Endres+Hauser Liquiline CM442/CM444/CM448 Operating instructions

Type
Operating instructions
BA00444C/07/EN/15.12
71189205
Valid as of:
Softwareversion 01.03.01
Operating Instructions
Liquiline CM442/CM444/CM448
Universal four-wire multichannel controller
Commissioning
Documentation information
Warnings
The structure, signal words and color coding of the warnings follow the specifications of ANSI Z535.6
("Product safety information in product manuals, instructions and other collateral materials").
Symbols used
Structure of the safety
symbol
Meaning
DANGER
!
Causes (/consequences)
Possible consequences if ignored
Preventive measures
This symbol alerts you to a dangerous situation.
Failure to avoid this situation will result in serious or fatal injury.
WARNING
!
Causes (/consequences)
Possible consequences if ignored
Preventive measures
This symbol alerts you to a dangerous situation.
Failure to avoid this situation can result in serious or fatal injury.
CAUTION
!
Causes (/consequences)
Possible consequences if ignored
Preventive measures
This symbol alerts you to a dangerous situation.
Failure to avoid this situation can result in medium or minor
injury.
NOTICE
Cause/situation
Possible consequences if ignored
Action/note
This symbol alerts you to situations that can cause damage to
equipment or property.
Additional information, tips
Permitted or recommended
Forbidden or not recommended
Documents supplied
The Operating Instructions are split into several parts:
Commissioning (BA00444C)
All steps that you have to carry out once only during initial commissioning
Descriptions of menus
General settings
– Display/operation
Technical data
Operation & settings (BA00450C)
Individual input settings
Sensor-specific settings
Calibration settings
Sensor-dependent diagnostics settings
Configuration of optional outputs
Current outputs
Alarm relays
–Relays
Additional functions
Limit contactor
– Controller
Cleaning programs
Data administration
Calibration (BA00451C)
Calibration menus
• Examples
Maintenance & diagnostics (BA00445C)
• Maintenance
Troubleshooting and diagnostics
Diagnostics menu
Troubleshooting instructions (troubleshooting and fault elimination)
Process-dependent errors
Accessories and spare parts
HART communication (BA00486C)
Onsite settings and installation instructions for HART
Description of the HART driver
All the Operating Instructions in all the available languages can be found on the CD-ROM
enclosed.
Liquiline CM442/CM444/CM448
4
Table of contents
1 Basic safety instructions . . . . . . 5
1.1 Requirements for personnel . . . . . . . . . . . . 5
1.2 Designated use . . . . . . . . . . . . . . . . . . . . . . 5
1.3 Occupational safety . . . . . . . . . . . . . . . . . . 5
1.4 Operational safety . . . . . . . . . . . . . . . . . . . . 6
1.5 Product safety . . . . . . . . . . . . . . . . . . . . . . 6
2 Identification. . . . . . . . . . . . . . . 7
2.1 Device designation . . . . . . . . . . . . . . . . . . . 7
2.2 Scope of delivery . . . . . . . . . . . . . . . . . . . . 7
2.3 Certificates and approvals . . . . . . . . . . . . . . 8
3 Device description . . . . . . . . . . 9
3.1 Housing closed . . . . . . . . . . . . . . . . . . . . . . 9
3.2 Housing open . . . . . . . . . . . . . . . . . . . . . . . 9
3.3 Device architecture . . . . . . . . . . . . . . . . . 10
4 Installation . . . . . . . . . . . . . . . 12
4.1 Incoming acceptance, transport, storage . . 12
4.2 Installation conditions . . . . . . . . . . . . . . . 13
4.3 Installation instructions . . . . . . . . . . . . . . 15
4.4 Post-installation check . . . . . . . . . . . . . . . 18
5 Wiring . . . . . . . . . . . . . . . . . . . 19
5.1 Quick wiring guide: remote operation . . . . 19
5.2 Opening the housing . . . . . . . . . . . . . . . . 22
5.3 Cable mounting rail . . . . . . . . . . . . . . . . . 22
5.4 Connecting the cable shield . . . . . . . . . . . 23
5.5 Cable terminals . . . . . . . . . . . . . . . . . . . . 24
5.6 Connecting the supply voltage . . . . . . . . . 25
5.7 Sensor connection . . . . . . . . . . . . . . . . . . 29
5.8 Connecting additional inputs, outputs or
relays . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.9 Connecting digital communication . . . . . . 35
5.10 Post-connection check . . . . . . . . . . . . . . . 39
6 Operation . . . . . . . . . . . . . . . . 40
6.1 Display and operating elements . . . . . . . . . 40
6.2 Operation concept . . . . . . . . . . . . . . . . . . 41
6.3 Configuration options . . . . . . . . . . . . . . . . 42
7 Commissioning . . . . . . . . . . . .45
