
USB-RS485 Converter
MODEL 3716A-20
USB-4W485-RJ45-T9P
(System Sacom Industry Corp.)
PC
Converter Connecting Cable
MODEL 3716A-21
USB Cable
(Accessory of MODEL 3716A-20
8
7
6
5
4
3
2
1
Power(9-28V)
0.5μm
5.0μm
A-GND
P-GND
ALM.ch1
ALM.ch2
ALM.COM
RS-485
DC-IN
- 4 -
1.4 Connection with the Distributor for the CRMS
Connecting this instrument with the distributor for the CRMS allows for operation with
the Clean Room Monitoring System. For more information on the distributor for the
CRMS, please contact your distributor or your KANOMAX service center.
Connection Method
The following Figure shows how to connect this instrument with the distributor for
the CRMS:
Distributor for the CRMS
(KANOMAX Japan Inc.)
PC
RS232c Cable
Shielded Ethernet CAT5 Cable
(Straight Type
8
7
6
5
4
3
2
1
Power(9-28V)
0.5μm
5.0μm
A-GND
P-GND
ALM.ch1
ALM.ch2
ALM.COM
RS-485
DC-IN
- 5 -
2. Getting Started
2.1 Power Supply
2.1.1 AC Adapter
Plug the AC adaptor in the inlet. The specification of the provided AC adapter is as
follows (When purchasing AC adapter for yourself, refer to 4. Main Specifications):
Input/Output SpecificationInput AC 100-240V (50-60Hz)
Output DC 12V 0.5A
Plug Specification
2.1.2 DC Power Supply
By using No.4 Pin Power (9-28V) and No.5 Pin P-GND of the analog terminal,
DC power can be supplied. (Note: Pay careful attention to the connecting order.)
0.3um
ALM.COM
ALM.ch2
ALM.ch1
P-GND
Power(9-28V)
A-GND
0.5um
1
2
3
4
5
6
7
8
3718-A Analog Terminal 3719-A Analog Terminal
Power also can be supplied through the RS485 connecting terminal. Connecting the
CRMS distributor and RS485 to supply power to the instrument.
For more details on the distributor for the CRMS, please contact your distributor
or your KANOMAX service center.
Inner Diameter 2.1mm
ALM.COM
ALM.ch2
ALM.ch1
P-GND
Power(9-28V)
A-GND
0.5um
1
2
3
4
5
6
7
8
RS-485
DC-IN
5.0um
- 6 -
2.2 Parameter Settings
Before starting a measurement, the parameter settings must be configured.
Prior to the parameter settings, please read the software operation manual.
Before setting the parameters, connect the 3718-A/3719-A, a power supply for
3718-A/3719-A, the USB-RS485 Converter (converter, connecting cable) and a PC
that has the software installed.
Sampling time settings: The settable range is 1-60 seconds, at a rate of 1.0 second.
All data compiled within the configured sampling time will be
transmitted via the analog output (4-20mA).
This sampling time is shared between 2 channels.
* The factory setting is 1 second.
Output range settings: Configure the range of the particle count and concentration.
The analog output will be 4 to 20mA within the configured
range. The following figure shows this setting range, which is
settable for each channel.
* The factory setting is 01 (0-10CNT).
Setting
Code
Output Range
Count /
Concentration
01
0-10CNT
Count
02
0-100CNT
03
0-1,000CNT
04
0-10,000CNT
11
0-10CNT/cf
Concentration
12
0-100CNT/cf
13
0-1,000CNT/cf
14
0-1,000,000CNT/cf
21
0-353CNT/m3
22
0-3,530CNT/m3
23
0-35,300CNT/m3
24
0-35,300,000CNT/m3
Alarm Settings: Set the measurement range (0-100%) in 1% increments.
An alarm setting can be configured per 2 channels. If the detected
particle counts exceeds the configured alarm level of concentration,
the alarm signal will be sent. For more details on the alarm settings,
refer to 2.4 Connecting the Analog Output Terminal.
