BA Liquiline CM442/CM444/CM448

Endres+Hauser BA Liquiline CM442/CM444/CM448 Operating instructions

  • Hello! I am an AI chatbot trained to assist you with the Endres+Hauser BA Liquiline CM442/CM444/CM448 Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Products Solutions Services
Operating Instructions
Liquiline CM442/CM444/CM448
Universal four-wire multichannel controller
Field device
BA00444C/07/EN/31.23-00
71625985
2023-08-25
Valid as of version
01.13.00
Liquiline CM442/CM444/CM448 Table of contents
Endress+Hauser 3
Table of contents
1 About this document ................ 5
1.1 Warnings ............................ 5
1.2 Symbols .............................. 5
1.3 Symbols on the device ................... 5
1.4 Documentation ........................ 6
2 Basic safety instructions ............ 7
2.1 Requirements of the personnel ............. 7
2.2 Intended use .......................... 7
2.3 Workplace safety ....................... 7
2.4 Operational safety ...................... 8
2.5 Product safety ......................... 8
3 Device description .................. 9
3.1 Housing closed ........................ 9
3.2 Standard device ........................ 9
3.3 Open?? device with sensor communication
module type 2DS Ex-i .................. 10
3.4 Slot and port assignment ................ 11
3.5 Terminal diagram ..................... 12
4 Incoming acceptance and product
identification ..................... 13
4.1 Incoming acceptance ................... 13
4.2 Product identification ................... 13
4.3 Scope of delivery ...................... 14
5 Mounting ......................... 15
5.1 Mounting requirements ................. 15
5.2 Mounting the measuring device ........... 16
5.3 Post-mounting check ................... 19
6 Electrical connection .............. 20
6.1 Connection conditions .................. 20
6.2 Connecting the measuring device .......... 20
6.3 Connecting the sensors ................. 27
6.4 Connecting additional inputs, outputs or
relays .............................. 30
6.5 Connecting PROFIBUS DP or Modbus RS 485 .34
6.6 Hardware settings ..................... 38
6.7 Ensuring the degree of protection .......... 39
6.8 Post-connection check .................. 40
7 System integration ................ 41
7.1 Web server .......................... 41
7.2 Service interface ...................... 43
7.3 Fieldbus systems ...................... 44
8 Operation options ................. 45
8.1 Overview ............................ 45
8.2 Access to operating menu via local display ... 46
8.3 Configuration options .................. 47
9 Commissioning .................... 49
9.1 Function check ....................... 49
9.2 Power-up ........................... 49
9.3 User definable screens ................. 50
9.4 Basic setup .......................... 51
10 Operation ......................... 52
10.1 Display ............................. 52
10.2 General settings ....................... 54
10.3 Current inputs ........................ 68
10.4 Outputs ............................. 68
10.5 Binary inputs and outputs ............... 78
10.6 Additional functions ................... 84
11 Calibration ...................... 117
12 Diagnosis and troubleshooting .... 118
12.1 General troubleshooting ................ 118
12.2 Diagnostic information on local display ..... 119
12.3 Diagnostic information via web browser .... 119
12.4 Diagnostic information via fieldbus ........ 119
12.5 Adapting the diagnostic information ...... 119
12.6 Overview of diagnostic information ....... 121
12.7 Pending diagnostic messages ............ 125
12.8 Diagnostics list ...................... 126
12.9 Event logbook ....................... 126
12.10 Simulation ......................... 129
12.11 Device test .......................... 130
12.12 Resetting the measuring device .......... 131
12.13 Device information ................... 132
12.14 Firmware history ..................... 135
13 Maintenance .................... 139
13.1 Cleaning ........................... 139
14 Repair ........................... 141
14.1 General notes ....................... 141
14.2 Spare parts ......................... 141
14.3 Return ............................. 141
14.4 Disposal ........................... 141
15 Accessories ...................... 142
15.1 Device-specific accessories .............. 142
15.2 Communication-specific accessories ....... 148
15.3 Service-specific accessories .............. 149
15.4 System components ................... 150
15.5 Other accessories ..................... 151
Table of contents Liquiline CM442/CM444/CM448
4 Endress+Hauser
16 Technical data ................... 152
16.1 Input .............................. 152
16.2 Digital inputs, passive ................. 153
16.3 Current input, passive ................. 153
16.4 Output ............................ 153
16.5 Digital outputs, passive ................ 155
16.6 Current outputs, active ................. 155
16.7 Relay outputs ........................ 156
16.8 Protocol-specific data .................. 157
16.9 Energy supply ....................... 160
16.10 Performance characteristics ............. 162
16.11 Environment ........................ 163
16.12 Mechanical construction ............... 165
Index ................................. 167
Liquiline CM442/CM444/CM448 About this document
Endress+Hauser 5
1 About this document
1.1 Warnings
Structure of information Meaning
LDANGER
Causes (/consequences)
If necessary, Consequences of
non-compliance (if applicable)
Corrective action
This symbol alerts you to a dangerous situation.
Failure to avoid the dangerous situation will result in a fatal or serious
injury.
LWARNING
Causes (/consequences)
If necessary, Consequences of
non-compliance (if applicable)
Corrective action
This symbol alerts you to a dangerous situation.
Failure to avoid the dangerous situation can result in a fatal or serious
injury.
LCAUTION
Causes (/consequences)
If necessary, Consequences of
non-compliance (if applicable)
Corrective action
This symbol alerts you to a dangerous situation.
Failure to avoid this situation can result in minor or more serious injuries.
NOTICE
Cause/situation
If necessary, Consequences of
non-compliance (if applicable)
Action/note
This symbol alerts you to situations which may result in damage to
property.
1.2 Symbols
Symbol Meaning
Additional information, tips
Permitted or recommended
Recommended
Not permitted or not recommended
Reference to device documentation
Reference to page
Reference to graphic
Result of a step
1.3 Symbols on the device
Symbol Meaning
Reference to device documentation
Do not dispose of products bearing this marking as unsorted municipal
waste. Instead, return them to the manufacturer for disposal under the
applicable conditions.
About this document Liquiline CM442/CM444/CM448
6 Endress+Hauser
1.4 Documentation
The following manuals complement these Operating Instructions and are available on the
product pages on the Internet:
Brief Operating Instructions for Liquiline CM44x, KA01159C
Operating Instructions for Memosens, BA01245C
Software description for Memosens inputs
Calibration of Memosens sensors
Sensor-specific diagnostics and troubleshooting
Operating Instructions for HART communication, BA00486C
Onsite settings and installation instructions for HART
Description of HART driver
Guidelines for communication via fieldbus and web server
HART, SD01187C
PROFIBUS, SD01188C
Modbus, SD01189C
Web server, SD01190C
EtherNet/IP, SD01293C
PROFINET, SD02490C
Liquiline CM442/CM444/CM448 Basic safety instructions
Endress+Hauser 7
2 Basic safety instructions
2.1 Requirements of the personnel
Installation, commissioning, operation and maintenance of the measuring system may
be carried out only by specially trained technical personnel.
The technical personnel must be authorized by the plant operator to carry out the
specified activities.
The electrical connection may be performed only by an electrical technician.
The technical personnel must have read and understood these Operating Instructions
and must follow the instructions contained therein.
Faults at the measuring point may only be rectified by authorized and specially trained
personnel.
Repairs not described in the Operating Instructions provided must be carried out only
directly at the manufacturer's site or by the service organization.
2.2 Intended use
2.2.1 Non-hazardous environment
Liquiline CM44x is a multichannel controller for connecting digital sensors with
Memosens technology in non-hazardous environments.
The device is designed for use in the following applications:
Food and beverages
Life science
Water and wastewater
Chemical industry
Power stations
Other industrial applications
2.2.2 Hazardous environment
Pay attention to the information in the relevant documents relating to safety
instructions (XA).
2.2.3 Non-intended use
Use of the device for any purpose other than that described, poses a threat to the safety of
people and of the entire measuring system and is therefore not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.
2.3 Workplace safety
As the user, you are responsible for complying with the following safety conditions:
Installation guidelines
Local standards and regulations
Regulations for explosion protection
Electromagnetic compatibility
The product has been tested for electromagnetic compatibility in accordance with the
applicable international standards for industrial applications.
The electromagnetic compatibility indicated applies only to a product that has been
connected in accordance with these Operating Instructions.
Basic safety instructions Liquiline CM442/CM444/CM448
8 Endress+Hauser
2.4 Operational safety
Before commissioning the entire measuring point:
1. Verify that all connections are correct.
2. Ensure that electrical cables and hose connections are undamaged.
3. Do not operate damaged products, and protect them against unintentional operation.
4. Label damaged products as defective.
During operation:
If faults cannot be rectified:
products must be taken out of service and protected against unintentional operation.
LCAUTION
Programs not switched off during maintenance activities.
Risk of injury due to medium or cleaning agent!
Quit any programs that are active.
Switch to the service mode.
If testing the cleaning function while cleaning is in progress, wear protective clothing,
goggles and gloves or take other suitable measures to protect yourself.
2.5 Product safety
2.5.1 State-of-the-art
The product is designed to meet state-of-the-art safety requirements, has been tested, and
left the factory in a condition in which it is safe to operate. The relevant regulations and
international standards have been observed.
2.5.2 IT security
We only provide a warranty if the device is installed and used as described in the
Operating Instructions. The device is equipped with security mechanisms to protect it
against any inadvertent changes to the device settings.
IT security measures in line with operators' security standards and designed to provide
additional protection for the device and device data transfer must be implemented by the
operators themselves.
Liquiline CM442/CM444/CM448 Device description
Endress+Hauser 9
3 Device description
3.1 Housing closed
1 2
3
4
5
6
A0025813
 1 Mounted on a post
1 Display
2 Weather protection cover (optional)
3 Navigator
4 Sensor cable or current output cable
5 Power supply cable
6 Soft keys, assignment depends on menu
3.2 Standard device
3.2.1 Housing open
1 2 3
6
78
10
11
5
4
9
A0039719
 2 Example of a four-channel device with an open display cover (without wiring)
1 Display cable 7 Threaded bolt for protective ground connection
2 Basic module 8 Extension power unit with internal cable
3 Extension modules (optional) 9 M12 connectors for sensor connection
(optional)
4 Shock protection, dummy cover and end cover 10 Distributor terminals for user-defined use 1)
5 Extension backplane 11 Storage slot for SD card
6 Cable mounting rail
1) Example: you want to loop the signal from the alarm relay to a siren and a lamp. The terminals on the
alarm relay only accommodate one cable. Route the signal from the alarm relay to a terminal on the
distributor block. The block's terminals are all interconnected. You therefore have 3 additional terminals on
this block from where you can carry the signal forward to the consumer (siren, lamp etc.). The signal can
be multiplied in this way.
Device description Liquiline CM442/CM444/CM448
10 Endress+Hauser
3.3 Open?? device with sensor communication module
type 2DS Ex-i
1 2 3 5
7
89
10
11
6
4
A0045639
 3 Example of a field device with sensor communication module type 2DS Ex-i and open display cover
(without wiring)
1 Display cable 7 Cable mounting rail
2 Basic module 8 Threaded bolt for protective ground connection
3 Extension modules (optional) 9 Extension power unit
4 Separation element (pre-installed) 10 Distributor terminals for user-defined use 1)
5 Sensor communication modules 2DS Ex-i 11 Storage slot for SD card
6 Shock protection, dummy cover and end cover
1) Example: you want to loop the signal from the alarm relay to a siren and a lamp. The terminals on the
alarm relay only accommodate one cable. Route the signal from the alarm relay to a terminal on the
distributor block. The block's terminals are all interconnected. You therefore have 3 additional terminals on
this block from where you can carry the signal forward to the consumer (siren, lamp etc.). The signal can
be multiplied in this way.
Liquiline CM442/CM444/CM448 Device description
Endress+Hauser 11
3.4 Slot and port assignment
BASE2-E
Slots
1
2
1
2
1
2
1
2
Ports
2DS 2DS 2DS 2AI 2AO 4R
1
2
Sensor 1
Sensor 1
Sensor 1
Sensor 2
Sensor 2
Sensor 2
1
2
1
2
1
2
1
2
3
4
 4 Slot and port assignment of the hardware modules
 5 Slot and port assignment on the
display
Inputs are assigned to measuring channels in the
ascending order of the slots and ports.
Adjacent example:
"CH1: 1:1 pH glass" means:
Channel 1 (CH1) is slot 1 (basic module) : Port 1 (input
1), pH glass sensor
Outputs and relays are named according to their
function, e.g. "current output", and are displayed in
ascending order with the slot and port numbers
Device description Liquiline CM442/CM444/CM448
12 Endress+Hauser
3.5 Terminal diagram
The unique terminal name is derived from:
Slot no. : Port no. : Terminal
Example, NO contact of a relay
Device with 4 inputs for digital sensors, 4 current outputs and 4 relays
Base module BASE2-E (contains 2 sensor inputs, 2 current outputs)
2DS module (2 sensor inputs)
2AO module (2 current outputs)
4R module (4 relays)
1 2 34
43 42 41
Relay 1
Relay 2
Relay 3
Relay 4
Slot
Slot 4 Port 1 Pin 41
BASE2-E 2DS 4R
2AO
43 42 41
43 42 41
43 42 41
43 42 41
A0039621
 6 Creating a terminal diagram using the example of the NO contact (terminal 41) of a relay
Liquiline CM442/CM444/CM448 Incoming acceptance and product identification
Endress+Hauser 13
4 Incoming acceptance and product
identification
4.1 Incoming acceptance
1. Verify that the packaging is undamaged.
Notify the supplier of any damage to the packaging.
Keep the damaged packaging until the issue has been resolved.
2. Verify that the contents are undamaged.
Notify the supplier of any damage to the delivery contents.
Keep the damaged goods until the issue has been resolved.
3. Check that the delivery is complete and nothing is missing.
Compare the shipping documents with your order.
4. Pack the product for storage and transportation in such a way that it is protected
against impact and moisture.
The original packaging offers the best protection.
Make sure to comply with the permitted ambient conditions.
If you have any questions, please contact your supplier or your local Sales Center.
4.2 Product identification
4.2.1 Nameplate
The nameplate provides you with the following information on your device:
Manufacturer identification
Order code
Extended order code
Serial number
Firmware version
Ambient conditions
Input and output values
Activation codes
Safety information and warnings
Degree of protection
Compare the information on the nameplate with the order.
4.2.2 Identifying the product
Product page
www.endress.com/cm442
www.endress.com/cm444
www.endress.com/cm448
Interpreting the order code
The order code and serial number of your product can be found in the following locations:
On the nameplate
In the delivery papers
Incoming acceptance and product identification Liquiline CM442/CM444/CM448
14 Endress+Hauser
Obtaining information on the product
1. Go to www.endress.com.
2. Page search (magnifying glass symbol): Enter valid serial number.
3. Search (magnifying glass).
The product structure is displayed in a popup window.
4. Click the product overview.
A new window opens. Here you fill information pertaining to your device,
including the product documentation.
Manufacturer address
Endress+Hauser Conducta GmbH+Co. KG
Dieselstraße 24
70839 Gerlingen
Germany
4.3 Scope of delivery
The scope of delivery comprises:
1 multichannel controller in the version ordered
1 mounting plate
1 wiring label (attached at the factory to the inside of the display cover)
1 printed copy of the Brief Operating Instructions in the language ordered
Separation element (pre-installed on hazardous area version type 2DS Ex-i)
Safety instructions for the hazardous area (for hazardous area version type 2DS Ex-i)
If you have any queries:
Please contact your supplier or local sales center.
Liquiline CM442/CM444/CM448 Mounting
Endress+Hauser 15
5 Mounting
5.1 Mounting requirements
5.1.1 Dimensions
H 5 6 7 8 I
GF
E
D
C
A
1 2 3 4 B
162 (6.38)
14 (0.55)
Ø4 (0.16)
Ø9 (0.35)
199 (7.38) 128 (5.04)
237 (9.33)
194 (7.64)
76 (2.99)
95 (3.74)
A0012396
 7 Dimensions of field housing in mm (in)
5.1.2 Mounting plate
80 (3.15)
190 (7.48)
3 (0.12)
4 x 6.5 (0.26)
170 (6.69)
125 (4.92)
A0012426
 8 Mounting plate. Engineering unit: mm (in)
Mounting Liquiline CM442/CM444/CM448
16 Endress+Hauser
5.1.3 Weather protection cover
NOTICE
Effect of climatic conditions (rain, snow, direct sunlight etc.)
Impaired operation to complete transmitter failure are possible!
Always use the weather protection cover (accessory) when installing the device
outdoors.
320 (12.6)
270 (10.6)
300 (11.8)
80 (3.15)
170 (6.69)
170 (6.69)
80 (3.15)
8 x Ø 7 (0.28)
A0012428
 9 Dimensions in mm (in)
5.2 Mounting the measuring device
5.2.1 Post mounting
You require the post mounting kit (optional) to mount the unit on a pipe, post or
railing (square or circular, clamping range 20 to 61 mm (0.79 to 2.40")).
1234
56
7
8
A0033044
 10 Post mounting
1 Weather protection cover (optional) 5 Spring washers and nuts (post mounting kit)
2 Post mounting plate (post mounting kit) 6 Pipe or railing (circular/square)
3 Spring washers and nuts (post mounting kit) 7 Mounting plate
4 Pipe clamps (post mounting kit) 8 Threaded rods (post mounting kit)
Liquiline CM442/CM444/CM448 Mounting
Endress+Hauser 17
A0033045
 11 Post mounting
A0025885
 12 Attach the device and click it into place
1. Place the device on the mounting plate.
2. Slide the device downwards in the guide on the mounting rail until it clicks into place.
5.2.2 Rail mounting
1 2 34
56
7
8
9
A0012668
 13 Rail mounting
1 Weather protection cover (optional) 6 Pipe or railing (circular/square)
2 Post mounting plate (post mounting kit) 7 Mounting plate
3 Spring washers and nuts (post mounting kit) 8 Threaded rods (post mounting kit)
4 Pipe clamps (post mounting kit) 9 Screws (post mounting kit)
5 Spring washers and nuts (post mounting kit)
A0025886
 14 Rail mounting
A0027803
 15 Attach the device and click it into place
1. Place the device on the mounting plate.
2. Slide the device downwards in the guide on the mounting rail until it clicks into place.
Mounting Liquiline CM442/CM444/CM448
18 Endress+Hauser
5.2.3 Wall mounting
A0012686
 16 Installation clearance in mm (in)
1
3
4
2
A0027798
 17 Wall mounting
1 Wall
24 drill holes 1)
3 Mounting plate
4 Screws Ø 6 mm (not part of scope of supply)
1)The size of the drill holes depends on the wall plugs used. The wall plugs and screws must be provided by the
customer.
A0027799
 18 Wall mounting
A0027797
 19 Attach the device and click it into place
1. Place the device on the mounting plate.
2. Slide the device downwards in the guide on the mounting rail until it clicks into place.
5.2.4 Disassembly (for conversion, cleaning etc.)
NOTICE
The device can be damaged if dropped
When pushing the housing out of the holder, secure the housing to prevent it from
falling. If possible, ask a second person to help you.
Liquiline CM442/CM444/CM448 Mounting
Endress+Hauser 19
A0025890
 20 Disassembling
A0025891
 21 Disassembling
1. Hold down the latch.
2. Push up the device to remove it from the holder.
3. Remove the device towards the front.
5.3 Post-mounting check
1. After installation, check the transmitter for damage.
2. Check whether the transmitter is protected against precipitation and direct sunlight
(e.g. by the weather protection cover).
Electrical connection Liquiline CM442/CM444/CM448
20 Endress+Hauser
6 Electrical connection
6.1 Connection conditions
6.1.1 Via HART (e.g. via HART modem and FieldCare)
1
3
2
USB
31
32
250 ... 500 Ω
A0039620
 22 HART using modem
1 Device module Base2-L, -H or -E: current output 1 with HART
2HART modem for connection to PC, e.g. Commubox FXA191 (RS232) or FXA195 1) (USB)
3 HART handheld terminal
1) Switch position "on" (substitutes the resistor)
6.2 Connecting the measuring device
LWARNING
Device is live!
Incorrect connection may result in injury or death!
The electrical connection may be performed only by an electrical technician.
The electrical technician must have read and understood these Operating Instructions
and must follow the instructions contained therein.
Prior to commencing connection work, ensure that no voltage is present on any cable.
NOTICE
The device does not have a power switch!
Provide a protected circuit breaker in the vicinity of the device at the place of
installation.
The circuit breaker must be a switch or power switch, and must be labeled as the circuit
breaker for the device.
At the supply point, the power supply must be isolated from dangerous live cables by
double or reinforced insulation in the case of devices with a 24 V supply voltage.
6.2.1 Opening the housing
NOTICE
Pointed or sharp tools
The use of unsuitable tools can cause scratches on the housing or damage to the seal, and
therefore negatively affect the leak-tightness of the housing!
Do not use any sharp or pointed objects, e.g. a knife, to open the housing.
Only use a suitable Phillips screwdriver.
/