7.1 Function check . . . . . . . . . . . . . . . . . . . . 45
7.2 Setting the bus address via the hardware . 45
7.3 Switching on the unit . . . . . . . . . . . . . . . 46
7.4 Basic setup . . . . . . . . . . . . . . . . . . . . . . . 47
7.5 Display . . . . . . . . . . . . . . . . . . . . . . . . . . 48
8 Technical data . . . . . . . . . . . . .51
8.1 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
8.2 Output . . . . . . . . . . . . . . . . . . . . . . . . . . 52
8.3 Current outputs, active . . . . . . . . . . . . . . 53
8.4 Relay outputs . . . . . . . . . . . . . . . . . . . . . 54
8.5 Protocol-specific data . . . . . . . . . . . . . . . 55
8.6 Power supply . . . . . . . . . . . . . . . . . . . . . 57
8.7 Performance characteristics . . . . . . . . . . . 58
8.8 Environment . . . . . . . . . . . . . . . . . . . . . . 59
8.9 Mechanical construction . . . . . . . . . . . . . 60
9 Control drawing. . . . . . . . . . . .62
Index . . . . . . . . . . . . . . . . . . . .63
Liquiline CM442/CM444/CM448 Basic safety instructions
Endress+Hauser 5
1 Basic safety instructions
1.1 Requirements for personnel
Installation, commissioning, operation and maintenance of the measuring system must only be
carried out by trained technical personnel.
The technical personnel must be authorized to perform the tasks by the owner-operator.
The electrical connection may only be established by an electrical technician.
The technical personnel must have read and understood these Operating Instructions and must
follow the instructions they contain.
Faults at the measuring point may only be rectified by authorized and properly trained personnel.
Repairs not described in the enclosed Operating Instructions may only be carried out directly at
the manufacturer's or by the Service Organization.
1.2 Designated use
Liquiline CM44x is a multichannel controller for connecting digital sensors with Memosens technology
in non-hazardous locations.
The device is designed for use in the following applications:
Water and wastewater
Power stations
Chemical industry
Industrial wastewater treatment plants
If the device is used for any purpose other than that described, this poses a threat to the safety of people
and the entire measuring system and is thus not permitted.
The manufacturer does not accept liability for damage caused by improper or non-designated use.
1.3 Occupational safety
As the user, you are responsible for observing the following safety regulations:
Installation guidelines
Local standards and regulations
Electromagnetic compatibility
This device has been tested for electromagnetic compatibility in accordance with the applicable
European standards for industrial applications.
The electromagnetic compatibility indicated only applies to a device that has been connected in
accordance with the instructions in
these Operating Instructions.
Basic safety instructions Liquiline CM442/CM444/CM448
6Endress+Hauser
1.4 Operational safety
Prior to commissioning the entire measuring point, check that all connections are correct. Make
sure that electric cables and hose connections are not damaged.
Do not commission damaged products. Protect them against unintentional startup. Label and
identify the damaged product as defective.
If the faults cannot be eliminated, take the products out of service and protect them against
unintentional startup.
CAUTION
!
The cleaning system is not switched off during calibration or maintenance activities
Risk of injury due to medium or cleaning agent
If a cleaning system is connected, switch it off before removing a sensor from the medium.
If you are not switching off the cleaning system because you wish to test the cleaning function, wear
protective clothing, goggles and gloves or take other appropriate measures.
1.5 Product safety
The transmitter is designed to meet state-of-the-art safety requirements, has been tested and left the
factory in a condition in which it is safe to operate.
The applicable regulations and European standards have been taken into account.
Liquiline CM442/CM444/CM448 Identification
Endress+Hauser 7
2 Identification
2.1 Device designation
2.1.1 Nameplate
Nameplates can be found:
On the outside of the housing
On the packaging (adhesive label, portrait format)
On the inside of the display cover
The nameplate provides you with the following information on your device:
Manufacturer identification
Order code
Extended order code
Serial number
Firmware version
Input and output values
Protection class
• Environment
Activation codes
Safety notices and warnings
Compare the data on the nameplate with your order.