*The factory setting is 0 (for both CH1 and CH2).
Address Settings: Set the address of this instrument. The settable range is 0-512.
The addresses are used to identify each instrument when using
multiple instruments.
*The factory setting is 0.
Sensor Communication Settings: Configure the communication setting of the
instrument.
Configure the communication settings between this
instrument and upper-layer PC/CRMS distributor.
- 7 -
Baud rate: Select from 4800, 9600, 19200, and 38400bps.
Data bitSelect 7 bits or 8 bits.
ParitySelect between Even, Odd or None.
*The communication settings with the distributor for the CRMS
is fixed: 9600bps, 7 bits and Even. The factory setting is the
same as the above.
Error Output Settings: Set the current output ON/OFF in the event that flow rate
errors or LD errors are detected.
ON: Current output varies according to error contents.
OFF: Measurement data will be output even if an error is
detected.
*Configure the appropriate error output settings based on the
system to connect.
- 8 -
2.3 Connecting the Vacuum Source
3718-A/3719-A measurement requires a connection with an external vacuum source.
The performance of the vacuum source should meet with the required specifications.
For more details, refer to 4. Main Specifications.
To confirm the condition of the vacuum source, check the STATUS lamp on the main
unit or the output current from No. 1/No. 2 of the analog output terminal.
If the output current from the No.1 or No.2 terminal is out of the range of 1-1.8mA or
3-3.8mA, the vacuum source is normal. On the contrary, if output current from No.1
or No.2 terminal is in the range of 1-1.8mA or 3-3.8mA, the vacuum source is bad.
*This output is active only when the error output setting is ON (2.2 Parameter Settings).
2.4 Connecting the Analog Output Terminal
The connector of the analog output terminal of this instrument has an automatic lock
function. When inserting and removing the lead, push the lock button (the
orange-colored part) on the connector to unlock it. To lock it simply turn the lock
button off. To connect the analog output terminal, remove the coating on the tip of
the lead and insert it.
The connector of the analog terminal is removable. However, we advise keeping it on
in order to avoid losing it.
Details of 3718-A analog output terminal Details of 3719-A analog output terminal
Lock
Button
Connecting
part
Lock
Button
Connecting
part
ALM.COM
ALM.ch2
ALM.ch1
P-GND
Power(9-28V)
A-GND
0.5um
1
2
3
4
5
6
7
8
RS-485
DC-IN
5.0um
0.3um
ALM.COM
ALM.ch2
ALM.ch1
P-GND
Power(9-28V)
A-GND
0.5um
1
2
3
4
5
6
7
8
- 9 -
The following Figures show the details of the Analog Output terminal.
Terminal
Number
Signal Name
(3718-A)
Functions
1
0.3μm
1Current output terminal of particle count or
concentration for 0.3μm channel measurement.
The output range is 4-20mA.
2This output terminal is also used when flow rate errors
or LD errors are detected.
Current output range of flow rate error is 1-1.8mA.
Current output range of LD error is 2-2.8mA.
Current output range when both flow rate and LD
errors occur at the same time is 3-3.8mA.
*This output is active only when the error output
setting is ON (2.2 Parameter Setting).
2
0.5μm
1Current output terminal of particle count or
concentration for 0.5μm channel measurement.
The output range is 4-20mA.
2This output terminal is also used when the flow rate
errors or LD errors are detected.
Current output range of flow rate error is 1-1.8mA.
Current output range of LD error is 2-2.8mA.
Current output range when both flow rate and LD
errors occur at the same time is 3-3.8mA.
*This output is active only when the error output
setting is ON (2.2 Parameter Setting).
3
A-GND
Common terminal (-) of current output
4
Power9-28V
Connecting terminal of DC power supply (+)
5
P-GND
Connecting terminal of DC power supply (GND)
6
ALM.ch1
Output terminal of the particle count for 0.3μm channel
measurement or concentration for alarm detection.
