cobas s 201 system

Roche cobas s 201 system User manual

  • Hello! I am an AI chatbot trained to assist you with the Roche cobas s 201 system User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Operator’s Manual
Version 1.1
For use with the
cobas® TaqScreen DPX Test
cobas® TaqScreen West Nile Virus Test
cobas® TaqScreen MPX Test
cobas® TaqScreen MPX Test, version 2.0
cobas s 201 system
Configuration E
cobas s 201 system Configuration E
Roche Diagnostics
2 Operator’s Manual · Version 1.1
Document information
Revision history
Edition notice This manual is for users of the cobas s 201 system.
Every effort has been made to ensure that all the information contained in this
manual is correct at the time of printing. However, Roche Diagnostics reserves the
right to make any changes necessary without notice as part of ongoing product
development.
Any customer modification to the instrument will render the warranty or service
agreement null and void.
Software updates are performed by Roche service representatives.
Intended use The cobas s 201 system is intended for use in testing of donor specimens for
infectious agents using polymerase chain reaction (PCR) amplification technology
and 5' nuclease real-time detection technology.
The cobas s 201 system is intended to be used by laboratory professionals trained in
laboratory techniques and on the use of the system.
This manual is not to be used for applications performed using versions of the PDM
software earlier than the 4.0 series or versions of the AMPLILINK software earlier
than the 3.3 series.
Copyright © 2014, Roche Diagnostics International Ltd. All rights reserved.
Trademarks The following trademarks are acknowledged:
ROCHE, AMPERASE, AMPLILINK, AMPLIPREP, TAQMAN, TAQSCREEN,
COBAS, and COBAS S are trademarks of Roche.
HP and STORAGEWORKS are trademarks of Hewlett-Packard Company.
MICROLAB is a trademark of Hamilton Bonaduz AG.
MICROSOFT and WINDOWS are trademarks of Microsoft Corporation.
ORACLE is a trademark of Oracle Corporation.
Manual
Version Software Version Revision Date Changes
1.0 PDM software 4.0 series
AMPLILINK software 3.3
series
September 2014 Original document
1.1 PDM software 4.0 series
AMPLILINK software 3.3
series
July 2015 Caution notes regarding use of
the Abort Advanced option
were added
Roche Diagnostics
Operator’s Manual · Version 1.1 3
cobas s 201 system Configuration E
Declaration of Conformity The cobas s 201 system is comprised of the following individual components, each of
which complies with the requirements of the Directive 98/79/EC of the European
Parliament and of the Council of 27 October 1998 on In-vitro Diagnostic Medical
Devices. Compliance is provided by means of the following individual Declarations of
Conformity.
21 CFR Part 11 The Pooling and Data Management Software (PDM SW) of the cobas s 201 system is
in compliance with Title 21 CFR Part 11 of the Code of Federal Regulations. These
regulations address the FDA’s requirements on electronic records and electronic
signatures in the United States.
Contact addresses
Pooling and Data Management
(PDM)
Roche Molecular Systems, Inc.
1080 US Highway 202 South
Branchburg, NJ 08876
USA
Hamilton MICROLAB STAR IVD /
STARlet IVD pipettor
Hamilton Company
Via Crusch 8
CH-7402 Bonaduz
Switzerland
COBAS
®
AmpliPrep instrument Roche Molecular Systems, Inc.
1080 US Highway 202 South
Branchburg, NJ 08876
USA
COBAS
®
TaqMan
®
analyzer Roche Molecular Systems, Inc.
1080 US Highway 202 South
Branchburg, NJ 08876
USA
Roche Molecular Systems, Inc.
1080 US Highway 202 South
Branchburg, NJ 08876
USA
Roche Diagnostics GmbH
Sandhofer Strasse 116
68305 Mannheim
Germany
cobas s 201 system Configuration E
Roche Diagnostics
4 Operator’s Manual · Version 1.1
Roche Diagnostics
Operator’s Manual · Version 1.1 5
cobas s 201 system Configuration E
Table of contents
Document information 2
Contact addresses 3
Table of contents 5
Preface 7
Where to find information 7
How to use this manual 8
Conventions used in this manual 10
System Description Part A
1 Safety
Safety classifications A–5
Safety precautions A–6
Safety summary A–7
2 Introduction
System overview A–15
Pipetting A–17
Preparation (extraction) A–18
Amplification and detection A–19
3 Hardware and network components
Overview A–23
Laboratory information system (LIS) A–24
Firewall provided by Roche A–24
PDM server A–24
Printer A–24
Pooling and Data Management workstation A–25
Hamilton MICROLAB STAR IVD / STARlet IVD
pipettor A–26
Test m odu l e A–2 6
Checking communication A–28
4 Pooling procedures
Pipetting overview A–31
Pooling algorithm A–32
Pipetting Primary Pools of 1 A–36
Pipetting Primary Pools of 6 A–38
Pipetting Simultaneous Pools A–40
Pipetting Primary Pools of 24 A–47
Pipetting Short Run Pools of 24 A–48
Two-step pooling for Primary Pools of 96 A–49
Two-step pooling for Primary Pools of 480 A–52
DPX Library Plate pooling A–55
Pipetting Secondary Pools of 1 A–57
Pipetting Secondary Pools of 4 A–60
Pipetting Secondary Pools of 6 A–61
Pipetting Secondary Pools of 12 A–63
Pipetting Secondary Pools of 48 A–65
Flexible loading A–67
Error handling A–68
Final Pool Size A–70
Loading Capacity A–71
5 Result assignment
Overview A–87
Batch evaluation (Controls Review tab) A–90
Pool evaluation (Pools Review tab) A–90
Donor sample evaluation (Donor Review tab) A–100
6 PDM software introduction
Overview A–109
Login service A–110
Electronic signature A–113
Screen layout A–114
Sorting and sizing columns A–117
Generating reports A–118
7 Pooling Wizard
Overview A–121
Batch Run tab A–123
Plate Run tab A–128
Pipetting Progress A–130
Including UDECs in a pooling run A–132
Interrupting a pooling run A–133
Manual Pipetting tab A–135
Maintenance tab A–138
8 Pooling Manager
Overview A–141
Query tab A–143
Details tab A–147
Worklist tab A–167
AmpliLink Status tab A–171
Reports A–174
9 Data Manager
Overview A–201
Pools to review tab A–204
Batches to review tab A–210
Query tab A–212
Details tab A–217
Worklist tab A–234
Reports A–243
10 AMPLILINK software
Overview A–263
Starting up, shutting down, and restarting A–264
Organization of the work area A–268
Working with lists A–280
Messages window A–288
Overview window A–292
Systems window A–295
Roche Diagnostics
6 Operator’s Manual · Version 1.1
cobas s 201 system Configuration E
Orders window A–324
Memo Pad window A–328
Operation Part B
11 Workflows
Introduction B–5
Safety information B–5
Two-step pooling for PP96 and PP480 B–6
DPX Library Plate workflow B–22
Aliquot Plate workflow B–40
One-step pooling for Primary Pools containing less
than 96 samples B–44
Manual pipetting B–58
12 Operating instructions
Safety information B–69
Workstations B–70
Hamilton MICROLAB STAR IVD / STARlet IVD
pipettor B–71
COBAS
®
AmpliPrep instrument B–74
COBAS
®
TaqMan
®
analyzer B–82
Reviewing test results B–87
Accepting and transmitting complete results B–89
Maintenance Part C
13 Service
Overview C–5
Hamilton MICROLAB STAR IVD / STARlet IVD
pipettor C–5
COBAS
®
AmpliPrep instrument C–6
COBAS
®
TaqMan
®
analyzer C–6
Troubleshooting Part D
14 Pooling run errors
Overview D–5
Loading and unloading errors D–6
Pipetting errors D–14
Workstation unresponsive D–17
15 Missing orders or results
Order download D–21
Result upload D–23
16 PDM software error messages
Overview D–29
List of PDM software messages D–30
17 AMPLILINK software error messages
Troubleshooting D–73
List of messages D–76
18 Test module flags
General flag description D–93
List of all AMPLILINK software flags D–96
COBAS
®
AmpliPrep instrument flag remarks D–100
COBAS
®
TaqMan
®
analyzer flag remarks D–105
COBAS
®
TaqMan
®
analyzer flags without a flag
remark D–116
Administrator’s guide Part E
19 PDM software administration
Pooling Manager E–5
Data Manager E–9
Configuration E–18
User Admin E–31
Report Manager E–39
20 PDM database maintenance
Archiving PDM data E–55
Archive Viewer E–59
Backing up PDM data E–73
21 AMPLILINK software administration
AMPLILINK software E–77
Glossary Part F
Glossary F–3
Index Part G
Index G–3
Revisions Part H
Revisions H–3
Roche Diagnostics
Operator’s Manual · Version 1.1 7
cobas s 201 system Configuration E
Preface
The cobas s 201 system is intended for use in testing of donor specimens for
infectious agents using polymerase chain reaction (PCR) amplification technology
and 5' nuclease real-time detection technology.
The cobas s 201 system is intended to be used by laboratory professionals trained in
laboratory techniques and on the use of the system.
This manual contains descriptions of the cobas s 201 system hardware and software
components and operational and troubleshooting procedures.
Where to find information
The following documents are available for the cobas s 201 system. Each document
contains applicable safety and troubleshooting information.
Operator’s Manual Contains a general description of the hardware and network components, pooling
procedures, and result assignment algorithms. Also contains a detailed explanation of
the software components and workflow procedures. Material in the Operator’s
Manual is customized for required assays. The contents of each chapter in the
Operator’s Manual are described starting on the next page.
Hamilton MICROLAB STAR
IVD / STARlet IVD Pipettor
Instrument Manual
Contains a detailed description of the Hamilton MICROLAB STAR IVD / STARlet
IVD pipettor, including technical specifications, pipetting operations, and hardware
components. Describes operation and maintenance procedures.
COBAS
®
AmpliPrep Instrument
Instrument Manual
Contains a detailed description of the COBAS
®
AmpliPrep instrument, including
technical specifications, sample preparation overview, and hardware components.
Describes operation and maintenance procedures.
COBAS
®
TaqMan
®
Analyzer
Instrument Manual
Contains a detailed description of the COBAS
®
TaqMan
®
analyzer, including
technical specifications, amplification and detection methodology, and hardware
components. Describes operation and maintenance procedures.
Package Inserts Contain test-specific instructions for use including specimen collection, storage, and
pooling procedures, materials required, and performance characteristics.
NOTICE
This manual is not to be used for applications performed using versions of the PDM
software earlier than the 4.0 series or versions of the AMPLILINK software earlier than the
3.3 series.
NOTICE
o Some of the tests identified for use with the cobas s 201 system may not be approved
for use in all countries. Contact a Roche representative to determine which tests are
commercially available in your country.
o Primary Pools of 24 are not available in the United States.
o Users of the cobas s 201 system tests should consult the test-specific package inserts
concerning use of the Docking Station.
cobas s 201 system Configuration E
Roche Diagnostics
8 Operator’s Manual · Version 1.1
How to use this manual
To help you find information quickly, there is a table of contents at the beginning of
the book and each chapter. In addition, a complete index can be found at the end.
The cobas s 201 system Configuration E Operator’s Manual contains the following
chapters:
Preface Summarizes the conventions used in this manual.
Safety Summarizes safety precautions necessary when operating the system.
Introduction Describes the cobas s 201 system major processing steps.
Hardware and network
components
Describes the cobas s 201 system hardware and network components.
Pooling procedures Describes the general algorithm for assigning pooling methods. Contains pipetting
diagrams to show pipetting patterns for each method.
Result assignment Explains how PDM software assigns batch, pool, and donor sample results.
PDM software introduction Provides a description of all PDM software components and describes common
functions.
Pooling Wizard Describes the Pooling Wizard software, which is used to operate the Hamilton
MICROLAB STAR IVD / STARlet IVD pipettor.
Pooling Manager Describes the Pooling Manager software, which is used to review pooling run status.
Data Manager Describes the Data Manager software, which is used to review and accept results, and
to transmit accepted results to the LIS.
AMPLILINK software Describes the AMPLILINK software, which is used to operate the COBAS
®
AmpliPrep instrument and COBAS
®
TaqMan
®
analyzer.
Workflows Contains instructions for routine operation of the cobas s 201 system.
Operating instructions Provides detailed instructions for starting up and loading the instruments and
reviewing, accepting, and transmitting results to the LIS.
Service Summarizes steps required for regular service of the cobas s 201 system hardware
components.
Pooling run errors Lists errors that may occur during a pooling run and provides recommended actions
for resolving the problems.
Missing orders or results Describes problems that may occur downloading orders and uploading results.
Contains instructions for determining order and result status and retransmitting
missing data.
o Keep this Operator’s Manual in a safe place to ensure that it is not damaged and
remains available for use.
o This Operator’s Manual must be easily accessible at all times.
Roche Diagnostics
Operator’s Manual · Version 1.1 9
cobas s 201 system Configuration E
PDM software error messages Lists all PDM software error messages and provides recommended actions.
AMPLILINK software error
messages
Lists all AMPLILINK software error messages and provides recommended actions.
Test module flags Lists all flags that can be generated by the COBAS
®
AmpliPrep instrument or
COBAS
®
TaqMan
®
analyzer and provides causes and recommended actions.
PDM software administration Describes the PDM software functions accessible to the laboratory administrator.
PDM database maintenance Explains how to archive and backup the PDM database.
AMPLILINK software
administration
Describes the AMPLILINK software functions accessible to the laboratory
administrator.
Glossary Contains definitions for specialized terms used in this manual.
Index Provides a list of reference pages for key terms in this manual.
cobas s 201 system Configuration E
Roche Diagnostics
10 Operator’s Manual · Version 1.1
Conventions used in this manual
This section explains the formatting conventions used in this manual.
Text conventions To convey information readily and consistently, certain text conventions are used
throughout this manual. These text conventions are as follows:
Visual cues are used to help locate and interpret information in this manual quickly.
This section explains formatting conventions used in this manual.
Symbols The following symbols are used:
Text Convention Used for
Numbered lists
Numbered procedures to be implemented sequentially during an
operation:
1. Display the Donor Review tab.
2. Select the required status filter.
Italics
Designating another section of this manual to be consulted.
e
For details, see Symbols on page 10
Bold
Terms in the user interface like tab or button names or options are
printed in bold letters.
1. Select the Pooling Wizard Maintenance tab.
2. Select Weekly Maintenance, then click Start.
>
Separates command options in a multi-level command:
Choose File > Shut Down from the menu bar.
Symbol Used for
a Start of procedure
End of procedure
o List item
e
Cross-reference
f Call-up (software reference)
Tip
Danger alert symbol
Warning alert symbol
Caution and general safety alert symbol
Electrical and electronic equipment marked with this symbol are
covered by the European directive WEEE.
The symbol denotes that the equipment must not be disposed of in
the municipal waste system.
Global Trade Item Number. Used on the type plate of the instrument.
Roche Diagnostics
Operator’s Manual · Version 1.1 11
cobas s 201 system Configuration E
Screenshots The screen representations shown in this publication are for illustrative purposes
only. The screens do not necessarily show valid data.
Abbreviations The following abbreviations are used.
µL Microliters (10
-6
liters)
AFI Absolute fluorescent increase
ALR Above linear range
ASTM American Society for Testing and Materials
B19 Parvovirus B19
BOD Beginning-of-day
BSCU Blood screening control unit
BSL-2 Biosafety level 2
CAC Cycle number of annealing temperature change
CAN Canada
CAP COBAS
®
AmpliPrep instrument
CFR Code of Federal Regulations (USA)
cLLD Capacitive liquid level detection
CLSI Clinical Laboratory Standards Institute
cm Centimeters
CO-RE Compression-induced O-ring expansion
CPU Central processing unit
CSA Canadian Standards Association
CSV Comma-separated values
Ct Cycle threshold
CTM COBAS
®
TaqMan
®
analyzer
CTS Clinical trial support
DNA Deoxyribonucleic acid
DPC Duplex positive control (positive qualitative control for HAV
and low positive quantitative control for B19)
DPX Duplex
DPX test cobas ® TaqScreen DPX Test
DSLM Double sigmoid Levenberg-Marquardt (equation)
DVD Digital versatile disc
EC European community
EN European normative
EOD End-of-day
FCC Federal Communications Commission
FTP File transfer protocol
GB Gigabyte
GHz Gigahertz
HAV Hepatitis A Virus
HBV Hepatitis B Virus
HCV Hepatitis C Virus
HIV Human immunodeficiency Virus
Hz Hertz
cobas s 201 system Configuration E
Roche Diagnostics
12 Operator’s Manual · Version 1.1
IC Internal control
IEC International Electrotechnical Commission
in Inches
IP Intermediate Plate
IU International units
kg Kilogram
LAN Local area network
lb Pounds (weight)
LED Light emitting diode
LIS Laboratory information system
LLD Liquid level detection
LMOM Levenberg-Marquardt outlier method
LP Library Plate
MB Megabytes (10
6
bytes)
MGP Magnetic glass particles
mL Milliliters (10
-3
liters)
mm Millimeter (10
-3
meters)
MMx Master Mix
MPC Multiplex positive control (positive qualitative control for
HIV-1 M, HBV, HCV)
MPX Multiplex (also appears on some screens as an abbreviation
for cobas® TaqScreen MPX Test)
MPX test cobas® TaqScreen MPX Test
MPX v2.0 test cobas® TaqScreen MPX Test, version 2.0
MPX2 Appears on some screens as an abbreviation for
cobas® TaqScreen MPX Test, version 2.0
ms Milliseconds
NEG Negative
OD Optical density
PCR Polymerase chain reaction
PDF Portable document format
PDM Pooling and data management
pipettor Hamilton MICROLAB STAR IVD / STARlet IVD pipettor
PlateIP Intermediate Plate Run for the DPX LP workflow
PlateLP Library Plate Run
Platen Batch Run for the DPX LP workflow
pLLD Pressure LLD
POS Positive
PPn Primary Pools of n
QS Quantitation standard
RAM Random-access memory
ResP1 Resolution Pooling
RF Radio frequency
RFI Relative fluorescence increase
RMEC Roche-manufactured external control
Roche Diagnostics
Operator’s Manual · Version 1.1 13
cobas s 201 system Configuration E
,
rPPn Repeat Primary Pools of n
rResP1 Repeat Resolution Pooling
rSPn Repeat Secondary Pools of n
RT Reverse transcription
R-tip Reagent tip
RW Read write
S/N Serial number
SPn Secondary Pools of n
SPU Sample processing unit
STAR Hamilton MICROLAB STAR IVD pipettor (in certain cases,
Hamilton MICROLAB STARlet IVD pipettor)
STARlet Hamilton MICROLAB STARlet IVD pipettor
StepMOM Internal step correction algorithm
S-tip Sample tip
TC Thermal cycler
TDF Test definition file
TND Target not detected
TS TaqScreen
UDEC User-defined external control
UL Underwriters Laboratories Inc.
UPS Uninterruptible power supply
US United States
USB Universal serial bus
UV Ultraviolet
VVolt
VA Volt-Ampères (power consumption)
VAC Volts (alternating current)
WWatt
WEEE Waste from electric and electronic equipment
WNV West Nile Virus
WNV test cobas ® TaqScreen West Nile Virus Test
cobas s 201 system Configuration E
Roche Diagnostics
14 Operator’s Manual · Version 1.1
1 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A–3
2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A–13
3 Hardware and network components . . . . . . . . . . . . . . . . . . . . . A–21
4 Pooling procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A–29
5 Result assignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A–85
6 PDM software introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . A–107
7 Pooling Wizard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A–119
8 Pooling Manager . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A–139
9 Data Manager . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A–199
10 AMPLILINK software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A–261
System Description
A
Roche Diagnostics
Operator’s Manual · Version 1.1 A-3
cobas s 201 system Configuration E 1 Safety
Tabl e of con tents
Safety
This chapter provides general safety information for working with the cobas s 201
system. Specific safety instructions are provided in the manual where required. For a
description of safety labels on the instruments and instructions for disposal of the
instruments, refer to the specific Instrument Manuals. For additional safety
information refer to the test-specific package inserts.
Safety classifications ................................................................................................... A–5
Safety precautions ...................................................................................................... A–6
Operator qualification .......................................................................................... A–6
Safe and proper use of the instruments ............................................................... A–6
Safety summary .......................................................................................................... A–7
Warning messages ................................................................................................ A–7
Biohazardous materials .................................................................................. A–7
Optical safety .................................................................................................. A–7
Caution messages ................................................................................................. A–8
Mechanical safety ........................................................................................... A–8
Proper sample identification .......................................................................... A–8
Correct use of pipette tips .............................................................................. A–9
Proper sample and control loading ............................................................... A–9
Reagents and other working solutions .......................................................... A–9
Cross contamination of samples .................................................................. A–10
Notices ................................................................................................................ A–11
Moving parts ................................................................................................. A–11
Proper operation .......................................................................................... A–11
Proper loading .............................................................................................. A–11
Third-party software .................................................................................... A–11
Windows display settings ............................................................................. A–11
Software viruses ............................................................................................ A–12
In this chapter
Chapter
1
Roche Diagnostics
A-4 Operator’s Manual · Version 1.1
1 Safety cobas s 201 system Configuration E
Tabl e of con tents
Roche Diagnostics
Operator’s Manual · Version 1.1 A-5
cobas s 201 system Configuration E 1 Safety
Safety classifications
Safety classifications
This section explains how precautionary information is presented in this manual.
The safety precautions and important user notes are classified according to the
ANSI Z535.6 Standard. Familiarize yourself with the following icons and their
meanings:
These symbols and signal words are used for specific hazards:
Important information that is not safety relevant is indicated by the following
symbol:
The safety alert symbol by itself without a signal word is used to promote awareness to
hazards that are generic or to direct the reader to safety information provided elsewhere in
the document.
DANGER
DANGER
Indicates a hazardous situation which, if not avoided, will result in death or serious injury.
WARNING
WARNING
Indicates a hazardous situation which, if not avoided, can result in death or serious injury.
CAUTION
CAUTION
Indicates a hazardous situation which, if not avoided, can result in minor or moderate
injury.
NOTICE
NOTICE
Indicates a hazardous situation which, if not avoided, can result in damage to equipment.
Tip
Indicates additional information on correct use of the instrument or useful tips.
Roche Diagnostics
A-6 Operator’s Manual · Version 1.1
1 Safety cobas s 201 system Configuration E
Safety precautions
Safety precautions
Operator qualification
Operators are required to have a sound knowledge of relevant guidelines and
standards, as well as the information and procedures contained in the Operator’s
Manual and Instrument Manuals.
o Do not perform operation and maintenance unless you have been trained.
o Carefully follow the procedures specified in the Operator’s Manual and
Instrument Manuals for the operation and maintenance of the instruments.
o Leave maintenance, installation, or service that is not described in the Operator’s
Manuals to trained Roche Service personnel.
o Follow standard laboratory practices, especially when working with biohazardous
material.
Safe and proper use of the instruments
Personal protective equipment o Be sure to wear appropriate protective equipment, including, but not limited to,
safety glasses with side shields, fluid resistant lab coat, and approved disposable
gloves.
o Wear a face shield if there is a chance of splash or splatter.
Accuracy/precision of measured
results
o Each laboratory must verify that instrument and reagent performance meets the
published specifications.
Correct use o Use the system only for preparing liquid samples with the provided reagents as
defined in the test-specific package insert.
Operating conditions o Operation outside of the specified ranges may lead to incorrect results or
malfunction of the instruments (see technical specifications in the Instrument
Manuals).
o Keep the Operator’s Manual in a safe place to ensure that it is not damaged and
remains available for use. This manual must be easily accessible at all times.
Particular attention must be paid to the following safety precautions. If these safety
precautions are ignored, the operator may suffer serious or fatal injury. Each precaution is
important.
/