2.1.2 Serial number and order code
The order code and serial number of your device can be found in the following locations:
On the nameplate
On the front page of these Operating Instructions
In the delivery papers
To find out what device version you have, enter the order code indicated on the nameplate in the
search screen at the following address:
www.products.endress.com/order-ident
2.2 Scope of delivery
1 controller in the version ordered
1 mounting plate
1 wiring label (attached at the factory to the inside of the display cover)
1 CD with Operating Instructions
1 printed copy of the "Commissioning" section of the Operating Instructions in the language ordered
If you have any queries, please contact your supplier or local sales center.
Identification Liquiline CM442/CM444/CM448
8Endress+Hauser
2.3 Certificates and approvals
2.3.1 CE mark: Declaration of Conformity
With this declaration, the manufacturer guarantees that the product conforms to the regulations of
European Directive 2004/108/EC and Low Voltage Directive 2006/95/EC. This is proven by
observing the standards listed in the Declaration of Conformity.
2.3.2 cCSAus
The product meets the requirements of "CLASS 2252 05 - Process Control Equipment" and
"CLASS 2252 85 - Process Control Equipment - Certified to US Standards".
2.3.3 FM/CSA (CM442 only)
FM/CSA Cl. 1, Div. 2
2.3.4 MCERTS
Certification has been requested for CM442
Liquiline CM442/CM444/CM448 Device description
Endress+Hauser 9
3 Device description
3.1 Housing closed
3.2 Housing open
a0012681
Fig. 1: Liquiline mounted on a post
1
2
3
4
5
6
Display
Weather protection cover (optional)
Navigator
Sensor or current output cable
Power supply cable
Soft keys, assignment depends on menu
a0016165
Fig. 2: Example of a CM444 with an open display cover (without wiring)
1
2
3
4
5
Storage slot for SD card
Basic module
Extension modules (optional)
Shock protection, dummy cover and end cover
Extension backplane
6
7
8
9
10
Cable mounting rail
Threaded bolt for protective ground connection
Extension power unit
M12 connectors for sensor connection (optional)
Display cable
Device description Liquiline CM442/CM444/CM448
10 Endress+Hauser
3.3 Device architecture
3.3.1 Slot and port coding
The electronic configuration follows a modular concept:
Several slots are available for electronics modules.
These slots are numbered consecutively in the housing. Slots 0 and 1 are always reserved for the basic
module.
Each electronics module has one or more inputs or outputs or relays, referred collectively to as
"ports".
Ports are numbered consecutively for each electronics module and are automatically recognized by
the software.
a0015819-en
Fig. 3: Slot and port coding of hardware and on-screen display
Liquiline CM442/CM444/CM448 Device description
Endress+Hauser 11
The outputs and relays are named after their function, e.g. "Current output", and are displayed in
ascending order with the slot and port numbers
Example, å3:
If "Current output 2:1" is displayed, this means:
Slot 2 (e.g. AOR module) : Port 1 (current output 1 of AOR module)
Inputs are assigned to measuring channels in ascending order "Slot:Port number"
Example:
If "CH1: 1:1 pH glass" is displayed, this means:
Channel 1 (CH1) is slot 1 (basic module) : port 1 (input 1) and a pH glass sensor is connected here
Consequently, this results in the following unique terminal name:
Slot No. : Port No. : Terminal
3.3.2 Terminal diagram
Example for creating a terminal diagram:
Device with 4 inputs for digital sensors, 4 current outputs and 4 relays
Basic module BASE-E (contains 2 sensor inputs, 2 current outputs)
Module 2DS (2 sensor inputs)
Module 2AO (2 current outputs)
Module 4R (4 relays)
a0015979
Fig. 4: Creating a terminal diagram taking the example of the NO contact (terminal 41) of a relay
Installation Liquiline CM442/CM444/CM448
12 Endress+Hauser
4 Installation
4.1 Incoming acceptance, transport, storage
Make sure the packaging is not damaged!
Inform your supplier of any damage to the packaging. Please keep the damaged packaging until any
issues have been resolved.
Make sure the contents are not damaged!
Inform your supplier of any damage to the contents. Please keep the damaged goods until any issues
have been resolved.
Check the scope of delivery against the delivery papers and your order to ensure it is complete and
nothing is missing.
Pack the product in such a way as to protect it reliably against impact and moisture for storage and
transportation. Optimum protection is provided by the original packaging materials. In addition, the
permitted ambient conditions must be observed (see Technical data).
If you have any queries, please contact your supplier or local sales center.
