BodyGuard Set Instrucons
Your safety in using the BodyGuard
TM
Infusion System is our utmost priorty. Please
only use the device according to the manufacturer’s operators manual or other
approved literature. In the unlikely event of experiencing an issue or probelm with
your system, please help us to invesgate the cause by following these important
steps:
1. Whenever possible retain the system and its original packaging (or make a
note of the Lot Number found on the upper le of the set packaging) so that it
can be returned to the manufacturer for invesgaon.
2. Record details of events or acvity at the me of the issue including details of
any alarms acvated on the pump immediately before or aer the event and
any other circumstances that may have contributed to the issue. No maer
how small or insignicant a detail may seem it may form an important part of
the invesgaon.
3. Contact your healthcare provider to report the event and arrange for the
collecon and return of the set.
MicroSet/BodySet instrucons for use administraon set with
male luer-lock to be used only with BodyGuard
TM
Infusion Pumps
1. Remove protecve cover and insert spike into soluon container.
2. If appropriate/preferred (oponal on sets incorporang manual prime valve)
manually prime the admin set using the priming valve with lter (if present)
orientated arrow up unl set fully primed.
3. Alternavely prime using the pumps automac priming feature aer loading the
set (see direcons in the BodyGuard
TM
operator manual) ensuring the key(s) is
located correctly.
4. Aach set securely to the paent’s access device.
5. Prior to start of the infusion refer to BodyGuard
TM
operator manual for applica-
ble precauonary instrucon. Start infusion when ready.
Warnings:
• Do not use if packaging is open, damaged or protectors are missing/detached.
• This system should not be used for the administraon of blood, blood products,
nitroglycerin or soluons that are signicantly light sensive. (This does not
apply to TPN whilst in the set, usually the reservoir (bag) is covered if in doubt
please contact your healthcare provider.)
• Air embolism can cause death or serious injury to the paent. Do not connect
paent to the BodyGuard
TM
Infusion System unl all air has been cleared from
the uid path.
• Puncturing set components may cause air embolism.
• Biological contaminaon can result from re-using single use disposable items or
failing to follow good ascepc technique. Properly discard single use disposables
aer use.
• Cauon: Component damage may occur if not installed properly. Ensure all
connectors are securely aached.
Important: Tubing must be installed so as uid ows in the direcon of arrows
inside the pump door.