BD BodyGuard 2CH™ Twin Channel Infusion Pump TPN Mode Quick start guide

Type
Quick start guide
Twin Channel Infusion Pump

Quick user guide
CME McKinley Device Training is RCN Accredited
The Multi-Function Keypad
Keys ‘0’ to ‘9’ are used to enter numeric values during programming.
Keys ‘2’ & ‘0’ are also used to scroll up and down respectively in menus.
Powers the pump on and off. Press and hold until
conrmation beep sounds to power off.
Starts/restarts the infusion.
Conrms programming or selection during set up.
Stops the infusion when required.
Deletes numbers/settings.
Takes user back up through the programming screens.
Allows set to be primed via the pump.
During an infusion repeatedly press to access
current infusion information: volume infused/to
be infused, battery status and tube temperature).
Allows the user to move between channels.

When an Alert Activates:
Infusion continues, 3 beeps are heard. A message on the display
screen alternates with the infusion running screen until the end
of the battery or infusion.
Low/end battery the battery has approximately 30 minutes power
remaining. Connect pump to charger or change battery pack.
Program nearly complete – the programme/infusion has
approximately 30 minutes until complete. Prepare to change
bag and re-programme or close down the device.
When an Alarm Activates:
Infusion stops. LED indicator light turns red and the alarm sound
continuously until either the pump is paused or the problem is
rectied. Always note the condition before pressing the STOP/NO
button to silence the alarm.
Air/upstream occlusion – activates when either the upstream
segment of the tubing is trapped/clamped or when more than
the pre-set allowed volume of air has passed through the pump.
Silence the alarm, check the upstream tubing is not occluded.
Down occlusion – indicates that an occlusion has blocked the line
or access device causing pressure build up in the set. The clamp
may be closed or set trapped/kinked, which you should resolve. If
successful the pump automatically resumes infusion. If you cannot
detect and resolve the cause the access device may be blocked.
Silence the alarm and contact your Clinician immediately.
Refer over the page for manual and pump priming instructions.
Pump unattended – pump has been left ‘stopped’ or unattended
(may have programmed but forgotten to press Start). Alarm
sounds after 2 minutes to indicate that the pump is not infusing.
Restart pump – a system malfunction has occurred. Press the
INFO key and record the error code. Some errors can be resolved by
turning the pump off and on again. If when the pump powers on
the message is displayed again, discontinue using the equipment,
contact your Clinician/Provider immediately and report the
error code.
End program – alerts the user that the programme/infusion is
now complete.
Missing keythe administration set is loaded incorrectly.
Check that the key is positioned correctly inside the pump
before continuing.
Wrong TPN Data – the programme entered by the user cannot be
performed as it conicts with either the operating parameters of
the pump or the clinician/provider congurable settings (which
require coded user access). Consult your clinician/provider for advice.
Clinical contact:
Phone number:
E-mail contact:
k The information contained in this guide is a summary only.
Refer to Operating Manual for full operating instructions.
k Screen information and sequences displayed are based on
pump default settings and numerical values shown are
examples only. You must refer to local policy/procedure
for specic guidance on pump settings/set-up, protocols
and use of accessories (e.g. cannula and administration lines).
k When using the pump always follow screen prompts and
before pressing keys to proceed ensure selections made
correspond with what is required.
k The information contained in this guide is applicable up
to software versions TWINTPN02. NOTE: With previous
software versions some text wording/screen information
may vary slightly.
CME McKinley UK Ltd
Kincraig Business Park,
Kincraig Road,
Blackpool FY2 0PJ
Tel: 01253 894646
Fax: 01253 896648
email: info@cme-mckinley.co.uk
www.cme-mckinley.co.uk
CME McKinley/7.1.0/TPN/July10
A1704-Pocket guide BG 2CH TPN Mode AW.indd 1 7/7/10 10:13:13
Twin Channel Infusion Pump

Quick user guide
CME McKinley Device Training is RCN Accredited
The Multi-Function Keypad
Keys ‘0’ to ‘9’ are used to enter numeric values during programming.
Keys ‘2’ & ‘0’ are also used to scroll up and down respectively in menus.
Powers the pump on and off. Press and hold until
conrmation beep sounds to power off.
Starts/restarts the infusion.
Conrms programming or selection during set up.
Stops the infusion when required.
Deletes numbers/settings.
Takes user back up through the programming screens.
Allows set to be primed via the pump.
During an infusion repeatedly press to access
current infusion information: volume infused/to
be infused, battery status and tube temperature).
Allows the user to move between channels.

When an Alert Activates:
Infusion continues, 3 beeps are heard. A message on the display
screen alternates with the infusion running screen until the end
of the battery or infusion.
Low/end battery the battery has approximately 30 minutes power
remaining. Connect pump to charger or change battery pack.
Program nearly complete – the programme/infusion has
approximately 30 minutes until complete. Prepare to change
bag and re-programme or close down the device.
When an Alarm Activates:
Infusion stops. LED indicator light turns red and the alarm sound
continuously until either the pump is paused or the problem is
rectied. Always note the condition before pressing the STOP/NO
button to silence the alarm.
Air/upstream occlusion – activates when either the upstream
segment of the tubing is trapped/clamped or when more than
the pre-set allowed volume of air has passed through the pump.
Silence the alarm, check the upstream tubing is not occluded.
Down occlusion – indicates that an occlusion has blocked the line
or access device causing pressure build up in the set. The clamp
may be closed or set trapped/kinked, which you should resolve. If
successful the pump automatically resumes infusion. If you cannot
detect and resolve the cause the access device may be blocked.
Silence the alarm and contact your Clinician immediately.
Refer over the page for manual and pump priming instructions.
Pump unattended – pump has been left ‘stopped’ or unattended
(may have programmed but forgotten to press Start). Alarm
sounds after 2 minutes to indicate that the pump is not infusing.
Restart pump – a system malfunction has occurred. Press the
INFO key and record the error code. Some errors can be resolved by
turning the pump off and on again. If when the pump powers on
the message is displayed again, discontinue using the equipment,
contact your Clinician/Provider immediately and report the
error code.
End program – alerts the user that the programme/infusion is
now complete.
Missing keythe administration set is loaded incorrectly.
Check that the key is positioned correctly inside the pump
before continuing.
Wrong TPN Data – the programme entered by the user cannot be
performed as it conicts with either the operating parameters of
the pump or the clinician/provider congurable settings (which
require coded user access). Consult your clinician/provider for advice.
Clinical contact:
Phone number:
E-mail contact:
k The information contained in this guide is a summary only.
Refer to Operating Manual for full operating instructions.
k Screen information and sequences displayed are based on
pump default settings and numerical values shown are
examples only. You must refer to local policy/procedure
for specic guidance on pump settings/set-up, protocols
and use of accessories (e.g. cannula and administration lines).
k When using the pump always follow screen prompts and
before pressing keys to proceed ensure selections made
correspond with what is required.
k The information contained in this guide is applicable up
to software versions TWINTPN02. NOTE: With previous
software versions some text wording/screen information
may vary slightly.
CME McKinley UK Ltd
Kincraig Business Park,
Kincraig Road,
Blackpool FY2 0PJ
Tel: 01253 894646
Fax: 01253 896648
email: info@cme-mckinley.co.uk
www.cme-mckinley.co.uk
CME McKinley/7.1.0/TPN/July10
A1704-Pocket guide BG 2CH TPN Mode AW.indd 1 7/7/10 10:13:13
Doc Ref: CME Medical/2CH/TPN QUG/v8/Sept 2014 - QUG12
Twin Channel Infusion Pump

Quick user guide
CME McKinley Device Training is RCN Accredited
The Multi-Function Keypad
Keys ‘0’ to ‘9’ are used to enter numeric values during programming.
Keys ‘2’ & ‘0’ are also used to scroll up and down respectively in menus.
Powers the pump on and off. Press and hold until
conrmation beep sounds to power off.
Starts/restarts the infusion.
Conrms programming or selection during set up.
Stops the infusion when required.
Deletes numbers/settings.
Takes user back up through the programming screens.
Allows set to be primed via the pump.
During an infusion repeatedly press to access
current infusion information: volume infused/to
be infused, battery status and tube temperature).
Allows the user to move between channels.

When an Alert Activates:
Infusion continues, 3 beeps are heard. A message on the display
screen alternates with the infusion running screen until the end
of the battery or infusion.
Low/end battery the battery has approximately 30 minutes power
remaining. Connect pump to charger or change battery pack.
Program nearly complete – the programme/infusion has
approximately 30 minutes until complete. Prepare to change
bag and re-programme or close down the device.
When an Alarm Activates:
Infusion stops. LED indicator light turns red and the alarm sound
continuously until either the pump is paused or the problem is
rectied. Always note the condition before pressing the STOP/NO
button to silence the alarm.
Air/upstream occlusion – activates when either the upstream
segment of the tubing is trapped/clamped or when more than
the pre-set allowed volume of air has passed through the pump.
Silence the alarm, check the upstream tubing is not occluded.
Down occlusion – indicates that an occlusion has blocked the line
or access device causing pressure build up in the set. The clamp
may be closed or set trapped/kinked, which you should resolve. If
successful the pump automatically resumes infusion. If you cannot
detect and resolve the cause the access device may be blocked.
Silence the alarm and contact your Clinician immediately.
Refer over the page for manual and pump priming instructions.
Pump unattended – pump has been left ‘stopped’ or unattended
(may have programmed but forgotten to press Start). Alarm
sounds after 2 minutes to indicate that the pump is not infusing.
Restart pump – a system malfunction has occurred. Press the
INFO key and record the error code. Some errors can be resolved by
turning the pump off and on again. If when the pump powers on
the message is displayed again, discontinue using the equipment,
contact your Clinician/Provider immediately and report the
error code.
End program – alerts the user that the programme/infusion is
now complete.
Missing keythe administration set is loaded incorrectly.
Check that the key is positioned correctly inside the pump
before continuing.
Wrong TPN Data – the programme entered by the user cannot be
performed as it conicts with either the operating parameters of
the pump or the clinician/provider congurable settings (which
require coded user access). Consult your clinician/provider for advice.
Clinical contact:
Phone number:
E-mail contact:
k The information contained in this guide is a summary only.
Refer to Operating Manual for full operating instructions.
k Screen information and sequences displayed are based on
pump default settings and numerical values shown are
examples only. You must refer to local policy/procedure
for specic guidance on pump settings/set-up, protocols
and use of accessories (e.g. cannula and administration lines).
k When using the pump always follow screen prompts and
before pressing keys to proceed ensure selections made
correspond with what is required.
k The information contained in this guide is applicable up
to software versions TWINTPN02. NOTE: With previous
software versions some text wording/screen information
may vary slightly.
CME McKinley UK Ltd
Kincraig Business Park,
Kincraig Road,
Blackpool FY2 0PJ
Tel: 01253 894646
Fax: 01253 896648
email: info@cme-mckinley.co.uk
www.cme-mckinley.co.uk
CME McKinley/7.1.0/TPN/July10
A1704-Pocket guide BG 2CH TPN Mode AW.indd 1 7/7/10 10:13:13
Kincraig Business Park, Kincraig Road, Blackpool FY2 0PJ
Tel: 01253 894646 Fax: 01253 896648 info@cmemedical.co.uk
Twin Channel Infusion Pump

Quick user guide
CME McKinley Device Training is RCN Accredited
The Multi-Function Keypad
Keys ‘0’ to ‘9’ are used to enter numeric values during programming.
Keys ‘2’ & ‘0’ are also used to scroll up and down respectively in menus.
Powers the pump on and off. Press and hold until
conrmation beep sounds to power off.
Starts/restarts the infusion.
Conrms programming or selection during set up.
Stops the infusion when required.
Deletes numbers/settings.
Takes user back up through the programming screens.
Allows set to be primed via the pump.
During an infusion repeatedly press to access
current infusion information: volume infused/to
be infused, battery status and tube temperature).
Allows the user to move between channels.