Liquiline CM442/CM444/CM448 Installation
Endress+Hauser 13
4.2 Installation conditions
4.2.1 Dimensions
a0012396
Fig. 5: Field housing dimensions
Installation Liquiline CM442/CM444/CM448
14 Endress+Hauser
4.2.2 Mounting plate
4.2.3 Weather protection cover (optional)
NOTICE
Effect of climatic conditions (rain, snow, direct sun etc.)
Impaired operation to complete transmitter failure
When installing outside, always use the weather protection cover (accessory).
a0012426
Fig. 6: Mounting plate
a0012428
Fig. 7: Weather protection cover for CM44x
Liquiline CM442/CM444/CM448 Installation
Endress+Hauser 15
4.3 Installation instructions
You will require the post mounting kit to mount the unit on a pipe, post or railing (circular or square,
clamping range 20 to 61 mm (0.79 to 2.40")). The post mounting kit can be ordered as an option.
4.3.1 Post mounting
a0012665
Fig. 8: Post mounting (exploded view)
1
2
3
4
Weather protection cover (optional)
Post mounting plate (post mounting kit)
Spring washers and nuts (post mounting kit)
Pipe clamps (post mounting kit)
5
6
7
8
Spring washers and nuts (post mounting kit)
Pipe or post (circular/square)
Mounting plate
Threaded rods (post mounting kit)
a0012666
Fig. 9: Post mounting
a0012667
Fig. 10: Attaching Liquiline, clicking into place
Installation Liquiline CM442/CM444/CM448
16 Endress+Hauser
4.3.2 Railing mounting
a0012668
Fig. 11: Railing mounting (exploded view)
1
2
3
4
5
Weather protection cover (optional)
Post mounting plate (post mounting kit)
Spring washers, nuts (post mounting kit)
Pipe clamps (post mounting kit)
Spring washers, nuts (post mounting kit)
6
7
8
9
Pipe or railing (circular/square)
Mounting plate
Threaded rods (post mounting kit)
Screws (post mounting kit)
a0012669
Fig. 12: Railing mounting
a0012670
Fig. 13: Attaching Liquiline, clicking into place
Liquiline CM442/CM444/CM448 Installation
Endress+Hauser 17
4.3.3 Wall mounting
Mount the controller in such a way that the wall support surface is the size of the rear housing panel at
the very least.
a0012686
Fig. 14: Necessary spacing
a0012684
Fig. 15: Wall mounting
1Wall
2 4 drill holes1)
3 Mounting plate
4 Screws Ø 6 mm (not part of
scope of supply)
1) The size of the drill holes depends on the wall plugs used. The wall plugs and screws must be provided by the customer.
a0012685
Fig. 16: Attaching Liquiline, clicking into
place
Installation Liquiline CM442/CM444/CM448
18 Endress+Hauser
4.3.4 Disassembly (for conversion, cleaning etc.)
NOTICE
The controller can be damaged if dropped
When pushing the housing out of the holder, secure the housing in such a way that you do not drop
it. If possible, ask a second person to help you.
4.4 Post-installation check
After installation, check the transmitter for damage.
Check whether the transmitter is protected against moisture and direct sunlight (e.g. by the weather
protection cover).
a0013154
Fig. 17: Disassembly
1. Hold down the latch
2. Push up the housing to remove it from the holder
a0013155
Fig. 18: Disassembly
3. Remove the controller from the front
Liquiline CM442/CM444/CM448 Wiring
Endress+Hauser 19
5Wiring
WARNING
!
The device is live!
Incorrect wiring can result in injury or fatality
The electrical connection may only be established by an electrical technician.
The electrical technician must have read and understood these Operating Instructions and must
follow the instructions they contain.
Prior to beginning any wiring work, make sure voltage is not applied to any of the cables.
5.1 Quick wiring guide: remote operation
5.1.1 Via HART (e.g. via HART modem and FieldCare)
a0015608
Fig. 19: HART via modem
1 Device module Base-L, -H or -E: current output 1 with HART
2 HART modem for connecting to PC, e.g. Commubox FXA191 (RS232) or FXA1951) (USB)
3 HART handheld terminal
1) Switch set to "on" (substitutes the resistor)
Wiring Liquiline CM442/CM444/CM448
20 Endress+Hauser
5.1.2 Via PROFIBUS DP
a0015874
Fig. 20: PROFIBUS DP
TTerminating resistor
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68

Endres+Hauser Liquiline CM442/CM444/CM448 Operating instructions

Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI