BD BodyGuard Infusion Pumps Set Instructions Quick start guide

Type
Quick start guide
BodyGuard Set Instrucons
Your safety in using the BodyGuard
TM
Infusion System is our utmost priorty. Please
only use the device according to the manufacturers operators manual or other
approved literature. In the unlikely event of experiencing an issue or probelm with
your system, please help us to invesgate the cause by following these important
steps:
1. Whenever possible retain the system and its original packaging (or make a
note of the Lot Number found on the upper le of the set packaging) so that it
can be returned to the manufacturer for invesgaon.
2. Record details of events or acvity at the me of the issue including details of
any alarms acvated on the pump immediately before or aer the event and
any other circumstances that may have contributed to the issue. No maer
how small or insignicant a detail may seem it may form an important part of
the invesgaon.
3. Contact your healthcare provider to report the event and arrange for the
collecon and return of the set.
MicroSet/BodySet instrucons for use administraon set with
male luer-lock to be used only with BodyGuard
TM
Infusion Pumps
1. Remove protecve cover and insert spike into soluon container.
2. If appropriate/preferred (oponal on sets incorporang manual prime valve)
manually prime the admin set using the priming valve with lter (if present)
orientated arrow up unl set fully primed.
3. Alternavely prime using the pumps automac priming feature aer loading the
set (see direcons in the BodyGuard
TM
operator manual) ensuring the key(s) is
located correctly.
4. Aach set securely to the paent’s access device.
5. Prior to start of the infusion refer to BodyGuard
TM
operator manual for applica-
ble precauonary instrucon. Start infusion when ready.
Warnings:
Do not use if packaging is open, damaged or protectors are missing/detached.
This system should not be used for the administraon of blood, blood products,
nitroglycerin or soluons that are signicantly light sensive. (This does not
apply to TPN whilst in the set, usually the reservoir (bag) is covered if in doubt
please contact your healthcare provider.)
Air embolism can cause death or serious injury to the paent. Do not connect
paent to the BodyGuard
TM
Infusion System unl all air has been cleared from
the uid path.
Puncturing set components may cause air embolism.
Biological contaminaon can result from re-using single use disposable items or
failing to follow good ascepc technique. Properly discard single use disposables
aer use.
Cauon: Component damage may occur if not installed properly. Ensure all
connectors are securely aached.
Important: Tubing must be installed so as uid ows in the direcon of arrows
inside the pump door.
BodyGuard ColourVision 323
TM
BodyGuard 323
TM
BodyGuard 2CH
TM
Twin Channel
Infusion Pumps
CME Medical UK Limited
Kincraig Business Park
Kincraig Road
Blackpool
FY2 0PJ
Tel: 01253 206700
Fax: 01253 896648
cmemedical.co.uk
QUG27/BodyGuard Pump Priming Instrucons/v1/May 2016
  • Page 1 1