When an Alert Activates:
Infusion continues, 3 beeps are heard. A message on the display
screen alternates with the infusion running screen until the end
of the battery or infusion.
Low/end battery – the battery has approximately 30 minutes power
remaining. Connect pump to charger or change battery pack.
Program nearly complete – the programme/infusion has
approximately 30 minutes until complete. Prepare to change
bag and re-programme or close down the device.
When an Alarm Activates:
Infusion stops. LED indicator light turns red and the alarm sound
continuously until either the pump is paused or the problem is
rectied. Always note the condition before pressing the STOP/NO
button to silence the alarm.
Air/upstream occlusion – activates when either the upstream
segment of the tubing is trapped/clamped or when more than
the pre-set allowed volume of air has passed through the pump.
Silence the alarm, check the upstream tubing is not occluded.
Down occlusion – indicates that an occlusion has blocked the line
or access device causing pressure build up in the set. The clamp
may be closed or set trapped/kinked, which you should resolve. If
successful the pump automatically resumes infusion. If you cannot
detect and resolve the cause the access device may be blocked.
Silence the alarm and contact your Clinician immediately.
Refer over the page for manual and pump priming instructions.
Pump unattended – pump has been left ‘stopped’ or unattended
(may have programmed but forgotten to press Start). Alarm
sounds after 2 minutes to indicate that the pump is not infusing.
Restart pump – a system malfunction has occurred. Press the
INFO key and record the error code. Some errors can be resolved by
turning the pump off and on again. If when the pump powers on
the message is displayed again, discontinue using the equipment,
contact your Clinician/Provider immediately and report the
error code.
End program – alerts the user that the programme/infusion is
now complete.
Missing keythe administration set is loaded incorrectly.
Check that the key is positioned correctly inside the pump
before continuing.
Wrong TPN Datathe programme entered by the user cannot be
performed as it conicts with either the operating parameters of
the pump or the clinician/provider congurable settings (which
require coded user access). Consult your clinician/provider for advice.
Clinical contact:
Phone number:
E-mail contact:
k The information contained in this guide is a summary only.
Refer to Operating Manual for full operating instructions.
k Screen information and sequences displayed are based on
pump default settings and numerical values shown are
examples only. You must refer to local policy/procedure
for specic guidance on pump settings/set-up, protocols
and use of accessories (e.g. cannula and administration lines).
k When using the pump always follow screen prompts and
before pressing keys to proceed ensure selections made
correspond with what is required.
k The information contained in this guide is applicable up
to software versions TWINTPN02. NOTE: With previous
software versions some text wording/screen information
may vary slightly.
CME McKinley UK Ltd
Kincraig Business Park,
Kincraig Road,
Blackpool FY2 0PJ
Tel: 01253 894646
Fax: 01253 896648
email: info@cme-mckinley.co.uk
www.cme-mckinley.co.uk
CME McKinley/7.1.0/TPN/July10
A1704-Pocket guide BG 2CH TPN Mode AW.indd 1 7/7/10 10:13:13
CME Medical Device and Advanced User training is RCN accredited
and Skills for Health Quality Marked on all the guides.
The Multi-Function Keypad
Keys ‘0’ to ‘9’ are used to enter numeric values during programming.
Keys ‘2’ & ‘0’ are also used to scroll up and down respectively in menus.
Features of the Set
Each set has an opaque pumping
segment which contains a ‘key’
shaped block that must be tted
with the protrusions pointing to
the right hand side, the key can
only be loaded one way so uid
ows in the direction of the arrows inside the pump door. Each set
has anti-siphon/anti-reux protection and a micron particulate
lter, incorporating air-elimination.
k A 1.2 lter set can be primed manually or via pump
k A 0.2 lter set can only be primed via the pump
Manual Priming
1. Remove spike cover from the giving set and spike uid bag
2. Orientate the lter so that the arrow is pointing upwards
3. Gently apply pressure to the priming button to start priming
4. Maintain the lter vertically until uid is visible at the luer
lock connector
5. Connect to access device/catheter
Pump Priming and Loading of the Set
1. Open pump door by pulling down latch
2. Insert line pumping section into the pump and press
into place
3. Close pump door and press ON/OFF to power up the pump
4. Press PRIME button
5. Either programme in priming volume required or use default
priming volume
6. Press START to
commence priming
7. On completion of priming
make sure that all air
is removed, re-prime if
necessary
8. Connect line to access
device/catheter
Key
Filter
Priming button
Programming the Pump in TPN Mode
1. Power On
Switch pump ON using the ON/OFF key
Press the START/OK key to conrm the programme or hold the
STOP/NO key to select programme required.
The ashing LED light indicates the channel you are programming.
To swap to other channel use the vk channel key.
Example shown: 2000mls over 12 hours, 1 hour ramp down
2. Volume to be Infused
Enter the volume you want to infuse
from the bag and press OK to conrm.
3. Total Time
Enter the ‘Total Time’ you want the
infusion to run over in hours and
minutes and press OK to conrm.
4. Up Time (Ramp Up)
If required enter the ramp up
duration period (in hours (hh) and
minutes (mm) or press OK to proceed
without ramp up.
The pump will gradually ‘Ramp Up’ the infusion rate over the
selected duration.
5. Down Time (Ramp Down)
If required enter the ramp down
duration period (in hours (hh) and
minutes (mm)) or press OK to
proceed without ramp up.
The pump will gradually ‘Ramp Down’ the infusion rate over the
selected duration.
6. Programme Summary
The programme summary screen displays the infusion parameters
you have just conrmed during programming or that are pre-set/
locked on the pump.
Scroll ‘up’ and down’ to review the
summary using the ‘2’ and ‘0’ keys.
Press ‘OK’ when ready to start.
During Infusion
The pump is now infusing.
The vertical text on the left indicates
either; ‘Flat’, ‘Up’ or ‘Down’ to signify
which section of the infusion is
currently running. In this example the
rate is Flat (Level) at 173mls per hour. Pump delivering is also indicated
by a ashing green LED light above the screen.
k To run channel 2, press channel key and follow previous
steps for the infusion required. When both channels
running the screen will split in half and display both rates.
k If you are running a 25 Steps or a Continuous programme
on the second channel, please refer to the relevant Quick
Guide for set up sequence.
Keypad Lock
Allows locking operation of the keypad to prevent tampering and/
or inadvertent key presses/power off.
To activate - press and hold the INFO key until the graphic lls left
to right and beeps
To deactivate - press and hold the INFO key until the graphic
empties right to left and beeps.
Resuming your Infusion
If an infusion has been stopped part
way through and the pump powered
off, switching the pump power back
on displays:
k Press ‘OK’ to resume your
infusion at the point it was stopped or interrupted.
k Press ‘NO’ to repeat the whole infusion/programme
a new infusion.
Volume 2000ml
Enter and press OK
www.CME-infusion.com
BodyGuard 2
channel 121
Channel 1
TPN
T Total Time 12:00
P
N Press OK key
T Volume 2000ml
P Max Rate: 173ml/h
N Press OK to start
T Up Time: 0:00
P
N Press OK key
F
L Rate
A 173ml/h
T
T Down Time: 1:00
P
N Press OK key
T Resume Press OK
P
N Repeat Press NO
These screens
display
automatically
with pump
information
Air Sensor
Pumping
Segment
Pump Door
Key
A1704-Pocket guide BG 2CH TPN Mode AW.indd 2 7/7/10 10:13:14