The specification of the connection point of this terminal is
60V 0.4A or less. Use it within this specification.
7
ALM.ch2
Output terminal of the particle count for 0.5μm channel
measurement or concentration for alarm detection.
The specification of the connection point of this terminal is
60V 0.4A or less. Use it within this specification.
8
ALM.COM
Common terminal (-) of the alarm output terminal
- 10 -
Terminal
Number
Signal Name
(3719-A)
Functions
1
0.5μm
1Current output terminal of particle count or
concentration for 0.5μm channel measurement.
The output range is 4-20mA.
2This output terminal is also used when the flow rate
errors or LD errors are detected.
Current output range of flow rate error is 1-1.8mA.
Current output range of LD error is 2-2.8mA.
Current output range when both flow rate and LD
errors occur at the same time is 3-3.8mA.
*This output is active only when the error output
setting is ON (2.2 Parameter Settings).
2
5.0μm
1Current Output terminal of particle count or
concentration for 5.0μm channel measurement.
The output range is 4-20mA.
2This output terminal is also used when the flow rate
errors or LD errors are detected.
Current output range of flow rate error is 1-1.8mA.
Current output range of LD error is 2-2.8mA.
Current output range when both flow rate and LD
errors occur at the same time is 3-3.8mA.
*This output is active only when the error output
setting is ON (2.2 Parameter Settings).
3
A-GND
Common terminal (-) of the current output
4
Power9-28V
Connecting terminal of DC power supply (+)
5
P-GND
Connecting terminal of DC power supply (GND)
6
ALM.CH1
Output terminal of particle count for 0.5μm channel
measurement or concentration for alarm detection.
The specification of the connection point of this terminal is
60V 0.4A or less. Use it within this specification.
7
ALM.CH2
Output terminal of particle count for 5.0μm channel
measurement or concentration for alarm detection.
The specification of the connection point of this terminal is
60V 0.4A or less. Use it within this specification.
8
ALM.COM
Common terminal (-) of the alarm output terminal
- 11 -
2.5 Connecting RS485 Connector
When setting the parameters of 3718-A/3719-A, connect an RS485 connector and PC
with the optional extra, USB-RS485 converter (converter + connecting cable).
Activate the provided settings software to set the parameters of this instrument.
Both 3718-A and 3719-A are connectable to the Clean Room Monitoring System
CRMSusing an RS485 connector.
2.6 Connecting the Zero Filter
To confirm the cleanness of the optical system, conduct a zero check before each
measurement.
Connect the tube side of the zero filter to the Adapter 1 of this instrument.
The filters must be handled with care.
If air leakage occurs between the filter and this instrument, zero checks cannot be
conducted properly.
Pin Number
RS485 Connector Signal Name
1
RS485 TXD +
2
RS485 TXD -
3
RS485 RXD +
4
RS485 RXD -
5, 6
DC power supply (-) terminal (P-GND)
7, 8
DC power supply (+) terminal (DC 9-28V)
*Not in use when the power is supplied
through the AC adapter or the analog output
terminal.
RS485 Connector
Zero Filter
1
2
3
4
5
6
7
8
STATUS
PARTICLE
MODEL 3719-A
LASER PARTICLE SENSOR
- 12 -
3. Measurements
Make sure that all procedures stated in 2.Getting Started in this manual have been
completed.
To perform measurements, take in sample air directly from the inlet of the
3718-A/3719-A or take in sample air with the provided Tygon tube connected to the
inlet.
Turn the power on 3718-A/3719-A to start a measurement using the configured
parameters.
If you have a problem with your instrument, such as the condition of the STATUS lamp,
the output current value of the analog output terminal and alarm output status, please
refer to 5.Troubleshooting in this manual first to find possible cause(s) and solution(s).
If that does not solve the problem, please contact your distributor or your KANOMAX
service center for repair.
- 13 -
4. Main Specifications
Product Name
Laser Particle Sensor
Model
3718-A
3719-A
Particle Size
Distribution
0.3μm, 0.5μm
0.5μm, 5.0μm
Rated Flow
28.3L/min (1CFM)
Optical Source
Laser Diode
Counting Efficiency
Compliant with ISO21501-4JIS B 9921
Max Detectable
Concentration
500,000CNT/ft3
Setting Parameter
Sampling Time1 to 60 seconds
Output Range Setting (To be configured for each particle size)
Count: 0-10CNT, 0-100CNT, 0-1,000CNT, 0-10,000CNT
Concentration: 0-10CNT/cf, 0-100CNT/cf, 0-1,000CNT/cf,
0-1,000,000CNT/cf
0-353CNT/ m3, 0-3,530CNT/m3,
0-35,300CNT/m3, 0-35,300,000CNT/m3
Alarm Setting: 0%-100
Address Setting: 0-512
Error Output Setting: ON/OFF
Sampling Pump
External vacuum>51kPa
Interface
RS485, Analog Output4-20mA
Communication
RS485
Setting Software
Japanese, English
Power Supply
DC Power Supply DC 9V28V
AC Adapter AC 100-240V (50-60Hz) , DC 12V
Operating
Environment
10302085%RHNon Condensing
Storing Environment
Temperature: -10 to 50Humidity: 95%RH or less
Non Condensing
Dimension
133(W)× 95(D)× 80(H) mm
Weight
1.0kg
Standard
Accessories
User's Manual (English)(1),User's Manual (Japanese)(1),
AC adapter (1), Zero filter (1), Adapter 1(1),Adapter 2(1),
Tygon tube 1 (1), Setting software DVD(3718/19-40) (1)
Optional Extras
USB-RS485 Converter 3716A-20 (Converter + Converter
Connecting Cable), Converter Connecting Cable 3716A-21,
RS232c-RS485 Converter 3716A-22 (Converter + Converter
Connecting Cable), Isokinetic Probe 3905-07(1),Isokinetic
Suction Probe3718-01(1),Tygon tube 1 (1),Tygon tube 2 (1)
- 14 -
5. Troubleshooting
Problems
Possible Cause(s) / Solution(s)
Page
No power.
Defective adapter
Replace the adapter.
5
The lead of the DC power supply is not
connected correctly.
Connect to the lead correctly.
5
Poor connection to the CRMS
Reconnect to the CRMS.
11
STATUS lamp is
not lit.
No connection to the power supply
Connect to the power supply.
5
The lamp is damaged.
Return the instrument for repair.
15
STATUS lamp lights
up in red.
The vacuum source is not operating or not
capable of operating.
Operate the vacuum source.
Replace the vacuum source with a better one.
8
The vacuum source has piping leakage or a
clogged pipe.
Inspect the vacuum source for leaks or
clogs. Clear any clogs of the pipe.
——
The internal piping is clogged.
Return the instrument for repair.
15
An internal part is damaged.
Return the instrument for repair.
15
PARTICLE lamp
is not lit when
detecting particles
Internal part failure
Return the instrument for repair.
15
Analog output
(4~20mA) has an
output error.
Wiring error of the analog output
Rewire the analog output correctly.
8
When the error output setting is ON, current
output will be 4mA or less according to the
error contents.
7
Internal circuit failure
Return the instrument or repair.
15
Output error of the
alarm terminal
Wiring error of the analog output
Rewire correctly.
8
Internal circuit failure
Return the instrument for repair.
15
Clean Room
Monitoring System
(CRMS) suffers from
a communication/
measurement data
error.
Wiring error
Confirm the wiring of RS485 cable.
11
Internal circuit failure
Return the instrument for repair.
15
PC software setting error
Contact your distributor or your KANOMAX
service center for repair.
15
- 15 -
6Warranty and After-sales Service
KANOMAX Limited Warranty
The limited warranty set below is given by KANOMAX with respect to the KANOMAX brand
Airborne Particle Counter, its attachment parts including Probe and other accessories
(hereafter referred to as “PRODUCT”) that you have purchased. PRODUCT you have purchased
shall be the only one that the limited warranty stated herein applies to.
