Roche-Manufactured External Controls (RMECs)
Each batch requires Roche-manufactured external controls (RMECs).
The number of RMECs required is test-specific.
WNV testing encompasses a single analyte. One negative RMEC and one
positive RMEC must be pipetted for each batch. During pipetting, an
aliquot of the negative control is transferred to the S-tube in position 23 of
each SK24 rack. Then an aliquot of the positive control is mixed with a
second aliquot of the negative control (obtained from a second tube of
negative control to dilute the positive control) in the S-tube in position 24
of each SK24 rack (Figure 2.2).
Pipetting of RMECs always occurs before sample pipetting. This
allows the operator to correct any control pipetting errors before
sample pipetting begins.
RMECs are always placed in the last positions in each SK24 rack so
that the entire testing process, from sample preparation through
amplification and detection, is monitored by control samples.
User-Defined External Controls (UDECs)
The cobas s 201 system allows up to five user-defined external controls
(UDECs) to be assigned to each test. UDEC requirements, including the
control name, barcode pattern, lot number, expiration date, and position
of the UDEC in the SK24 rack, are specified by the laboratory
administrator.
Once UDECs are assigned for a particular test, the operator can determine
whether to include them in a pipetting run.
If included, UDECs are always pipetted in the first SK24 rack.
UDECs are identified in the Roche PDM Pooling Manager and Roche
PDM Data Manager screens and reports.
Figure 2.2
WNV RMECs in an SK24 Rack
Negative
Positive
Control
Control