Your PRODUCT, when delivered to you in new condition in its original container, is warranted
against defects in materials or workmanship as follows: for a period of three (3) years from the
date of original purchase, defective parts or a defective PRODUCT returned to your sales
representative, as applicable, and proven to be defective upon inspection, will be exchanged
for a new or comparable rebuilt parts, or a refurbished PRODUCT as determined by your sales
representative. Warranty for such replacements shall not extend the original warranty period
of the defective PRODUCT.
This limited warranty covers all defects encountered in normal use of the PRODUCT, and does
not apply to the following cases:
(1) Use of parts or supplies other than the PRODUCT sold by your sales representative, which
cause damage to the PRODUCT or cause abnormally frequent service calls or service
problems.
(2) If any PRODUCT has its serial number or date altered or removed.
(3) Loss of damage to the PRODUCT due to abuse, mishandling, improper packaging by the
owner, alteration, accident, electrical current fluctuations, failure to follow operating,
maintenance or environmental instructions prescribed in the PRODUCT's instruction
manual provided by KANOMAX, or service performed by a party other than KANOMAX.
NO IMPLIED WARRANTY, INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR
A PARTICULAR PURPOSE, APPLIES TO THE PRODUCT AFTER THE APPLICABLE PERIOD OF THE EXPRESS
LIMITED WARRANTY STATED ABOVE, AND NO OTHER EXPRESS WARRANTY OR GUARANTY, EXCEPT AS
MENTIONED ABOVE, GIVEN BY ANY PERSON OR ENTITY WITH RESPECT TO THE PRODUCT SHALL BIND
KANOMAX. KANOMAX SHALL NOT BE LIABLE FOR LOSS OF STORAGE CHARGES, LOSS OR CORRUPTION OF
DATA, OR ANY OTHER SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES CAUSED BY THE USE OR
MISUSE OF, OR INABILITY TO USE, THE PRODUCT, REGARDLESS OF THE LEGAL THEORY ON WHICH THE
CLAIM IS BASED, AND EVEN IF KANOMAX HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMAGES.
IN NO EVENT SHALL RECOVERY OF ANY KIND AGAINST KANOMAX BE GREATER IN AMOUNT THAN THE
PURCHASE PRICE OF THE PRODUCT SOLD BY KANOMAX AND CAUSING THE ALLEGED DAMAGE. WITHOUT
LIMITING THE FOREGOING, THE OWNER ASSUMES ALL RISK AND LIABILITY FOR LOSS, DAMAGE OF, OR
INJURY TO THE OWNER AND THE OWNER'S PROPERTY AND TO OTHERS AND THEIR PROPERTY ARISING
OUT OF USE OR MISUSE OF, OR INABILITY TO USE, THE PRODUCT NOT CAUSED DIRECTLY BY THE
NEGLIGENCE OF KANOMAX. THIS LIMITED WARRANTY SHALL NOT EXTEND TO ANYONE OTHER THAN THE
ORIGINAL PURCHASER OF THE PRODUCT, OR THE PERSON FOR WHOM IT WAS PURCHASED AS A GIFT,
AND STATES THE PURCHASER'S EXCLUSIVE REMEDY.
- 16 -
After-sales Service
If the PRODUCT is malfunctioning, please see “Troubleshooting” page 13 to find possible
causes first.
Repair parts are retained for a minimum period of five (5) years after production cessation of
the PRODUCT. This storage period of repair parts is considered as the period during which
KANOMAX can provide repair service.
For more information, please contact your sales representative. When you call, please have
the following information at hand:
(1) PRODUCT name;
(2) Model number;
(3) Serial number;
(4) Probe number;
(5) Description of Symptom, and;
(6) Date of purchase
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21

Kanomax 3718-A User manual

Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI